K Number
K133140

Validate with FDA (Live)

Date Cleared
2013-11-15

(29 days)

Product Code
Regulation Number
866.3480
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV antigen from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader.

Device Description

The BD Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in respiratory specimens. The patient specimen is mixed in a prefilled unitized tube containing RV Reagent C and added to the test device. RV Reagent C contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Flu A+B test device.

The specimen is mixed and added to the test device where influenza A or influenza B viral antigens bind to anti-influenza antibodies conjugated to detector particles on the BD Flu A+B test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of influenza A or B viral antigens.

The BD Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone addresses non-specific signal generation and is not labeled on the test device. The remaining zone is used to measure the assay background and is also not labeled.

The BD Flu A+B assay incorporates an active negative control feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eye is unable to accurately perform the subtraction of the nonspecific signal.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a modification to the BD Veritor™ System Flu A+B assay, specifically regarding the ability to test excess processed sample in the BD Veritor™ RSV clinical assay. This document does NOT contain a study that proves the device (BD Veritor™ System Flu A+B assay) meets acceptance criteria for its primary function. Instead, it focuses on demonstrating substantial equivalence for a modification related to cross-testing with another BD product (RSV assay).

Therefore, I cannot fulfill all your requests as the complete performance study data and acceptance criteria for the primary influenza detection function are not detailed in this submission. The document states that performance characteristics for the Flu A+B assay were established during specific influenza seasons, implying that studies were done, but the results and detailed acceptance criteria for those original studies are not provided here.

However, I can extract information related to the context of its performance and the type of information typically found in such submissions.

Here's a breakdown of what can be derived from the provided text, and where information is missing:


1. A table of acceptance criteria and the reported device performance

The provided document does not list specific acceptance criteria or detailed performance data for the BD Veritor™ System Flu A+B assay. It mentions that performance characteristics were "established" during certain influenza seasons but does not present the actual data (e.g., sensitivity, specificity, PPA, NPA values).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions that performance characteristics were established during:

  • January through March of 2011 for nasopharyngeal (NP) washes/aspirates.
  • January through April of 2012 for NP swabs in transport media.

It states these periods were when specific influenza viruses (A/2009 H1N1, A/H3N2, B/Victoria lineage, and B/Yamagata lineage) were predominant, according to CDC reports. This implies the data was collected prospectively during active influenza seasons in the United States.

The sample size for the test set is NOT specified in this document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is NOT specified in this document. The ground truth method is implied to be viral culture or an FDA-cleared molecular assay (see point 7), but details on the experts involved in interpreting results or establishing ground truth are absent.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is NOT specified in this document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is a diagnostic device, not an AI-assisted interpretation device. The BD Veritor™ System Reader uses a proprietary algorithm to interpret test results by subtracting non-specific signals, but it does not involve human readers in the interpretation process in the way an MRMC study for imaging interpretation would. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and not mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself (BD Veritor™ System Flu A+B) operates in a "standalone" fashion in terms of interpretation, meaning the BD Veritor™ System Reader interprets the test strip results algorithmically without human visual interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document states:

  • "A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay."
  • For the RSV assay (which the modification allows cross-testing with), it states: "It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay."

This strongly indicates that viral culture or FDA-cleared molecular assays were used as the ground truth.

8. The sample size for the training set

The document describes a modification to an existing device and does not provide information on the training set size for the original device's development or for the algorithmic interpretation by the BD Veritor™ System Reader.

9. How the ground truth for the training set was established

This information is NOT specified in this document. As it's a 510(k) submission for a modification, details on the original device's development, including training set ground truth establishment, are typically not included unless directly relevant to the modification.

{0}------------------------------------------------

Becton, Dickinson and Company BD Diagnostics 10865 Road to the Cure, Suite 200 San Diego, CA 92121 www.ho com (858) 812-8500 名 (858) 812-8505 €7

K133140

Image /page/0/Picture/2 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular emblem on the left and the letters "BD" on the right. The emblem features a stylized human figure with arms raised inside a circle with radiating lines above, resembling a sun. The letters "BD" are in a bold, sans-serif font.

Helping all people live healthy lives

NOV 1 5 2013

510(K) SUMMARY

SUBMITTED BY:

BECTON, DICKINSON AND COMPANY 10865 Road to the Cure, Suite 200 San Diego, CA 92121 Tel: (858) 795 7890 Fax: (858) 795 7885

CONTACT NAME: Gregory P. Payne, RAC, Director Regulatory Affairs

DATE PREPARED: September 26, 2013

BD Veritor™ System Flu A+B assay DEVICE TRADE NAME:

DEVICE COMMON NAME: Influenza virus serological reagents

DEVICE CLASSIFICATION: 21 CFR 866.3330

PREDICATE DEVICES : BD Veritor™ System for Rapid Detection of Flu A+B POC (K120049, K121797, K132256, K132692 )

INTENDED USE :

Clinical Kit

The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal wash, aspirate and swab in transport media samples from symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influenza A viral antigens can be distinquished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as

{1}------------------------------------------------

the sole basis for treatment or other patient management decisions. The test is not intended to detect influenza C antigens.

Performance characteristics for influenza A and B nasopharyngeal (NP) washes/aspirates were established during January through March of 2011 when influenza viruses A/2009 H1N1, A/H3N2, B/Victoria lineage, and B/Yamagata lineage were the predominant influenza viruses in circulation according to the Morbidity and Mortality Weekly Report from the CDC entitled "Update: Influenza Activity-United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine." Performance characteristics may vary against other emerging influenza viruses.

Performance characteristics for influenza A and B NP swabs in transport media were established during January through April of 2012 when influenza viruses A/2009 H1N1. A/H3N2, B/Victoria lineage, and B/Yamagata lineage were the predominant influenza viruses in circulation according to the Morbidity and Mortality Weekly Report from the CDC entitled "Update: Influenza Activity-United States, 2011-2012 Season, and Composition of the 2012-2013 Influenza Vaccine." Performance characteristics may vary against other emerging influenza viruses.

If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to the state or local health department for testing. Virus culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

DEVICE DESCRIPTION :

The BD Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in respiratory specimens. The patient specimen is mixed in a prefilled unitized tube containing RV Reagent C and added to the test device. RV Reagent C contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Flu A+B test device.

The specimen is mixed and added to the test device where influenza A or influenza B viral antigens bind to anti-influenza antibodies conjugated to detector particles on the BD Flu A+B test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of influenza A or B viral antigens.

The BD Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard

{2}------------------------------------------------

neqative control zone addresses non-specific signal generation and is not labeled on the test device. The remaining zone is used to measure the assay background and is also not labeled.

Image /page/2/Figure/1 description: The image shows a diagram of a test strip with labels on the left and a visual representation of the test strip on the right. The labels on the left indicate the different lines on the test strip, including "Positive Ctrl C", "Flu B test line L4", "Flu A test line L3", "Negative Ctrl L2", and "Background". The visual representation on the right shows a test strip with four distinct lines, corresponding to the labeled lines on the left.

The BD Flu A+B assav incorporates an active negative control feature in each test to identify and compensate for sample-related. nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eye is unable to accurately perform the subtraction of the nonspecific signal.

DEVICE COMPARISON:

The modified device differs from the currently marketed BD Veritor™ System Flu A+ B in the following way:

The RV Reagent C solution used in both the BD Veritor Flu A+B and RSV clinical kits is identical in terms of chemical composition and volume. The reagent volume supplied in the unitized reagent tubes is approximately 100 uL and the assay test procedure specifies a volume of 300 µL of the liguid sample be added to the tube. After mixing, three drops (~80-100 µL) of the processed sample is added to the test device. As there is sufficient processed sample remaining, if desired, the remainder can be tested in another device cleared for use whith the same sample processing reagent and procedure. If the operator receives a message indicating an invalid test from the BD Veritor™ System Reader, the user is directed to retest the processed sample. The labeling has been changed to reflect instructions to the user that the remaining sample may alternatively be tested in a BD Veritor RSV clinical kit if the sample was collected from someone meeting the criteria in the RSV intended use of less than 20 years of age.

SUBSTANTIAL EQUIVALENCE:

The modified device BD Veritor™ System Flu A+ B is substantially equivalent to the current legally marketed device, BD Veritor™ System Flu A+B clinical assay. Modifications are being made to the labeling to allow testing of remaining processed specimen in the BD Veritor™ RSV clinical assay. This modfication does not change the intended use of the device or the fundamental scientific technology.

{3}------------------------------------------------

Additions are as follows:

ChangePotential Impact of Change
RV Reagent C is identical in both the BDVeritor Flu A+B and RSV clinical kits.Instructions have been added to the BDVeritor™ Flu A+B clinical product insert toinform users that the excess processedsample may be tested using the BDVeritor™ RSV clinical kit if the samplewas collected from someone meeting thecriteria in the RSV intended use of lessthan 20 years of age.This allows users to test excess processedflu samples for the presence of RSV.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services USA. The logo consists of a symbol resembling a stylized human figure with three flowing lines extending from its arm, all enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WObb-Gh09 Silver Spring, MI) 20993-0002

November 15, 2013

BECTON. DICKINSON AND COMPANY GREGORY P. PAYNE. RAC DIRECTOR. REGULATORY AFFAIRS AND QUALITY SYSTEMS 10865 ROAD TO THE CURE. SUITE 200 SAN DIEGO CA 92121

Re: K133140

Trade/Device Name: BD Veritor™ System RSV clinical assay Regulation Number: 21 CFR 866.3480 Regulation Name: Respiratory syncytial virus serological reagents Regulatory Class: I Product Code: GQG Dated: September 26, 2013 Received: October 17. 2013

Dear Mr. Payne:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Payne

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sally AFEDivat-S

Sally Hojvat, M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: BD Veritor" System RSV Clinical Assay

Indications for Use:

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV antigen from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader.

Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of Center for Devices and Radiological Health (CDRH)

Tamara V: Feldblyum -S 2013.11.13 13:01:31 -05'00'

Page 1 of 1

§ 866.3480 Respiratory syncytial virus serological reagents.

(a)
Identification. Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.