K Number
K132708

Validate with FDA (Live)

Date Cleared
2013-12-20

(113 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Biomac Fiducial Marker is indicated for use to radiographically mark soft tissue for future surgical procedures.

Device Description

The BiomarC Fiducial Marker is a sterile, pyrogen free, single patient use, carbon/metallic composite discrete marker that is visible on standard radiographs and Magnetic Resonance Imaging (MRI). The marker can be delivered with the preloaded delivery device system or through commercially available, compatible delivery devices chosen by the user.

AI/ML Overview

This document is a 510(k) Pre-market Notification Summary for the BiomarC Fiducial Marker. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study proving a device meets specific performance acceptance criteria for a new clinical efficacy claim. Therefore, much of the requested information regarding clinical study design and results is NOT applicable or available in this document.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not present specific quantitative acceptance criteria or detailed performance metrics in the way one would expect for a study proving clinical efficacy (e.g., sensitivity, specificity, accuracy for a detection algorithm). Instead, the "performance" discussed relates to a qualitative assessment of technological characteristics and safety.

Feature/CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Technological EquivalenceEquivalent to predicate devices (BiomarC Tissue Marker (K063193) and EchoTip® Ultra Fiducial Needle (K111895))."The technological characteristics are equivalent to the predicate devices."
Risk AssessmentRisks associated with modifications are acceptable."A Failure Modes and Effects Analysis (FMEA) was performed in order to assess the risks associated with the modifications introduced." (Conclusion implicitly acceptable as the device received clearance).
BiocompatibilityBiocompatibility is acceptable."A biocompatibility... evaluation confirmed that the modified device, BiomarC Fiducial Marker, was substantially equivalent to the predicate devices."
VisibilityVisible on standard radiographs and Magnetic Resonance Imaging (MRI)."A... visibility... evaluation confirmed that the modified device, BiomarC Fiducial Marker, was substantially equivalent to the predicate devices." (Device description states: "carbon/metallic composite discrete marker that is visible on standard radiographs and Magnetic Resonance Imaging (MRI).")
MR Safety / CompatibilityMR Safety / Compatibility is acceptable."A... MR safety / compatibility... evaluation confirmed that the modified device, BiomarC Fiducial Marker, was substantially equivalent to the predicate devices."
Sterilization & PackagingSterilization and packaging maintain sterility and shelf life, equivalent to predicates."A... sterilization and packaging / shelf life adoption evaluation confirmed that the modified device, BiomarC Fiducial Marker, was substantially equivalent to the predicate devices."
Indication for UseTo radiographically mark soft tissue for future surgical procedures."The BiomarC Fiducial Marker is indicated for use to radiographically mark soft tissue for future surgical procedures." (Consistent with the stated indication).

2. Sample Size Used for the Test Set and Data Provenance

Not applicable. The document describes a substantial equivalence submission, not a study with a test set of patient data to assess a clinical performance metric. The "evaluations" mentioned are likely preclinical (e.g., material testing, MR compatibility testing) rather than human subject clinical trials.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This information is relevant for studies assessing diagnostic accuracy or image interpretation, which is not the primary focus of this 510(k) submission.

4. Adjudication Method for the Test Set

Not applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

Not applicable. This is not an AI-assisted device, nor does the submission describe an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a physical medical device (fiducial marker), not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the evaluations performed (biocompatibility, visibility, MR safety, sterilization, packaging), the "ground truth" would likely be based on:

  • Biocompatibility: In vitro and/or in vivo testing against established ISO standards and material specifications.
  • Visibility: Imaging phantom studies and comparison to predicate device visibility characteristics.
  • MR Safety/Compatibility: Testing according to ASTM standards for MR safety (e.g., heating, artifact, force).
  • Sterilization and Packaging: Adherence to ISO standards for sterilization and package integrity testing over shelf life.

These are not "clinical ground truths" derived from patient data in the typical sense.

8. The Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is not an algorithm being developed.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

{0}------------------------------------------------

510(k) SUMMARY

DEC 2 0 2013

Submitter's Name, Address and Date of Submission

Andrew Adams Director, Regulatory Affairs and Quality Assurance Carbon Medical Technologies, Inc. 1290 Hammond Road Saint Paul, MN 55110

Phone: 651-653-8512 Fax: 651-407-1975

Submitted: August 28, 2013

Device Name

Trade Name:BiomarC Fiducial Marker
Common Name:Tissue Marker
Classification Name:Implantable Clip (21 CFR 878.4300, NEU)

Predicate Devices

BiomarC Tissue Marker (K063193) EchoTip® Ultra Fiducial Needle (K111895)

Indication for Use

The BiomarC Fiducial Marker is indicated for use to radiographically mark soft tissue for future surgical procedures.

Device Description

The BiomarC Fiducial Marker is a sterile, pyrogen free, single patient use, carbon/metallic composite discrete marker that is visible on standard radiographs and Magnetic Resonance Imaging (MRI). The marker can be delivered with the preloaded delivery device system or through commercially available, compatible delivery devices chosen by the user.

Technological Characteristics and Performance

The technological characteristics are equivalent to the predicate devices. A Failure Modes and Effects Analysis (FMEA) was performed in order to assess the risks associated with the modifications introduced. A biocompatibility, visibility, MR safety / compatibility and sterilization and packaging / shelf life adoption evaluation confirmed that the modified device. BiomarC Fiducial Marker, was substantially equivalent to the predicate devices.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Carbon Medical Technologies Incorporated Mr. Andrew J. Adams Director, Regulatory Affairs & Quality Assurance 1290 Hammond Road Saint Paul, Minnesota 55110-5876

December 20, 2013

Re: K132708

Trade/Device Name: BiomarC Fiducial Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: November 15, 2013 Received: November 18, 2013

Dear Mr. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{2}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

For

Sincerely vours.

Radiological Health

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132708

. .. ... .. .. .. .. .. .. ..

Device Name BiomarC Fiducial Marker

Indications for Use (Describe)

The Biomac Fiducial Marker is indicated for use to radiographically mark soft tissue for future surgical procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: December 31, 2013

See PRA Statement on last page.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.