K Number
K132811

Validate with FDA (Live)

Date Cleared
2014-02-27

(171 days)

Product Code
Regulation Number
870.4210
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sams™ Soft-Flow Extended Aortic Cannula is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.

Device Description

The Sams™ Soft-Flow® Extended Aortic Cannula provides a conduit for the flow of patient blood within the extracorporeal circuit during cardiopulmonary bypass surgery. The or parent of the Sams™ Soft-Flow® Extended Aortic Cannula allows patient blood to re-enter the body after the blood has routed through an extracorporeal circuit for blood oxygenetion. The connila consists of a tubular conduit with a tip that can be positioned into the patient's anatomy (ascending aorta) to allow the blood to re-enter the vascular blood stream.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Sarns™ Soft-Flow® Extended Aortic Cannula, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance EvaluationAcceptance Criteria (Implicit)Reported Device Performance
Force at BreakSatisfy appropriate specifications (implied: at least equal to predicate device)Passed pre-determined acceptance criteria
Liquid and Air LeakSatisfy appropriate specifications (implied: no leaks)Passed pre-determined acceptance criteria
Ink AdhesionSatisfy appropriate specifications (implied: maintains legibility)Passed pre-determined acceptance criteria
HemolysisSatisfy appropriate specifications (implied: low or acceptable levels of hemolysis; straight tip design generally produces slightly lower pressure drop, thus less potential for hemolysis than predicate's angled tip)Passed pre-determined acceptance criteria
Pressure DropSatisfy appropriate specifications (implied: low resistance to flow; straight tip design generally produces slightly lower pressure drop)Passed pre-determined acceptance criteria
Exit VelocitySatisfy appropriate specifications (implied: comparable to predicate device)Passed pre-determined acceptance criteria
BiocompatibilityFull support of FDA General Program Memorandum #G95-1 / ISO 10993 (implied: no adverse biological reactions)Passed pre-determined acceptance criteria

Note on Acceptance Criteria: The document states that "performance studies were conducted to ensure...passed pre-determined acceptance criteria." However, the specific quantitative acceptance criteria for each test are not explicitly detailed in the provided text. They are implied to be "appropriate device performance specifications" and to ensure "no appreciable differences in performance between the subject devices and the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set as it refers to these as "bench studies" and "performance evaluations" rather than a clinical human-subject study. The data provenance is retrospective in the sense that the evaluations were done on already manufactured devices as part of a premarket notification (510(k)) process. The location of the testing is not explicitly stated, but it would have been conducted by Terumo Cardiovascular Systems Corporation or a contracted lab. There is no mention of country of origin for test data in the context of human subjects, as no clinical studies were performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This was a medical device regulatory submission for a physical device, and the "ground truth" was established through engineering and laboratory testing against predetermined specifications, not through expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. As noted above, this was a medical device regulatory submission for a physical device, not related to expert review of data that would require adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is an aortic cannula for cardiopulmonary bypass surgery. It is a physical medical device and does not involve AI or human "readers" in the context of diagnostic interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

Not applicable. This device is an aortic cannula, not a software algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance evaluations was established by engineering specifications and industry standards for medical devices (e.g., ISO 10993 for biocompatibility). These criteria are designed to ensure the device performs safely and effectively for its intended use, comparable to legally marketed predicate devices.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" in the context of this physical medical device. The "training" for such devices typically involves design iterations and prototyping, which are then subject to the performance evaluations described.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no "training set" in the AI or clinical study sense, this question does not apply. The analogous process for a physical device would be the iterative design and manufacturing process guided by established engineering principles and regulatory requirements.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with a "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the text "Cardiovascular Systems" in a smaller font size. The logo is simple and professional, and it is likely used to represent the company's brand.

K132811

FEB 2 7 2014

510(K) SUMMARY Sarns™ Soft-Flow® Extended Aortic Cannula

Date Prepared:

August 2013

Sponsor Information: Owner/Applicant/Submitter:

Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor, Michigan 48103 Phone: 1-800-262-3304 Fax: 410-398-6079 Registration No. 1828100

Contact Person:

Garry A. Courtney, MBA, RAC Corporate Director, Regulatory Affairs Phone: 1-800-262-3304 ext. 7486 Fax: 410-398-6079 Email: garry.courtney@terumomedical.com

Device Names/Classifications:

Device Trade Name:Sarns™ Soft-Flow® Extended Aortic Cannula
Device Common Name:Aortic cannula for cardiopulmonary bypass
Classification Name:Cardiopulmonary bypass vascular catheter, cannula, tubing
Regulation Number:21 CFR 870.4210
Classification:Class II
Product Code:DWF

Predicate Devices:

  • SarnsTM Low Jetting Aortic Arch Cannula, K934127 .
  • Sarns™ Xcoated Cannula, K083301

Device Description:

The Sams™ Soft-Flow® Extended Aortic Cannula provides a conduit for the flow of patient blood within the extracorporeal circuit during cardiopulmonary bypass surgery. The or parent of the Sams™ Soft-Flow® Extended Aortic Cannula allows patient blood to re-enter the body after the blood has routed through an extracorporeal circuit for blood oxygenetion. The connila consists of a tubular conduit with a tip that can be positioned into the patient's anatomy (ascending aorta) to allow the blood to re-enter the vascular blood stream.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the Terumo Cardiovascular Systems logo. The logo consists of a circle with the letter T inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the phrase "Cardiovascular Systems" in a smaller font size. The logo is simple and professional, and it is likely used to represent the company's brand.

Intended Use/Indications for Use:

The Sams™ Soft-Flow Extended Aortic Cannula is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.

Performance Evaluations:

Clinical studies involving patients are not necessary to demonstrate the safety and effectiveness. of the subject devices. Performance assessments for safety and effectiveness were eccomplished through bench studies that included the following evaluations:

  • force at break .
  • liquid and air leak .
  • ink adhesion #
  • hemolysis .
  • pressure drop 4
  • exit velocity .

Substantial Equivalence Comparison:

The information presented in this section depicts a comparison between the subject device, the modified Sarns™ Soft-Flow® Aortic Extended Cannula and the predicate device:

  • Sarns™ Low Jetting Aortic Arch Cannula, K934127 .
  • SarnsTM Xcoated Cannula, K083301 .

Comparison of Intended Use: .

The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate have the exact same intended use statements:

The modified and the unmodified devices are each indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.

Duration of Use:

The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate can both be used in procedures lasting up to 6 hours in duration.

Comparison of Labeling:

The labeling used for the modified device is similar to the labeling used with the predicate device, although some labeling is revised to provide greater clarity for the use. The revised instructions labeling accurately presents the directions that are necessary for the end-user to employ the device in a safe and effective manner. Terumo submits that the labeling complies with applicable regulations in those regions where the device is to be distributed.

Comparison of Operation and Technology:

The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate utilize the cxact same technologies and principles of operation. The technology of the modified ince is not impacted by the modifications made to the subject device.

Comparison of Design:

With respect to the design of the modified Sains™ Soft-Flow Extended Aortic Cannula, changes to the device have been made since the original submission - each of which Callina determined to maintain the integrity of the product without impacting the safety and effectiveness of the device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" are the words "Cardiovascular Systems" in a smaller, sans-serif font. The logo is black and white.

The Sarns™ Soft-Flow® Extended Cannula is 14.5 inches long, one-half inch longer than the 14 inch predicate device. This half inch difference in length is not clinically significant. Longer tubing allows clearance from and, thus improved visualization of, the surgical field, particularly in larger patients.

The Sarns™ Soft-Flow® Extended Cannula has a positional suture ring and depth . markings on the tubing. The positional suture ring and centimeter depth markings are used to gauge insertion depth of the cannula and provide a location to tighten the sutures around the cannula to hold it in place. These features are comparable in function to the predicate device which has a suture bulb to gauge insertion depth of the cannula and provide a location to tighten the sutures around the cannula to hold it inc place.

The Sarns" Soft-Flow "Extended Cannula has a straight tip rather than an angled tip like the predicate devices. A straight tip design generally produces a slightly lower pressure drop (the difference between the pressure entering the camula and the pressure leaving the cannula) compared to the same size angled tip. A lower pressure drop is desirable because there is less resistance to flow and thus less potential for hemolysis.

The material that is used for the luer-threaded port cap is changed from polypropylene to ABS plastic. Terumo Cardiovascular Systems maintains a complete profile of all biocompatibility data that fully supports the suggested elements of FDA General Program Memorandum #G95-1: Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part I: Evaluation and Testing,"


Comparison of Materials : ----------------------------------------------------------------------------------------------------------------------------------------------------

The materials of construction used in the modified Sarns™ Soft-Flow® Extended Aortic Cannula are the same generic materials that are used in the predicate device with the exception of a cap material change from polypropylene to ABS plastic.

Comparison of Performance:

Terunio Cardiovascular has conducted performance studies with the modified devices to ensure that they all continue to satisfy appropriate device performance specifications and to ensure that they satisfy customer needs. Specifically, performance studies were conduced to ensure force at break, liquid and air leak, hemolysis, pressure drop, exit velodity, ink adhesion, and biocompatibility testing passed pre-determined acceptance criteria. There are no appreciable differences in performance between the subject devices and the predicate devices. The nonclinical tests demonstrate the subject device is as safe and effective as the predicate devices.

Conclusion:

The information and data included in the 510(k) notice demonstrate the Sams™ Soft-Flow Extended Aortic Cannula is substantially equivalent to the predicate devices for perfusion of the ascending aorta during cardiopulmonary bypass for up to 6 hours of use.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Image /page/3/Picture/3 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The logo is simple and recognizable, conveying a sense of authority and trustworthiness.

February 27, 2014

Terumo Cardiovascular Systems Corporation Mr. Garry A. Courtney Corporate Director, Regulatory Affairs 6200 Jackson Rd. Ann Arbor, MI 48103

Re: K132811

Trade/Device Name: Sams Soft-Flow Extended Aortic Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: January 29, 2014 Received: January 30, 2014

Dear Mr. Courtney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subiect to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{4}------------------------------------------------

Page 2 - Mr. Garry A. Courtney

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

MADA

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" are the words "Cardiovascular Systems" in a smaller, sans-serif font.

SECTION 4 - INDICATION FOR USE Sarns™ Soft-Flow® Extended Aortic Cannula

510(k) Number: K132811

Sams™ Soft-Flow® Extended Aortic Cannula Device Name:

Indications for Use:

Device Intended Use - The device is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to six hours of use.

Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

or

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. FDA

15

:

§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.

(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).