K Number
K133047

Validate with FDA (Live)

Date Cleared
2014-12-18

(447 days)

Product Code
Regulation Number
862.3610
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RapidFRET Oral Fluid Assay for MDMA is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Methylenedioxymethamphetamine at 50 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only.

The RapidFRET Oral Fluid MDMA Calibrator Set and RapidFRET Oral Fluid MDMA Control Set are intended for use only with the RapidFRET Oral Fluid Assay for MDMA and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.

Device Description

The RapidFRET Oral Fluid Assay for MDMA is an In Vitro Diagnostic competitive immunoassay used to detect MDMA in human oral fluid. This is a ready-to-use homogenous system that involves energy transfer between an acceptor fluorophore labeled to an antibody and a donor fluorophore labeled to drug. The assay is based on competition between drug in the sample and drug labeled with the donor fluorophore for a fixed number of binding sites on the antibody reagent. When acceptor and donor fluorophores are brought into close proximity through a binding event, energy transfer occurs. The fluorescence resonance energy transfer (FRET) signal is measured at the wavelength of the acceptor fluorophore and is inversely proportional to the amount of drug in the sample. A Cutoff Calibrator is used to translate the sample measurement into a positive result. Controls are used to establish and monitor precision and accuracy.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about the RapidFRET Oral Fluid Assay for MDMA:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the expected results and agreement percentages reported.

Acceptance Criteria (Implied)Reported Device Performance
Precision & Analytical Sensitivity:
0% MDMA: Expect 100% NEG0% MDMA: 100% NEG (279/279)
25% MDMA: Expect 100% NEG25% MDMA: 100% NEG (279/279)
50% MDMA: Expect 100% NEG50% MDMA: 100% NEG (278/278)
75% MDMA: Expect 100% NEG75% MDMA: 100% NEG (279/279)
100% MDMA: Expect near 100% POS100% MDMA: 95% POS (266/279)
125% MDMA: Expect 100% POS125% MDMA: 100% POS (278/278)
150% MDMA: Expect 100% POS150% MDMA: 100% POS (263/263)
175% MDMA: Expect 100% POS175% MDMA: 100% POS (294/294)
200% MDMA: Expect 100% POS200% MDMA: 100% POS (278/278)
Correlation with MS Quantitation:
Agreement with MS for MS POS samples: High (ideally 100%)100% Agreement (RapidFRET POS when MS POS) (119/119)
Agreement with MS for MS NEG samples: High (ideally 100%)97% Agreement (RapidFRET NEG when MS NEG) (200/206*, where 206 = 200 NEG + 6+ POS)
Overall Accuracy (Implied)99% accuracy overall (for the correlation study with MS)
Cross-Reactivity & Analytical Specificity:
No unexpected cross-reactivity at "normal" physiological levelsMany compounds cross-reacted, with 12 doing so at 1000 ng/mL or less. This study identified what cross-reacts, rather than stating an acceptance criterion it achieved. However, for the second study of common substances, all listed items gave expected results (NEG with 25 ng/mL MDMA, POS with 75 ng/mL MDMA).

*Note: The 97% for MS NEG agreement is calculated from the provided table: 200 RapidFRET NEG when MS NEG, and 6+ RapidFRET POS when MS NEG (false positives). So, 200 / (200 + 6) = 200/206 ≈ 97%.

2. Sample Sizes and Data Provenance

  • Precision and Analytical Sensitivity Test Set:
    • Sample Size: 279 at 0%, 279 at 25%, 278 at 50%, 279 at 75%, 279 at 100%, 278 at 125%, 263 at 150%, 294 at 175%, 278 at 200% of cutoff. These were spiked oral fluid pools.
    • Data Provenance: Not explicitly stated, but likely laboratory-prepared samples. It doesn't specify country of origin or if it's retrospective/prospective.
  • Correlation with MS Quantitation Test Set:
    • Sample Size: 325 neat oral fluid samples.
    • Data Provenance: Collected from "volunteers potentially positive and negative for MDMA." This suggests prospective collection for the purpose of the study. Country of origin not specified, but the applicant's address is in California, USA, making it probable the study was conducted there.
  • Cross Reactivity and Analytical Specificity Test Set:
    • Sample Size:
      • Structurally related compounds: A library of over 170 compounds. Specific numbers of samples per compound are not given, but they were spiked into neat oral fluid aliquots (likely 0, 25, 75 ng/mL MDMA).
      • Common substances: Not specified per substance, but implied to be multiple (volunteers for some, spiked aliquots for others).
    • Data Provenance: Laboratory-prepared samples (spiked) and samples from volunteers.

3. Number of Experts and Qualifications for Ground Truth

  • Precision and Analytical Sensitivity: No human experts were explicitly mentioned for ground truth. The ground truth was established by the known concentrations of MDMA spiked into the oral fluid pools.
  • Correlation with MS Quantitation: The ground truth was established by GC/MS or LC/MS/MS results. No human experts are mentioned for interpreting these confirmatory tests, as they are analytical methods providing quantitative results.
  • Cross Reactivity and Analytical Specificity: No human experts were explicitly mentioned. The ground truth was based on the known concentrations of MDMA and cross-reactants spiked into the samples, or the known consumption of common substances.

4. Adjudication Method for the Test Set

  • Precision and Analytical Sensitivity: Not applicable, as samples were spiked at known concentrations.
  • Correlation with MS Quantitation: Not applicable. GC/MS or LC/MS/MS provides a definitive analytical result, not a judgment requiring adjudication.
  • Cross Reactivity and Analytical Specificity: Not applicable. Ground truth was based on known concentrations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic assay, not a medical imaging or interpretive device that typically involves human readers in this context. It performs a qualitative screen (positive/negative) based on a cut-off.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

  • Yes, the studies presented are all standalone performance studies. The RapidFRET Oral Fluid Assay for MDMA is an automated laboratory assay that performs a qualitative screen without human interpretation involved in the direct result generation. The performance metrics (precision, analytical sensitivity, correlation with MS) directly reflect the algorithm's performance.

7. Type of Ground Truth Used

  • Precision and Analytical Sensitivity: Known concentrations of spiked MDMA (analytical truth).
  • Correlation with MS Quantitation: Confirmed analytical results by GC/MS or LC/MS/MS (gold standard analytical truth).
  • Cross Reactivity and Analytical Specificity: Known concentrations of spiked compounds (analytical truth) and known consumption of common substances (experiential/known truth).

8. Sample Size for the Training Set

  • The document does not specify a separate training set or its sample size. This type of immunoassay device is typically developed through chemical and biological experimentation to establish reagent formulations and cutoff values, rather than through machine learning training on a large dataset in the way a modern AI algorithm would be. The experiments described appear to be validation/verification studies.

9. How Ground Truth for the Training Set Was Established

  • As no training set is explicitly mentioned, the establishment of ground truth for it is not described. The device's operational parameters (e.g., cutoff) would be derived from laboratory experiments during development.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

BIOPHOR DIAGNOSTICS, INC. NATHANIEL BUTLIN VICE PRESIDENT 1201 DOUGLAS AVE REDWOOD CITY CA 94063

December 18, 2014

Re: K133047

Trade/Device Name: RapidFRET Oral Fluid Assay for MDMA RapidFRET Oral Fluid MDMA Calibrator Set RapidFRET Oral Fluid MDMA Control Set Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: II Product Code: LAF, DLJ, DIF Dated: November 18, 2014 Received: November 19, 2014

Dear Dr. Nathaniel Butlin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director

Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K133047

Device Name

RapidFRET Oral Fluid Assay for MDMA RapdFRET Oral Fluid MDMA Calibrator Set RapidFRET Oral Fluid MDMA Control Set

Indications for Use (Describe)

The RapidFRET Oral Fluid Assay for MDMA is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Methylenedioxymethamphetamine at 50 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only.

The RapidFRET Oral Fluid MDMA Calibrator Set and RapidFRET Oral Fluid MDMA Control Set are intended for use only with the RapidFRET Oral Fluid Assay for MDMA and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for the RapidFRET Oral Fluid Assay for MDMA

December 15, 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is: K133047

807.92(a)(1): Contact Information

Name:Biophor Diagnostics, Inc.
Address:1201 Douglas AvenueRedwood City, CA 94063
Contact:Nathaniel G. Butlin, Ph.D.
Phone:650-367-4954
Fax:650-364-4985

807.92(a)(2): Device Name, Common Name and Classification

RapidFRET Oral Fluid Assay for MDMA (Enzyme Immunoassay for Methylenedioxymethamphetamine) RapidFRET Oral Fluid MDMA Calibrator Set (Clinical Toxicology Calibrator) RapidFRET Oral Fluid MDMA Control Set (Drug Mixture Control Materials)

ProductCodeClassRegulationSectionPanel
RapidFRET Oral Fluid Assay for MDMALAFII862.361091 - Toxicology
RapidFRET Oral Fluid MDMA Calibrator SetDLJII862.320091 - Toxicology
RapidFRET Oral Fluid MDMA Control SetDIFI862.328091 - Toxicology

807.92(a)(3): Identification of Legally Marketed Predicate Devices

CEDIA Methamphetamine OFT Assay and CEDIA Methamphetamine OFT Calibrators (K101753) Salivabuse Liquid Oral Fluid Control (K132688)

807.92(a)(4): Device Description

The RapidFRET Oral Fluid Assay for MDMA is an In Vitro Diagnostic competitive immunoassay used to detect MDMA in human oral fluid. This is a ready-to-use homogenous system that involves energy transfer between an acceptor fluorophore labeled to an antibody and a donor fluorophore labeled to drug. The assay is based on competition between drug in the sample and drug labeled with the donor fluorophore for a fixed number of binding sites on the antibody reagent. When acceptor and donor fluorophores are brought into close proximity through a binding event, energy transfer occurs. The fluorescence resonance energy transfer (FRET) signal is measured at the wavelength of the acceptor fluorophore and is inversely proportional to the amount of drug in the sample. A Cutoff

{4}------------------------------------------------

Calibrator is used to translate the sample measurement into a positive result. Controls are used to establish and monitor precision and accuracy.

807.92(a)(5): Intended Use

The RapidFRET Oral Fluid Assay for MDMA is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Methylenedioxymethamphetamine at 50 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only.

The RapidFRET Oral Fluid MDMA Calibrator Set and RapidFRET Oral Fluid MDMA Control Set are intended for use only with the RapidFRET Oral Fluid Assay for MDMA and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.

Similarities
ItemPredicateDevice
CEDIA Methamphetamine OFT assayand CEDIA Methamphetamine OFTCalibratorsK101753RapidFRET Oral Fluid Assay forMDMA and RapidFRET OralFluid MDMA Calibrator SetK133047
Intended UseAssay: Qualitative detection ofmethamphetamine based analyte inneat oral fluid.Calibrators: Calibration of d-methamphetamine for assay systemSame except analyte is MDMA
MethodologyHomogenous competitiveimmunoassaySame
Kit Components1 specific antibody reagent and 1drug conjugate reagentSame

807.92(a)(6): Technological Similarities and Differences to the Predicate

{5}------------------------------------------------

Differences
PredicateDevice
CEDIA Methamphetamine OFTRapidFRET Oral Fluid Assay for
ltemassay and CEDIAMDMA and RapidFRET Oral Fluid
Methamphetamine OFTMDMA Calibrator Set
CalibratorsK133047
K101753
AnalyteMethamphetamineMDMA
Oral fluid is collected with theNeat oral fluid is collected with
Oral-Eze Saliva Collection System.the RapidEASE Oral Fluid Collector
Sample CollectionThis device uses an absorbentvia direct expectoration. No
swab and diluent. Sample isdiluent is used and sample is
stored in plastic tube with snapstored in glass sample tube with
сар.inert screw cap.
Sample dilutionYesNo
120 ng/mL neat oral fluid using a
Neat Oral FluidCutoff40 ng/mL cutoff calibrator to50 ng/mL neat oral fluid.
account for sample dilution by
collection device.
PlatformMGC 240 AnalyzerRapidFRET Integrated Workstation
from Biophor Diagnostics
Reagent FormatLyophilized reagent withLiquid, ready to use
reconstitution buffer
Calibrator LevelsCalibrators at 0, 40, and 200Calibrators at 0 and 50 ng/mL.
ng/mL
Similarities and Differences
ItemPredicateSalivabuse Liquid Oral Fluid ControlK132688DeviceRapidFRET Oral Fluid MDMAControl SetK133047
Intended UseQuality control material for oral fluiddrugs of abuse assays to monitorsystem performanceSame
Control LevelsNegative, -60% cutoff, ±50% cutoff, -30% cutoff, ±25% cutoff, 2X cutoff,and 3x cutoff25 ng/mL and 75 ng/mL
AnalyteMultiple drugs of abuseMDMA
StabilityShelf: 12 months frozen orrefrigeratedOpen: 31 days refrigeratedShelf: 12 months refrigeratedOpen: 30 days refrigerated

807.92(b)(1): Brief Description of Study Data:

A series of studies were performed that evaluated the device performance characteristics including precision and analytical sensitivity, correlation with GC/MS and LC/MS/MS, cross reactivity, and

{6}------------------------------------------------

analytical specificity that are summarized below.

Precision and Analytical Sensitivity

Three lots of the RapidFRET Oral Fluid Assay for MDMA were analyzed, four times daily, for a minimum of 20 days. Negative oral fluid pools were spiked with MDMA at 0%, 25%, 50%, 75%, 100%, 125%, 150%, 175% and 200% of the cutoff level corresponding to approximately 0, 12.5, 25, 37.5, 50, 62.5, 75, 87.5 and 100 ng/mL. The aggregate data is summarized in the table below:

0%25%50%75%100%125%150%175%200%
POSO266278263294278
NEG279279278279130
z279279278279279278263294278
0%25%50%75%100%125%150%175%200%
POS0%0%0%0%95%100%100%100%100%
NEG100%100%100%100%5%0%0%0%0%
N279279278279279278263294278

The data indicate that the analytical sensitivity is between 75% and 125% of cutoff, and expected results were achieved at a 100% frequency.

Correlation with MS Quantitation

Neat oral fluid was collected with the RapidEASE Oral Fluid Collection Device from volunteers potentially positive and negative for MDMA. The samples (n=325) were randomized and blinded to the instrument operator and assayed using RapidFRET MDMA reagents. Following screening, positive and negative samples were sent for confirmatory testing. The summarized data are shown below.

n = 325MS POSMS NEG
RapidFRET POS1196+
RapidFRET NEG0200
% Agreement100%97%

Samples contained compounds at concentrations that are known to cross react including d-methamphetamine, damphetamine, I-amphetamine, 4-Methylethcathionone (4-MEC) and Methylone.

The data indicate that the RapidFRET Oral Fluid Assay for MDMA was accurate 99% of the time in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector.

Cross Reactivity and Analytical Specificity

A compound library of more than 170 different structurally related and unrelated compounds including metabolites, OTC and prescription medications and drugs of abuse was used to evaluate the device cross reactivity and specificity. Compounds were spiked at

30,000 ng/mL into neat oral fluid pool aliquots with 0 ng/mL, 25 ng/mL and 75 ng/mL MDMA, processed with the RapidEASE Collector, and tested with the RapidFRET MDMA assay. Those compounds that gave an unexpected result were further titrated to determine the concentration at which the cross-reacting compound yielded a result approximately equivalent to the cutoff. Twenty

{7}------------------------------------------------

nine structurally related compounds were determined to cross-react below 30,000 ng/mL in the absence of MDMA with twelve cross-reacting at 1000 ng/mL or less. An additional four compounds were determined to cross react below 30,000 ng/mL in the presence of 25 ng/mL MDMA.

CompoundLevel (ng/mL)0% MDMA†(0 ng/mL)50% MDMA†(25 ng/mL)150% MDMA†(75 ng/mL)
Structurally Related Compounds That Cross React in Neat Oral Fluid Pool with 0 ng/mL MDMA
(-) Ephedrine30,0009000 [0.6%]POSPOS
(R, 2R) Pseudoephedrine30,0002800 [1.8%]POSPOS
Amitriptyline30,0001900 [2.6%]POSPOS
Benzodioxolylbutanamine30,000390 [13%]POSPOS
Phenethylamine30,00015,500 [0.3%]POSPOS
d-Amphetamine30,00016,900 [0.2%]POSPOS
d-Ephedrine30,0004800 [1.0%]POSPOS
d,l-Amphetamine9,0006,500 [0.7%]ND1ND1
d,l-Ephedrine66,00011,000 [0.5%]ND1ND1
d-Methamphetamine30,0001200 [4.2%]POSPOS
Doxepin30,00011,000 [0.5%]POSPOS
Fenfluramine30,00037 [135%]POSPOS
Imipramine30,00026,000 [0.2%]POSPOS
Isoxsuprine30,000100 [50%]POSPOS
I-Amphetamine30,0002100 [2.4%]POSPOS
I-Methamphetamine30,00090 [56%]POSPOS
I-Phenylephrine30,00016,600 [0.3%]POSPOS
MBDB30,00042 [119%]POSPOS
MDA30,000130 [38%]POSPOS
MDEA30,00040 [125%]POSPOS
4-Methylethcathinone (4-MEC)30,0007,839 [0.6%]ND1ND1
Mephentermine30,000120 [42%]POSPOS
Methylone30,0003,783 [1.3%]ND1ND1
Nortriptyline30,00023,100 [0.2%]POSPOS
Phentermine30,0001000 [5%]POSPOS
PMA30,000350 [14%]POSPOS
PMMA30,00039 [128%]POSPOS
Dihydrobupropion30,000460 [10.9%]POSPOS
Verapamil60,0001,800 [2.7%]ND1ND1
Structurally Related Compounds That Cross React in Neat Oral Fluid Pool with 25 ng/mL MDMA
(+/-) Pseudoephedrine30,000NEG>20,000 [< 0.3%]POS
Cyclobenzaprine30,000NEG>10,000 [< 0.5%]POS
Fentanyl30,000NEG4,000 [1.3%]POS
Buproprion30,000NEG>20,000 [< 0.3%]POS

{8}------------------------------------------------

(1) ND – Not Determined. †Results are presented as either the RapidFRET MDMA screening result (POS / NEG) or the concentration in ng/mL of the cross-reactant that gives a Cutoff equivalent response.

A second study evaluated common substances such as foods and dental products as well as pH variations. HSA, ethanol, baking soda, whole blood, hemoglobin, hydrogen peroxide, sodium chloride, cholesterol, denture adhesive, ascorbic acid, bilirubin, IgA, IgG and IgM were spiked into neat oral fluid pool aliquots that contained either 25 ng/mL of MDMA. Neat oral fluid pool was titrated to pH values of 5, 6, 7, 8 and 9, spiked with MDMA to 25 ng/mL or 75 ng/mL and assayed with the RapidFRET MDMA Assay. The effects of antiseptic mouthwash, cough syrup, cranberry juice, orange juice, tooth paste, chewing tobacco, cigarettes, chewing gum, hard candy, teeth whitening strips, cola, water, antacid, coffee and tea were evaluated by asking volunteers to use a specific item and provide an oral fluid sample. These samples were then spiked with MDMA to 25 ng/mL or 75 ng/mL, processed with a RapidEASE Collector and assayed with the RapidFRET MDMA device. All compounds at the listed concentrations gave a NEG result when spiked with 25 ng/mL MDMA and a POS result when spike with 75 ng/mL MDMA.

807.92(b)(3): Conclusions

The RapidFRET Oral Fluid Assay for MDMA including the RapidFRET Oral Fluid Negative and Cutoff MDMA Calibrators, the RapidFRET Oral Fluid Negative and Positive MDMA Controls and the RapidEASE Oral Fluid Collector were determined to be safe and effective for their intended use.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).