K Number
K133033

Validate with FDA (Live)

Device Name
HIDREX PSP1000
Manufacturer
Date Cleared
2015-04-08

(559 days)

Product Code
Regulation Number
890.5525
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Tap-Water-Iontophoresis device is intended to treat hyperhidrosis (pathological sweating) affecting hands, feet, and underarms. Any other use or usage beyond this scope is considered use and may have dangerous consequences.

Device Description

HIDREX iontophoresis devices are primarily intended for treating hyperhidrosis (excessive sweating) of hands and/or feet. Provided the optional axillary applicators are utilized, the system can also be used for treating axillary hyperhidrosis. The system consists of a control unit, power supply/safety wall adapter, and application accessories (treatment cover towel, treatment electrode, axillary sponge cushions). The control unit includes connection-jackets for accessories, control-buttons to choose the therapy-mode, a main display, and LED indicators. The device offers both DC and pulsed current output with adjustable pulse-width.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Hidrex PSP1000 iontophoresis device, claiming substantial equivalence to two predicate devices: MD-1A (K964208) and Drionic (K831320). The document focuses on comparing the Hidrex PSP1000 with these predicates to establish that its differences do not raise new questions of safety or effectiveness.

Here's an analysis of the requested information based on the provided text:

1. Table of acceptance criteria and reported device performance

The document does not explicitly present a table of "acceptance criteria" in the format of specific thresholds for performance metrics. Instead, it compares the technical specifications and features of the Hidrex PSP1000 with those of its predicate devices to argue for substantial equivalence. The "performance" is described in terms of technical capabilities rather than clinical outcomes.

Here's a summary of the Hidrex PSP1000's technical performance metrics as reported:

Performance MetricReported Device Performance (Hidrex PSP1000)
DC current outputmax. 60 V, max. 35 mA (automatically regulated)
Pulsed current outputmax. 60 V, max. 35 mA, 9.9 kHz, 5 output modes (pulse-width: 50%, 60%, 70%, 80%, 90%) (automatically regulated)
Polarity reversalManual
Automatic current regulationYes
TimerYes
DisplayLCD-multi character display
Microprocessor controlledYes (with built-in self-test)
ControlsSoft sensor buttons
MeterDigital Monitor
Signal typeMonophasis square, at DC: pulsed square signal, selectable in increments of 10 from 50% to 100%
Patient leakage current (AC)+000.9 μΑ
Patient leakage current (AC SFC)+000.4 μΑ
Patient leakage current (DC)+000.0 μΑ
Patient leakage current (NAT)+004.2 μΑ
Energy output to patient (10 minutes)4.704 Ws
Energy output to patient (15 minutes)7.056 Ws

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document makes no mention of a "test set" for a clinical study to prove the device meets acceptance criteria related to efficacy or clinical outcomes. The studies mentioned are "non-clinical tests" to demonstrate electrical safety, software validation, performance, electromagnetic compatibility, and compliance with recognized standards (IEC 60601-1, IEC 60601-1-2). These are engineering and safety tests, not clinical performance studies with a sample size of patients.

The document mentions "Extensive trials showed that the efficiency of pulsed current could be increased dramatically" and "The additional benefit of the pulsed-current is demonstrated in several clinical trials." However, no details about these clinical trials (sample size, design, provenance) are provided within this 510(k) summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The document discusses non-clinical engineering and safety tests, not clinical studies requiring expert-established ground truth for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set for performance or efficacy is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The device is a medical device for treating hyperhidrosis, not an AI-assisted diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The Hidrex PSP1000 is a physical medical device (iontophoresis device), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests described (electrical safety, software validation, EMC), the "ground truth" would be established by compliance with international standards (IEC 60601-1, IEC 60601-1-2) and internal specifications. There is no mention of clinical ground truth (like pathology or outcomes data) for evaluation as this is a 510(k) submission focused on substantial equivalence rather than de novo clinical efficacy.

8. The sample size for the training set

Not applicable. The device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles a stylized caduceus or a symbol representing health and medicine. The emblem features a series of lines or strokes that form a shape reminiscent of a bird or a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 8, 2015

Hidrex Gmbh % Stefanie Bankston Official Correspondent 2611 Shark Circle Texas City, TX 77591

Re:

K133033 Trade/Device Name: Hidrex PSP1000 Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis Device Regulatory Class: Class III Product Code: EGJ Dated: February 23, 2015 Received: February 27, 2015

Dear Ms. Bankston,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your device for use with

{1}------------------------------------------------

specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact:

Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857

As you are aware, there are concerns relating to the fact that no drug is currently labeled for administration via an iontophoresis device. The Agency currently is evaluating this public health concern regarding the safety and effectiveness of this route of administration of drugs, and in the near future will inform manufacturers of certain additional steps the Agency believes are necessary to address this concern.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Felipe Aquel -S

  • Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K133033

Device Name HIDREX PSP1000

Indications for Use (Describe)

This Tap-Water-Iontophoresis device is intended to treat hyperhidrosis (pathological sweating) affecting hands, feet, and underarms. Any other use or usage beyond this scope is considered use and may have dangerous consequences.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for HIDREX. The logo is in blue and orange. To the right of the HIDREX logo are three additional logos: MED, VET, and REHA.

K133033

510(k) SUMMARY

July 24, 2014

Office of Device Evaluation U.S. Food & Drug Administration

Dear Madame/Sir;

In accordance with Section 510(k) of the Federal Food & Drug and Cosmetic Act, and in conformance with 21 CFR Part 807, pre-market notification is hereby made of the intention of HIDREX GmbH to introduce into interstate commercial distribution the iontophoresis device HIDREX PSP1000.

Applicant: HIDREX GmbH Otto-Hahn-Str. 12 D-42579 Heiligenhaus Phone_: +49 (0)2056 98 11 0 Fax +49 (0)2056 98 11 31 Email: info@hidrex.de

User fee Organization Number:348050

Contact Person:Mr. Andreas KaemperPhone: +49 (0)2056 98 11 0Email: kaemper@hidrex.de
US-agent:Mrs. Stefanie D. BankstonBEO MedConsulting Berlin GmbH3001 Ferndale DrLeague City TX 77573Phone: 713-483 46 17Email: s.bankston@beoberlin.com

Device:

  • Proprietary: HIDREX PSP1000 a.
  • b. Common Name: Iontophoresis device, Other uses
  • Regulation description: lontophoresis device C.
  • ರ. Device Class: III
  • Regulation Number: 890.5525 e.
  • Review Panel: Physical Medicine f.
  • இ. Product Code: EGJ

Predicate Device Information:

We claim substantial equivalence for the subject device in intended use, design and function to the predicate device MD-1A (K964208) by R.A. Fischer CO. CORP and to the predicate device Drionic (K831320) by General Medical Company.

Intended Use / Indication for Use:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for HIDREX. The word "HIDREX" is written in blue, except for the "I", which is orange and has an orange circle below it. To the right of the main logo are three smaller logos, stacked vertically, with the words "MED", "VET", and "REHA" in blue and green, each preceded by a small orange graphic.

This Tap-Water-lontophoresis device is intended to treat hyperhidrosis (pathological sweating) affecting hands, feet and underarms.

Any other use or usage bevond this scope is considered un-intended use and may have dangerous consequences.

Scientific Concept:

lontophoresis is a physical process in which ions flow diffusively in a medium driven by an electric field. Iontophoresis is to be distinguished from the carriage of uncharged molecules by diffusive fluxes of other molecules, especially of solvent molecules.

ln the treatment of hyperhidrosis, tap water is often the chosen solution for mild and medium forms.

Since 1968, tap water iontophoresis has been employed as the method of choice for treating palmoplantar hyperhidrosis. Special electrodes also allow treatment of axillary hyperhidrosis. Tap water iontophoresis also can extend symptom -free intervals in dyshidrotic palmar eczema. The mechanism action is most likely a functional disturbance of the secretory mechanism of eccrine acini. During the induction phase, treatments are carried out once daily. Current direction may be switched before each treatment or, even better, kept constant until one side, preferably the dominant hand on the anode, is sweating normally. Then polarity is switched until both hands are adequately treated. During the weekly maintenance therapy, current direction is switched before each treatment. The most comfortable means of iontophoretic treatment employs pulsed direct current of high frequency (5-10 kHz) which is better tolerated than continuous direct current. Side effects are minimal and transient. Only slight skin irritation or sensations of discomfort may occur during treatment. Electric burns and shocks can be avoided by following routine precautions. Contraindications for tap water iontophoresis are metallic implants, such as cardiac pacemakers, or orthopaedic joint or bone implants, if they are within the electric circuit.

Device Description:

HIDREX iontophoresis devices are primarily intended for treating hyperhidrosis1 (excessive sweating) of hands and/or feet. Provided the optional axillary applicators are utilized, the system can also be used for treating axillary hyperhidrosis.

Control unit:The control unit is connected with thepower supply/safety wall adapter. Theunit includes all connection-jackets toattach the accessories for thetreatment. It includes the control-buttons to choose the therapy-mode.Dimension: 190 x 137 x 49 mmImage: Control unit
Power supply / safety wall adapterthat is connected with the control unit(secondary plug). The primary plug isconnected to the AC-socket in thewall.Image: Power supply

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for HIDREX, a company that specializes in medical, veterinary, and rehabilitation equipment. The logo is composed of the word "HIDREX" in blue, with the "I" in orange and a circle below it. To the right of the main logo are the words "MED", "VET", and "REHA", each with a small orange graphic to the left.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for HIDREX. The logo is split into two parts. The first part shows the word "HIDREX" in blue, except for the "I", which is orange and has an orange dot below it. The second part shows three words stacked on top of each other: "MED", "VET", and "REHA", each preceded by an orange crescent shape with a small circle inside. The words "MED" and "REHA" are in blue, while "VET" is in green.

Image /page/6/Figure/1 description: The image shows two medical devices for hand and feet treatment. The first device is a treatment cover towel, which is an applicator for hand and feet treatment. The towel covers the treatment electrodes and has to be watered with tap water before the treatment starts, and its dimensions are 8.1" x 12.2". The second device is a treatment electrode, which is an accessory for hand and feet treatment, and its dimensions are 4.5" x 11.2 x 2.4".

Control unit

Image /page/6/Figure/3 description: The image shows two diagrams of a device with labels. The first diagram shows the side of the device with the therapy system's main ON/OFF switch labeled as '1'. The second diagram shows the back of the device with a 12V DC connector labeled as '3' and jacks for connecting the dual connecting cable set labeled as '2'.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for HIDREX, which is written in blue with an orange exclamation point. To the right of the logo are three smaller logos, MED, VET, and REHA. Each of these logos has an orange semi-circle with a small circle above the letters, which are blue for MED and REHA, and green for VET.

Image /page/7/Figure/1 description: The image shows a control panel for a device labeled HIDREX PSP1000 MED. The panel features a main display, several LED indicators for user settings, amperes, pulsed current, and dive mode. There are also buttons for adjusting time and dose, with labels indicating '+' and '-' functions for each. The panel's layout includes indicators and buttons arranged in a user-friendly manner.

1 DisplayMain Display to show therapeutic settings
2 ButtonAdjustment to lower the dose
3 ButtonAdjustment to raise the dose
4 ButtonAdjustment to decrease the therapeutic time-span
5 ButtonAdjustment to increase the therapeutic time-span
6 LED-Indicator3 Indicators to show user settings / application mode
7 LED Indicator1 Indicator to show therapeutic mode activated: current
8 LED Indicator1 Indicator to show therapeutic mode activated: pulsed current
9 LED Indicator1 Indicator to show 'ready for treatment' (electric circuit closed)

Current Density of Applicators and Electrodes

ApplicatorConductiveArea[sqin]Max.Densityat 30mA[mA/sqin]PeakDensityat 35mA[mA/sqin]
Hand-Feet, Hard Shell Caseindividual Water Surface112.130.260.31
Hand-Feet, Ergonomic Trayindividual Water Surface90.880.330.38
Hand-Feet, individual Electrode122.130.240.28
Hand-Feet, individual Cover Towel98.80.300.35
Axillary, Sponge Cushions36.750.810.95
Axillary, individual Electrode8.253.634.24

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX replaced by an orange exclamation point. To the right of the HIDREX logo are three additional logos, stacked vertically. The top logo is MED, the middle logo is VET, and the bottom logo is REHA. Each of these logos has an orange semi-circle to the left of the text.

Function:

The HIDREX treatment concept comprises two treatment phases:

Phase 1: The initial phase (therapy initiation) is conducted under a doctor's supervision with the doctor's or the patient's device. During this stage, patients learn to administer treatments. For therapy initiation, three weekly treatments of approximately 15 minutes each should be scheduled (not more than one treatment per day). Sweat secretion will normalize after approximately 10 treatments.

Phase 2: Long term treatment (maintenance therapy) is necessary because the HIDREX treatment effect is reversible. Patients should conduct maintenance therapy sessions by themselves at home and with their own unit. Depending on the severity of the condition, maintenance therapy involves one to three weekly sessions of approximately 15 minutes each.

The Hidrex PSP1000 with changeable pulse-width allows the innovative adjustment of the pulse-intensity.

Extensive trials showed that the efficiency of pulsed current could be increased dramatically. The effectiveness is not reduced at all compared to direct current when the pulse-width is 90%. The possible pulse-widths are 50%, 60%,70%, 80%, 90% and 100%

Application:

Image /page/8/Picture/8 description: The image shows two different scenarios of iontophoresis treatment. In the first scenario, a person is placing their feet on two orange pads that are connected to a blue device with wires. The device displays the number 15.30. In the second scenario, a person is placing their hands on the same type of pads, which are also connected to the same blue device.

feet

hands

Performance data:

DC current output(automaticallyregulated)max. 60 Vmax. 35 mA
Pulsed currentoutput(automaticallyregulated)max. 60 Vmax. 35 mA9.9 kHz

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX being an orange exclamation point. To the right of the HIDREX logo are three additional logos. The first is MED, the second is VET, and the third is REHA. Each of these logos has an orange semi-circle to the left of the text.

Comparison to legally marketed device MD-1A (K964208) by R.A. Fischer CO. CORP (Substantial Equivalence):

The Chart below summarizes the similarities and differences:
DEVICEHIDREX PSP1000SE-DEVICEMD-1A (K964208)
Indication for useThis Tap-Water-lontophoresisdevice is intended to treathyperhidrosis (pathologicalsweating) affecting hands, feet andunderarms.The MD-1a / MD-2 Galvanic unitsare indicated for use in the treatmentof palmar or plantar hyperhidrosis(abnormal sweating not related toexercise or resulting from anotherunderlying condition) using thetechnique of tapwater iontophoresis.
Performance
DC current outputmax. 60 Vmax. 35 mAdata not available0-10 and 0-50 mA
Pulsed currentoutputmax. 60 Vmax. 35 mA9.9 kHz5 output modes (pulse-width:50%, 60%, 70%, 80% and 90%)--
Polarity reversal (toalternate POS andNEG application)manualmanual
Automatic currentregulationYesOnly over-current protection
TimerYes--
DisplayLCD-multi character display todisplay all session settings--
Microprocessorcontrolled with built-in self-testYes--
Controls
Meterdigital Monitoranalog meter
Intensitysoft sensor buttonsanalog switch/knob
Output Jacks6mm insulated connectors6mm insulated connectors
Application-accessory
Feetplastic case with towel insideplastic tray / case with towel inside
Handsplastic tray / case with towel insideplastic tray / case with towel inside
Axillary applicatorsPads--
Dimensions
Control unit7,5" (W) x 2" (H) 5,4" (D)Approximately 1 lb10" (W) x 4" (H) 12" (D)Approximately 5 lb
Hard-shell-case13.4" x10.8" x 3.3"data not available
Treatment tray10.2" x 15.75" x 2.2"data not available
Axillary spongecushions2.9" x 3.45" x 1.3"data not available
Towel8.1" x 12.2"data not available
Electrodes (feet,hands)4.5" x 11.2" x 2.4"data not available
Electrodes(underarms)1.9" x 2.17" x 0.6"data not available
Conductive area [sqin]

rizes the similarities and diffe e Chart belo

{10}------------------------------------------------

HIDREXMED
VET
REHA
Hard shell Case112.13data not available
Treatment tray90.88data not available
Towel98.8data not available
Sponge Cushion36.75data not available
Electrode (feet, hands)122.13data not available
Electrode (underarms)8.25data not available
Density at 30mA[mA/sqin]
Hard shell Case0.26data not available
Treatment tray0.33data not available
Towel0.30data not available
Sponge Cushion0.81data not available
Electrode (feet, hands)0.24data not available
Electrode (underarms)3.63data not available
Stimulating device
Signal typeMonophasis square, at DC: pulsed square signal, selectable in increments of 10 from 50% to 100%data not available
Leakage current Type BFPatient leakage current AC:+000.9 μΑPatient leakage current AC SFC:+000.4 μΑPatient leakage current DC:+000.0 μΑPatient leakage current NAT:+004.2 μΑdata not available
Energy output to patienttreatment time: 10 MinutesW = 9,8 V * 0,0008 A * 600 s =4,704 Wstreatment time: 15 MinutesW = 9,8 V * 0,0008 A * 900 s =7,056 Ws1 min => 0,47 Ws = 0,47 Jdata not available
Power supply
externalinternal
100-240 V~ / 50-60 Hz100-240 V~ / 50-60 Hz
Input400mAdata not available
12 V=--
outputmax. 500mA--
6 VA--
Certification
CEYESNO
StandardsIEC 60601-1:2007?
GMPYesYes

Difference's Analysis

These technical differences were identified:

Subject device (sd)Predicate device (pd)
1.DC-current output max. 35mADC-current output 0-10 and 0-50mA
The pd has two output ranges because it uses an analog Ampere-meter. It is verylikely to make reading-mistakes using a wide-range-scale while operating in a low-current mode. Therefore, the manufacturer suggests to use the low-range-scale (0-10mA), while using the device in the low-current-mode. However, it is still likely tomake mistakes.The sd has a display that shows digital digits. No matter what the current-mode is, it
is unlikely to make reading-mistakes. Therefore there is no need to have separatedscale-ranges.
2.DC and pulsed-current outputOnly DC-current output
A pulse-width of 100% (no pause) is corresponding to the classical DC method.However, pulse-width-method makes the sensation for the user more present because ofthe physiological 10kHz frequency. The pulsed current output modes are more sensitive.So the risk of skin-irritation is reduced.
3.Automatic current regulation and over-current protectionOver-current protection
The pd, due to its basic concept with analog technology, has a manual currentadjustment. The patient/user has to read continuously the analog meter to stay withthe appreciated settings.The sd has a microprocessor that is able to watch physical conditions. The sd canautomatically adjust the current and keep the therapeutic settings based on thepatient's settings. Physical changes caused by the patient, like position ormovements, can automatically be compensated.
4.Timer
The pd has no timer. The patient has to measure and check the time span for thetreatment-session by himself.The sd has a timer that automatically starts the countdown when the treatmentstarts. The patient's benefit is the automatic ending of the treatment session. The riskfor an overdose is reduced.
5.DisplayAnalog meter
The pd has an analog Ampere-meter. It is very likely to make reading-mistakes usinga wide-range-scale while operating in a low-current mode. It is also necessary tohave god visibility to be able to read the relatively small numbers on the meter.Reading mistakes are likely.The sd has a display with big digital digits. The display shows the setting as well asthe current. The possibility to make reading mistakes is unlikely.
6.microprocessorAnalog technology
The pd operates based on analog technology.The sd operates based on digital technology. The advantage is to be able to checkand control the treatment-parameter during the session, to reduce the weight of thedevice and offer more adjustment for treatment-parameter.
7.Digital MonitorAnalog meter
The pd has an analog Ampere-meter. It is very likely to make reading-mistakes usinga wide-range-scale while operating in a low-current mode. It is also necessary tohave god visibility to be able to read the relatively small numbers on the meter.Reading mistakes are likely.
The sd has a display with big digital digits. The display shows the setting as well asthe current. The possibility to make reading mistakes is unlikely.
8.Soft sensor buttonsAnalog switch/knob
Analog switches and knobs are subject to corrosion. It is likely that their precisionand even functionality is reduced after a few years of use.The sd has foil-insulated micro-switches. They last much longer and it is unlikely thatthey loose their functionality at all.
9.Application accessory axillary-spongeNo axillary-sponge
Some patients suffer also from axillary hyperhidrosis. To offer effective treatmentalso for these patients the sd has specialized accessory.
10.External power supplyInternal Power supply
The pd has an internal power supply while the sd has an external power supply.There are no significant differences concerning safety or comfort. A break in thepower supply can be fixed more easily with an external one.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for HIDREX, which is written in blue with an orange exclamation point. To the right of the HIDREX logo are three other logos stacked on top of each other. The top logo is MED, the middle logo is VET, and the bottom logo is REHA. The text for each of these logos is in a different color, and each logo has an orange graphic to the left of the text.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX replaced by an orange exclamation point. To the right of the HIDREX logo are three additional logos, stacked vertically. The top logo is MED, the middle logo is VET, and the bottom logo is REHA. Each of these logos also has an orange graphic to the left of the text.

The above mentioned differences between the subject device and the predicate device are due to technological progress. The subject device and its modern technology increase the comfort and benefit for the patient in using the device. Misuse or misunderstanding by using an analog meter is impossible with the clear digital-display of the subject device. All settings are protocolled and displayed in the display. This avoids mistakes in the user-setting. The additional benefit of the pulsed-current is demonstrated in several clinical trials. The safety of the device was demonstrated in non-clinical tests according to recognized standards.

Comparison between materials and principles of use

The HIDREX PSP1000 as well as the predicate device MD-1A GALVANIC UNIT by R.A. Fischer Co. (K964208) use a standard housing made from ABS-Plastic. This housing material is non-toxic and used with many other devices. The same is true for the trays/cases that are used to apply the therapy. None of these materials is in direct contact with the skin for medical purpose. Both devices, the subject device as well as the predicate device, use standard-household-towels that are in direct contact with the patients skin.

The system components are the same. Both systems consist of a central-unit and application components. The subject device has in addition to the foot- and hand-applicator also an axillary applicator. Therefore the subject device offers more benefit to the user than the predicate device.

The principles of use are basically the same. The subject device uses a modern technology like LCD-display to display the setting etc., micro-buttons to adjust and digital controlled processes, while the predicate device has analog technology, like an analog meter to display the current, analog resistor-knob for adjustments and no digital control. Therefore the subject device offers higher comfort and control and advanced usability.

The performance of the subject device and the predicate device are similar concerning the standard dc-current-therapy. The subject device offers a pulsed current in addition, which makes the sensation for the user more present because of the physiological 10kHz frequency.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for HIDREX, which is written in blue letters except for the "I" which is orange. To the right of the logo are three other logos, MED, VET, and REHA. Each of these logos has a small orange circle to the left of the word, and the words are written in blue except for VET which is green.

In conclusion the differences between the subject device and the predicate device don't lead to any harm of the patient or third persons. Comfort and usability is increased with the subject device compared to the predicate device.

Comparison to legally marketed device Drionic (K831320) by General Medical Company (Substantial Equivalence):

LILA ATIALL RAINIL ARIIIIIIMIILAA FITA AIIIIIIIIIIIIIIIIII AILAANI
DEVICEHIDREX PSP1000SE-DEVICEDrionic (K831320)
Indication for useThis Tap-Water-lontophoresisThe Drionic device is indicated for
device is intended to treatuse in the treatment of hyperhidrosis
hyperhidrosis (pathological(abnormal sweating) affecting
sweating) affecting hands, feet andhands, feet and underarms. It uses
underarms.the technique of tap water
iontophoresis.
Performance
max. 60 Vhand or feet treatment: 9V
underarm treatment: 18V
DC current output
max. 35 mAhand or feet treatment: 6-14mA
underarm treatment: 1-2,5mA
max. 60 V--
Pulsed currentmax. 35 mA--
output9.9 kHz--
5 output modes (pulse-width: 50%,--
60%, 70%, 80% and 90%)
Polarity reversal
(to alternate POSmanualNo
and NEG
application)
Automatic current
regulationYesNo
TimerYesNo
DisplayLCD-multi character display toNo
display all session settings
Microprocessor
controlled withYesNo
built-in self-test
Controls
Meterdigital MonitorNo
Intensitysoft sensor buttonsanalog control wheel
Output Jacks6mm insulated connectorsNo
Application-
accessory
Feetplastic tray / case with towel insidedata not available
Handsplastic tray / case with towel insidedata not available
Axillary applicatorsPadsdata not available
Dimensions
sizeControl unit:hand or feet unit:
7,5" (W) x 2" (H) 5,4" (D)11,6" (W) x 2,2" (H) 4,9" (D)
underarm unit:
4,4" (W) x 0,6" (H) 4,3" (D)
weightApproximately 1 Ib0,825 lb or 0,374 lb
Hard-shell-case13.4" x10.8" x 3.3"data not available

The Chart below summarizes the similarities and differences:

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logos for HIDREX, MED, VET, and REHA. The HIDREX logo is in blue and orange, with the "I" in HIDREX being orange and having a circle below it. The MED logo is in blue and orange, the VET logo is in green and orange, and the REHA logo is in blue and orange.

Treatment tray10.2" x 15.75" x 2.2"data not available
Axillary spongecushions2.9" x 3.45" x 1.3"data not available
Towel8.1" x 12.2"data not available
Electrodes (feet,hands)4.5" x 11.2" x 2.4"data not available
Electrodes(underarms)1.9" x 2.17" x 0.6"data not available
Conductive area[sqin]
Hard shell Case112.13data not available
Treatment tray90.88data not available
Towel98.8data not available
Sponge Cushion36.75data not available
Electrode (feet,hands)122.13data not available
Electrode(underarms)8.25data not available
Density at 30mA[mA/sqin]
Hard shell Case0.26data not available
Treatment tray0.33data not available
Towel0.30data not available
Sponge Cushion0.81data not available
Electrode (feet,hands)0.24data not available
Electrode(underarms)3.63data not available
Stimulating device
Signal typeMonophasis square, at DC: pulsedsquare signal, selectable inincrements of 10 from 50% to 100%data not available
Leakage currentType BFPatient leakage current AC: +000.9μΑPatient leakage current AC SFC:+000.4 µAPatient leakage current DC: +000.0μΑPatient leakage current NAT: +004.2μΑdata not available
Energy output topatienttreatment time: 10 MinutesW = 9,8 V * 0,0008 A * 600 s =4,704 Wstreatment time: 15 MinutesW = 9,8 V * 0,0008 A * 900 s =7,056 Ws1 min => 0,47 Ws = 0,47 Jdata not available
Power supply
external100-240 V~/50-60 Hzinternal
Inputhand or feet treatment: 9Vunderarm treatment: 18V
400mAdata not available
outputhand or feet treatment: 9V
underarm treatment: 18V
max. 500mA--
6 VA--
Certification
CEYESNO
StandardsIEC 60601-1:2007No Standards
GMPYesdata not available

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX being an orange exclamation point. To the right of the HIDREX logo are three smaller logos, MED, VET, and REHA. Each of these logos has an orange semi-circle with a small circle inside of it to the left of the text.

Comparison between principles of use

The HIDREX PSP1000 has nearly the same system components as the predicate device Drionic by General Medical Company (K831320). Both systems consist of a unit for hand and foot treatment and an application component for underarm treatment.

The principles of use are basically the same. The subject device uses a modern technology like LCD-display to display the setting etc., micro-buttons to adjust and digital controlled processes, while the predicate device has old technology like an analog control wheel to adjust the intensity, a light to display the operating state and no digital control. Therefore the subject device offers higher comfort and control and an advanced usability.

The performance of the subject device and the predicate device are the same concerning the standard dc-current-therapy. In addition the subject device offers a pulsed current which makes the sensation for the user more present because of the physiological 10kHz frequency.

Difference's Analysis

These technical differences were identified:

Subject device (sd)Predicate device (pd)
1.DC-current output max. 35mAThe pd has two output ranges because it has no Ampere-meter.The sd has a display that shows digital digits. No matter what the current-mode is, itis unlikely to make reading-mistakes. Therefore there is no need to have separatedscale-ranges.DC-current output 6-14mA and 1-2,5mA
2.DC and pulsed-current outputA pulse-width of 100% (no pause) is corresponding to the classical DC method.However, pulse-width-method makes the sensation for the user more present because ofthe physiological 10kHz frequency. . The pulsed current output modes are moresensitive. So the risk of skin-irritation is reduced.Only DC-current output
3.Polarity reversal (to alternate POS andNEG application)The pd has no possibility to change the polarity.The sd provides a manual adjustable reversible polarity. The user is able to changebetween positive and negative application.--
4.Automatic current regulation and over-current protectionThe pd has a manual adjustment. The patient/user has to advance the control wheeluntil there is a comfortable tingle. During the treatment-session the patient/ user isresponsible for the intensity by himself without any technical help.No automatic regulation or protection

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX being an orange exclamation point. To the right of the HIDREX logo are three other logos, MED, VET, and REHA. Each of these logos has an orange semi-circle to the left of the text.

The sd has a microprocessor that is able to watch physical conditions. The sd canautomatically adjust the current and keep the therapeutic settings based on thepatient's settings. Physical changes caused by the patient, like position or
movements, can automatically be compensated. This reduces the risk of a harmfulhigh intensity.
5.Timer
The pd has no timer. The patient has to measure and check the time span for the
treatment-session by himself.
The sd has a timer that automatically starts the countdown when the treatment
starts. The patient's benefit is the automatic ending of the treatment session. The risk
for an overdose is reduced.
6.Display
The pd has no display or any other possibility to check setting and current.
The sd has a display with big digital digits. The display shows the setting as well as
the current. The possibility to make reading mistakes is unlikely.
7.MicroprocessorAnalog technology
The pd operates based on analog technology.
The sd operates based on digital technology. The advantage is to be able to check
and control the treatment-parameter during the session, to reduce the weight of the
device and offer more adjustment for treatment-parameter.
8.Digital Monitor
The pd has no digital monitor.
The sd has a display with big digital digits. The display shows the setting as well as
the current. The possibility to make reading mistakes is unlikely.
9.Soft sensor buttonsAnalog switch/knob
Analog switches and knobs are subject to corrosion. It is likely that their precision
and even functionality is reduced after a few years of use.
The sd has foil-insulated micro-switches. They last much longer and it is unlikely that
they loose their functionality at all.
10.External power supplyInternal Power supply
The pd has an internal power supply while the sd has an external power supply.
There are no significant differences concerning safety or comfort. A break in the
power supply can be fixed more easily with an external one.

The above mentioned differences between the subject device and the predicate device are due to technological progress. The subject device and its modern technology increases the comfort and benefit for the patient in using the device. Because of the automatic current regulation and protection the risk of a harmful high intensity and of an overdose is reduced. All settings are protocolled and displayed in the display. This avoids mistakes in the user-setting. The additional benefit of the pulsed-current is demonstrated in several clinical trials. The attached trial is an explanation as to why the higher levels of voltage and current do not affect safety and effectiveness of the subject device. Furthermore, the safety of the device was demonstrated in non-clinical tests according to recognized standards.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for HIDREX, with the "I" in HIDREX replaced by an orange exclamation point. To the right of the HIDREX logo are three additional logos, each with an orange circle and a word in blue or green. The words are MED, VET, and REHA.

In conclusion the differences between the subject device and the predicate device don't lead to any harm of the patient or third persons. Comfort and usability is increased with the subject device compared to the predicate device.

Non-clinical tests:

To demonstrate substantial equivalence we performed non-clinical tests according recognized standards:

  • To demonstrate electrical safety, software-validation and performance, non-clinical . tests were performed according to IEC 60601-1 Medical electrical equipment -General requirements for safety
  • To demonstrate safety of the software, non-clinical validation was performed . according to IEC 60601-1 Medical electrical equipment - General requirements for safety -- clause 14 Programmable electrical medical systems
  • . To demonstrate electromagnetic compatibility, non-clinical tests were performed according to IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.

Quality Assurance and Manufacturing Controls:

HIDREX GmbH operates to an established and certified quality management system according to ISO 13485 requirements.

Conclusion:

The subject device HIDREX PSP1000 is as safe, as effective and performs more comfortable than the predicate devices.

§ 890.5525 Iontophoresis device.

(a)
Iontophoresis device intended for certain specified uses —(1)Identification. An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined.(2)
Classification. Class II (performance standards).(b)
Iontophoresis device intended for any other purposes —(1)Identification. An iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.(2)
Classification. Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted:
(A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended;
(B) Testing of the ability of the device to maintain a safe pH level; and
(C) If used in the ear, testing of the device to demonstrate mechanical safety.
(ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following:
(A) A description and/or graphical representation of the electrical output;
(B) A description of the electrode materials and pH buffer;
(C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and
(D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning:

Warning: Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile.
(vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.