(49 days)
The Flower Ankle Plating Set is intended for use for fixation of the ankle in adults and adolescents (12-21) in whom the growth plates have fused, and particularly in osteopenic bone. Specifically,
- . Distal Medial and Lateral Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
- . Distal Tibia A-Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
- . Straight and Distal Lateral Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.
The Flower Ankle Plating Set consists of the following components and accessories: distal medial tibia plates, distal lateral tibia plates, distal tibia A-plate, straight fibula plate, distal lateral fibula plate, all made of pure titanium compliant with ASTM F67. The system accepts locking and non-locking screws cleared via K123562 and K131657. The device is provided with general purpose instruments.
Here's an analysis of the provided text regarding the Flower Ankle Plating Set, focusing on the requested information about acceptance criteria and study details:
Summary of Acceptance Criteria and Device Performance (Based on provided text)
The document focuses on demonstrating substantial equivalence to predicate devices rather than setting specific numeric acceptance criteria for clinical performance. The "acceptance criteria" are implied by adherence to recognized standards and an engineering analysis confirming mechanical strength, which are used to show the device functions "as intended" and is equivalent to previously cleared devices.
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Compliance with ASTM F67 | Met |
| Compliance with ISO 5832-2 | Met |
| Compliance with ASTM F136 | Met |
| Compliance with ISO 5832-3 | Met |
| Compliance with ISO 7153-1 | Met |
| Compliance with ISO 10993-1 | Met |
| Compliance with ISO 10993-5 | Met |
| Compliance with ISO 11137-1 | Met |
| Appropriate Mechanical Strength (Engineering Analysis) | "functioned as intended" and "demonstrate substantial equivalence with the cited predicate devices." |
Details of the Study:
The provided document describes a 510(k) Premarket Notification, which generally does not require clinical studies to demonstrate safety and effectiveness for devices seeking substantial equivalence to existing predicate devices. Instead, it relies on performance testing (benchtop, materials, biocompatibility) and engineering analysis.
-
Sample sized used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. No clinical test set data is presented. The "test set" refers to the components of the Flower Ankle Plating Set and materials that underwent physical, chemical, and biological testing according to the listed standards.
- Data Provenance: Not applicable for clinical data. The performance data is derived from laboratory testing against recognized standards for materials (e.g., ASTM, ISO) and an engineering analysis. The location of these labs or the specific origin of these test results (e.g., specific country) is not detailed beyond the standards themselves being international.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. There was no clinical "ground truth" established by experts in the context of this 510(k) submission, as it was not a clinical study requiring such ground truth for diagnostic or prognostic performance. The "ground truth" for the material and mechanical tests is defined by the passing criteria within the specified ASTM and ISO standards.
-
Adjudication method for the test set:
- Not applicable. As there were no human readers or clinical expert interpretations involved in establishing ground truth for a test set, no adjudication method was used. The adjudication for the standard tests would be the pass/fail criteria of those specific test methodologies.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC comparative effectiveness study was not performed. This device is a surgical implant (ankle plating set), not an AI-powered diagnostic or assistive tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This device is a surgical implant. "Standalone algorithm performance" is not relevant to this type of medical device submission.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- The "ground truth" for this device's submission is based on adherence to recognized material and mechanical engineering standards (ASTM, ISO) and an engineering analysis demonstrating mechanical strength. The "ground truth" for safety and effectiveness is established by demonstrating substantial equivalence to predicate devices that have already been cleared.
-
The sample size for the training set:
- Not applicable. This device does not involve a "training set" in the context of an algorithm or AI.
-
How the ground truth for the training set was established:
- Not applicable, as no training set was used for an algorithm.
{0}------------------------------------------------
NOV 1 8 2013
Page 1 of 3
K133102
510(k) SUMMARY
Flower Orthopedics Corporation's Ankle Plating Set
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Flower Orthopedics Corporation 7715 Crittenden Street, #413 Philadelphia, PA 19118
Phone: (267) 437 3063 (267) 437 3072 Facsimile:
Contact Person: Oliver Burckhardt, Chief Executive Officer Date Prepared: September 30, 2013
Name of Device
Flower Ankle Plating Set
Common or Usual Name/Classification Name
The Flower Ankle Plating Set consists of bone ankle plates classified under product code HRS (21 C.F.R. 888.3030, Single/multiple component metallic bone fixation appliance and accessories: Class II) and bone fixation screws classified under product code HWC (21 C.F.R. 888.3040, Smooth or threaded metallic bone fixation fastener; Class II).
Predicate Devices
Synthes USA, Synthes 2.7mm/3.5mm LCP Distal Fibula Plates (K083213) Synthes USA, Synthes Variable Angle LCP Ankle Trauma System (K120854) Synthes USA, Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System-Anterolateral Distal Tibia Plates (K121601) Flower Orthopedics Corporation, Flower Small and Medium Implants Set (K123562 and K131657)
Intended Use / Indications for Use
The Flower Ankle Plating Set is intended for use for fixation of the ankle in adults and adolescents (12-21) in whom the growth plates have fused, and particularly in osteopenic bone. Specifically.
- . Distal Medial and Lateral Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
- . Distal Tibia A-Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
- . Straight and Distal Lateral Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.
{1}------------------------------------------------
Device Description
The Flower Ankle Plating Set consists of the following components and accessories: distal medial tibia plates, distal lateral tibia plates, distal tibia A-plate, straight fibula plate, distal lateral fibula plate, all made of pure titanium compliant with ASTM F67. The system accepts locking and non-locking screws cleared via K123562 and K131657. The device is provided with general purpose instruments.
Technological Characteristics
The Flower Ankle Plating Set consists of the following components/configurations:
- Distal Medial Tibia Plates with a width of 11mm and a length range of 107mm-� 179mm:
- Distal Lateral Tibia Plates with a width of 11mm and a length range or 81mm-. 141mm;
- . Distal Tibia A-Plate with a width of 11mm and length of 65mm;
- Straight Fibula Plates with a width of 11mm and a length range of 45mm-86mm; and .
- Distal Lateral Fibula Plates with a width of 11mm and length range of 84mm-180mm. .
Performance Data
The Flower Ankle Plating Set was tested (worse case) according to the following standards:
- . ASTM F67, Standard Specification for Unalloyed Titanium for Surgical Implant Applications
- . ISO 5832-2. Implants for Surgery, Metallic Materials. Part 2: Unalloyed Titanium
- . ASTM F136, Standard Specification for Wrought Titanium- 6 Aluminum- 4 Vanadium ELI Allov for Surgical Implant Applications
- ISO 5832-3, Implants for Surgery. Metallic materials. Part 3: Wrought titanium 6-. aluminium 4-vanadium allov:
- ISO 7153-1. Surgical instruments Metallic materials Part 1: Stainless steel (ISO . 7153-1:1991, including Amendment 1:1999); German version EN ISO 7153-1:2000;
- . ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process;
- . ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity;
- ISO 11137-1. Sterilization of health care products Radiation Part 1: . Requirements for development, validation and routine control of a sterilization process for medical devices. (Sterility)
In addition, an engineering analysis was performed to demonstrate that the subject ankie plates provide appropriate mechanical strength for the claimed intended use.
{2}------------------------------------------------
K133102
In all instances, the Flower Ankle Plating Set functioned as intended and test results, as well as an engineering analysis, demonstrate substantial equivalence with the cited predicate devices.
Substantial Equivalence
The Flower Ankle Plating Set is substantially equivalent to the identified predicate devices. The subject devices have the same intended uses/indications, technological characteristics, and principles of operation as the predicate devices. An engineering analysis was performed to demonstrate that the components in the Flower Ankle Plating Set provide appropriate mechanical strength for the claimed intended use. Thus, the subject ankle plates are substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Flower Orthopedics Corporation % Ms. Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103 November 18, 2013
Re: K133102
Trade/Device Name: Flower Ankle Plating Set Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: September 30, 2013 Received: September 30, 2013
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Janice M. Hogan
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark Nighterson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
K133102 510(k) Number (if known):
Device Name: Flower Ankle Plating Set
Indications for Use:
The Flower Ankle Plating Set is intended for use for fixation of the ankle in adults and adolescents (12-21) in whom the growth plates have fused, and particularly in osteopenic bone. Specifically,
- Distal Medial and Lateral Tibia Plates are intended for fixation of osteotomies, . fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
- Distal Tibia A-Plates are intended to buttress partial articular fractures and bone . fragments of the distal tibia, and
- . Straight and Distal Lateral Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth L. Frank -S
Division of Orthopedic Devices
Page 1 of 1
41
N/A