K Number
K133676

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-07-11

(224 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ABX Pentra Micro ALB Control L/H is a quality control used to monitor the performance of ABX Pentra Micro ALBUMIN CP determination by immunoturbidimetry.

Device Description

The ABX Pentra Micro ALB Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with added chemicals, constituents of human origin, stabilizers and preservatives. The assigned values and precise confidence interval are given in an enclosed annex, ensuring control of the appropriate HORIBA ABX SAS methods on the HORIBA clinical chemistry analyzer. Each control level is provided in one vial of 10 ml.

AI/ML Overview

The provided text describes the ABX Pentra Micro ALB Control L/H, a quality control material used to monitor the performance of Microalbumin determination. The study described focuses on establishing the stability claims of this control rather than evaluating the performance of a diagnostic device in detecting a medical condition. Therefore, many of the requested categories in the prompt (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set size) are not applicable to this type of device and study.

Here's an analysis of the provided information concerning the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Shelf Life24 months at 2 to 8°C
Open Vial Stability90 days at 2 to 8°C
Performance Monitoring (Intended Use)Monitors the performance of ABX Pentra Micro ALBUMIN CP determination by immunoturbidimetry.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated as a separate "test set" for an accuracy study. The stability studies (shelf life and open vial) would inherently involve testing multiple aliquots over time following specific protocols. The number of samples/replicates used in these stability studies is not detailed.
  • Data Provenance: The study was conducted by HORIBA ABX SAS, located in Montpellier, France. The data appears to be prospective, collected for the purpose of demonstrating stability claims.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This is not applicable as the device is a quality control material and the study focuses on its stability. "Ground truth" in the context of diagnostic accuracy (e.g., pathology, clinical outcomes) is not relevant here. The "ground truth" for a control material involves its assigned values, which are determined internally by the manufacturer through rigorous calibration and value assignment processes.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This is not applicable as the study is not evaluating a diagnostic device's performance against a clinical ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This is a quality control material, not a diagnostic device involving human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is a quality control material; there is no algorithm or human-in-the-loop performance to evaluate in this context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this control material is its assigned values and corresponding ±3SD confidence ranges, which are determined by the manufacturer (HORIBA ABX SAS) by calculating the mean value obtained from multiple determinations. This is a form of internal validation and characterization of the control material's expected performance.

8. The sample size for the training set

This is not applicable. This is a quality control material, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable.

{0}------------------------------------------------

SECTION 006 : 510(k) Summary

ABX Pentra Micro ALB Control L/H 510(k) Number: K133676

510(k) Summary ABX Pentra Micro ALB Control L/H

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is: K133676

1.0 Submitter

Horiba ABX SAS Parc Euromédecine Rue du Caducée – BP 7290 34184 Montpellier cedex 4 FRANCE + (33) 4 67 14 18 43 Telephone: Fax: + (33) 4 67 14 15 17

Contact Persons

Caroline Ferrer, Regulatory Affairs Manager (caroline.ferrer@horiba.com) Christelle Calvo, Regulatory Affairs Support (christelle.calvo@horiba.com)

Date of Summary Preparation

July 08, 2014

2.0 Device Identification

The following control is for use in conjunction with the ABX PENTRA 400 instrument, cleared to market under K052007.

Trade/Proprietary Name:ABX Pentra Micro ALB Control L/H
Model Number:A11A01967
Common or Usual Name:Micro ALB Control
Device ClassClass I
Classification Panel:Clinical Chemistry
Regulation:21 CFR § 862.1660 Quality control material(assayed and unassayed)
Product Code:JJY; Multi-Analyte Controls, All Kinds (Assayed)

3.0 Device to Which Substantial Equivalence is Claimed

Liquichek Microalbumin Control Bio-Rad Laboratories Irvine, California 510 (k) Number: K072835

· HORIBA ABX SAS, FRANCE

Section 006-1

{1}------------------------------------------------

4.0 Description of Device

The control included in this submission is for use on the ABX PENTRA 400 (K052007) which is discrete photometric benchtop clinical chemistry analyzer. There is no any modifications to the cleared instrument for this particular assay.

The ABX Pentra Micro ALB Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with added chemicals, constituents of human origin, stabilizers and preservatives. The assigned values and precise confidence interval are given in an enclosed annex, ensuring control of the appropriate HORIBA ABX SAS methods on the HORIBA clinical chemistry analyzer. Each control level is provided in one vial of 10 ml.

5.0 Value Assignment

Assigned values were determined by calculating the mean value obtained from multiple determinations. The concentration of the constituent(s) is lot specific. Assigned values and the corresponding ±3SD confidence ranges are indicated in a specific annex. It is recommended that each laboratory establish its own acceptable ranges and use those provided only as guides. Laboratory established ranges may vary from those listed during the life of this control. Variations over time and between laboratories may be caused by differences in laboratory technique, instrumentation and reagents, or by manufacturer test method modifications.

6.0 Intended use

ABX Pentra Micro ALB Control L/H is a quality control used to monitor the performance of ABX Pentra Micro ALBUMIN CP determination by immunoturbidimetry.

7.0 Comparison of the new device with the Predicate Device

The new ABX Pentra Micro ALB Control L/H claims substantial equivalence to the Liquichek Microalbumin Control currently in commercial distribution (K072835).

Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.

{2}------------------------------------------------

Predicate device (K072835):New Device :
Device NameLiquichek MicroalbuminControlABX Pentra Micro ALB Control L/H
Differences
Commercialized byBio-Rad LaboratoriesHORIBA ABX SAS
Indication for Use :Liquichek Microalbumin Controlis intended for use as an assayedquality control material tomonitor the precision oflaboratory testing procedures forthe analytes listed in the packageinsert.ABX Pentra Micro ALB Control L/H is aquality control used to monitor theperformance of ABX Pentra MicroALBUMIN CP determination byimmunoturbidimetry.
AnalytesMicroalbumin and creatinineMicroalbumin
Similarities
Produced byBio-Rad LaboratoriesSame
Shelf Life2-8°C until expiration dateSame
Open stability90 days at 2 to 8°CSame
Format / PackagingIn liquid form in 10mL vials forboth level 1 and level 2Same
MatrixHuman urine with addedconstituents of human origin,chemicals, preservatives andstabilizers.Same

Table 1: Differencies and Similarities between predicate device and new device

8.0 Statement of Supporting Data

Real time stability studies were conducted to establish the shelf life and open vial stability claims. Acceptance criteria were met to support the product claims as follows:

Shelf Life: 24 months at 2 to 8°C. Open vial stability: 90 days at 2 to 8℃.

9.0 Conclusion

The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that the device is substantially equivalent to its predicate device.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 11, 2014

HORIBA ABX SAS C/O CAROLINE FERRER PARC EUROMEDECINE RUE DU CADUCEE MONTPELLIER 34184 FRANCE

Re: K133676

Trade/Device Name: ABX Pentra Micro ALB Control L/H Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I, Reserved Product Code: JJY Dated: June 9, 2014 Received: June 13, 2014

Dear Ms. Caroline Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Ms. Ferrer

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

hup://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours.

Courtney H. Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K133676

Device Name ABX Pentra Micro ALB Control L/I I

Indications for Use (Describe)

ABX Pentra Micro ALB Control L/H is a quality control used to monitor the performance of ABX Pentra Micro ALBUMIN CP determination by immunoturbidimetry.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Ruth A. Chesler -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.