Intended Use

Total Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer® Patient Specific Instruments System is to be used with Zimmer NexGen CR-Flex fixed bearing, Zimmer NexGen CR fixed bearing, Zimmer NexGen LPS-Flex fixed bearing, Zimmer NexGen LPS fixed bearing, Zimmer Gender Solutions Natural – Knee Flex fixed bearing, Zimmer Persona™ CR fixed bearing, Zimmer Persono™ PS fixed bearing and Zimmer Persona™ Trabecular Metal™ prostheses families only.

Unicompartmental Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Unicompartmental Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer® Patient Specific Instruments System is to be used with Zimmer Unicompartmental High Flex Knee System prostheses families only.

The Zimmer® Patient Specific Instruments are intended for single use only.

Device Description

The subject 'device Zimmer® Patient Specific Instruments System 5.4 is a modification to the predicate device Zimmer® Patient Specific Instruments System 5.0 (K121640) to accommodate the new compatible implant system Zimmer® Personalized Knee System Trabecular Metal™ components cleared via 510(k) K122745 and K121771. It is designed to assist a surgeon in the placement of total knee replacement components for Zimmer® Persono™ Trabecular Metal™ components. The system consists of a software device, branded as Zimmer® Patient Specific Instruments Planner (ZPSIP) and a hardware component, branded as Zimmer® Patient Specific Instruments (ZPSI).

AI/ML Overview

Here's an analysis of the provided text regarding the Zimmer Patient Specific Instruments System 5.4, addressing your specific questions:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary does not explicitly state specific acceptance criteria in a quantitative manner (e.g., target accuracy percentages, deviation limits). Instead, it relies on the concept of "substantial equivalence" to a predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Safety and effectiveness substantially equivalent to predicate device (Zimmer Patient Specific Instruments System 5.0 (K121640))."Testing verified that the accuracy and performance of the system is adequate to perform as intended.""Results of software verification and validation testing demonstrated the device's safety and effectiveness is substantially equivalent to the predicate device."
Accuracy and performance adequate for intended surgical assistance."Testing verified that the accuracy and performance of the system is adequate to perform as intended.""A detailed comparison shows the subject device is substantially equivalent in intended use, design, functionality, operating principles, materials and performance characteristics to the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Non-clinical tests using saw bones" for performance evaluation.

  • Sample Size for Test Set: Not specified in the provided text.
  • Data Provenance: The tests were "Non-clinical" and used "saw bones." This implies an in vitro or laboratory setting, not patient data. No country of origin is mentioned for this testing. It is a retrospective analysis of the device's performance under controlled conditions, not utilizing patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not provide information on the number of experts used or their qualifications for establishing ground truth specific to the saw bone testing. Since it's saw bone testing, the "ground truth" would likely be established by precise measurements or engineering specifications rather than expert medical interpretation.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method. Given the nature of saw bone testing, where physical measurements confirm accuracy, expert adjudication in the typical clinical sense would not be applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a multi-reader multi-case comparative effectiveness study was not mentioned. The evaluation primarily focuses on the device's standalone performance in a non-clinical setting and its substantial equivalence to a predicate, not on how it improves human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance evaluation was done. The "Non-clinical tests using saw bones" without mention of human interaction suggests a standalone evaluation of the device's accuracy and performance. The software device, Zimmer Patient Specific Instruments Planner (ZPSIP), generates a pre-surgical plan, and the hardware (ZPSI) transfers this plan. The testing verifies the system's ability to "perform as intended" in this capacity, i.e., accurate planning and guidance.

7. Type of Ground Truth Used

The ground truth for the saw bone testing would be based on physical measurements and engineering specifications against the planned surgical cuts/positions. For example, if the device is intended to guide a cut at a specific angle, the ground truth would be the actual measured angle of the cut made on the saw bone. It is not expert consensus, pathology, or outcomes data, as these are typically associated with human patient clinical data.

8. Sample Size for the Training Set

The document does not provide any information regarding a training set. This is a medical device clearance based on substantial equivalence, and typical AI/ML "training sets" are not explicitly mentioned as part of the evaluation. The software component, ZPSIP, implies some form of algorithmic processing, but the development and training details are not disclosed in this summary.

9. How the Ground Truth for the Training Set Was Established

As no training set is described, there is no information on how its ground truth was established.

{0}------------------------------------------------

Zimmer Patient Specific Instruments System 5.4 510(k) Premarket Notification

510(k) Summary

K133162

510(k) Summary

The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92

Company nameMaterialise N.V.
Establishment registration number3003998208
Street AddressTechnologielaan 15
CityLeuven
Postal code3001
CountryBelgium
Phone number+32 16 74 45 37
Fax number+32 16 39 66 06
Principal Contact personOliver Clemens
Contact titleQuality and Regulatory Officer
Contact e-mail addressRegulatory.Affairs@materialise.be
Additional contact personWim Claassen
Contact titleProduct Manager
Contact e-mail addressWim.Claassen@materialise.be

Submission date

The date of the Traditional 510(k) submission is September 30", 2013.

Submission information

Trade NameZimmer® Patient Specific InstrumentsZimmer® Patient Specific Instruments Planner
Common NameKnee prosthesis
Classification NameKnee joint patellofemorotibial polymer /metal /polymer semi-constrained cemented prosthesis
Primary product codeJWH (21 CFR 888.3560)
Subsequent product codesMBH, OOG and OIY

Predicate Devices

The predicate device to which substantial equivalence is claimed to:

Trade or property or model nameZimmer® Patient Specific Instruments System 5.0
510(k) numberK121640
Decision date12/05/2013

{1}------------------------------------------------

Zimmer Patient Specific Instruments System 5.4 510(k) Premarket Notification

Product codeJWH, OOG (21 CFR 888.3560)
ManufacturerMaterialise NVMaterialise USA, LLC

Device Description

The subject 'device Zimmer® Patient Specific Instruments System 5.4 is a modification to the predicate device Zimmer® Patient Specific Instruments System 5.0 (K121640) to accommodate the new compatible implant system Zimmer® Personalized Knee System Trabecular Metal™ components cleared via 510(k) K122745 and K121771. It is designed to assist a surgeon in the placement of total knee replacement components for Zimmer® Persono™ Trabecular Metal™ components. The system consists of a software device, branded as Zimmer® Patient Specific Instruments Planner (ZPSIP) and a hardware component, branded as Zimmer® Patient Specific Instruments (ZPSI).

Intended Use

Total Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer Patient Specific Instruments System is to be used with Zimmer NexGen CR-Flex fixed bearing, Zimmer NexGen CR fixed bearing, Zimmer NexGen LPS-Flex fixed bearing, Zimmer NexGen LPS fixed bearing, Zimmer Gender Solutions Natural – Knee Flex fixed bearing, Zimmer Persona™ CR fixed bearing, Zimmer Persona™ PS fixed bearing and Zimmer Persona™ Trabecular Metal™ prostheses families only.

Unicompartmental Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Unicompartmental Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer Patient Specific Instruments System is to be used with Zimmer Unicompartmental High Flex Knee System prostheses families only.

The Zimmer® Patient Specific Instruments are intended for single use only.

Functioning of the Device

The Zimmer® Patient Specific Instruments System 5.4 generates a pre-surgical plan based on MRI image data sets using the Zimmer Patient Specific Instruments Planner. The software device then is used pre-operatively by a qualified surgeon to inspect, fine-tune and approve the pre-surgical plan. Next, Zimmer Patient Specific Instruments are designed and manufactured based on the approved pre-surgical plan. Zimmer Patient Specific Instruments are patient specific templates which transfer the pre-operatively determined positioning of the chosen total knee replacement co the patient intra-operatively, assisting the surgeon in positioning and aligning the actual total knee replacement components by guiding and marking drill locations.

{2}------------------------------------------------

Zimmer Patient Specific Instruments System 5.4 510(k) Premarket Notification

510(k) Summary

Technological Characteristics

A detailed comparison shows the subject device is substantially equivalent in intended use, design, functionality, operating principles, materials and performance characteristics to the predicate devices.

Performance Data

Non-clinical tests using saw bones have been performed to assess the safety and effectiveness of the subject device. Testing verified that the accuracy and performance of the system is adequate to perform as intended. Results of software verification and validation testing demonstrated the device's safety and effectiveness is substantially equivalent to the predicate device.

Summary

The characteristics that determine the functionality and performance of the Zimmer Patient Specific Instruments System 5.4, are substantially equivalent to the device cleared under K121640. The Zimmer Patient Specific Instrument System will be manufactured in compliance with FDA and ISO quality system requirements.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an abstract eagle or bird-like figure with three curved lines representing wings or feathers.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 25, 2014

Materialise N.V. Mr. Oliver Clemens Quality and Regulatory Officer Technologielann 15 Leuven, Belgium 3001

Re: K133162

Trade/Device Name: Zimmer Patient Specific Instruments System 5.4 (Zimmer Patient Specific Instruments Planner, Zimmer Patient Specific Instruments) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II

Product Code: JWH, MBH, OOG, OIY Dated: January 23, 2014 Received: January 27, 2014

Dear Mr. Clemens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Oliver Clemens

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Vincent

  • for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K133162

Device Name: Zimmer Patient Specific Instruments System 5.4 (Zimmer Patient Specific Instruments Planner, Zimmer Patient Specific Instruments)

Indications for Use:

Total Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer® Patient Specific Instruments System is to be used with Zimmer NexGen CR-Flex fixed bearing, Zimmer NexGen CR fixed bearing, Zimmer NexGen LPS-Flex fixed bearing, Zimmer NexGen LPS fixed bearing, Zimmer Gender Solutions Natural – Knee Flex fixed bearing, Zimmer Persona™ CR fixed bearing, Zimmer Persono™ PS fixed bearing and Zimmer Persona™ Trabecular Metal™ prostheses families only.

Unicompartmental Knee Replacement

The Zimmer® Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Unicompartmental Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Zimmer® Patient Specific Instruments System is to be used with Zimmer Unicompartmental High Flex Knee System prostheses families only.

The Zimmer® Patient Specific Instruments are intended for single use only.

Page 1 of 2

{6}------------------------------------------------

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley Ph.D.
Division of Orthopedic Devices

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K133162