K Number
K132841

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-01-23

(135 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.

Device Description

Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein. The major modifications are: 1) extended selection of interchangeable components manufactured by additional suppliers; 2) additional contract manufacturer for the device: 3) Gamma Irradiation as additional sterilization method.

AI/ML Overview

The Zyno Medical Administration Set is an intravenous administration set used to deliver fluids from a container to a patient's vascular system.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Category)Reported Device Performance
Safety & Effectiveness
BiocompatibilitySuccessfully met acceptance criteria.
PyrogenicitySuccessfully met acceptance criteria.
Chemical TestingSuccessfully met acceptance criteria.
Microbial IngressSuccessfully met acceptance criteria.
DEHP TestingSuccessfully met acceptance criteria.
Sterilization TestingSuccessfully met acceptance criteria.
Functional Performance
Bench Testing (physical & mechanical)Successfully met acceptance criteria.
Integration Testing (with Z-800 series infusion pumps)Successfully met acceptance criteria.
Aging TestingSuccessfully met acceptance criteria.
Material & Design Standards
Tubing MaterialApproved Standard PVC (Same as predicate)
Single UseYes (Same as predicate)
SterileYes (Same as predicate)
ISO 8536-4:2010 CompliantYes (Same as predicate)
ISO 10993 CompliantYes (Same as predicate)
Operation ModeFor use with Z-800 series of infusion pumps or gravity feed (Same as predicate)

Conclusion of Nonclinical Testing: "All nonclinical testing results successfully demonstrated that the Zyno Medical Administration Sets performed as intended. Conclusions drawn from the nonclinical testing is that the testing data demonstrated that the device is as safe, as effective, and performs as well as the legally marketed device cleared in K120685."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each of the nonclinical tests (Biocompatibility, Pyrogenicity, Chemical, Microbial Ingress, DEHP, Sterilization, Bench, Integration, and Aging Testing). It only states that "extensive verification and validation testing" was performed and that "All results successfully met acceptance criteria."

The data provenance is not specified regarding country of origin, nor is it explicitly stated whether the data was retrospective or prospective. Given these are bench and lab-based tests for a medical device administration set, they are inherently prospective controlled tests conducted by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. The nonclinical tests appear to rely on established testing methods and standards (e.g., ISO standards, FDA guidance documents) rather than subjective expert consensus for establishing ground truth. Compliance with these standards is the metric, not expert opinion.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of data by multiple experts. Since no human clinical studies were performed, and the testing was nonclinical (bench, lab-based), an adjudication method is not applicable and therefore not mentioned.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Zyno Medical Administration Sets." The submission focuses on demonstrating substantial equivalence through nonclinical testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable to the Zyno Medical Administration Sets. This device is a physical medical administration set, not an AI algorithm or software device. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply. The device's performance stands alone in its physical and functional testing.

7. The Type of Ground Truth Used

The ground truth for the nonclinical tests was based on established industry standards and regulatory guidance. For example:

  • Biocompatibility: Likely ISO 10993 series.
  • Sterilization: Compliance with relevant sterility assurance levels.
  • Physical/Mechanical Testing: Adherence to performance specifications to ensure proper function and safety.
  • Integration Testing: Successful operation with the Z-800 series infusion pumps, demonstrating compatibility and correct fluid delivery.
  • Pyrogenicity, Chemical, Microbial Ingress, DEHP, Aging: Compliance with specific test methods and limits defined in relevant standards and guidance documents (e.g., "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]" and "Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions").

8. The Sample Size for the Training Set

This question is not applicable. The Zyno Medical Administration Set is a physical medical device, not a machine learning or AI model that requires a "training set."

9. How the Ground Truth for the Training Set was Established

This question is not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

KI32841

JAN 2 3 2014

SPECIAL 510(K) SUMMARY

Submitted By:

Chaoyoung Lee, PhD President / CEO Zyno Medical. LLC 177 Pine Street Natick, MA 01760, USA Fax: 508-650-2006 Phone: 508-650-2008 x 103 Email: cylee@zynomed.com

Company Contact:

Mei Zhang, PhD Director of Engineering Zyno Medical. LLC 177 Pine Street Natick, MA 01760, USA Fax: 508-650-2006 Phone: 508-650-2008 x 205 Email: mei.zhang@zynomed.com

Date Originated: September 6, 2013 Establishment Registration No: 3006575795 Device Proprietary Name: Zyno Medical Administration Sets Device Common, Usual or Classification Name: Intravascular Administration Set Device Class: Class II Panel: 80 General Hospital Procodes: FPA 880.5440

Predicate Device

Zyno Medical Administration Sets (K120685, September 2012)

I. INTENDED USE

Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.

The intended use of the modified device is the same as the predicate device.

Page 1 of 3

{1}------------------------------------------------

II. DEVICE DESCRIPTION AND COMPARISON

Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.

Zyno Medical LLC is submitting this Special 510(k): Zyno Medical Administration Sets to request a modification to our currently marketed administration sets. The major modifications are: 1) extended selection of interchangeable components manufactured by additional suppliers; 2) additional contract manufacturer for the device: 3) Gamma Irradiation as additional sterilization method. Verification and validation results support that the device is as safe and effective as the predicate device. We believe this application is eligible for the Special 510 (k) process since the same fundamental scientific technology and intended use as the predicate device. The principles of operation, intended use and the applications remain the same.

III.SUMMARY OF NONCLINICAL TESTING

Zyno Medical performed extensive verification and validation testing with the Zyno Medical Administration Set with modifications listed in above section II. Testing was completed in accordance with Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)] issued on July 11, 2008. In planning for bench testing to ensure proper functions and safety when used with the intended infusion pumps. Guidance for Industry and FDA Staff - Total Product Life Cycle: Jnfusion Pump - Premarket Notification [510(k)] Submissions, issued on April 23, 20 10 was used as a reference.

Below is a summary of the nonclinical testing and evaluation performed. All results successfully met acceptance criteria:

  • A. Biocompatibility testing
  • B. Pyrogenicity testing
  • C. Chemical testing
  • Microbial ingress evaluation D.
  • E. DEHP testing
  • F. Sterilization testing
  • G. Bench testing, including integration testing with Z-800 series infusion pumps, physical and mechanical testing
  • H. Aging testing

Conclusion of Nonclinical Testing

All nonclinical testing results successfully demonstrated that the Zyno Medical Administration Sets performed as intended. Conclusions drawn from the nonclinical

{2}------------------------------------------------

testing is that the testing data demonstrated that the device is as safe, as effective, and performs as well as the legally marketed device cleared in K120685.

IV. CLINICAL STUDY

Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Zyno Medical Administration Sets.

V. STATEMENT OF EQUIVALENCE

The Zyno Medical Administration Sets are substantially equivalent to the predicate devices, based on comparisons of the device classifications, intended use, and technological characteristics. Verification and validation tests confirmed the suitability of the devices for their intended uses. The test results did not raise any safety or performance questions. and confirmed that the Zyno Medical Administration Sets are substantially equivalent to the predicate device.

The equivalency matrix (Table 1) compares the modified device with the predicate device (K120685).

Table 1: Equivalency Matrix

ParameterZyno Administration Set(Submitted in this package)Zyno Administration Set(K120685)
Device TypeAdministration SetAdministration Set
Intended useTo administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.To administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.
Tubing materialApproved Standard PVCApproved Standard PVC
Single use?YesYes
SterileYesYes
ISO 8536-4:2010compliant?YesYes
ISO 10993compliant?YesYes
Operation modeFor use with Z-800 series of infusion pumps or gravity feedFor use with Z-800 series of infusion pump or gravity feed

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with healthcare, with a double helix-like design.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 23, 2014

Zyno Medical LLC C/O Mei Zhang Director of Engineering 177 Pine Street Natick, MA 01760

Re: K132841

Trade/Device Name: Zyno Medical Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 20, 2013 Received: December 24, 2013

Dear Ms. Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

{4}------------------------------------------------

Page 2 - Ms. Zhang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Image /page/4/Picture/7 description: The image shows the name "Kwame O. Ulmer-S" in a bold, sans-serif font. The text is arranged in two lines, with "Kwame O." on the top line and "Ulmer-S" on the bottom line. There is some noise and artifacts around the text, which may be due to the image quality or processing.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132841

Device Name

Zyno Medical Administration Sets

Indications for Use (Describe)

Zyno Medical Administration Set is a device used to administer to a patient's vascular system through a neede or a catheter inserted into a vein.

Type of Use (Select one or both, as applicable)

[x] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY . : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally signed by Richard C. Chapman Date: 2014.01.22 14:07:30 -05'00' Page 1 of 2 PSC Publishing Scriptions (301) 443-6740 EF FORM FDA 3881 (1/14)

{6}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.