K Number
K132901

Validate with FDA (Live)

Device Name
AIRSOFT F15
Manufacturer
Date Cleared
2014-06-09

(266 days)

Product Code
Regulation Number
868.5905
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AirSoft F15 is a noninvasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as continuous positive airway pressure (CPAP) or bilevel system.

The AirSoft F15 is:
• to be used by patients (weighing >66 lb (30 kg)) for whom positive airway pressure therapy has been prescribed
• intended for single-patient reuse in the home environment and multipatient reuse in the hospital/institutional environment.

Device Description

The AirSoft F15 Mask system is an externally placed mask covering the mouth and the nose of the patient. It provides a seal such that positive pressure from a positive pressure source is directed to the patient's nose and/or mouth. It is held in place with an adjustable headgear that straps the mask to the face.
AirSoft F15 can be cleaned by the patient in the home environment and reprocessed by professionals in the hospital/institutional environment.
AirSoft F15 is intended to be used under the conditions and purposes indicated in the labelling provided with the product.
AirSoft F15 is a prescription device supplied non-sterile.

AI/ML Overview

The provided text is a 510(k) summary for the ResMed AirSoft F15, a vented full-face mask. It describes the device, its intended use, and a comparison to predicate devices, focusing on non-clinical performance data to demonstrate substantial equivalence.

Here's the information requested, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a separate section. Instead, the performance evaluations are broadly stated as demonstrating "substantial equivalence" to predicate devices. The key performance areas evaluated and their outcomes are:

Performance AreaAcceptance Criteria (Implicit)Reported Device Performance
Pressure-Flow Characteristics & Flow ImpedanceIdentical to predicate device (Quattro Air) in labelled performance range.Identical to predicate devices.
CO2 Washout Performance (Physical & Functional Dead Space)Substantially equivalent to predicate device (Ultra Mirage FFM). Adequate venting to flush out expired CO2.Substantially equivalent to predicate devices.
BiocompatibilityComplies with ISO 10993-1, -3, -5, -6, -10 and FDA guidance #G95-1 for materials in contact with patient/gas pathway.Evaluated for genotoxicity, cytotoxicity, implantation, sensitization, and irritation. Implied compliance.
Mechanical Integrity & PerformanceWithstands normal use and reasonable abuse scenarios, storage temperature, humidity, and transportation shock & vibration.Tested and confirmed to withstand these conditions.
Cleaning & Reuse ValidationMaintained intended function after validated disinfection protocols and indicated number of cycles for single-patient home reuse and multi-patient hospital/institutional reuse.Confirmed to continue functioning as intended.
Overall Substantial EquivalenceSame intended use, similar technological and performance characteristics as predicates.Substantially equivalent to nominated predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" and its sample size in the context of clinical trials or human subject testing. The performance data presented appears to be derived from non-clinical bench testing. The data provenance is not explicitly stated beyond being "extensive testing performed." It is implied to be internal testing conducted by ResMed Ltd in Australia ("ResMed Ltd, 1 Elizabeth Macarthur Drive, Bella Vista NSW 2153 Australia"). The data is retrospective in the sense that it was collected and analyzed to support the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the evaluations were primarily non-clinical bench tests and did not involve a test set requiring expert-established ground truth in the context of medical image interpretation or clinical outcomes. The "ground truth" for these engineering tests would be derived from physical measurements and established engineering standards.

4. Adjudication Method for the Test Set

This information is not applicable as the performance evaluations were non-clinical bench tests and did not involve human interpretation or a need for adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Use of Full Face masks with CPAP or Bilevel therapy equipment is proven technology and is well accepted by the medical community. So was the case for the predicate devices, bench testing was sufficient to demonstrate substantial equivalence to the predicate devices."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

This question is not applicable to this device. The ResMed AirSoft F15 is a physical medical device (a mask), not an AI algorithm. Therefore, "standalone" performance in the context of AI is not relevant.

7. The Type of Ground Truth Used

For the non-clinical tests, the "ground truth" was established through:

  • Engineering and physical measurements: For pressure-flow characteristics, flow impedance, CO2 washout, mechanical integrity.
  • ISO standards and FDA guidance: For biocompatibility evaluations (e.g., ISO 10993-1, FDA guidance #G95-1).
  • Validated protocols: For cleaning and reuse validation.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical mask and does not involve a training set as would be used for an AI algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

K132901 JUN 0 9 2014

RESMED

AirSoft F15 Traditional 510(k)


510(k) SUMMARY[As required by 21 CFR 807.92(c)]
Date SentMay 14th, 2014
Company Name/OwnerResMed Ltd1 Elizabeth Macarthur DriveBella Vista NSW 2153 Australia
Submitter NameMr. Jean-Nicolas BoudaudRegulatory Affairs Manager, Patient Interface+61 2 88841000 (phone)+61 2 88842004 (fax)Jean.boudaud@resmed.com.au
Correspondent Details/Official ContactMr. Jim CassiVice President - Quality Assurance Americas9001 Spectrum Center BlvdSan Diego CA 92123 USATel: (858) 836 6081
Trade/Device NameAirSoft F15
Device Common NameVented Full Face Mask
Regulation Number21 CFR 868.5905
Regulation NameNoncontinuous Ventilator (IPPB)
Regulatory ClassClass 2
Product CodeBZD
Predicate DevicesQuattro Air (K123979)Ultra Mirage FFM (K023244)
DescriptionThe AirSoft F15 Mask system is an externally placed mask coveringthe mouth and the nose of the patient. It provides a seal such thatpositive pressure from a positive pressure source is directed to thepatient's nose and/or mouth. It is held in place with an adjustableheadgear that straps the mask to the face.AirSoft F15 can be cleaned by the patient in the home environmentand reprocessed by professionals in the hospital/institutionalenvironment.AirSoft F15 is intended to be used under the conditions andpurposes indicated in the labelling provided with the product.AirSoft F15 is a prescription device supplied non-sterile.
Intended UseThe AirSoft F15 is a noninvasive accessory used for channelingairflow (with or without supplemental oxygen) to a patient from apositive airway pressure (PAP) device such as continuous positiveairway pressure (CPAP) or bilevel system.
The AirSoft F15 is:
• to be used by patients (weighing >66 lb (30 kg)) for whompositive airway pressure therapy has been prescribed
• intended for single-patient reuse in the home environmentand multipatient reuse in the hospital/institutionalenvironment.
Intended UsecomparisonComparison with predicate Quattro Air (K123979)The new device and the predicate Quattro Air mask have identicalintended uses. Both are intended to be used with Positive AirPressure therapy equipment, in the same environments for use andfor an identical patient population.
TechnologicalCharacteristicscomparisonComparison with the predicate Quattro Air (K123979)AirSoft F15 broadly reuses the technological characteristics and designfeatures of the previously cleared ResMed Quattro Air Full Face Mask
• Both devices have very similar physical properties and operatingprinciple
• Both masks are offered in various sizes and are provided withvery similar adjustable headgears. They aim to fit the samepatient population
• Both masks incorporate diffuse type vent holes to providecontinuous air leak to flush out and minimize the amount of CO2re-breathed by the patient. The design of the mask componentsis such that the incorporation of these vent-holes does notinterfere with the intended performance of the mask
• Both masks are equipped with a common elbow component,which can freely rotate through 360 degrees and includes ananti-asphyxia valve (AAV) to enable the patient to breathefresh air in the event that airflow from the flow generator isimpeded
• Both masks connect to a conventional air delivery hosebetween the mask and the positive airway-pressure sourcevia standard conical connectors (ref: ISO 5356-1:2004). Theswivel connector can also rotate freely through 360 degrees
The main differences between the predicate Quattro Air (K123979)and the new AirSoft F15 mask system are:
• the sealing interface - the predicate device utilizes aconventional silicone cushion interface whereas the newAirSoft F15 mask system utilizes an alternate conformingmaterial
• the design of certain components (e.g. frame) and howcomponents interface with each other (e.g. retention features,
10th September. 2013510(k) Page # 35

510(k) Page # 34

{1}------------------------------------------------


·

.

{2}------------------------------------------------

location and shape of these interfaces)

In addition, the AirSoft F15 Mask system was developed within a risk management process in accordance to ISO 14971:2007, Medical devices - Application of risk management to medical devices. Both devices are designed and manufactured under the same 21 CFR 820 compliant Quality system.

Performance Data Comparison with the predicate Quattro Air (K123979) Performance data is provided to demonstrate that the AirSoft F15 design choices do not impact the fundamental scientific concept nor the therapeutic effects of the new device when compared to the predicate.

Both the new mask and the predicate device are designed to operate on the same Full Face Mask ResMed flow generator settings. The pressure-flow characteristics and flow impedance of both devices are identical in the labelled performance range for each device.

Comparison with the predicate Ultra Mirage FFM (K023244) CO2 washout performance testing of the AirSoft F15 demonstrates that it is substantially equivalent to the predicate Ultra Mirage FFM (K023244).

Non-Clinical test data

Extensive testing was performed pre and post cleaning or disinfection of the AirSoft F15 mask to demonstrate that the new device is substantially equivalent to the predicate Quattro Air (K123979) when used in accordance with the supplied instructions for use.

Pressure-flow characteristics and through impedance of the new mask were tested for comparison to the predicate devices and the relevant published data to ensure clinicians are able to prescribe the appropriate therapy when using the new device.

The CO2 performance (physical and functional dead space) of the new device was also tested and demonstrated substantially equivalent to the predicates. This ensures the new mask design provides adequate venting to flush out the expired CO2.

As was the case with the predicates, the AirSoft F15 materials not previously cleared by FDA were subject to appropriate biocompatibility evaluations according to ISO 10993-1. Materials used the construction of components that:

  • contact the heated humidified gas pathway have been . classified as permanent "external communicating devices" (with tissue/bone/dentin)
  • contact the patient during the therapy have been classified as permanent "skin contact"

As relevant and to support the biocompatibility evaluation of each mask component, following biological effects (selected in accordance with FDA guidance #G95-1) were assessed:

10th September, 2013

{3}------------------------------------------------

RESMED

  • . Genotoxicity (ISO 10993-3) Cytotoxicity (ISO 10993-5) . Implantation (ISO 10993-6) . Sensitization (ISO 10993-10) . Irritation (ISO 10993-10) . Mechanical integrity and performance of the new device was tested to simulated normal use and reasonable abuse scenarios. The device was also tested to demonstrate that it can withstand the effects of storage temperature, humidity and transportation shock & vibration. Validation of cleaning and reuse was completed to establish that the device can be reused by a single patient (at home), or by several patients in the hospital/institutional environment. Following validated disinfection protocols and after the number of cycles indicated in the published data for the AirSoft F15 mask, it was confirmed that the device continued to function as intended. The device was shown to be substantial equivalent to the predicate devices. Clinical Data Use of Full Face masks with CPAP or Bilevel therapy equipment is proven technology and is well accepted by the medical community. So was the case for the predicate devices, bench testing was sufficient to demonstrate substantial equivalence to the predicate devices. The new AirSoft F15 Mask System is substantially equivalent to the Substantial Equivalence Conclusion nominated predicate devices: it has the same intended use; .
    • it has similar technological and performance characteristics to . the predicates.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services, featuring three curved lines or shapes.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

June 9, 2014

ResMed Ltd. c/o Jim Cassi Vice President-Quality Assurance Americas ResMed Corp. 9001 Spectrum Center Boulevard San Diego CA 92123

Re: K132901

Trade/Device Name: AirSoft F15 Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: June 4th, 2014 Received: June 4th, 2014

Dear Mr. Cassi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Cassi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

Sincerely yours,

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Image /page/5/Figure/7 description: The image shows the FDA logo with the text "Tejashri Purohit-Sheth, M.D." written on both sides of the logo. The text on the right side of the logo also includes "Clinical Deputy Director DAGRID/ODE/CDRH FOR". The logo is made up of three letters, F, D, and A, with the letters being stylized and interconnected.

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

AirSoft F15 Traditional 510(k)

Indication for Use

510(k) Number (if known): K132901 Device Name: AirSoft F15 Indication for Use

The AirSoft F15 is a noninvasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as continuous positive alrway pressure (CPAP) or bilevel system.

The AlrSoft F15 is:

  • . to be used by patients (weighing >66 lb (30 kg)) for whom positive airway pressure therapy has been prescribed
  • . intended for single-patient reuse in the home environment and multipatient reuse in the hospital/institutional environment.
Prescription UseX
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

Page 1 of_ 1

Image /page/6/Picture/11 description: The image shows a logo with the letters FDA. The letters are stylized with a triple-line effect, giving them a bold and somewhat retro appearance. The logo is in black and white, with the letters standing out against the background.

Anya C. Harry -S 2014.06.05 11:02:46 -04'00'

10" September 2013

510(k) Page# 15

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).