K Number
K133637

Validate with FDA (Live)

Date Cleared
2013-12-18

(41 days)

Product Code
Regulation Number
866.3328
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Alere™ Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results are presumptive and should be confirmed by cell culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions.

Device Description

The Alere™ Influenza A & B Test is an immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect influenza type A and B nucleoprotein antigens in nasal swab specimens. These antibodies and a control protein are immobilized onto a membrane support as three distinct lines and are combined with other reagents/pads to construct a Test Strip.

Nasal swab samples are added to a Coated Reaction Tube to which an extraction reagent has been added. An Alere™ Influenza A & B Test Strip is then placed in the Coated Reaction Tube holding the extracted liquid sample. Test results are interpreted at 10 minutes based on the presence of pink-to-purple colored Sample Lines. The yellow Control Line turns blue in a valid test.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and study details for the Alere™ Influenza A & B Test:

The Alere™ Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens. It's intended to aid in the rapid differential diagnosis of influenza A and B viral infections.


Acceptance Criteria and Device Performance

The provided document details the performance of the Alere™ Influenza A & B Test against viral culture as the ground truth. While explicit "acceptance criteria" are not stated as target thresholds in this document, the reported performance metrics are presented against the comparison method. We can infer the implied acceptance would be that the device demonstrates adequate diagnostic accuracy (sensitivity and specificity) to be useful for its intended purpose.

Here's the performance as reported in the clinical study:

MetricInfluenza Type A (Alere Test vs. Culture)Influenza Type B (Alere Test vs. Culture)
Sensitivity93.8% (95% CI: 83.2, 97.9%)77.0% (95% CI: 67.4, 85.0%)
Specificity95.8% (95% CI: 93.5, 97.3%)98.0% (95% CI: 96.1, 99.0%)
Invalid Rate1.9% (95% CI: 1.0%, 3.5%) - overall1.9% (95% CI: 1.0%, 3.5%) - overall

Study Details

2. Sample Size and Data Provenance

  • Test Set Sample Size: A total of 470 prospective nasal swab specimens were evaluated.
  • Data Provenance:
    • Country of Origin: The study was conducted at seven U.S. trial sites.
    • Retrospective/Prospective: The study was a "multi-center, prospective, clinical study conducted at seven U.S. trial sites during the 2008-2009 respiratory season."

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth (viral culture). Viral culture itself is a laboratory gold standard, and its interpretation would typically be performed by trained laboratory personnel or microbiologists.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for discordant results between the Alere™ Influenza A & B Test and viral culture for the primary clinical study. However, for the 19 samples where the Alere™ test was negative but culture was positive for influenza B, an investigational RT-PCR assay was used. "Ten (10) of these samples were negative for influenza B by PCR," suggesting a form of secondary confirmation for these specific discrepancies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance. This device is a rapid diagnostic test (immunochromatographic assay), not an AI-assisted diagnostic tool for image interpretation or complex data analysis that would typically involve human readers interpreting results.

6. Standalone Performance

Yes, the clinical study directly assesses the standalone performance (algorithm only, without human-in-the-loop performance) of the Alere™ Influenza A & B Test. The reported sensitivity, specificity, and invalid rate are for the device itself against the viral culture.

7. Type of Ground Truth Used

The primary ground truth used for the clinical study was viral culture.
For discordant Influenza B results (Alere negative, culture positive), an investigational RT-PCR assay was used as a secondary confirmation method.

8. Sample Size for the Training Set

The document does not specify a separate "training set" sample size in the context of device development for this immunochromatographic assay. Such devices are typically developed through analytical studies and optimization of reagents, rather than machine learning models that require distinct training sets. The "Analytical Sensitivity" section involved testing various concentrations of influenza viruses, which could be considered part of the development and optimization process, but not a typical "training set" in the AI sense.

9. How Ground Truth for the Training Set Was Established

As there is no explicit "training set" in the context of a machine learning algorithm, the concept of establishing ground truth for a training set does not directly apply here. The analytical studies (Analytical Sensitivity, Analytical Reactivity, Analytical Specificity) used well-characterized viral strains and microorganisms, where their presence and concentration served as the known "ground truth" for evaluating the immunoassay's performance characteristics. This would involve laboratory-established concentrations and identification of these biological agents.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K133637

SUBMITTER

Alere Scarborough, Inc. 10 Southgate Road Scarborough, ME 04074 Establishment Registration Number: 1221359

CONTACT PERSON

Angela Drysdale (207) 730-5737 (Office) (207) 730-5767 (FAX) Angela.drysdale@alere.com (email)

DATE PREPARED

TRADE NAME Alere™ Influenza A & B Test

COMMON NAME Not Applicable

CLASSIFICATION NAME

Influenza virus serological reagents (per 21 CFR 866.3330)

CLASSIFICATION Class I

PRODUCT CODE GNX

PANEL Microbiology

PREDICATE DEVICE Alere™ Influenza A & B Test, K103610

{1}------------------------------------------------

DEVICE DESCRIPTION

The Alere™ Influenza A & B Test is an immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect influenza type A and B nucleoprotein antigens in nasal swab specimens. These antibodies and a control protein are immobilized onto a membrane support as three distinct lines and are combined with other reagents/pads to construct a Test Strip.

Nasal swab samples are added to a Coated Reaction Tube to which an extraction reagent has been added. An Alere™ Influenza A & B Test Strip is then placed in the Coated Reaction Tube holding the extracted liquid sample. Test results are interpreted at 10 minutes based on the presence of pink-to-purple colored Sample Lines. The yellow Control Line turns blue in a valid test.

INTENDED USE

The Alere™ Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results are presumptive and should be confirmed by cell culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions.

COMPARISON TO THE PREDICATE

The Alere™ Influenza A & B Test under consideration in this special 510(k) submission is exactly the same as the currently 510(k) cleared Alere™ Influenza A & B Test in all aspects except for the addition of Influenza A H7N9 to the analytical reactivity table. There have been no other modifications to the test, the fundamental scientific technology of the test has not been altered. Both tests use lateral flow immunochromatographic technology. Both tests are rapid immunoassays that employ specific antibodies immobilized onto solid phases to capture and visualize influenza nucleoprotein antigens.

PERFORMANCE SUMMARY

CLINICAL STUDY

Alere™ Influenza A & B Test Performance vs. Viral Culture – Prospective Study

The clinical performance of the Alere™ Influenza A & B Test was established in a multi-center, prospective, clinical study conducted at seven U.S. trial sites during the 2008-2009 respiratory season.

A total of 470 prospective nasal swab specimens, collected from patients of all ages presenting with influenzalike symptoms, were evaluated in the Alere™ Influenza A & B Test and compared to viral culture. Forty-four percent (44%) of the population tested was < 5 years of age, 31% was 5 - < 18 years of age, and 25% was > 18 years. A/H3 and A/H1 were the predominant influenza A subtypes observed during the times the specimens were collected.

Alere™ Influenza A & B Test performance versus viral culture, including 95% confidence intervals, is detailed below.

{2}------------------------------------------------

Alere™ Influenza A & B Test Performance vs. Culture

Influenza Type A

Culture +Culture -
Alere +451863
Alere -3412415
48430478
Culture +Culture -
Alere +65873
Alere -19*386405
84394478

Sensitivity: 93.8% (45/48) (95% CI: 83.2, 97.9%) Specificity: 95.8% (412/430) (95% CI: 93.5, 97.3%)

Influenza Type B

Sensitivity: 77% (65/84) (95% CI: 67.4, 85.0%) Specificity: 98.0% (386/394) (95% CI: 96.1, 99.0%)

  • The nineteen samples that tested positive on culture for influenza B, but were negative on the Alere™ Influenza A & B Test, were also tested on an investigational RT-PCR assay. Ten (10) of these samples were negative for influenza B by PCR.

The rate of invalid results was 1.9% (9/487) with 95% Cl: 1.0%, 3.5%.

ANALYTICAL STUDIES

ANALYTICAL SENSITIVITY

The Alere™ Influenza A & B Test limit of detection (LOD or C35), defined as the concentration of influenza virus that produces positive Alere™ Influenza A & B Test results approximately 95% of the time, was identified by evaluating different concentrations of 2 subtypes of live influenza A and 2 strains of live influenza B in the Alere™ Influenza A & B Test: Multiple operators tested each concentration of the four influenza strains multiple times. The concentrations identified as the LOD (or Cos) levels for each strain tested are listed below.

Influenza SubtypeConcentration(TCID50/ml)# Detected perTotal Tests% Detected
Influenza A/HongKong/8/68$2.37 \times 10^4$64/6697%
Influenza A/PuertoRico/8/34$3.16 \times 10^5$37/4288%
Influenza B/Malaysia/2506/2004$3.00 \times 10^6$19/2095%
Influenza B/Lee/40$4.20 \times 10^5$19/2095%

ANALYTICAL REACTIVITY

The influenza A and B strains listed tested positive in the Alere™ Influenza A & B Test at the concentrations specified. Although the specific influenza strains causing infection in humans can vary year to year, all contain the conserved nucleoproteins targeted by the Alere™ Influenza A & B Test! Performance characteristics of the Alere™ Influenza A & B Test for detecting influenza A virus from human specimens were established when H1 and H3 subtypes were prevalent. Performance characteristics of the test when other influenza A virus subtypes are emerging as human pathogens have not been established.

Influenza Strain

Flu A/Port Chalmers/1/73 (H3N2) Flu A/WS/33 (H1N1) Flu A/Aichi/2/68 (H3N2) Flu A/Malaya/302/54 (H1N1) Flu A/New Jersey/8/76 (H1N1) Flu A/Denver/1/57 (H1N1) Flu A/Victoria/3/75 (H3N2)

Concentration

5.6 x 105 TCID50/ml 5.0 x 104 TCID50/ml 3.0 x 104 TCID50/ml 6.0 x 105 TCID50/ml 2.8 x 105 TCID50/ml 8.9 x 103 TCID50/ml 1.8 x 104 TCID50/ml

{3}------------------------------------------------

1.5 x 105 TCID50/ml Flu A/Solomon Islands/3/2006 (H1N1) Flu A/Brisbane/10/07 (H3N2) 2.5 x 106 EIU50/ml Flu A/PuertoRico/8/34 (H1N1) 5.6 x 105 TCID50/ml Flu A/Wisconsin/67/2005 (H3N2) 1.3 x 105 TCID50/ml Flu A/Hong Kong/8/68 (H3N2) 7.9 x 103 TCID50/ml Flu A/California/04/2009 (H1N1) 1.4 x 105 TCID50/ml Flu A/ANHUI/1/2013 (H7N9) 8.7 x 106 EID50/ml Flu B/Florida/02/2006 1.4 x 104 TCID50/ml Flu B/Florida/04/2006 7.1 x 104 TCID50/ml Flu B/Florida/07/04 8.5 x 104 TCID50/ml Flu B/Malaysia/2506/04 1.5 x 106 TCID50/ml Flu B/Panama/45/90 1.7 x 104 TCID50/ml Flu B/R75 5.0 x 105 TCID50/ml Flu B/Russia/69 2.2 x 106 TCID50/ml 1.0 x 105 TCID50/ml Flu B/Taiwan/2/62 Flu B/Mass/3/66 1.5 x 105 TCID50/ml Flu B/Lee/40 1.8 x 105 TCID50/ml

The Alere™ Influenza A & B Test was used to test 55 archived respiratory patient specimens, confirmed to be positive for the 2009 H1N1 influenza virus by an FDA cleared RT-PCR assay. Overall, the Alere™ Influenza A & B Test detected 45% (25/55) of the RT-PCR assay positive specimens. The detection rate was 94% (16/17) with the higher titer specimens and 24% (9/38) with the lower titer specimens.

Although this test has been shown to detect the Flu A/California/04/2009 (H1N1) and A/Anhui/1/2013 (H7N9) viruses cultured from positive human specimens, the performance characteristics of this device with fresh (non-frozen) human specimens infected with these two influenza viruses have not been established. The Alere™ Influenza A & B test can distinguish between influenza A and B viruses, but it does not differentiate seasonal influenza A virus from influenza A 2009 H1N1 or influenza A H7N9. The ability to detect human infection with the 2009 H1N1 or H7N9 influenza virus in clinical specimens is unknown.

ANALYTICAL SPECIFICITY (CROSS-REACTIVITY)

To determine the analytical specificity of the Alere™ Influenza A & B Test, 54 commensal and pathogenic microorganisms (38 bacteria, 15 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following microorganisms were negative when tested at concentrations ranging from 108 to 1010 cells/ml, CFU/ml (bacteria), 105 to 108 TCIDso/ml or CEIDso/ml (viruses), and 10° cells/ml (veast).

BacteriaVirusesYeast
Acinetobacter calcoaceticusAdenovirus type 1Candida albicans
Bacteroides fragilisAdenovirus type 7
Bordetella pertussisCoronavirus OC43
Chlamydia pneumoniaeCoronavirus 229E
Corynebacterium diphtheriaCoxsackievirus B4
Enterococcus faecalisCytomegalovirus (CMV) (Herpes V)
Escherichia coliEpstein Barr Virus
Gardnerella vaginalisHuman metapneumovirus
Haemophilus influenzaeMeasles (Edmonston)
Klebsiella pneumoniaeMumps (Enders)
Lactobacillus caseiParainfluenza 1

{4}------------------------------------------------

Lactobacillus plantarum Parainfluenza 2 Legionella pneumophila Parainfluenza 3 Respiratory Syncytial Virus type B Listeria monocytogenes Moraxella catarrhalis Rhinovirus type 1A Mycobacterium avium Mycobacterium intracellulare Mycobacterium tuberculosis Mycoplasma pneumoniae Neisseria gonorrhoeae Neisseria meningitidis Neisseria sicca Neisseria subflava Proteus vulgaris Pseudomonas aeruginosa Serratia marcescens Staphylococcus aureus Staphylococcus aureus (Cowan protein A producing strain) Staphylococcus epidermidis Streptococcus, Group A Streptococcus, Group B Streptococcus, Group C Streptococcus, Group F Streptococcus, Group G Streptococcus mutans Streptococcus pneumoniae Streptococcus salivaris Streptococcus sanguis

INTERFERING SUBSTANCES

The following substances, naturally present in respiratory specimens or that may be artificially introduced into the nasal cavity or nasopharynx, were evaluated in the Alere™ Influenza A & B Test at the concentrations listed below and were found not to affect test performance. Whole blood (1%) did not interfere with the interpretation of negative Alere™ Influenza A & B Test results, but did interfere with the interpretation of influenza A LOD (or C53) positive samples. Therefore, visibly bloody samples may not be appropriate for use in this test.

SubstanceConcentration
3 OTC nasal sprays10%.
3 OTC mouthwashes10%
3 OTC throat drops10%
4-acetamidophenol10 mg/ml
Acetylsalicylic acid20 mg/ml
Albuterol20 mg/ml
Chlorpheniramine5 mg/ml
Dexamethasone5 mg/ml
Dextromethorphan10 mg/ml
Diphenhydramine5 mg/ml
Doxylamine succinate1 mg/ml

{5}------------------------------------------------

Flunisolide3 mg/ml
Guaiacol glycerol ether20 mg/ml
Mucin1%
Mupirocin250 µg/ml
Oxymetazoline10 mg/ml
Phenylephrine10 mg/ml
Phenylpropanolamine20 mg/ml
Rebetol® (Ribavirin)500 ng/ml
Relenza® (Zanamivir)20 mg/ml
Rimantadine500 ng/ml
Tamiflu® (Oseltamivir)100 mg/ml
Tobramycin40 mg/ml
Triamcinolone14 mg/ml

REPRODUCIBILITY

A reproducibility study of the Alere™ Influenza A & B Test was conducted by operators from 3 sites using panels of blind coded randomized specimens containing negative (below the limit of detection), low positive (at the limit of detection), and moderate positive (above the limit of detection) influenza A and B viral samples: Participants tested each sample multiple times on 5 different days. The detection rates for the influenza A moderate positive, low positive, and high negative samples were 99.2% (119/120), 94.2% (113/120) and 9.2% (11/120), respectfully. The detection rates for the influenza B moderate positive, low positive, and high negative samples were 99.2% (119/120), 96.7%(116/120) and 7.5% (9/120), respectfully. All of the negative samples (118) generated negative test results.

Signed __

Date _

Angela Drysdale

VP Regulatory and Clinical Affairs – Infectious Disease

Alere Scarborough, Inc.

  1. Dowdle, W.R. Kendal, A.P., and Noble, G.R. 1980. Influenza Virus, p 836-884. Manual of Clinical Microbiology, 3rd edition, In Lennette, et. Al (ed.). American Society for Microbiology, Washington, D.C

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image contains the words "Public Health Service" in a bold, sans-serif font. The text is arranged vertically, with "Public Health" on the first line and "Service" on the second line. The words are black against a white background.

Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with outstretched arms, representing care and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 18, 2013

ALERE SCARBOROUGH, INC. ANGELA DRYSDALE VP OF REGULATORY AND CLINICAL AFFAIRS - INFECTIOUS DISEASE 10 SOUTHGATE ROAD SCARBOROUGH ME 04074

Re: K133637

Trade/Device Name: Alere™ Influenza A & B Test Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza virus serological reagents Regulatory Class: I Product Code: GNX Dated: November 4, 2013 Received: November 27, 2013

Dear Ms. Drysdale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2-Ms. Drysdale

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sally A

Sally Hojvat, M.Sc., Ph.D Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications For Use

510(k) Number (if known): K133637

Device Name: Alere™ Influenza A & B Test

Intended Use: The Alere™ Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasal swab specimens collected from symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results are presumptive and should be confirmed by cell culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Tamara V. Feldblyum -S 2013.12.16 15:32:37 -05'00'

§ 866.3328 Influenza virus antigen detection test system.

(a)
Identification. An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.