K Number
K132658

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-11-27

(93 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Weck® Ligating Clips are intended for use in procedures involving vessels or anatomic structures for which the surgeon determines ligating clips are the best choice. Surgeons should select the size, type and material of the clip based upon their experience, judgment and needs.

Device Description

Weck® Metal Ligating Clips are single-use, non-absorbable, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. Each product line possesses a unique design necessary for compatibility with its appropriate Weck® ligating clip applier. The clips are available in a range of sizes, allowing the end user to ligate a wide range of vessels and tissue structures. Weck® Metal Ligating Clips are manufactured from medical grade titanium, tantalum or stainless steel alloys and are provided prepackaged in color-coded cartridges, which are provided as single-use, sterile devices.

Accessories to the Weck® Metal Ligating Clips include manual clip appliers and removers for use in both general open and endoscopic procedures. Both the appliers and removers are multiple use, non-sterile devices that require cleaning and sterilization prior to each use.

AI/ML Overview

The Weck® Metal Ligating Clips are non-absorbable, non-active implantable devices used for vessel or tissue ligation in general surgical procedures. The clips are made from medical-grade titanium, tantalum, or stainless steel alloys.

Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
"MR Conditional" Claim"MR Conditional" for environments of 3-Tesla or less
Safety and Compatibility in MR EnvironmentDemonstrated safe use in 3-Tesla or less MR environments
Equivalence to Predicate DevicesSubstantially equivalent in performance to K982313 and K841547
No New Safety/Effectiveness IssuesModifications do not introduce new issues of safety and effectiveness

2. Sample Size and Data Provenance

  • Test Set Sample Size: Not explicitly stated. The document mentions "Non-clinical performance testing" was conducted, but specific sample sizes for this testing are not provided.
  • Data Provenance: The studies were non-clinical performance testing conducted by the manufacturer, Teleflex Medical, Incorporated, to support a 510(k) submission to the FDA in the USA. This is considered prospective, manufacturer-conducted testing for regulatory approval.

3. Number of Experts and Qualifications for Ground Truth

  • Not applicable. The ground truth for this device (MR Conditional claim, material compliance, and substantial equivalence) was established through non-clinical performance testing and comparison to existing predicate devices, rather than expert interpretation of medical images or data.

4. Adjudication Method

  • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation of data where consensus among experts is needed. This submission relies on objective engineering and material testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic aids or AI systems that assist human readers in tasks like image interpretation, and the Weck® Metal Ligating Clips are a physical implantable device.

6. Standalone Performance Study

  • Yes, a standalone study was done in the sense that the device's "MR Conditional" performance was tested independently of human intervention. The "Non-clinical performance testing" was conducted on the device itself to determine its behavior in an MR environment. However, this is not "algorithm only" performance as often seen in AI/software substantial equivalence discussions, but rather direct device performance testing.

7. Type of Ground Truth Used

  • Engineering and Material Standards: The ground truth for the "MR Conditional" claim was based on established engineering standards (ASTM F2503-08) and FDA guidance for medical devices in MR environments.
  • Comparison to Predicate Devices: The ground truth for overall safety and effectiveness was established by demonstrating substantial equivalence to previously cleared predicate devices (K982313 and K841547), implying that if the new device performs similarly to devices already deemed safe and effective, it is also safe and effective.
  • Biological Compatibility: Patient-contacting materials compliance with ISO10993-1 provides ground truth for biocompatibility.

8. Sample Size for the Training Set

  • Not applicable. This device is not an AI/ML algorithm that requires a training set. The performance data is based on physical testing of the device itself.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as no training set was used.

{0}------------------------------------------------

K132658 Page 1/3

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Weck® Metal Ligating Clips

NOV 2 7 2013

A. Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8065 919-433-4996 Fax:

B. Contact Person

. .

Ashlea Ricci, RAC Senior Regulatory Affairs Specialist

Lorraine DeLong Manager RA/QE Surgical

C. Date Prepared

August 23, 2013

D. Device Name

Trade Name:Weck® Metal Ligating Clips
Common Name:Implantable Clips
Classification Name:Clips, Implantable

E. Device Description

Weck® Metal Ligating Clips are single-use, non-absorbable, non-active implantable devices designed for use in general surgical procedures that require vessel or tissue ligation. Each product line possesses a unique design necessary for compatibility with its appropriate Weck® ligating clip applier. The clips are available in a range of sizes, allowing the end user to ligate a wide range of vessels and tissue structures. Weck® Metal Ligating Clips are manufactured from medical grade titanium, tantalum or stainless steel alloys and are provided prepackaged in color-coded cartridges, which are provided as single-use, sterile devices.

Accessories to the Weck® Metal Ligating Clips include manual clip appliers and removers for use in both general open and endoscopic procedures. Both the appliers and removers are multiple use, non-sterile devices that require cleaning and sterilization prior to each use.

{1}------------------------------------------------

F. Indications for Use

Weck® Ligating Clips are intended for use in procedures involving vessels or anatomic structures for which the surgeon determines ligating clips are the best choice. Surgeons should select the size, type and material of the clip based upon their experience, judgment and needs.

G. Contraindications

Weck® Ligating Clips are not intended for use as a contraceptive tubal occlusion device.

Weck® Ligating Clips are contraindicated for use in ligating the renal artery during laparoscopic donor nephrectomies.

H. Environmental Conditions

Weck® Metal Ligating Clips are "MR Conditional" up to and including 3-Tesla MR environments.

I. Substantial Equivalence

The proposed Weck® Metal Ligating Clips are substantially equivalent to the predicate devices:

Predicate DeviceManufacturer510(k) No.Date Cleared
Horizon™ Ligating ClipTeleflex Medical(Weck)K98231308/10/1998
Hemoclip® Traditional LigatingClipTeleflex Medical(Weck)K84154708/14/1984

J. Comparison To Predicate Devices

The proposed Weck® Metal Ligating Clips have the same technology and functional characteristics as the predicate system. The proposed modifications include the addition of magnetic resonance (MR Conditional) claims, addition of a contraindication for renal donor nephrectomy, addition of cleaning and sterilization instructions for reusable instrumentation, two line extensions, a change in packaging materials, and a change in the sterilization method of the Horizon™ Ligating Clips.

K. Materials

All patient contacting materials are in compliance with ISO10993-1.

L. Technological Characteristics

A comparison of the technological characteristics of the proposed Weck® Metal Ligating Clips and the predicates has been performed. The results of this comparison demonstrate that the Weck® Metal Ligating Clips are equivalent to the marketed predicate devices.

{2}------------------------------------------------

K132658 Page 3/3

Traditional 510(k) Section 7 - Summary of Safety and Effectiveness

Weck® Metal Ligation Clips

M. Performance Data

Non-clinical performance testing was conducted to support the "MR Conditional" labeling claim in accordance with ASTM F2503-08. Standard Practice for Marketing Medical Devices and Other Items for Safety in the Magnetic Resonance Environment as well as the FDA quidance document. Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment, dated August 21, 2008. The results of these tests demonstrate that the Weck® Ligating Clips are "MR Conditional" and may be safely used in environments of 3-Tesla or less.

N. Conclusion

Based upon the comparative test results, the proposed Weck® Metal Ligating Clips are substantially equivalent in performance to the predicate devices cleared to market via 510(k) K982313 and K841547. The modifications made to the proposed Weck® Metal Ligating Clips do not introduce any new issues of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, clutching a caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Teleflex Medical Incorporated Ms. Ashlea Ricci, RAC Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, North Carolina 27709

November 27, 2013

Re: K132658

Trade/Device Name: Weck® Metal Ligation Clips Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: August 23, 2013 Received: August 30, 2013

Dear Ms. Ricci:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Ashlea Ricci, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

For

Sincerely yours,

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices . Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

Page 1 of 1

510(k) Number:

-K132658

Device Name:

Weck® Metal Ligation Clips

Indications for Use:

Weck® Ligating Clips are intended for use in procedures involving vessels or anatomic structures for which the surgeon determines ligating clips are the best choice. Surgeons should select the size, type and material of the clip based upon their experience, judgment and needs.

Prescription Use ਸ (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krause -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132658

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.