K Number
K133289

Validate with FDA (Live)

Date Cleared
2013-12-09

(45 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

Device Description

The Ellipse PRECICE System is composed of the PRECICE nail (supplied sterile), locking screws, surgical instruments and an external remote controller (ERC). The nail is available in tibia or femur models with various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE nail is supplied sterile by gamma radiation while the locking screws and PRECICE specific accessories are supplied non-sterile and must be sterilized prior to use. The nail contains an enclosed rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing.

AI/ML Overview

The Ellipse PRECICE® System, specifically the 8.5 mm PRECICE Nail and its associated 3.5 mm diameter locking screw, obtained 510(k) clearance (K133289) based on its substantial equivalence to a previously cleared device, the Ellipse Intramedullary Limb Lengthening System (K131677). This clearance was supported by non-clinical testing.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test DescriptionApplicable Test StandardAcceptance Criteria (Implied by Equivalence)Reported Device Performance
PRECICE Nail, Static Four Point BendASTM F1264-03Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Nail, Dynamic Four Point BendASTM F1264-03Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Nail, Static Torque to FailureASTM F1264-03Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Locking Screw, Static Three Point BendASTM F1264-03Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Locking Screw, Dynamic Three Point BendASTM F1264-03Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Locking Screw, Axial Pullout StrengthASTM F543-07Equivalent to predicate (K131677)Met equivalence criteria
PRECICE Locking Screw, Torque to FailureASTM F543-07Equivalent to predicate (K131677)Met equivalence criteria
Device functionality and verificationNoneFunctionally equivalent to predicateMet equivalence criteria
  • Note: The document explicitly states that "Substantial equivalence is based on similar indications for use, designs, and on in vitro testing performed. Where specific dimensional differences exist, bench testing has shown that these differences do not present new risks." This implies that the acceptance criteria for these tests were to demonstrate performance at least equivalent to, or not presenting new risks compared to, the predicate device (K131677). The reported performance is that the tests were performed and the device was found to be "as safe, effective, and performs as well as the predicate device (K131677)."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a distinct "test set" in terms of patient data. The evaluation was primarily based on mechanical testing and design functionality/performance testing of the device components (nail and screws). Therefore, the "sample size" would refer to the number of devices or components tested in these benchtop studies. This information is not provided in the given text.
  • Data Provenance: The data provenance is non-clinical bench testing. The country of origin of the data is not explicitly stated, but the company is U.S.-based (Irvine, CA), and the submission is to the U.S. FDA, suggesting the testing was likely conducted in the U.S. or at a facility that meets U.S. standards. The data is retrospective in the sense that it's laboratory testing, not prospective patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This question is not applicable to this submission. The "ground truth" for this medical device clearance was established through objective engineering standards and mechanical testing, not expert consensus on patient data. No human experts were used to establish ground truth for a "test set" in the context of clinical performance.

4. Adjudication Method for the Test Set

This question is not applicable as there was no "test set" requiring human adjudication as would be found in a clinical study for diagnostic devices. The acceptance was based on direct comparison of mechanical properties and design features to a predicate device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This device is a mechanical implant, not a diagnostic imaging device that would typically undergo such a study. The evaluation focused on mechanical and functional equivalence.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

No, a standalone algorithm performance study was not done. This device is a mechanical implant, not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this submission was based on objective performance metrics from mechanical testing conducted according to ASTM standards (F1264-03, F543-07) and a comparison to the known performance and design of a legally marketed predicate device (K131677). Essentially, the ground truth was "does this new device perform at least as well as the predicate device on established engineering tests, and does it introduce any new risks?"

8. The Sample Size for the Training Set

This question is not applicable. There was no "training set" in the context of machine learning or AI algorithms. The submission is for a medical device (intramedullary nail).

9. How the Ground Truth for the Training Set was Established

This question is not applicable for the same reason as #8.

{0}------------------------------------------------

November 2013

Ellipse PRECICE® System

DEC 9 2013

510(k) Summary

November 11, 2013

  1. Company: Ellipse Technologies, Incorporated 13900 Alton Parkway, Suite 123 Irvine, CA 92618

Contact: John McIntvre

Vice President, RA/QA/CA Phone: (949) 837-3600 x203 Fax: (949) 837-3664

    1. Proprietary Trade Name: Ellipse PRECICE® System
      Common Name: Intramedullary Nail 3.
  1. Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020)

Product Code: HSB (Rod, Fixation, Intramedullary and Accessories) રો

Product Description: The Ellipse PRECICE System is composed of the PRECICE nail (supplied 6. sterile), locking screws, surgical instruments and an external remote controller (ERC). The nail is available in tibia or femur models with various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE nail is supplied sterile by gamma radiation while the locking screws and PRECICE specific accessories are supplied non-sterile and must be sterilized prior to use. The nail contains an enclosed rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing.

  1. Indications: The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

  2. Substantial equivalence: Documentation that includes mechanical test results and detailed comparison to the predicate devices demonstrates that the Ellipse PRECICE System is substantially equivalent to the following 510(k) cleared device:

. Ellipse Intramedullary. Limb Lengthening System (K 131677)

{1}------------------------------------------------

and the control of the control of the control of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of the c

Substantial equivalence is based on similar indications for use, designs, and on in vitro testing performed. Where specific dimensional differences exist, bench testing has shown that these differences do not present new risks.

The 8.5 mm PRECICE Nail and the predicate device have intended use. Specifically, to lengthen the femur or tibia. The 8.5 mm diameter nail and the predicate are available in the same application, screw hole configurations, stroke lengths, and overall lengths. The 8.5 PRECICE Nail has the same materials, technological characteristics and principles of operation as that of the predicate. Both devices are inserted into the intramedullary canal of the femur or tibia and secured with locking screws. Both devices are adjusted non-invasively by the Ellipse external remote controller (ERC). The differences between the modified PRECICE System and the predicate device are as follows:

  • . Addition of an 8.5 mm diameter nail which has a tapered 10.7 mm proximal end for compatibility with the 5.0 mm locking screws and implantation tools.
  • . Addition of a 3.5 mm diameter locking screw, which will be used to secure the distal end of the 8.5 mm implant to the bone.

Data relied upon to determine substantial equivalence of the PRECICE System with the device modifications described in this submission to the cleared PRECICE System include the following:

  • . Mechanical testing
    ·

  • Design functionality and performance testing .
    The following specific tests have been performed in order to establish equivalence to the predicate devices:

Test DescriptionApplicable Test Standard .
PRECICE Nail, Static Four Point BendASTM F1264-03
PRECICE Nail, Dynamic Four Point BendASTM F1264-03
PRECICE Nail, Static Torque to FailureASTM F1264-03
PRECICE Locking Screw, Static Three Point BendASTM F1264-03
PRECICE Locking Screw, Dynamic Three Point BendASTM F1264-03
PRECICE Locking Screw, Axial Pullout StrengthASTM F543-07
PRECICE Locking Screw, Torque to FailureASTM F543-07
Device functionality and verificationNone

Based on the non-clinical testing performed, the 8.5 mm PRECICE Nail in this Special 510(k) is as safe, effective, and performs as well as the predicate device (K131677).

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, with three wave-like lines extending from the right side.

Public Health Service

Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 9, 2013

Ellipse Technologies, Incorporated Mr. John McIntyre Vice President, Regulatory, Quality, and Clinical Affairs 13900 Alton Parkway, Suite 123 Irvine, California 92618

Re: K133289

Trade/Device Name: Ellipse PRECICE® System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Codes: HSB Dated: November 11, 2013 Received: November 12, 2013

Dear Mr. McIntyre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. John Meintyre

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resources/orYowIndustry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statemant on last page.

510(k) Number (if known) K133289

Device Nama Ellipse PRECICE® System

TAR

Indications for Use (Describe)

The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 601 Subpart D)

[] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Hisatth (CDRH) (Signature)

Casey L Hanley, Ph.D.

Division of Orthopedic Devices

FORM FDA 3881 (8/13)

Pago for (Page 1/1)

PRC Publishing Borden (180 AD-4040 C

N/A