K Number
K132729

Validate with FDA (Live)

Device Name
ISO-GARD MASK
Date Cleared
2014-01-09

(128 days)

Product Code
Regulation Number
868.5430
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ISO-Gard® Mask is intended to be used to scavenge waste anesthetic gases from patients during recovery from general anesthesia and to provide supplemental oxygen.

The ISO-Gard® Mask helps to reduce the amount of anesthetic agents released to the work environment of the healthcare worker.

Device Description

The ISO-Gard Mask system is an oxygen delivery mask that actively scavenges waste anesthetic gases (WAGS) exhaled by patients recovering from surgery in the Post-Anesthetic Care Unit (PACU). Vacuum/suction for scavenging of WAGS is provided by the institution's regulated vacuum source. The proposed device allows for the delivery of supplemental / therapeutic oxygen to patients to aid in their recovery while reducing the amount of patient expelled waste anesthetic agents released to the work environment of the healthcare workers. The mask can be used with or without suction / vacuum to function as a standard oxygen mask with an ETCO2 monitoring port.

AI/ML Overview

ISO-Gard® Mask Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

TestTest ObjectiveAcceptance CriteriaReported Device Performance
Oxygen DeliveryTo evaluate the oxygen delivery performance at variable oxygen flow rates and vacuum levels at standard Tidal Volumes of 500 ml without the use of N₂OThe delivered oxygen percentage using the ISO-Gard Mask must be equal to or greater than a standard medium concentration oxygen mask for all vacuum settingsThe device met this criterion.
ScavengingTo evaluate the scavenging performance at variable oxygen flow rates and vacuum levels at standard Tidal Volumes of 500 ml with N₂ON₂O levels must be lower than with a standard medium concentration oxygen maskThe device met this criterion.
ETCO₂To evaluate the ETCO₂ performance in simulated conditionsThe traces/waveforms during testing must be distinct and generated consistentlyThe device met this criterion.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the sample size used for the test set.

The data provenance is not explicitly stated as retrospective or prospective, nor does it specify the country of origin. However, the tests are described as nonclinical performance testing and simulated conditions, suggesting laboratory-based studies rather than patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of experts to establish ground truth for the test set. The performance tests appear to be based on objective measurements against established technical standards for oxygen delivery, scavenging efficacy, and ETCO2 waveform characteristics, rather than expert judgment.

4. Adjudication Method for the Test Set

Since there is no mention of experts or human interpretation in establishing the ground truth for the test set, there is no adjudication method described (e.g., 2+1, 3+1, none). The tests rely on quantifiable measurements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was conducted. The device is a physical mask for gas scavenging and oxygen delivery, not an imaging or diagnostic AI tool that would typically involve human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This question is not applicable. The ISO-Gard® Mask is a medical device for gas delivery and scavenging, not an algorithm or AI system. Therefore, a "standalone algorithm performance" study is not relevant. The performance tests evaluate the physical functioning of the device.

7. Type of Ground Truth Used

The ground truth for the performance tests appears to be based on objective technical standards and measurements:

  • For Oxygen Delivery: Comparison against "a standard medium concentration oxygen mask" and measurements of delivered oxygen percentage.
  • For Scavenging: Measurement of N₂O levels compared against "a standard medium concentration oxygen mask."
  • For ETCO₂: Evaluation of generated traces/waveforms for distinctness and consistency.

8. Sample Size for the Training Set

The document does not mention a training set. This is expected as the ISO-Gard® Mask is a physical medical device and not an AI or machine learning system that requires training data.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no mention of a training set for this device.

{0}------------------------------------------------

K132729

510(k) Submission Tab 007 - Section 011 - 510(k) Summary ISO-Gard® Mask

JAN - 9 2014

510(k) SUMMARY

Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8050 919-433-4996 Fax:

Contact Person

Angela Bouse Senior Regulatory Affairs Specialist

Date Prepared

December 11, 2013

Device Name

Trade Name:ISO-Gard® Mask
Classification Name:Apparatus, gas-scavenging
Product Code:CBN
Regulation Number:868.5430
Classification:II
Classification Panel:Anesthesiology

Predicate Device

This submission demonstrates substantial equivalence to the predicate device ISO-GARD ClearAir Mask - K123176

Device Description

The ISO-Gard Mask system is an oxygen delivery mask that actively scavenges waste anesthetic gases (WAGS) exhaled by patients recovering from surgery in the Post-Anesthetic Care Unit (PACU). Vacuum/suction for scavenging of WAGS is provided by the institution's regulated vacuum source. The proposed device allows for the delivery of supplemental / therapeutic oxygen to patients to aid in their recovery while reducing the amount of patient expelled waste anesthetic agents released to the work environment of the healthcare workers. The mask can be used with or without suction / vacuum to function as a standard oxygen mask with an ETCO2 monitoring port.

{1}------------------------------------------------

Indications for Use

The ISO-Gard® Mask is intended to be used to scavenge waste anesthetic gases from patients during recovery from general anesthesia and to provide supplemental oxygen.

The 1SO-Gard® Mask helps to reduce the amount of anesthetic agents released to the work environment of the healthcare worker.

Patient Population

Patients recovering from general anesthesia in the PACU.

Environments of use

The environment of use is - Post-operative Care Units (PACU) in hospital, sub-acute facilities.

Contraindications

None

Substantial Equivalence

The proposed device is substantially equivalent to the predicate devices:

ComparativeCharacteristicsPredicateK123176ISO-GARD® ClearAir™ MaskProposedISO-Gard®Mask
Classification NameApparatus, gas scavengingSame
Product Code / CFRCBN868.5430SecondaryCCK – Gaseous-Phase Carbon Dioxide GasAnalyzer868.1400Same
Indications for UseThe ISO-GARD® ClearAir™ Mask is intendedto be used to scavenge waste anesthetic gasesfrom patients during recoveryfrom general anesthesia and to providesupplemental oxygen.The ISO-GARD® ClearAir™ Mask helps toreduce the amount of anesthetic agents released tothe work environment of the healthcare worker.Same
Trade NameISO-GARD® ClearAir™ MaskISO-Gard® Mask
Environment of UseHospital, sub-acute facilities PACUSame
Patient PopulationPatients recovering from general anesthesiaSame
and may need supplemental oxygenAdults
ContraindicationsNoneSame
Basic ComponentsMaskOxygen delivery tubingVacuum (scavenging) tubingMask Manifold controlling oxygen deliveryand scavengingSame
Design, Features, and Specifications
MaskFlexible oxygen mask with sealing foamSame
Method to hold mask on patient for sealElastic band / strapSame
Tubing to deliver oxygenStandard oxygen tubingSame
Connects to ETCO2 monitorYesSame
Connector to sampling lineStandard female luer lockSame
Method of separating gas flowsDivided manifold for separating vacuum and oxygen delivery and then a separate sampling portSame
Safety features
Excess negative pressureContains entrainment valves if the negative pressure from vacuum is too greatValves are one-way flapper/diaphragm valves that open with minimal negative pressure or flowSame
Excess positive pressureContains entrainment valves if patient's inhalation is greater than the supply of the oxygenSame
Method to assist in sealingFoam pad around bridge of nose to assist in sealing of the maskSame
Method to separate oxygen delivery from scavengingMask manifold body is a divided adapter which has an oxygen inlet and a scavenging outletSame
Oxygen sourceWall oxygenSame
Vacuum sourceWall vacuumSame
Port for sampling end tidal CO2Port connector on exhalation side of Mask Manifold adapterSame
Typical oxygen delivered flow ratesUp to 10 lpmSame
Oxygen at various Oxygen flow rates and Vacuum settingDelivered oxygen equal or greater than oxygen concentration maskSame
Mask sizesAdultSame
PerformanceNoneSame
StandardsSame
Shelf LifeNo shelf life1 year shelf life
Patient Contacting Materials
MaskPVCSame
Star-Lumen OxygenTubePVCSame
Connector, Star-Lumen Oxygen TubePVCSame
Gasketing Foamw/AdhesiveNatural ester foam with acrylic pressure sensitiveadhesiveSame
Tethered CapThermoplastic ElastomerSame
One-way inhalationvalvesPolyisopreneSame
One-way Valve BodyPolystyrene, Trans BlueSame
Oxygen DeliveryPort AdaptorPolypropyleneSame
White Elastic StrapWhite Polyester/PolyisopreneSame
Mask ManifoldPolystyreneSame
Suction/ExhalationPortPolystyreneSame
Oxygen PortConcentratorPVCSame

{2}------------------------------------------------

.

{3}------------------------------------------------

Comparison to Predicate Device

The proposed ISO-Gard Mask is substantially equivalent to the predicate device with respect to indications for use, technology, materials, and construction. The proposed change is to add a one year shelf life to the ISO-Gard Mask labeling.

. Indications for Use -

The indications for use are identical to the predicate.

Technology and construction -●

The proposed device design, drawings, components, accessories, materials, packaging and product configurations remain unchanged. The proposed change is to add a one year shelf life to the ISO-Gard Mask labeling.

Environment of use -.

The environment of use is identical to the predicate.

{4}------------------------------------------------

. Patient Population -

The patient population is identical to the predicate.

. Materials -All patient contacting materials are identical to the predicate.

. Performance Testing

Nonclinical Performance Testing Summary

TestTest ObjectiveAcceptance Criteria
Oxygen DeliveryTo evaluate the oxygendelivery performance atvariable oxygen flow ratesand vacuum levels atstandard Tidal Volumes of500 ml without the use ofN₂OThe delivered oxygenpercentage using the ISO-GardMask must be equal to orgreater than a standardmedium concentration oxygenmask for all vacuum settings
ScavengingTo evaluate thescavenging performanceat variable oxygen flowrates and vacuum levels atstandard Tidal Volumes of500 ml with N₂ON₂O levels must be lower thanwith a standard mediumconcentration oxygen mask
ETCO₂To evaluate the ETCO₂performance in simulatedconditionsThe traces/waveforms duringtesting must be distinct andgenerated consistently

Conclusion

The ISO-Gard Mask has the same indications for use, technological characteristics, and constructions as the predicate. Performance test results demonstrate that the proposed device does not raise new questions of safety and effectiveness and because an acceptance criteria has been met, the device can be found substantially equivalent.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 9, 2014

Teleflex Medical, Inc. Angela Bouse Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709

Re: K132729

Trade Name: ISO-Gard® Mask Regulation Number: 21 CFR 868.5430 Regulation Name: Gas-scavenging apparatus Regulatory Class: Class II Product Code: CBN Dated: December 11, 2013 Received: December 12, 2013

Dear Ms. Bouse:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Ms. Bouse

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/6/Picture/8 description: The image contains a signature and some text. The text includes the name "Tejashri Purohit-Sheth, M.D.", followed by the title "Clinical Deputy Director". The acronym "DAGRID" is also present. The word "FOR" is at the bottom right.

Erin Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use Statement

Page I of I

510(k) Number:

(To be assigned)

Device Name:

ISO-Gard® Mask

Indications for Use:

The ISO-Gard® Mask is intended to be used to scavenge waste anesthetic gases from patients during recovery from general anesthesia and to provide supplemental oxygen.

The ISO-Gard® Mask helps to reduce the amount of anesthetic agents released to the work environment of the healthcare worker.

Prescription Use XX (Part 21 CFR 801 Subpart D) or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K132729

Nayan J. Patel 2014.01.07 15:25:50 05'00'

Teleflex Medical, Inc

§ 868.5430 Gas-scavenging apparatus.

(a)
Identification. A gas-scavenging apparatus is a device intended to collect excess anesthetic, analgesic, or trace gases or vapors from a patient's breathing system, ventilator, or extracorporeal pump-oxygenator, and to conduct these gases out of the area by means of an exhaust system.(b)
Classification. Class II (performance standards).