K Number
K133019

Validate with FDA (Live)

Date Cleared
2015-02-20

(512 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fiberion Ophthalmic Endolaser Probes are in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached

Device Description

Fiberion endolaser probes are designed to be connected to ophthalmic laser photocoagulators and to transmit laser energy inside the patient's eye. One end of the probe contains a connector for attachment to the laser unit, the other, a tip section to be introduced inside the eye. Device components are an input connector, a glass fiber optic protected by a buffer coating, a protective tubing, an aluminum handle with a medical grade stainless steel needle. The illuminating endolaser probe has an additional illumination fiber optic.

AI/ML Overview

This document describes the Fiberion Ophthalmic Endolaser Probe. Below is an anaylsis of its acceptance criteria and the study that proves the device meets the criteria.

Note: The provided document is a 510(k) summary for a medical device (Fiberion Ophthalmic Endolaser Probe). It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria, test sets, or expert evaluations as one might find for an AI/ML-based diagnostic device. Therefore, many of the requested fields cannot be directly answered from this document. The responses below are based on the information available and acknowledge the limitations.

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (Bench/Lab Testing)Reported Device Performance
Laser Power Transmission: Equivalent to predicate deviceFound to be equivalent to predicate device.
Beam Divergence: Equivalent to predicate deviceFound to be equivalent to predicate device.
Beam Density Homogeneity: Equivalent to predicate deviceFound to be equivalent to predicate device.
Illumination Intensity (for illuminating models): Equivalent to predicate deviceFound to be equivalent to predicate device.
Illumination Area Diameter (for illuminating models): Equivalent to predicate deviceFound to be equivalent to predicate device.
Safe Delivery of Intended Output Power: Verified performance"Fiberion Ophthalmic Endolaser Probes were tested for safe delivery of the intended output power."
Sterility: Fulfil requirements"Found to fulfil the requirements."
Biocompatibility (sections in contact with patient): Fulfil requirements"Found to fulfil the requirements."
EO-ECH Residues: EO < 1.25 micrograms/device, ECH < 5 micrograms/deviceEO levels determined to be < 1.25 micrograms per device, ECH levels < 5 micrograms per device.
Endotoxins: < 0.2 Endotoxin units/deviceEndotoxin levels determined to be < 0.2 Endotoxin units per device.

2. Sample size used for the test set and the data provenance

The document describes bench and laboratory testing, not a clinical study on patient data.

  • Sample size for test set: Not specified in the document for the bench and lab tests. It's implied that "Fiberion Ophthalmic Endolaser Probes" (plural) were tested, but the exact number of units or replicates is not given.
  • Data provenance: Not applicable in the context of patient data. The tests were performed on the manufactured devices themselves (bench and lab testing).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The "ground truth" for this device's performance is based on physical and chemical measurements (e.g., laser power, chemical residues, sterility) against established standards or equivalence to a predicate device, not on expert medical interpretation of images or clinical outcomes.

4. Adjudication method for the test set

Not applicable. This was not a study requiring human adjudication, but rather objective technical and laboratory measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an ophthalmic endolaser probe, which is a surgical tool, not an AI-based diagnostic or imaging device. Therefore, an MRMC study related to human readers and AI assistance is irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance of this device is based on:

  • Physical measurements (e.g., laser power, beam divergence, illumination intensity) against defined specifications or comparison to a predicate device's measured performance.
  • Chemical analysis (e.g., EO-ECH residues, endotoxin levels) against regulatory limits and standards.
  • Microbiological testing (sterility) against established standards.

8. The sample size for the training set

Not applicable. This is a hardware medical device and does not involve machine learning or a training set in the context of an algorithm.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping heads or faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 20, 2015

Emtron Elektronik Ve Mekanik Sanayi Ve Ticaret Lim Dr. Mehmet Melek President Bebek Yolu Sokagi 23/3 Etiler Istanbul, 34337 TR

Re: K133019

Trade/Device Name: Fiberion Ophthalmic Endolaser Probe Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: January 9, 2015 Received: January 12, 2015

Dear Dr. Melek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -S

for Malvina B. Eydelman, MD Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K133019

Device Name Fiberion Ophthalmic Endolaser Probe

Indications for Use (Describe)

Fiberion Ophthalmic Endolaser Probes are in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary EMTRON™ Corporation Fiberion™ Ophthalmic Endolaser Probes

Submitter's Name, Address, Telephone Number, E-mail, Contact Person and Date Prepared

EMTRON Elektronik ve Mekanik Sanayi ve Ticaret Limited Sirketi Bebek Yolu Sokaqi 23/3 Etiler 34337 lstanbul Turkey +90 (212) 257 1787 melek@emtron.com Dr. Mehmet Melek Date Prepared: February 20, 2015

Device Information

Trade Name: Fiberion Ophthalmic Endolaser Probes Classification Name: Laser, Ophthalmic, Accessory CFR Section: 886.4390 Class II Product Code: HQF

Predicate Devices

Fiberion Ophthalmic Endolaser Probes is substantially equivalent in intended use and/or method of operation to other currently legally marketed laser probes of IRIDEX Corporation's IRIS Medical EndoProbe (K022228).

Device Description

Fiberion endolaser probes are designed to be connected to ophthalmic laser photocoagulators and to transmit laser energy inside the patient's eye. One end of the probe contains a connector for attachment to the laser unit, the other, a tip section to be introduced inside the eye. Device components are an input connector, a glass fiber optic protected by a buffer coating, a protective tubing, an aluminum handle with a medical grade stainless steel needle. The illuminating endolaser probe has an additional illumination fiber optic.

List of Fiberion Ophthalmic Endolaser Probes

SMA-S: Standard SMA connector, straight 0,9mm (20 Gauge) tip
SMA-A: Standard SMA connector, angled 0,9mm (20 Gauge) tip
E-SMA-S: Extended SMA connector, straight 0,9mm (20 Gauge) tip
E-SMA-A: Extended SMA connector, angled 0,9mm (20 Gauge) tip
E-CYL-S: Cylindrical connector, straight 0,9mm (20 Gauge) tip
E-CYL-A: Cylindrical connector, angled 0,9mm (20 Gauge) tip
E-906-S: Extended SMA 906 connector, straight 0,9mm (20 Gauge) tip
E-906-A: Extended SMA 906 connector, angled 0,9mm (20 Gauge) tip
U-SMA-S: SMA connector with long handle, straight 0,9mm (20Gauge) tip
U-SMA-A: SMA connector with long handle, angled 0,9mm (20Gauge) tip
SMA-400-S: Standard SMA connector, 400 micron fiber, straight 0,9mm (20Gauge) tip
SMA-400-A: Standard SMA connector, 400 micron fiber, angled 0,9mm (20Gauge) tip
STA-S: ST connector with long ferrule, straight 0,9mm (20 Gauge) tip
STA-A: ST connector with long ferrule, angled 0,9mm (20 Gauge) tip
SMA-23-S: Standard SMA connector, straight 0,6mm (23 Gauge) tip
SMA-23-A: Standard SMA connector, angled 0,6mm (23 Gauge) tip
E-SMA-23-S: Extended SMA connector, straight 0,6mm (23 Gauge) tip
E-SMA-23-A: Extended SMA connector, angled 0,6mm (23 Gauge) tip
E-CYL-23-S: Cylindrical connector, straight 0,6mm (23 Gauge) tip
E-906-23-S: Extended SMA 906 connector, straight 0,6mm (23 Gauge) tip
E-906-23-A: Extended SMA 906 connector, angled 0,6mm (23 Gauge) tip
U-SMA-23-S: SMA connector with long handle, straight 0,6mm (23Gauge) tip
U-SMA-23-A: SMA connector with long handle, angled 0,6mm (23Gauge) tip
SMA-400-23-S: Standard SMA connector, 400 micron fiber,straight 0,6mm (23 Gauge) tip
STA-23-S: ST connector with long ferrule, straight 0,6mm (23 Gauge) tip
STA-23-A: ST connector with long ferrule, angled 0,6mm (23 Gauge) tip
SMA-25-S: Standard SMA connector, straight 0,5mm (25 Gauge) tip
SMA-25-A: Standard SMA connector, angled 0,5mm (25 Gauge) tip
E-SMA-25-S: Extended SMA connector, straight 0,5mm (25 Gauge) tip
E-SMA-25-A: Extended SMA connector, angled 0,5mm (25 Gauge) tip
E-906-25-S: Extended SMA 906 connector, straight 0,5mm (25 Gauge) tip
E-906-25-A: Extended SMA 906 connector, angled 0,5mm (25 Gauge) tip
U-SMA-25-S: SMA connector with long handle, straight 0,5mm (25Gauge) tip
STA-25-S: ST connector with long ferrule, straight 0,5mm (25 Gauge) tip
STA-25-A: ST connector with long ferrule, angled 0,5mm (25 Gauge) tip
SMA-27-S: Standard SMA connector, straight 0,4mm (27 Gauge) tip
E-SMA-27-S: Extended SMA connector, straight 0,4mm (27 Gauge) tip
E-906-27-S: Extended SMA 906 connector, straight 0,4mm (27 Gauge) tip
STA-27-S: ST connector with long ferrule, straight 0,4mm (27 Gauge) tip
ILL-SMA-S: Standard SMA connector, straight 0,9mm (20 Gauge) tip
ILL-SMA-A: Standard SMA connector, angled 0,9mm (20 Gauge) tip
ILL-E-SMA-S: Extended SMA connector, straight 0,9mm (20 Gauge)tip
ILL-E-SMA-A: Extended SMA connector, angled 0,9mm (20 Gauge) tip
ILL-E-906-S: Extended SMA 906 connector, straight 0,9mm (20Gauge) tip
ILL-E-906-A: Extended SMA 906 connector, angled 0,9mm (20Gauge) tip
ILL-U-SMA-S: SMA connector with long handle, straight 0,9mm (20Gauge) tip
ILL-U-SMA-A: SMA connector with long handle, angled 0,9mm (20Gauge) tip
ILL-STA-S: ST connector with long ferrule, straight 0,9mm (20 Gauge) tip
ILL-STA-A: ST connector with long ferrule, angled 0.9mm (20 Gauge) tip

{5}------------------------------------------------

Intended Use

Fiberion Ophthalmic Endolaser Probes are intended for use in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination function is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached.

{6}------------------------------------------------

Substantial equivalence

Application for 510KSubstantial equivalence to
Fiberion Ophthalmic Endolaser ProbesIRIDEX Corporation's IRIS Medical
EndoProbe (K022228)
Light transmission for photocoagulationLight transmission for photocoagulation
Aluminum handpieceAluminum handpiece
Stainless steel needleStainless steel needle
Glass optical fiberGlass optical fiber
Thermoplastic rubber jacketVinyl jacket
ETO sterilized.ETO sterilized.

Differences between the Fiberion Ophthalmic Endolaser Probes and the predicate device

Some terminology differences exist in the intended use description due to trademark issues, but both devices' intended use is exactly the same.

Some models of the predicate device are not available in the Fiberion line. Fiberion probes offer a range of connectors that have been validated with the ophthalmic lasers listed in the instructions for use. The probes have been used internationally for years without safety or laser compatibility problems.

The protective tubing which does not come in contact with the patient is of a different material. Fiberion probes use thermoplastic rubber meeting FDA food grade requirements.

Hence, none of these differences are cause for safety or effectiveness concerns.

Bench testing

Fiberion Ophthalmic Endolaser Probes and the predicate device were tested for laser power transmission, beam divergence, beam density homogeneity, and additionally for illuminating models, illumination intensity and illumination area diameter; and were found to be equivalent. Fiberion Ophthalmic Endolaser Probes were tested for safe delivery of the intended output power.

Laboratory testing

Fiberion Ophthalmic Endolaser Probes were tested for sterility and their sections in contact with the patient for biocompatibility, EO-ECH residues and endotoxins and were found to fulfil the requirements. EO levels were determined to be < 1.25 micrograms per device, ECH levels < 5 micrograms per device, and Endotoxin levels < 0.2 Endotoxin units per device.

Conclusion

Fiberion Ophthalmic Endolaser Probes shares similar indications for use, materials, and similar performance characteristics as, and thus are substantially equivalent to. IRIDEX Corporation's IRIS Medical EndoProbe (K022228).

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.