(85 days)
The Diagnostic Ultrasound System Xario 100 Model TUS-X100 and TUS-X100S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, and musculo-skeletal (both conventional and superficial).
The Xario 100 Model TUS-X100 and TUS-X100S are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex, and sector array with frequency ranges between approximately 3 MHz to 10 MHz.
The provided text is a 510(k) Summary for a Toshiba Xario 100 diagnostic ultrasound system. This document outlines the general characteristics of the device and its intended use, but does not contain specific information about acceptance criteria, device performance studies, sample sizes for test or training sets, expert qualifications, adjudication methods, or MRMC studies.
The document states:
- "Design Control Activities including risk management following the ISO14971, verification/validation testing and Acoustic Output testing (UD3, 2004) were conducted through bench testing are included in this submission. This documentation includes testing which demonstrates that the requirements for the features have been met." (Section 15. SAFETY and 16. TESTING)
This statement indicates that verification and validation testing was performed, and requirements were met, but it does not provide the details of those tests, the acceptance criteria, or the reported performance. The 510(k) summary is a high-level overview and directs to the full submission for detailed documentation.
Therefore, I cannot extract the specific information you requested as it is not present in the provided text. The document confirms that testing was done to demonstrate that the device meets requirements, but the details of what those requirements are and the results of the tests are not included here.
{0}------------------------------------------------
K133277
Page 1 of 3
510(k) SUMMARY
-
- SUBMITTER'S NAME: Toshiba America Medical Systems, Inc.
-
- ADDRESS: 2441 Michelle Drive Tustin, CA. 92780-2068
-
- ESTABLISHMENT REGISTRATION: 2020563
CONTACT PERSON: 4. Charlemagne Chua
Manager, Regulatory Affairs (714) 669-7896
-
- U.S AGENT:
Paul Biggins Director, Requlatory Affairs (714) 730-5000
- U.S AGENT:
-
- Date Prepared: October 23, 2013
TRADE NAME(S): 7.
Diagnostic Ultrasound System Xario 100 Model TUS-X100 and TUS-X100S, SW V1.0
8. COMMON NAME:
System, Diagnostic Ultrasound
9. DEVICE CLASSIFICATION:
Class II
Ultrasonic Pulsed Doppler Imaging System -- Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [per 21 CFR 892.1570]
10. PREDICATE DEVICE:
| Product | Marketed by | 510(k) Number | Clearance Date |
|---|---|---|---|
| Xario 200(TUS-X200 and TUS-X200S), v1.0 | Toshiba AmericaMedical Systems | K131507 | August 28, 2013 |
{1}------------------------------------------------
11. REASON FOR SUBMISSION:
New device.
12. DEVICE DESCRIPTION:
The Xario 100 Model TUS-X100 and TUS-X100S are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex, and sector array with frequency ranges between approximately 3 MHz to 10 MHz.
13. SUMMARY OF INTENDED USES:
The Diagnostic Ultrasound System Xario 100 Model TUS-X100 and TUS-X100S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both aduit and pediatric), peripheral vascular, and musculo-skeletal (both conventional and superficial).
14. SUBSTANTIAL EQUIVALENCE:
This device is substantially equivalent to the Xario 200 Diagnostic Ultrasound System, K131507, marketed by Toshiba America Medical Systems. The Xario 100 Model TUS-X100 and TUS-X100S, SW Version 1.0, functions in a manner similar to and is intended for the same use as the predicate device. The subject device is a compact diagnostic ultrasound system by implementing latest technologies.
A comparison table is included in this submission detailing the similarities and differences between the predicate device and the subject device.
15. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 (2005), IEC 60601-2-37 (2007), IEC 62304 (2006), AIUM RTD2-2004 Output Display and ISO 10993-1 standards.
16. TESTING
Design Control Activities including risk management following the ISO14971, verification/validation testing and Acoustic Output testing (UD3, 2004) were conducted through bench testing are included in this submission. This documentation includes testing which demonstrates that the requirements for the features have been met.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
{2}------------------------------------------------
Additionally, testing of this device was conducted in accordance with the applicable standards published by the International Electromechnical Commission (IEC) for Medical Devices.
:
. .
:
.
. . . . .
ﯿﺪ
. .
.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue ocument Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 17, 2014
Toshiba Medical Systems Corporation % Mr. Charlemagne Chua Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K133277
Trade/Device Name: Xario 100, TUS-X100 and TUS-X100S V1.0 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 23, 2013 Received: October 24, 2013
Dear Mr. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Xario 100 TUS-X100 and Xario 100 TUS-X100S Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
| PSU-30BT | PVU-375BT | PVU-674MV |
|---|---|---|
| PVU-78IVT | PLU-704BT | PLU-1005BT |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2-Mr. Chua
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/lygustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safets/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133277
Device Name
Xario 100 TUS-X 100 and TUS-X100S, v1.0
Indications for Use (Describe)
The Diagnosuc Ultrasound System Xario 100 Model TUS-X100S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: feal, abdominal, pediatric, small organs, trans-rectal, nevnatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, and musculo-skeletal (both conventional and superficial).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | ||
|---|---|---|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | ||
| Page 001 of 008 | ||
| FORM FDA 3881 (9/13) | Page 1 of 2 | PSC Publishing Services (301) 443-6740 EF |
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{6}------------------------------------------------
510(k) Premarket Notification Xario 100 V1.0 Diagnostic Ultrasound System
System: Xario 100 TUS-X100 TUS-X100S VI.O
Transducer:
2018年08月10日 10:00:00 PM
Intended Use: Diagnostic ultrasound imaging or thuid flow analysis of the human body as follows:
Clinical Application Mode of Operation
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | TIII | AdvancedDynamicFlow | Power | CIII | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | N | 2 | N | N | N | N | 5.6.7.8 | |||||
| Fetal | N | N | N | N | 2.3 | N | N | N | N | 5.6.7.8 | ||
| Abdominal | N | N | N | N | N | 2.3 | N | N | N | N | 5.6.7.8 | |
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | N | N | N | N | N | 2.3 | N | N | N | N | 5.6.7.8 | |
| Small Organ (Note 1) | N | N | N | N | N | 2 | N | N | N | N | 5.6.7.8.9 | |
| Neonatal Cephalic | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Adult Cephalic | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Trans-rectal | N | N | N | N | 2 | N | N | N | 4.5.7 | |||
| Trans-vaginal | N | N | N | N | 2 | N | N | N | 4.5.7 | |||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | 2 | N | N | N | 5.6.7.8.9 | |||
| Musculo-skeletal(Superficial) | N | N | N | N | 2 | N | N | N | 5,6,7,8.9 | |||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | N | N | N | N | N | 3 | N | N | N | 4.7 | ||
| Cardiac Pediatric | N | N | N | N | N | 3 | N | N | N | 4.7 | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vascularOther (Specify) | N | N | N | N | 2 | N | N | N | 5.6.7.8.9 |
N = new indication: P = previously cleared by FDA: E = added under this appendix
Note I Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/PW1D
Note 3 Combined mode includes B/M: B/PWD: BDF/RWD: BDF/MDF: BDF/MDF/I/WD: 2D/CWD: BDF/CWD
-
Note 4 TDI
Note 5 ApliPure -
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Prescription Use Only (Per 21 CFR 801.109)
Page 002 of 008
{7}------------------------------------------------
、
510(k) Premarket Notification
Xario 100 V1.0 Diagnostic Ultrasound System
Svstem: _ Xario 100 TUS-X100,TUS-X100S VI.0 Transduccr:_ PSU-30BT
Intended Use: Diagnostic ultrasound imaging or Nuid flow analysis of the human body as follows: .
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler(Specify) | Combined* | THI | AdvancedDynamicFlow | Power | CII | 4D | Other[Note] | |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Adult Cephalic | N | N | N | N | N | 3 | N | N | N | 7 | ||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | N | N | N | N | N | 3 | N | N | N | 4.7 | ||
| Cardiac Pediatric | N | N | N | N | N | 3 | N | N | N | 4.7 | ||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vascular | ||||||||||||
| Other (Specify) |
N = new indication: P = previously cleared by FDA: F: = added under this appendix
Note I Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/PWD
Note 3 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/PWD: 2D/CWD: BDF/CWD
Nole 4 TDI
Note 5 ApliPure
Note 6 ApliPurc Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Preseription Use Only (Per 21 CFR 801.109)
Page 003 of 008
510(k): Xarto 100 (TUS-X100 and TUS-X100S), v1.0 Page 464 of 3827
{8}------------------------------------------------
: • ••
510(k) Premarket Notification
Xario 100 VI.0 Diagnostic Ultrasound System
System: __Xario_100 TUS-X100,TUS-X100S V1.0 Transducer:____PV11-375BT
Intended Use: Diagnostic ultrasound imaging or Huid How analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | Other[Note] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify) | THI | AdvancedDynamicFlow | Power | CIII | 4D | ||
| Ophthalmic | ||||||||||||
| Fetal | N | N | N | N | 2 | N | N | N | 5.6.7.8 | |||
| Abdominal | N | N | N | N | 2 | N | N | N | 5.6.7.8 | |||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | N | N | N | N | 2 | N | N | N | 5.6.7.8 | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vascular | ||||||||||||
| Other (Specify) | ||||||||||||
N = new indication: P = previously cleared by FDA: F: = added under this appendix
Note 1 Small organ includes thyroid. breast and testicle. Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/IVWD Note 3 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF/PWD: 2D/CWD: BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 ApliPure Plus Note 7 Precision Imaging Note 8 Differential THI Note 9 Elastography
Preseription Use Only (Per 21 CFR 801.109)
Page 004 of 008
510(k): Xario 100 (TUS-X100 and TUS-X100S), v1.0 Page 465 of 3827
{9}------------------------------------------------
. . ...
510(k) Premarket Notification Xario 100 V1.0 Diagnostic Ultrasound System
System: Xario 100 TUS-X100, TUS-X100S V1.0 Transducer:____ PVU-674MV
Intended Use: Diagnostic ultrasound imaging or thrid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | THI | AdvancedDynamicFlow | Power | CHI | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | N | N | N | N | 2 | N | N | N | N | 5.6.7.8 | ||
| Abdominal | N | N | N | N | 2 | N | N | N | N | 5.6.7.8 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | N | N | N | N | 2 | N | N | N | N | 5.6.7.8 | ||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vascular | ||||||||||||
| ()ther (Specify) |
N = new indication: P = previously cleared by FDA: F: = added under this appendix
Note 1 Small organ includes thyroid. breast and testicle. Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/IVWD Note 3 Combined mode molades B/M: B/PW1): BDF/MDF: BDF/MDF: BDF/MDF/PWD: 2D/CWD: BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 ApliPurc Plus Note 7 Precision Imaging Note 8 Differential TIII Note 9 Elastography
Prescription Use Only (Per 21 CFR 801.109)
Page 005 of 008
.
510(k): Xario 100 (TUS-X100 and TUS-X1008), v1.0 Page 466 of 3827
{10}------------------------------------------------
、..
510(k) Premarket Notification
Xario 100 V1.0 Diagnostic Ultrasound System
System: Xario 100 TUS-X100 TUS-X100S V1.0 PVU-78IVT Transducer:____
Intended Use: Diagnostic ultrasound imaging or Nuid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | THI | DynamicFlow | Power CHI | 4D | Other[Note] |
| Ophthalmic | |||||||||||
| Fetal | |||||||||||
| Abdominal | |||||||||||
| Intra-operative (Abdominal) | |||||||||||
| Intra-operative (Neuro) | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | |||||||||||
| Small Organ (Specify) (1) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Trans-rectal | N | N | N | N | 2 | N | N | N | 4.5.7 | ||
| Trans-vaginal | N | N | N | N | 2 | N | N | N | 4.5.7 | ||
| Trans-urethral | |||||||||||
| Trans-esoph. (non-Card.) | |||||||||||
| Musculo-skeletal(Conventional) | |||||||||||
| Musculo-skeletal (Superficial) | |||||||||||
| Intravascular | |||||||||||
| Other (Specify) | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Intravascular (Cardiac) | |||||||||||
| Trans-esoph. (Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Other (Specify) | |||||||||||
| Peripheral vascular | |||||||||||
| Other (Specify) |
N = new indication: P = previously cleared by I'DA: F: = added under this appendix
Note I Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M: B/PW1D: BIDF/PW1D: BDF/MDF: BDF/MDF/I'W1)
Note 3 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/PWD: 2D/CWD: BDF/CWD
- Note 4 TDI Note 5 ApliPure
- Note 6 ApliPure Plus
- Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Prescription Use Only (Per 21 CFR 801.109)
Page 006 of 008
510(k): Xario 100 (TUS-X100 and TUS-X100S), v1.0 Page 467 of 3827
{11}------------------------------------------------
.,
.
510(k) Premarket Notification
Xario 100 VI.0 Diagnostic Ultrasound System
System: Xario 100 TUS-X100,TUS-X100S VI.O Transducer:_ Pl.U-704BT
Intended Use: Diagnostic ultrasound imaging or Nuid Now analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | THI | AdvancedDynamicFlow | Power | CII | 4D | Other[Note] | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | |||||||||||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Specify) (1) | N | N | N | N | 2 | N | N | N | 5.6.7.8 | ||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | 2 | N | N | N | 5.6.7.8 | ||||
| Musculo-skeletal (Superficial) | N | N | N | N | 2 | N | N | N | 5.6.7.8 | ||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vascular | N | N | N | N | 2 | N | N | N | 5.6.7.8 | ||||
| Other (Specify) |
N = new indication: P = previously cleared by FI)A; F: = added under this appendix
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF: BDF/MDF/PWD Note 3 Combined mode includes B/M: B/PWD; BDF/PWD: BDF/MDF/PWD: 2D/CWD: BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 ApliPure Plus Note 7 Precision Imaging Note 8 Differential TI Il Note 9 Elastography
Preseription Use Only (Per 21 CFR 801.109)
Page 007 of 008
510(k): Xario 100 (TUS-X100 and TUS-X100S), v1.0 Page 468 of 3827
{12}------------------------------------------------
・ .. .
System: Xario 100 TUS-X100, JUS-X100S V1.0 Transducer: Pl.U-1005BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | TIII | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | N | N | N | N | N | 2 | N | N | N | 5.6.7,9 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | 2 | N | N | N | 5,6,7,9 | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | 2 | N | N | N | 5.6.7.9 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| 'Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vascular | N | N | N | N | N | 2 | N | N | N | 5.6.7.9 | ||
| Other (Specify) |
N = new indication: P = previously cleared by FDA: I: = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle,
Note 2 Combined mode includes B/M: B/PWD: BDF:MDF: BDF:MDF: BDF:MDF: BDF:MDF:1WD
Note 3 Combined mode includes B/M: B:PW1): BDF/MDF: BDF/MDF/W12: 2D/CW12: BDF/CW/D
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Prescription Use Only (Per 21 CFR 801.109)
Page 008 of 008
N/A