K Number
K133258

Validate with FDA (Live)

Device Name
TRUFREEZE SYSTEM
Manufacturer
Date Cleared
2014-06-19

(239 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The truFreeze is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures.

The truFreeze is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions.

Device Description

The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the treatment area via a small, low pressure, open tipped catheter. The truFreeze System consists of (1) a console and (2) a disposable spray kit.

Console:
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the- shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A wireless remote control provides alternative timer control from a distance in the treatment room. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction.

Disposable spray kit:
The truFreeze disposable spray kit consists of 5 individually packaged sterile single-use catheters ( 7 Fr Straight Tip Catheter and one Catheter Introducer.) and 5 individually packaged cryogen decompression tubes (CDTs) (each containing one Dual Lumen 20 Fr CryoDecompression Tube (CDT), Connector, and Suction Tubing.). The catheter is flexible and capable of retroflex in a scope. The CDT and accessory tubes are included for use with the on-board suction system.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the truFreeze® System based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Bench Test CharacteristicAcceptance Criteria (Predicate IceSense3 K102360)Reported Device Performance (Proposed truFreeze® System K133258)Comments from Study
Ability to reach equivalent dose (using same freeze/thaw/freeze cycle)2min/2min/2min = 2.1-2.5cm depth2min/2min/2min = 1.8-2.4cm depthBoth capable of producing equivalent depth of ice
3min/2min/3min = 2.6-3.0cm depth3min/2min/3min = 2.4-3.0cm depth
4min/4min/4min = 3.1-4.0cm depth4min/4min/4min = 3.0-3.6cm depth
Output Temperature at tip-170 degree C-196 degree CBoth achieve temperatures low enough to destroy malignant tissue
Equivalent temperature distribution using the radial distance from the 0°C isotherm (border of ice ball) to the -20°C, -40°C and -50°C isotherms@ freeze depth of 22mm at 0°C: -20°C isotherm at 16.9mm depth -40°C isotherm at 13.4mm depth@ freeze depth of 21.87mm at 0°C: -20°C isotherm at 15.5mm depth -40°C isotherm at 11.8mm depthBoth can produce lethal ice at comparable depths

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states that "limited bench testing was conducted to support demonstration of substantial equivalence." It does not specify the sample sizes (e.g., number of tests, number of trials) for each bench test performed. Given the nature of bench testing, the data provenance would be laboratory-controlled (not patient data). The document refers to the data as "bench testing," implying a prospective experimental setup in a lab environment. The country of origin of the data is not mentioned but can be inferred as the USA, as the applicant is CSA Medical in Lexington, MA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. Bench testing typically does not involve human experts establishing ground truth in the same way clinical studies do. The "ground truth" for these tests would be the physical measurements taken (e.g., ice ball dimensions, temperature readings) by the testing equipment and personnel.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided. Bench testing focuses on objective physical measurements, not subjective evaluations that would require adjudication among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not conducted. The document describes bench testing to demonstrate substantial equivalence to a predicate device, not a study involving human readers or AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study of an algorithm was not done. The "device" in this context is a cryosurgical tool (hardware with integrated software), not an AI algorithm. The performance evaluation was of the physical device's capabilities (ice ball formation, temperature output) in a lab setting.

7. The Type of Ground Truth Used

For the bench testing, the ground truth was based on physical measurements of:

  • Ice ball depth (in cm) under different freeze/thaw/freeze cycles.
  • Output temperature at the catheter tip (in degrees C).
  • Radial distance of specific isotherms (-20°C, -40°C) from the 0°C isotherm at a given freeze depth.

These are objective, quantitative measurements taken directly from the device's operation.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. The truFreeze® System is a physical cryosurgical device, not a machine learning model that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above (no training set for a physical device).

{0}------------------------------------------------

Image /page/0/Picture/0 description: This image contains the text 'K133258' at the top, which appears to be a document or identification number. Below that, it indicates 'Page 1 of 3', suggesting this is the first page of a three-page document. The image also contains the number '2014' on the left side.

JUN 1 9 2014 csomedical.com

CSA Medical Inc. 91 Hartwell Avenue Lexington, MA 02421 781,538,4720 re 781.538.4730 lax

510K Summary TruFreeze® System

Applicant Establishment Registration Number Contact Person

  • Summary Date Proprietary Name Classification Classification Name Regulation Number Classification Predicate Device
    CSA Medical 3004534508 Sherrie Coval-Goldsmith VP RA/QA CSA Medical 91 Hartwell Ave Lexington, Ma 02142 Phone: 781-538-7447 Fax: 781-538-4730 sgoldsmith@csamedical.com June 9, 2014 truFreeze® System Class II Cryosurgical Unit, Cryogenic Surgical Device 21 CFR 878.4350 Product Code GEH K102360 IceSense3

Device Description

The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the treatment area via a small, low pressure, open tipped catheter. The truFreeze System consists of (1) a console and (2) a disposable spray kit.

Console:

The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the- shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A wireless remote control provides alternative timer control from a distance in the treatment room. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction.

{1}------------------------------------------------

Disposable spray kit:

The truFreeze disposable spray kit consists of 5 individually packaged sterile single-use catheters ( 7 Fr Straight Tip Catheter and one Catheter Introducer.) and 5 individually packaged cryogen decompression tubes (CDTs) (each containing one Dual Lumen 20 Fr CryoDecompression Tube (CDT), Connector, and Suction Tubing.). The catheter is flexible and capable of retroflex in a scope. The CDT and accessory tubes are included for use with the on-board suction system.

Intended Use/Indications for Use

The truFreeze System is similar in its Intended Use as the predicate device (K102360 IceSense3). Both devices describe the ablation of benign and malignant tissue in general terms.

The requirement to use active or passive venting is in the cleared truFreeze Instructions for Use document (K113021) and adding it to the intended use/indications for use statement highlights the importance of venting and will enhance safe use of the device. Therefore, the revised intended use/indications for use statement raises no new issues of safety or effectiveness.

The truFreeze is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures.

The truFreeze is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions.

Technical and Operational Characteristics

The truFreeze System is similar in design, operational and technological characteristics as the predicate device and supports that no new safety concerns are being raised by change in intended use/indications for use statement and thus raises no new issues of safety or effectiveness.

Summary of Testing

The truFreeze System was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility testing, software testing, animal testing, biocompatibility and sterilization testing. Since the change in:the, truFreeze, system limited to the Indications for Use, limited bench testing swas conducted to support

{2}------------------------------------------------

demonstration of substantial equivalence in performance from the truFreeze to its predicate device. A summary table of the bench testing is provided below:

Bench Testing Comparison Table
Bench TestProposed truFreeze SystemPredicate IceSense3Comments
Characteristic(K133258)(K102360)
Ability to reachequivalent dose (usingsame freeze/thaw/freezecycle)2min/2min/2min =1.8-2.4cm3min/2min/3min=2.4-3.0cm4min/4min/4min=3.0-3.6cm2min/2min/2min =2.1-2.5cm3min/2min/3min=2.6-3.0cm4min/4min/4min=3.1-4.0 cmBoth capable of producingequivalent depth of ice
Output Temperature attip-196 degree C-170 degree CBoth achieve temperatureslow enough to destroymalignant tissue
Equivalent temperaturedistribution using theradial distance from the0° C isotherm (borderof ice ball) to the -20°C,-40°C and -50°Cisotherms@ freeze depth of 21.87mmat 0°C-20°C isotherm at 15.5mmdepth-40°C isotherm at 11.8mmdepth@ freeze depth of 22mm at0°C-20°C isotherm at 16.9mmdepth-40°C isotherm at 13.4mmdepthBoth can produce lethalice at comparable depths

Rationale For Substantial Equivalence

The Intended Use/Indications for Use statement and technological characteristics of the truFreeze System and the predicate device (K102360 IceSense3) were compared. The Intended Use/Indications for Use statement of the two devices had equivalent general claims and do not raise new questions of safety and performance. Bench testing demonstrated that both devices can achieve equivalently cold temperatures in equivalent time periods to produce equivalent amounts of ice. These temperatures are cold enough to destroy unwanted benign or malignant tissue. This proposed change to the Indications for Use is well supported by a review of the peer reviewed clinical literature and are similar to those of the predicate device with the same general intended use of cryogenic destruction of benign or malignant tissue during surgical procedures.

The truFreeze device requires either active or passive venting for safe use. This is described within the previously cleared Instructions for Use document for the truFreeze device. The addition of this statement to the Intended Use/Indications for Use statement enhances the safe use of the device.

Conclusion

Based on the label and technology comparison as well as the performance testing, the truFreeze device is substantially equivalent to the predicate device listed above.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Fleath Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 19, 2014

CSA Medical Ms. Sherrie Coval-Goldsmith Vice President, Regulatory Affairs/Quality Assurance 91 Hartwell Avenue Lexington, Massachusctts 02142

Re: K133258

Trade/Device Name: truFreeze" System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: May 15, 2014 Received: May 16. 2014

Dear Ms. Coval-Goldsmith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any reathardthe Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Sherrie Coval-Goldsmith

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S.

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement befow.

510(k) Number (if known) K133258

Device Name truFreeze@ System

Indications for Use (Describe)

The truPreeze is intended for cryogenic destruction of ther active or passive venting during surgical procedures.

The truFreeze is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

SE SERVICE CARD FOR CONSTITUTION FOR FOR FOR FOR FOR MARK CONTROLLER CONSULTION CONTROLL

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4

The burden time for this collection of Information is estimated to average 79 hours per response, including the I he burden instructions, search existing data sources, gather and maintain the catale and camplete lime to review instructions, Search examing duto ocenose, general bath burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@lda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

:

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.