K Number
K132994

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-12-05

(72 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CoreLink ANODYNE™ Anterior Cervical Plate System is intended for anterior fixation of the cervical spine. Indications for use include the temporary stabilization of the anterior spine during the evolution of cervical fusions in patients with degenerative disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, or scollosis), pseudarthrosis, and/or failed previous fusions. The intended levels for treatment range from C2 - T1.

Device Description

The Anodyne™ Anterior Cervical Plate System is comprised of an assortment of cervical plates and screws that act to stabilize the spine during the intervertebral fusion process. The cervical plate has a rotatable anti-backout "lock" for each screw position to prevent back-out of the screw.

The plate is available in single and double plate configurations with multiple lengths ranging from 13 mm – 22 mm (1 level) and 26 mm – 40 mm (2 level) for the single plates. The double plates include 1 level (13 mm - 30 mm), 2 level (26 mm - 46 mm), 3 level (46 mm - 70 mm), and 4 level (60 mm -100 mm). The screws are available in various lengths from 12 mm - 20 mm, with major thread diameter options of 4.6 mm or 5.2 mm.

AI/ML Overview

The Anodyne™ Anterior Cervical Plate System is a medical device intended for anterior fixation of the cervical spine. The device's acceptance criteria and the study proving it meets these criteria are outlined in the 510(k) summary.

1. Acceptance Criteria and Reported Device Performance:

The acceptance criteria for the Anodyne™ Anterior Cervical Plate System are based on the mechanical performance of the device, specifically its ability to withstand static and dynamic loads and torsion, as compared to predicate devices. The study demonstrates that the device performs equivalently to previously cleared devices.

Acceptance CriteriaReported Device Performance
Mechanical Performance (Static and Dynamic Compression per ASTM 1717): Ensure the device can withstand static and dynamic compression forces without failure, matching or exceeding predicate devices.The Anodyne™ Anterior Cervical Plate System's mechanical performance in static and dynamic compression was found to be equivalent to predicate devices.
Mechanical Performance (Static Torsion per ASTM F1717): Ensure the device can withstand static torsional forces without failure, matching or exceeding predicate devices.The Anodyne™ Anterior Cervical Plate System's mechanical performance in static torsion was found to be equivalent to predicate devices.
Materials: Use of Ti-6Al-4V alloy per ASTM F136.The device uses Ti-6Al-4V alloy per ASTM F136, matching the material specification.
Substantial Equivalence: Demonstrate substantial equivalence to predicate devices in terms of intended use, design, materials, mechanical safety, and performance.The device was determined to be substantially equivalent to predicate devices in all specified aspects.

2. Sample Size and Data Provenance for the Test Set:

  • Sample Size: Not explicitly stated as a number of devices tested. Instead, the testing refers to "the ANODYNE™ Anterior Cervical Plate System" as the subject of the tests, implying a representative sample of its components (plates and screws across various sizes and configurations).
  • Data Provenance: The tests are described as "Non-Clinical Test Summary" and were likely conducted in a controlled laboratory environment. There is no information regarding the country of origin of the data or whether it was retrospective or prospective, as these are non-clinical (mechanical) tests rather than clinical studies on human subjects.

3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set):

Not applicable. This device underwent non-clinical mechanical testing, not clinical studies involving human patients or expert interpretation of medical images. Therefore, no medical experts were involved in establishing ground truth for a test set.

4. Adjudication Method for the Test Set:

Not applicable. As described above, there was no clinical test set requiring adjudication. The assessment was based on objective mechanical testing results against industry standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. The document explicitly states: "No clinical studies were performed." Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted.

6. Standalone Performance Study:

No. Since no clinical studies were performed, there was no standalone algorithm performance study (the device is a physical medical implant, not an AI or digital diagnostic tool). The "standalone performance" in this context refers to the device's mechanical integrity as tested in a lab setting, independent of human interaction during real-world use, which was indeed evaluated through the non-clinical tests.

7. Type of Ground Truth Used:

The ground truth for the non-clinical tests was established by recognized industry standards (ASTM F1717) for mechanical testing of spinal implants. The performance of the Anodyne™ system was compared against the established mechanical properties and performance of legally marketed predicate devices.

8. Sample Size for the Training Set:

Not applicable. The device is a physical medical implant, not an AI algorithm. There is no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for a physical medical implant.

{0}------------------------------------------------

510(k) Summary for the Anodyne™ Anterior Cervical Plate System K132994/S001

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Anodyne™ Anterior Cervical Plate System

1 -GENERAL INFORMATION

Date Prepared:December 02, 2013
Trade Name:Anodyne™ Anterior Cervical Plate System
Common Name:Anterior cervical plateDEC 0 5 2013
Classification Name:Spinal intervertebral fixation orthosis
Class:II
Product Code:KWQ
CFR section:21 CFR section 888.3060
Device panel:Orthopedic
Legally MarketedAnodyne™ Anterior Cervical Plate System (K121514)UNIPLATE Anterior Cervical Plate System (K042544/ K082273/K100070)
Predicate Device:CSLP Anterior Cervical Plate (Synthes - K000536)
Submitter:CoreLink, LLC7606 Forsyth BlvdClayton, MO 63105
Contact:J.D. Webb1001 Oakwood BlvdRound Rock, TX 78681512-388-0199 Tele512-692-3699 Faxe-mail: jdwebb@orthomedix.net

2. DEVICE DESCRIPTION

The Anodyne™ Anterior Cervical Plate System is comprised of an assortment of cervical plates and screws that act to stabilize the spine during the intervertebral fusion process. The cervical plate has a rotatable anti-backout "lock" for each screw position to prevent back-out of the screw.

The plate is available in single and double plate configurations with multiple lengths ranging from 13 mm – 22 mm (1 level) and 26 mm – 40 mm (2 level) for the single plates. The double plates include 1 level (13 mm - 30 mm), 2 level (26 mm - 46 mm), 3 level (46 mm - 70 mm), and 4 level (60 mm -100 mm). The screws are available in various lengths from 12 mm - 20 mm, with major thread diameter options of 4.6 mm or 5.2 mm.

Change from Predicate:

This 510(k) is submitted in order to gain clearance for the single plate configuration.

Materials:

Ti-6Al-4V alloy per ASTM F136

{1}------------------------------------------------

3. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES

The Anodyne™ Anterior Cervical Plate System is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances.

4. INTENDED USE

The Corelink ANODYNE™ Anterior Cervical Plate System is intended for anterior fixation of the cervical spine. Indications for use include the temporary stabilization of the anterior spine during the evolution of cervical fusions in patients with degenerative disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, or scollosis), pseudarthrosis, and/or failed previous fusions. The intended levels for treatment range from C2 - T1.

5. NON-CLINICAL TEST SUMMARY

The following tests were conducted:

  • Static and dynamic compression per ASTM 1717 .
  • Static torsion per ASTM F1717 .

The results of this testing indicate that the ANODYNE™ Anterior Cervical Plate System is equivalent to predicate devices.

6. CLINICAL TEST SUMMARY

No clinical studies were performed

7. CONCLUSIONS NONCLINICAL AND CLINICAL

CoreLink, LLC considers the ANODYNE™ Anterior Cervical Plate System to be equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities of operation, technology, materials and indications for use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

December 5, 2013

CoreLink, LLC % Mr. J.D. Webb OrthoMedix Group, Incorporated 1001 Oakwood Boulevard Round Rock. Texas 78681

Re: K132994

Trade/Device Name: Anodyne™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: October 7, 2013 Received: October 9, 2013

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. J.D. Webb

forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson -S

  • Mark N. Melkerson Director Division of Orthonedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K132994/S001

Device Name

Anodyne™ Anterior Cervical Plate System

Indications for Use (Describe)

The CoreLink ANODYNE™ Anterior Cervical Plate System is intended for anterior fixation of the cervical spine. Indications for use include the temporary stabilization of the anterior spine during the evolution of cervical fusions in patients with degeneralive disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, defined as kyphosis, or scoliosis), pseudarthrosis, and/or failed previous fusions. The intended levels for treatment range from C2 - T1.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

: :::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: : FOR FOR FDA USE ONLY .....

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.