(62 days)
The MAC™ 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analyze, display, and record information for adult and pediatric populations in a hospital, medical professional's facility, clinics, physician`s office or outreach centers.
NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years.
The MACTM 2000 ECG analysis system provides the following modes of operation:
- Resting ECG mode .
- . Arrhythmia mode
- Exercise mode for exercise stress testing (optional) .
- RR analysis mode for RR interval analysis (optional) .
The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis. Arrhythmia detection is provided for the convenience of automatic documentation.
Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MACTM 2000 ECG analysis system can print and display multiple leads of ECG data. The device is capable of acquiring 2 additional ECG leads beyond what is needed for classical 12 lead ECG acquisition.
The MACTM 2000 ECG analysis system provides ECG measurement in both resting and stress mode. The device also supports arrhythmia as a mode of operation.
The Hook up advisor feature in MAC™ 2000 ECG analysis system advises the users of poor lead quality based on noise measurement.
The MACT14 2000 ECG Analysis System delivers multiple leads of ECG on full-size reports and includes an alphanumeric keyboard for patient demographics and other data entry, an integrated 7' color display, and an integrated thermal writer. The thermal writer will print real time continuous waveform, alphanumeric data and non real time reports.
The device has an optional feature of transmission and reception of ECG data, downloading of orders and patient demographics to and from a central ECG cardiovascular information system and supports exporting of ECG records in PDF. It also has an optional internal memory and removable storage to store resting ECG records.
The patient information can be entered with the help of optional bar code reader.
Clinical Trials Data Guard and audit trail options are also available to support electronic record requirements.
MAC™ 2000 ECG Analysis system can be used as a portable unit also.
Here's a breakdown of the acceptance criteria and study information for the MACTM 2000 ECG Analysis System, based on the provided text:
Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly state specific quantitative acceptance criteria for the MACTM 2000 ECG Analysis System. Instead, it emphasizes substantial equivalence to predicate devices and compliance with voluntary standards.
However, based on the statements, the implicit acceptance criteria are related to:
- Performance: Meeting all design requirements and performance claims.
- Functionality: Acquiring, analyzing, displaying, and recording ECG data in various modes (resting, arrhythmia, exercise, RR analysis).
- Usability: Providing features like alphanumeric keyboard, integrated display, thermal writer, hook-up advisor, and options for data transmission/reception.
- Safety: Compliance with relevant safety standards.
- Reliability: Demonstrating reliable operation.
- Substantial Equivalence: Being as safe and effective as the predicate devices.
Since no specific quantitative metrics are provided for the acceptance criteria, a table directly comparing them to reported device performance cannot be created with numerical values. The text only states that the device "meets all design requirements and performance claims."
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Implicit) | Reported Device Performance Statement |
|---|---|
| Performance: Device effectively acquires, analyzes, displays, and records ECG data. | "The results demonstrate that MAC™ 2000 ECG analysis system meets all design requirements and performance claims." |
| Functionality: Supports various ECG modes (resting, arrhythmia, exercise, RR analysis). | "...provides the following modes of operation: Resting ECG mode, Arrhythmia mode, Exercise mode for exercise stress testing (optional), RR analysis mode for RR interval analysis (optional)." |
| Usability: Provides effective interface for data entry, display, and printing. | "delivers multiple leads of ECG on full-size reports and includes an alphanumeric keyboard for patient demographics and other data entry, an integrated 7' color display, and an integrated thermal writer." |
| Safety: Complies with relevant safety standards. | "MACTM 2000 ECG analysis system comply with voluntary standards as detailed in Section 09, 15, 16, 17 and 18 of this premarket submission." |
| Reliability: Device operates reliably. | "Verification and Testing activities establish the performance, functionality, usability, safety, and reliability characteristics of MAC™ 2000 ECG analysis system." |
| Substantial Equivalence: As safe and effective as predicate devices. | "GE Healthcare considers the MAC™ 2000 ECG analysis system to be as safe and as effective as the predicate device, and the performance to be substantially equivalent to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a patient sample size for any clinical testing. It mentions "Testing on unit level, system level, as well as usability and safety parameters."
- Data Provenance: The document does not mention the country of origin of any data or whether it was retrospective or prospective. The study detailed appears to be a simulated use setting for verification testing rather than a clinical trial on patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- The document does not mention the use of experts to establish ground truth for a test set related to interpretation or diagnosis. The testing described focuses on engineering verification, not clinical validation based on expert review.
4. Adjudication Method for the Test Set
- Since there's no mention of expert review or clinical ground truth establishment, no adjudication method (like 2+1 or 3+1) is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, MAC™ 2000 ECG analysis system, did not require clinical studies to support substantial equivalence." This means there was no study comparing human readers with and without AI assistance to measure an effect size.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- The document refers to the "MACTM 2000 ECG Analysis System" as a device that "aquires, analyzes, displays, and records information." The analysis component is embedded in the device. The "Summary of Non-Clinical Tests" and "Summary of Simulated Use Setting" focus on "Design verification... several protocols that include electrical, mechanical, safety testing, reliability, and system design verification." This suggests testing of the integrated system and its analytical capabilities in a controlled engineering environment. However, it does not detail a standalone algorithm performance study in terms of diagnostic accuracy or specific metrics for its "interpretive analysis" option. The focus is on the device's overall functionality and safety, and its equivalence to predicate devices, rather than the isolated performance of its analytical algorithms against a clinical ground truth.
7. Type of Ground Truth Used
- For the described verification testing, the "ground truth" was effectively design requirements and established performance claims, as well as compliance with voluntary standards. There is no mention of clinical ground truth such as pathology, expert consensus, or outcomes data used for the "analysis system" part of the device. The verification focused on whether the device functions as designed and meets engineering specifications and regulatory standards.
8. Sample Size for the Training Set
- The document does not mention any training set size because it did not involve the development or validation of a new AI model with a training phase detailed in this submission. The device uses "the same fundamental technology as its predicate device MAC1600 ECG Analysis System" for acquiring, analyzing, recording, displaying, and printing ECG data. This suggests reliance on established algorithms, rather than newly trained ones requiring a dedicated training set as described in modern AI/ML submissions.
9. How the Ground Truth for the Training Set Was Established
- As no training set is mentioned in the context of this submission, the document does not provide information on how ground truth was established for a training set.
{0}------------------------------------------------
ﺮ
્રિ
Premarket Notification 510(k) Summary As required by section 807.92
MACTM 2000 ECG Analysis System
GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
| Date: | [4th Nov 2013] |
|---|---|
| Submitter: | Wipro GE Healthcare Private Ltd.4, kadugodi industrial areabangalore, INDIA 560067 |
| Primary Contact Person: | Sanjukta DebRoyGE Healthcare.Phone : +91 9632211022Fax :+9180 28411645 |
| Secondary Contact Person: | Douglas KentzGE HealthcarePhone : 1-414 581-8987Fax : 1-414 362-2585 |
| Device:Trade Name: | MACTM 2000 ECG Analysis System |
| Common/Usual Name: | Electrocardiograph |
| Classification Names: | Electrocardiograph,Programmable Diagnostic Computer,Transmitter receiver Electrocardiograph, Telephone |
| Product Code: | DPS,DQK,DXH |
| Regulation No: | 21 CFR 870.2340,21 CFR 870.1425,21 CFR 870.2920 |
| Predicate Device(s): | MAC 1600 ECG Analysis SystemMAC 5500 ECG Analysis SystemMAC 800 ECG Analysis System |
| Regulatory Class | II |
| Prescription Status | Prescription Device |
| Device Classification Panel | Cardiovascular |
| Device Description: | The MACTM 2000 ECG analysis system can print and display multipleleads of ECG data. The device is capable of acquiring 2 additional ECGleads beyond what is needed for classical 12 lead ECG acquisition. |
{1}------------------------------------------------
page 2 of 4 K133622
The MACTM 2000 ECG analysis system provides ECG measurement in both resting and stress mode. The device also supports arrhythmia as a mode of operation.
The Hook up advisor feature in MAC™ 2000 ECG analysis system advises the users of poor lead quality based on noise measurement.
The MACT14 2000 ECG Analysis System delivers multiple leads of ECG on full-size reports and includes an alphanumeric keyboard for patient demographics and other data entry, an integrated 7' color display, and an integrated thermal writer. The thermal writer will print real time continuous waveform, alphanumeric data and non real time reports.
The device has an optional feature of transmission and reception of ECG data, downloading of orders and patient demographics to and from a central ECG cardiovascular information system and supports exporting of ECG records in PDF. It also has an optional internal memory and removable storage to store resting ECG records.
The patient information can be entered with the help of optional bar code reader.
Clinical Trials Data Guard and audit trail options are also available to support electronic record requirements.
MAC™ 2000 ECG Analysis system can be used as a portable unit also.
Indication for Use:
The MACTM 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analvze, display, and record information for adult and pediations in a hospital, medical professional's facility, clinics, physician`s office or outreach centers.
NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years.
The MACTM 2000 ECG analysis system provides the following modes of operation:
- Resting ECG mode .
- . Arrhythmia mode
- Exercise mode for exercise stress testing (optional) .
- RR analysis mode for RR interval analysis (optional) .
The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis. Arrhythmia detection is provided for the convenience of automatic documentation.
Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
{2}------------------------------------------------
MAC™ 2000 ECG Analysis System uses the same fundamental technology as Technology: its predicate device MAC1600 ECG Analysis System in acquiring, analyzing, recording, displaying and printing ECG data for both adult and pediatric populations.
The basic system prints 6 or 12 leads of ECG and provides optional transmission and reception of ECG data to and from a central ECG cardiovascular information system. The system can be upgraded with software options such as stress testing, interpretive analysis, storage and communication options which is similar to predicate device.
The intended use of the predicate device differs from the proposed device on verbiage as it has been modified to increase clarity and also on the offering of RR analysis mode of operation which is not available for sale in United States.
MACTM 2000 ECG Analysis System is similar to MAC 5500 Resting ECG Analysis Systems K073625 in the technology of downloading orders and patient demographics from a central ECG cardiovascular information system.
MACTM2000 ECG Analysis System makes use of the similar technology of exporting ECG reports in PDF and WiFi communication from MAC800 Resting ECG Analysis System K090212.
For more on the predicate device comparison refer to section 12 of this 510K submission.
Performance Standards
The MACTM 2000 ECG Analysis System complies with the voluntary standards as detailed in Section 9 of this submission.
Sterilization
Not Applicable
ي
{3}------------------------------------------------
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
Verification and Testing activities establish the performance, functionality. usability, safety, and reliability characteristics of MAC™ 2000 ECG analysis system.
MACTM 2000 ECG analysis system comply with voluntary standards as detailed in Section 09, 15, 16, 17 and 18 of this premarket submission.
The following quality assurance measures were applied to the development of the system:
- Risk Analysis ◆
- Requirements Reviews .
- . Design Reviews
Summary of Simulated Use Setting:
The Design verification of MAC"131 2000 ECG analysis system has been divided into several protocols that include electrical, mechanical, safety testing, reliability, and system design verification.
The performance testing included testing on unit level, system level, as well as usability and safety parameters.
The results of the Design verification testing protocols have been documented in Section 18 of this 510(k) application.
The results demonstrate that MAC"N 2000 ECG analysis system meets all design requirements and performance claims.
The subject of this premarket submission, MAC™ 2000 ECG analysis system , did not require clinical studies to support substantial equivalence.
Conclusion:
GE Healthcare considers the MAC"M 2000 ECG analysis system to be as safe and as effective as the predicate device, and the performance to be substantially equivalent to the predicate device.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Food and Drug Administration
January 23, 2014
Wipro GE Healthcare Douglas Kentz 8200 West Tower Ave Milwaukee, WI 53223 US
Re: K133622
Trade/Device Name: MAC 2000 ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK, DXH Dated: November 20, 2013 Received: November 22, 2013
Dear Douglas Kentz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Douglas Kentz
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen P., Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K133622
.
Indications for Use
510(k) Number (if known):
Device Name: MAC™ 2000 ECG analysis system
Indications for Use:
The MAC™ 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analyze, display, and record information for adult and pediatric populations in a hospital, medical professional's facility, clinics, physician's office or outreach centers.
NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years.
The MACTM 2000 ECG analysis system provides the following modes of operation:
- . Resting ECG mode
- . Arrhythmia mode
- . Exercise mode for exercise stress testing (optional)
- . RR analysis mode for RR interval analysis (optional)
The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis.
Arrhythmia detection is provided for the convenience of automatic documentation. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
Prescription Use_ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -S
Date: 2014.01.23
15:35:48-05'00'
Vol. 1 Section 4 Page 2 of 2
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).