K Number
K133194

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-06-26

(252 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicated for use via a lateral / anterolateral surgical approach above the great vessels in the treatment of the thoracolumbar spine (T1-L5). The Eminent Spine Red Diamond Rattlesnake Lumbar Plating System is indicated for use via an anterior surgical approach below the bifurcation of the great vessels in the treatment of the lumbosacral (L1-S1) spine. The Diamond Rattlesnake System is intended to provide temporary stabilization as an adjunct to fusion using a lateral approach in skeletally mature patients for the treatment of the following acute and chronic instabilities and deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudarthrosis, spondylolysis, spondylolisthesis, deformity (scoliosis, kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation and subluxation), tumor, or failed previous spine surgery.

Device Description

The Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System consists of a variety of plates and screws. All components are fabricated and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136. The plates are manufactured in a variety of configurations with lengths of 21 to 33mm in a single level plate and 64 to 76mm in a two level plate. Additionally, the non-locking screws are provided in diameters of 5.5mm with lengths of 25 - 50mm. The responsible surgeon will determine the correct size of the implant in accordance with the size of the individual patient. The Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System utilizes a variety of standard orthopedic instruments to assist in the placement of the devices.

AI/ML Overview

The document describes a 510(k) submission for the Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System. This is a spinal intervertebral body fixation orthosis, a medical device used for temporary stabilization as an adjunct to fusion in the lumbar and thoracolumbar spine.

It's important to understand that a 510(k) submission is for demonstrating substantial equivalence to a predicate device, not for proving novel clinical claims or efficacy through extensive clinical studies in the same way a PMA (Pre-Market Approval) process would. Therefore, the "study that proves the device meets the acceptance criteria" in this context refers to non-clinical performance testing designed to show the new device performs at least as well as the predicate device(s) in specific mechanical properties.

Here's the breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Reference to Predicate)Reported Device Performance (Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System)
Meet or exceed performance of predicate devices in:Meets or exceeds the performance of the predicate device(s)
Static axial compression (per ASTM F1717)Results show satisfactory performance
Dynamic axial compression (per ASTM F1717)Results show satisfactory performance
Static torsion (per ASTM F 1717)Results show satisfactory performance

Note: The document explicitly states "The results of these studies show that the subject Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System meets or exceeds the performance of the predicate device, and the device was therefore found to be substantially equivalent." The specific numerical acceptance criteria (e.g., minimum load, cycles to failure) are not detailed in this summary, but are implied to be derived from the predicate device performance and the ASTM F1717 standard.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not explicitly stated. For non-clinical (mechanical) testing of medical devices like this, sample sizes are typically determined by statistical validation requirements for the specific test (e.g., 5-10 samples per test condition to achieve statistical significance). The document only mentions "Nonclinical testing was performed..."
  • Data Provenance: Not specified. This is non-clinical lab testing, not human data. The testing would have been conducted in a laboratory setting, likely in the US (where Eminent Spine, LLC and Secure BioMed Evaluations are located).
  • Retrospective or Prospective: Not applicable as this is non-clinical mechanical testing, not a clinical study involving human subjects or data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This question is not applicable in this context. "Ground truth" and "experts" as described (e.g., radiologists) are relevant for clinical studies, particularly those involving image interpretation or clinical diagnosis. For a spinal implant's mechanical testing, the "ground truth" is established by adherence to recognized engineering standards (like ASTM F1717) and the physical measurements obtained during the tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This question is not applicable for non-clinical mechanical testing. Adjudication methods are used in clinical studies to resolve discrepancies in expert interpretations or clinical outcomes.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable. This describes a spinal implant, not an AI software or imaging system that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable. This device is a physical spinal implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this device's performance is based on established engineering principles and standards for mechanical testing (ASTM F1717). The performance of the device is compared against the known or expected performance of legally marketed predicate devices under these standardized test conditions.

8. The sample size for the training set

  • This question is not applicable. There is no "training set" in the context of mechanical testing for a spinal implant. Training sets are typically used for machine learning or AI models.

9. How the ground truth for the training set was established

  • This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

2014 2 6 2014

In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System is provided below.

Date Summary PreparedJune 11, 2014
Manufacturer/Distributor/Spons orEminent Spine, LLC7200 N 1-35 Building #1Georgetown, TX 46037Phone 512-868-5980Fax 512-864-1462
510(k) ContactSecure BioMed EvaluationsLinda Braddon, Ph.D.7828 Hickory Flat HighwaySuite 120Woodstock, GA 30188770-837-2681 (direct)855-MED-DEV1 (office)LGB@SecureBME.com
Trade NameBlack / Red Diamond Rattlesnake Lumbar Plating System
Common NameSpinal intervertebral body fixation orthosis
Code -ClassificationKWQ 21 CFR 888.3060: Class II
Predicate DevicesMedtronic Z-Plate (K922543)Acromed M-2 Plate (K972718)Nuvasive Lateral Plate System (K061789, K082070, K091071)
Predicate Device - 2 HoleLateral PlateNuvasive Lateral Plate System K091071
Predicate - Anterior PlateMedtronic Z-Plate (K922543)Acromed M-2 Plate (K972718)
Device DescriptionThe Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating Systemconsists of a variety of plates and screws. All components are fabricatedand manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136.The plates are manufactured in a variety of configurations with lengths of21 to 33mm in a single level plate and 64 to 76mm in a two level plate.Additionally, the non-locking screws are provided in diameters of 5.5mmwith lengths of 25 - 50mm. The responsible surgeon will determine thecorrect size of the implant in accordance with the size of the individualpatient.The Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating Systemutilizes a variety of standard orthopedic instruments to assist in theplacement of the devices.
Intended UseThe Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicatedfor use via a lateral / anterolateral surgical approach above the great vessels in thetreatment of the thoracolumbar spine (T1-L5). The Eminent Spine Red DiamondRattlesnake Lumbar Plating System is indicated for use via an anterior surgicalapproach below the bifurcation of the great vessels in the treatment of thelumbosacral (L1-S1) spine. The Diamond Rattlesnake System is intended to providetemporary stabilization as an adjunct to fusion using a lateral approach in skeletallymature patients for the treatment of the following acute and chronic instabilitiesand deformities: degenerative disc disease (DDD, defined as back pain of discogenicorigin with degeneration of the disc confirmed by patient history and radiographicstudies), pseudarthrosis, spondylolysis, spondylolisthesis, deformity (scoliosis,kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation andsubluxation), tumor, or failed previous spine surgery.
Technological CharacteristicsAs was established in this submission, the subject Eminent Spine Black / RedDiamond Rattlesnake Lumbar Plating System is substantially equivalent tothe predicate systems in terms of design, intended use, materialcomposition, function, and range of sizes.
Non-Clinical PerformanceTesting ConclusionNonclinical testing was performed to demonstrate that the Eminent SpineBlack / Red Diamond Rattlesnake Lumbar Plating System is substantiallyequivalent to other predicate devices. The following testing was performed:Static and dynamic axial compression testing per ASTM F1717 Static torsion testing per ASTM F 1717 The results of these studies show that the subject Eminent Spine Black / RedDiamond Rattlesnake Lumbar Plating System meets or exceeds theperformance of the predicate device, and the device was therefore found to be substantially equivalent.
Substantial EquivalenceSummary (Conclusion)The information presented demonstrates the substantial equivalency of theEminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System.

{1}------------------------------------------------

.

:

.

:

.

)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 26, 2014

Eminent Spine, LLC % Linda Braddon, Ph.D. Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188

Re: K133194

Trade/Device Name: Black / Red Diamond Rattlesnake Lumbar Plating System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: May 22, 2014 Received: May 28, 2014

Dear Dr. Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Linda Braddon, Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please-note the requlation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133194

Device Name

Black / Red Diamond Rattlesnake Lumbar Plating System

Indications for Use (Describe)

The Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicated for use via a lateral / anterolateral surgical approach above the great vessels in the thoracolumbar spine (TI-LS). The Eminent Spine Red Diamond Rattlesnake Lumbar Plating System is indicated for use via an anterior surgical approach below the great vessels in the treatment of the lumbosacral (LI-SI) spine. The Diamond Rattlesnake System is intended to provide temporary stabilization as an adjunct to fusjon using a lateral approach in skeletally mature patients for the following acute and chronic instabilities and deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), poeudarthrosis, spondylolisthesis, deformity (scollosis, kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation and subluxation), tumor, or failed previous spine surgery.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

.. . . .

A. W. FABER

Division of Orthopedic Devices

:

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

้า

N/A