K Number
K133516

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-04-30

(166 days)

Product Code
Regulation Number
868.1400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.

Device Description

Flexicare's CO2 monitoring line is single patient use, small diameter tubing intended to be connected to a luer lock connector on a patient interface to allow for gas sampling from a patients exhaled breath by gas sampling equipment. The Gas monitoring/analysing device will feature a pump, which draws expired breath from the patient through the monitoring line and into the gas analyser. Flexicare's CO2 monitoring line is available with or without a 0.45um hydrophobic filter, which prevents the transfer of water vapour down the monitoring line and into the gas sampling equipment. Flexicare's CO2 monitoring lines feature clear construction to aid visual inspection of the device. The CO2 monitoring line is a single patient use device, designed for use with one patient over a single course of treatment. The CO2 monitoring line is supplied in a sealed poly bag in a non sterile state and is not to be sterilized. The Flexicare CO2 monitoring lines feature 2M & 3M length tubes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Flexicare CO2 monitoring line, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

TestStandard / Pre-determined Acceptance CriteriaReported Flexicare Device Performance
Visual inspectionPre-determined Acceptance CriteriaAll samples pass
Resistance to flowPre-determined Acceptance CriteriaAll samples pass
Dimensional inspection. Length, ID, ODPre-determined Acceptance CriteriaAll samples pass
LeakagePre-determined Acceptance CriteriaAll samples pass
Performance - CO2 monitoringPre-determined Acceptance CriteriaAll samples pass
Tensile testingPre-determined Acceptance CriteriaAll samples pass
Luer connectors - Dimensional, leakage, connection, stress crackingISO 594-1 / BS EN 20594-1:1994, ISO 594-2 (BS EN 1707:1997)All samples pass
Accelerated age testingASTM F1980All samples pass
Cytotoxicity, Irritation, sensitizationBS EN ISO 10993-10:2010, BS EN ISO 10993-5:2009All samples pass
GenotoxicityBS EN ISO 10993-3:2009All samples pass

Specific Resistance to Flow Performance Details:

  • CO2 monitoring line (without filter):
    • Flexicare: 8.8 mbar at 100ml/min flow, 30.2 mbar at 300ml/min flow
    • Predicate: 15.2 mbar at 100ml/min flow, 51.2 mbar at 300ml/min flow
  • CO2 monitoring line (with filter):
    • Flexicare: 30.0 mbar at 100ml/min flow, 95.7 mbar at 300ml/min flow
    • Predicate: 33.4 mbar at 100ml/min flow, 114.5 mbar at 300ml/min flow

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific number of samples used for each test (the sample size). However, it consistently states "All samples pass" for each test.

The data provenance is from Flexicare Medical Limited, United Kingdom, as indicated by the manufacturer's details and the report being dated in the UK. The tests described are part of a premarket notification (510(k)) submission, meaning they were conducted specifically for this regulatory purpose. The nature of these tests (performance, dimensional, biocompatibility etc.) implies they are prospective in the sense that they were designed and executed to evaluate the new device against established criteria and a predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests performed are primarily engineering and laboratory-based assessments of physical and functional characteristics against defined standards and criteria, rather than expert-driven clinical evaluations requiring ground truth establishment by medical experts.

4. Adjudication Method for the Test Set

This information is not applicable/not provided. The tests are objective measurements against standards or functional comparisons, not subjective clinical assessments requiring adjudication. The results are reported as "All samples pass," indicating direct conformance to acceptance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The device is a CO2 monitoring line, which is a passive tubing device. The comparison studies focused on physical and functional equivalence to a predicate device, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. The device does not involve any software or electronic components, therefore no algorithm is present to be evaluated.

7. The Type of Ground Truth Used

The "ground truth" for the device's performance is established by:

  • Pre-determined Acceptance Criteria: These are internal specifications or performance targets set by Flexicare.
  • International Standards: e.g., ISO 594-1, ISO 594-2, ASTM F1980, BS EN ISO 10993-3, -5, -10.
  • Comparison to a Predicate Device: The performance of the Flexicare device (e.g., resistance to flow, CO2 detection capability, leakage, tensile strength) was directly compared to the Intersurgical predicate device (K122075) to demonstrate similar or better performance.

For the CO2 monitoring performance, the ground truth was whether the device could consistently detect CO2 from beneath the nostrils, similar to the predicate device. The document states: "Both manufacturers' devices detected CO2 consistently, proving substantial equivalence."

8. The Sample Size for the Training Set

This information is not applicable/not provided. The device is a simple mechanical tubing product and does not involve machine learning or AI, thus there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "Flexicare". The logo is in black and white and features a stylized "F" with two curved lines above it. The word "Flexicare" is written in a bold, sans-serif font to the right of the "F". Below the word "Flexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.

510(k) Summary

Trade name

C.F.R Section: Device tvpe: Product Code:

Manufacturer:

Regulatory Affairs Contact:

Christopher Watkins Quality Requlatory & Technical Director Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom

CO2 monitoring line & CO2 monitoring line with filter

Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Telephone:

Date Summary Prepared:

Common Name:

Classification:

Predicate Device

Description:

00 44 1443 471 593

22nd April 2014

CO2 monitoring line

Class 2

K133516

сск

21 CFR 868.1400.

Flexicare Medical Limited

Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom

Intersurgical - CO2 Monitoring line and CO2 monitoring line with filter, which was cleared for marketing by 510(k) No K122075

Flexicare's CO2 monitoring line is single patient use, small diameter tubing intended to be connected to a luer lock connector on a patient interface to allow for gas sampling from a patients exhaled breath by gas sampling equipment. The Gas monitoring/analysing device will feature a pump, which draws expired breath from the patient through the monitoring line and into the gas analyser.

Flexicare's CO2 monitoring line is available with or without a 0.45um hydrophobic filter, which prevents the transfer of water vapour down the monitoring line and into the gas sampling equipment.

Flexicare's CO2 monitoring lines feature clear construction to aid visual inspection of the device.

The CO2 monitoring line is a single patient use device, designed for use with one patient over a single course of treatment. The CO2 monitoring line is supplied in a sealed poly bag in a non sterile state and is not to be sterilized.

The Flexicare CO2 monitoring lines feature 2M & 3M length tubes.

APR 3 0 2014

1

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.

Part Numbers:

Part NumberProduct description
010-700UCO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 2M
010-705UCO2 MONITORING LINE WITH FEMALE/FEMALE CONNECTORS - 2M
010-710UCO2 MONITORING LINE WITH MALE/FEMALE CONNECTORS - 2M
010-700XUCO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 3M
010-700AUCO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 2M - WITH FILTER
010-705AUCO2 MONITORING LINE WITH FEMALE/FEMALE CONNECTORS - 2M - WITH FILTER
010-710AUCO2 MONITORING LINE WITH MALE/FEMALE CONNECTORS - 2M - WITH FILTER
010-700XAUCO2 MONITORING LINE WITH MALE/MALE CONNECTORS - 3M - WITH FILTER

Sizing

Part NumberLength (mm)Tube OD (mm)Tube ID (mm)
010-700U20002.71.2
010-705U20002.71.2
010-710U20002.71.2
010-700XU30002.71.2
010-700AU20002.71.2
010-705AU20002.71.2
010-710AU20002.71.2
010-700XAU30002.71.2

Intended Use:

CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.

Substantial Equivalence

Flexicare's CO2 monitoring line & CO2 monitoring line with filter have the same intended use as predicate device:

Intersurgical - CO2 Monitoring line and CO2 monitoring line with filter, which was cleared for marketing by 510(k) No K122075

Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are single patient use devices.

Neither Flexicare's CO2 monitoring lines or Intersurgical CO2 monitoring lines are life supporting or life sustaining.

Neither Flexicare's CO2 monitoring lines or Intersurgical CO2 monitoring lines use software, and are not electronically driven.

Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are supplied non- sterile in individual poly bags.

Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are designed for the same intended use in the same intended conditions.

Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines consist of 3-4 components made from injection molded & extruded plastics.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for "Flexicare". The logo is black and white and features a stylized "F" with two curved lines. The word "Flexicare" is written in a bold, sans-serif font to the right of the "F". Below the word "Flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.

Both Flexicare's CO2 monitoring lines and Intersurgical CO2 monitoring lines are supplied either with or without 0.45µm hydrophobic filters. Flexicare's hydrophobic filter appeared to be identical to the Intersurgical filters in appearance and product marking. It is likely they are sourced from the same supplier.

From the performance testing it was discovered both Flexicare's and the predicate device tubes are manufactured from flexible PVC, and that both devices feature male and female luers with the option of a hydrophobic filter.

The technological characteristics of Flexicare's CO2 monitoring line and Intersurgical's CO2 monitoring line are very similar. These similarities are evident in the comparison table below.

Flexicare - CO2 monitoring lineK133516Intersurgical - CO2 Monitoring LineK122075
Intended use
TargetPopulationAny patient from which gas monitoring isrequiredAny patient from which gas monitoring is required
Indications for,useCO2 Monitoring lines are intended toconnect from a patient interface CO2sampling port to the expired gas monitor.The CO2 monitoring lines are intended to connectfrom a CO2 sampling port to the expired gasmonitor
Place of useHospitalsHospitals
Product labelingCO2 monitoring lineCO2 monitoring line
Single use or re-usable?Single useSingle use
Design andperformance
Resistance toflow8.8 mbar at 100ml/min flow30.2 mbar at 300ml/min flow15.2 mbar at 100ml/min flow51.2 mbar at 300ml/min flow
Leakage<1.0ml/min<1.0ml/min
Connectors2x luer lock connectors - male or female2x luer lock connectors- male or female
Outer diameter2.7mm3.05mm
Inner diameter1.2mm1.47mm
Lengths available2.0m, 3.0m0.7m, 1.8m, 2.45m, 3.0m
StandardsISO 594-2 (BS EN 1707:1997) , ISO 10993-3&5ISO 594-2 (BS EN 1707:1997) , ISO 10993-3&5
ProcessesExtrusion, injection moulding, adhesivebondingExtrusion, injection moulding, adhesive bonding
ColoursNone - colourlessNone - colourless
EnergyUsed/delivered:Gas is pulled away from the patient by apump in the gas sampling deviceGas is pulled from one end of the tube to the otherby a pump in the gas sampling device
Compatability :Designed for use with gas monitoringdevice (for example a Capnograph) withluer connections to gas sampling tubingDesigned for use with gas monitoring device (forexample a Capnograph) with luer connections togas sampling tubing
Materials :PVC (main line) and ABS (luer connectors)PVC (main line) and Polycarbonate (luerconnectors)
Biocompatibility:Compliant with ISO 10993-3&5Compliant with 10993-3&5
Sterility:Non sterileNon sterile
StorageStore in a cool, dry place out of direct sunlightStore in a cool, dry place out of direct sunlight
ComponentinteractionsTube – intact skin contactConnector - intact skin contact ( whilst connecting to analyzing machineTube - intact skin contactConnector - intact skin contact ( whilst connecting to analyzing machine
Intended useAny patient from which gas monitoring is requiredAny patient from which gas monitoring is required
TargetPopulationAny patient from which gas monitoring is requiredAny patient from which gas monitoring is required
Indications foruseCO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.The CO2 monitoring lines are intended to connect from a CO2 sampling port to the expired gas monitor
Place of useHospitalsHospitals
Product labelingCO2 monitoring line + Hydrophobic filterCO2 monitoring line + Hydrophobic filter
Single use or re-usable?Single useSingle use
Design andperformance
Resistance toflow30.0 mbar at 100ml/min flow95.7 mbar at 300ml/min flow33.4 mbar at 100ml/min flow114.5 mbar at 300ml/min flow
Leakage<1.0ml/min<1.0ml/min
Connectors2 x luer lock connectors - male or female (plus luer lock male to luer lock female connection between monitoring line and filter)2 x luer lock connectors - male or female (plus luer lock male to luer lock female connection between monitoring line and filter)
Outer diameter2.7mm3.05mm
Inner diameter1.2mm1.47mm
Length2.0m, 3.0m0.7m, 1.8m, 2.45m, 3:0m
StandardsISO 594-2 (BS EN 1707:1997), ISO 10993-3&5ISO 594-2 (BS EN 1707:1997), ISO 10993-3&5
ProcessesExtrusion, injection moulding, adhesive bondingExtrusion, injection moulding, adhesive bonding
ColoursNone - colourlessNone - colourless
EnergyUsed/delivered:Gas is pulled away from the patient by a pump in the gas sampling deviceGas is pulled from one end of the tube to the other by a pump in the gas sampling device
Compatability :Designed for use with gas monitoring device (for example a Capnograph) with luer connections to gas sampling tubingDesigned for use with gas monitoring device (for example a Capnograph) with luer connections to gas sampling tubing
Materials :PVC (main line) and ABS (luer connectors)PVC (main line) and Polycarbonate (luer connectors)
Biocompatibility:Compliant with ISO 10993-3&5Compliant with ISO 10993-3&5
Sterility:Non sterileNon sterile
StorageStore in a cool, dry place out of direct sunlightStore in a cool, dry place out of direct sunlight
ComponentinteractionsTube – intact skin contactConnector - intact skin contact (whilst connecting to analyzing machineTube - intact skin contactConnector - intact skin contact (whilst connecting to analyzing machine
Filter - intact skin contact (whilstconnecting to analyzing machineFilter - intact skin contact (whilst connecting toanalyzing machine

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "Flexicare". The logo has a stylized "F" on the left side, followed by the word "lexicare" in a sans-serif font. Below the word "lexicare" is the text "TOTAL QUALITY - TOTAL CARE" in a smaller font.

:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Flexicare. The logo consists of a stylized letter "f" followed by the word "lexicare" in a serif font. Below the word "lexicare" is the phrase "TOTAL QUALITY : TOTAL CARE" in a smaller, sans-serif font.

From the table above it is evident that the CO2 monitoring line devices from both manufacturers have very similar technological characteristics. Both devices feature the same number of components, have been made using the same manufacturing methods, have been manufactured from the same (or very similar) materials and critical dimensions (tube ID, Luer dimensions) are identical. Any differences between the Flexicare and Intersurgical devices are stated below with justifications.

Differences

Both devices are simple products, comprising of 3 components each (4 each if with filter). There are very little differences between the two devices. Below are the only differences discovered.

The luer connectors of the Intersurgical predicate device are Polycarbonate, while Flexicare's luers are ABS. This difference in material is not crucial as both materials are amorphous polymers and have similar properties including high clarity, hardness, high impact strength and very low mould shrinkage.

The only other difference between the devices is that the Intersurgical sample has a frosted finish tube, which is purely a surface finish difference.

TestStandard / Pre-determinedAcceptance CriteriaOutcome
Visual inspectionPre-determined Acceptance CriteriaAll samples pass
Resistance to flowPre-determined Acceptance CriteriaAll samples pass
Dimensional inspection. Length, ID, ODPre-determined Acceptance CriteriaAll samples pass
LeakagePre-determined Acceptance CriteriaAll samples pass
Performance - CO2 monitoringPre-determined Acceptance CriteriaAll samples pass
Tensile testingPre-determined Acceptance CriteriaAll samples pass
Luer connectors - Dimensional,leakage, connection, stress crackingISO 594-1 / BS EN 20594-1:1994ISO 594-2 (superseded by BS EN1707:1997)All samples pass
Accelerated age testingASTM F1980All samples pass
Cytotoxicity, Irritation, sensitizationBS EN ISO 10993-10:2010BS EN ISO 10993-5:2009All samples pass
GenotoxicityBS EN ISO 10993-3:2009All samples pass

Non Clinical Tests Performed:

All the tests that were carried out on Flexicare's CO2 monitoring line are listed in the table above.

The tests selected to be carried out in order to determine substantial equivalence would test the critical aspects of the each device's functionality.

The visual & dimensional inspections were carried out to ensure that the Flexicare device had the same number of components, the same general assembly, the same material type and was dimensionally the same as (or very similar to) the predicate device. Both manufacturers' devices were found to be very similar in appearance, assembly, materials and dimensions with critical dimensions such as tube ID being identical.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller font.

6

Resistance to flow testing was carried out to compare the level of resistance between Flexicare's CO2 monitoring line with Intersurgical's at the same flow rates. This aspect is critical as a higher resistance would restrict CO2 reaching the gas analyzer. Testing was carried out at 100ml/min and again at 300ml/min.

CO2 monitoring testing was carried out in order to assess whether the devices would detect CO2 from beneath the nostrils as this is the main purpose of the device. Samples were not assessed for % of CO2 detected as this depends on the patient-end device that the CO2 monitoring line is attached to. Both manufacturers' devices detected CO2 consistently, proving substantial equivalence.

Leak and tensile testing were also carried out as a leak or disconnection from sampling line to luer could result in less or no CO2 being drawn to the gas sampling device. Both Flexicare's and Intersurgical's devices passed the leak testing and withstood the pre-determined acceptance criteria of 40N tensile force without disconnection. This also proves substantial equivalence of the devices.

Conclusion

Flexicare's & Intersurgical's CO2 monitoring line & CO2 monitoring line with filter passed the performance testing when tested against methods and criteria from both pre-determined acceptance criteria test methods and relevant standards. The results of this testing show that the CO2 monitoring line & CO2 monitoring line with filter passes all performance tests that are critical to its substantially equivalent operation. In all tests the Flexicare device performed as well as or better than the Predicate device.

Flexicare's CO2 monitoring line & CO2 monitoring line with filter is substantially equivalent to, and performs as well as the marketed predicate device manufactured by Intersurgical.

Signature of contact Person:

Chris Watkins Quality Regulatory & Technical Director

euashano

Dated: 22-04-2014

-------End of 510k Summary----------------------------------------------------------------------------------------------------------------------------------------------------

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 30, 2014

Flexicare Medical Limited Christopher Watkins Ouality Regulatory and Technical Director Cynon Vallev Business Park Mountain Ash, Mid. Glamorgan CF45 4ER. United Kingdom

Re: K133516

Trade Name: CO2 Monitoring Line & CO2 Monitoring Line with Filter Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: March 26. 2014 Received: March 31, 2014

Dear Mr. Watkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mr. Watkins

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ticket Park Shah M.D.

Tejashri Purohit Sheth, MiD.

Clinical Deputy Director,

DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Flexicare. The logo consists of a stylized letter "f" with two curved lines above it, followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the tagline "TOTAL QUALITY - TOTAL CARE" in a smaller, sans-serif font.

Indications for Use

510(k) Number (if known): _K133516

Page 1 of 1

CO2 monitoring line & CO2 monitoring line with filter Device Name:

Indications for Use:

CO2 Monitoring lines are intended to connect from a patient interface CO2 sampling port to the expired gas monitor.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Todd D. Cour

§ 868.1400 Carbon dioxide gas analyzer.

(a)
Identification. A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.(b)
Classification. Class II (performance standards).