K Number
K132602

Validate with FDA (Live)

Date Cleared
2014-02-04

(168 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The single use dialyzer with pre-attached blood tubing set is indicated for use with the Fresenius 2008 Series Hemodialysis systems for the treatment of acute and chronic renal failure.

Device Description

The proposed device provides for the treatment of acute or chronic renal failure when used with the commercially available Fresenius 2008 Series hemodialysis systems. The device is a single use high flux (permeability) hollow-fiber dialyzer pre-attached to a blood tubing set.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the NxStage® Streamline Express for Fresenius 2008 Series Hemodialysis Systems. This submission focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study with a test set, expert ground truth, or a multi-reader, multi-case study. Therefore, many of the requested criteria cannot be directly extracted from the provided document.

Here's a breakdown of what can and cannot be answered:

1. A table of acceptance criteria and the reported device performance

The document provides a comparative table of technological characteristics and in-vitro performance between the proposed device and predicate devices. The acceptance criteria are implicitly that the proposed device should have "Same" or comparable performance to the predicate devices, as well as meeting general safety and performance standards for hemodialysis components.

ParameterAcceptance Criteria (Implicit: "Same" as predicate in K113023)Reported Device Performance (NxStage Streamline Express)
Dialyzer
Number of fibers10,900 ± 200Same
Fiber internal diameter200 μmSame
Fiber wall thickness30 μmSame
Fiber length23 cmSame
Effective surface area1.6 m²Same
Priming volume91 mlSame
Max. Transmembrane Pressure500 mmHgSame
In-vitro Performance
Pressure Drop Blood Compartment<100 mmHg at Qb=300mL/min, UF = 0 ml/minSame
Pressure Drop Dialysate Compartment17 mmHg at Qd = 200 ml/minSame
Ultrafiltration Flow Rates (Qb 200, TMP 100)60 ml/minSame
Ultrafiltration Flow Rates (Qb 200, TMP 300)86 ml/minSame
Clearance - Urea (Qb=100, Qd=50, UF=0)50 ml/minSame
Clearance - Vitamin B12 (Qb=100, Qd=50, UF=0)44 ml/minSame
Sieving Coefficients - Albumin<0.005Same
Sieving Coefficients - Urea1.0Same
Sieving Coefficients - Creatinine1.0Same
Sieving Coefficients - Vitamin B121.0Same
Blood Tubing Set
Patient connections (Male luer lock)Yes (as per K113023 & K080807)Same as K113023 & K080807
Blood tubing diameters (I.D.)4.55 - 4.8 mm (as per K113023 & K080807)Same as K113023 & K080807
Pump section I.D.8.0 mm (as per K113023 & K080807)Same as K113023 & K080807
Pump segment connectorsYes (as per K113023 & K080807)Same as K113023 & K080807
Heparin. line I.D.0.75 mm (as per K113023 & K080807)Same as K113023 & K080807
Saline lineYes (as per K113023 & K080807)Same as K113023 & K080807
Access lineYes (as per K113023 & K080807)Same as K113023 & K080807
Female luer connectorYes (as per K113023 & K080807)Same as K113023 & K080807
Female luer capYes (as per K113023 & K080807)Same as K113023 & K080807
Y-SiteYes (as per K113023 & K080807)Same as K113023 & K080807
Needless injection siteYes (as per K113023 & K080807)Same as K113023 & K080807
ClampsYes (as per K113023 & K080807)Same as K113023 & K080807
Priming volume212 cc (K113023) or 109-119 cc (K080807)201 cc
ReversoCleared as accessory K994306 / Included as standardReverso is cleared as blood tubing set accessory K994306

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily relies on bench testing and comparison to predicate devices, not on a clinical test set with patient data. No sample size for a test set of patient data, data provenance, or retrospective/prospective nature is mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a 510(k) submission for a medical device (dialyzer and tubing set), not an AI/software device that requires expert-established ground truth for a test set. The evaluation is based on engineering and performance specifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set with human interpretation requiring adjudication is discussed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device. The "performance" assessment is based on its physical and functional characteristics (e.g., clearance rates, pressure drops) through bench testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established through in-vitro performance testing and comparison to the known engineering specifications and performance of legally marketed predicate devices. For example, clearance data is measured directly in a lab setting, not through expert consensus or clinical outcomes for this type of submission.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that requires a "training set."

9. How the ground truth for the training set was established

Not applicable. No training set is involved.

{0}------------------------------------------------

NxStage Medical, Inc. NxStage® Streamline Express 510(k) Premarket Notification Submission

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92

Date:
A. DateDecember 20, 2013
B. Submitter's Information:Name:NxStage Medical, Inc.
Address:350 Merrimack StreetLawrence, MA 01843United States
FDA EstablishmentOwner/Operator Number:9045797
Contact Person:Mary Lou StroumbosDirector, Regulatory Affairs
Phone:(978) 687-4872
Fax:(978) 687-4750
Manufacturing Site:MEDIMEXICO, S. DE R.L. DE C.V.Av. Valle imperial No. 10523Parque industrial Valle SurTijuana, B.C., Mexico 22180
FDA EstablishmentRegistration Number:9616074
Sterilization Site:Steris CorporationIsomedix Services, Inc.1000 S. Sarah PlaceOntario, CA 91761
FDA EstablishmentRegistration Number:Contract sterilizer
C. Device Name:
Trade/Proprietary Name:NxStage Streamline Express for Fresenius 2008Series Hemodialysis Systems
Device:Set, tubing, blood, with and without anti-regurgitationvalve
Regulation Description:Hemodialysis System and Accessories

510(k) Premarket Notification NxStage Medical, Inc.

FEB - 4 2014

{1}------------------------------------------------

NxStage Medical, Inc. NxStage® Streamline Express 510(k) Premarket Notification Submission

Regulation MedicalSpecialty:Gastroenterology/Urology Devices
Review Panel:Gastroenterology/Urology
Product Code:KDIFJKKOC
Submission Type:510(k)
Regulation Number:876.5860
Device Class:II

D. Predicate Devices:

Streamline Airless System Set with Pre-Attached Dialyzerfor B.Braun Dialog Series Hemodialysis SystemsK113023
Blood Tubing SetsK080807
ReversoK994306

E. Substantial Equivalence:

The proposed NxStage dialyzer with pre-attached blood tubing set is substantially equivalent to the identified predicates.

F. Device Description/Indications for Use:

The proposed device provides for the treatment of acute or chronic renal failure when used with the commercially available Fresenius 2008 Series hemodialysis systems. The device is a single use high flux (permeability) hollow-fiber dialyzer pre-attached to a blood tubing set.

Indications for use:

The single use dialyzer with pre-attached blood tubing set is indicated for use with the Fresenius 2008 Series hemodialysis systems for the treatment of acute and chronic renal failure.

G. Technological Characteristics:

The proposed device has the same technological characteristics and is similar in design and configuration as compared to the predicate devices.

{2}------------------------------------------------

K132602
Page 3 of 6

Device Technological Characteristics Comparison Table
ParameterProposed DeviceStreamline Express for Fresenius 2008 SeriesHemodialysis SystemsPredicate DeviceStreamline Airless System Set withPre-Attached Dialyzer for B. Braun Dialog SeriesHemodialysis Systems(K113023)
Intended useThe single use dialyzer with pre-attached bloodtubing set is indicated for use with the Fresenius2008 Series Hemodialysis Systems for thetreatment of acute and chronic renal failure.The single use blood tubing set with pre-attacheddialyzer is indicated for use with the B. Braun DialogSeries Hemodialysis Systems for the treatment ofacute and chronic renal failure.
Principle of OperationRemoval of solutes via diffusion or convectionRemoval of solutes via diffusion or convection
Product configurationArterial and venous side ports on filter end caps(ports configured perpendicular to blood flow)Arterial and venous side ports on filter end caps(ports configured perpendicular to blood flow)
How suppliedPre-connected to the Streamline Blood Tubing Setwith standard Hansen connectorsPre-connected to the Streamline Blood Tubing Setwith standard Hansen connectors
Dialyzer
Number of fibersSame10,900 ± 200
Fiber internal diameterSame200 μm
Fiber wall thicknessSame30 μm
Fiber lengthSame23 cm
Effective surface areaSame1.6 m²
Priming volumeSame91 ml
Max. TransmembranePressureSame500 mmHg
In-vitro Performance
Pressure Drop BloodCompartment (mmHg)Same<100 mmHgat Qb=300mL/minUF = 0 ml/min
Pressure DropDialysate Compartment(mmHg)Same17 mmHg atQd = 200 ml/min
Ultrafiltration FlowRates at Specific BloodFlow Rates and TMPsSame60 ml/min (Qb 200, TMP 100)86 ml/min (Qb 200, TMP 300)(Qb in ml/min; TMP in mmHg)
Clearance data(ml/min)
UreaSame50 ml/min atQb = 100 ml/minQd = 50 ml/minUF = 0 ml/min
Vitamin B12Same44 ml/min atQb = 100 ml/minQd = 50 ml/minUF = 0 ml/min1
Sieving CoefficientsSame
Albumin<0.005
Urea1.0
Creatinine1.0
Vitamin B121.0

510(k) Premarket Notification
NxStage Medical, Inc.

Page 24

{3}------------------------------------------------

K132602
Page 4 of 6

{4}------------------------------------------------

K132602
Page 5 of 6

Device Technological Characteristics Comparison Table
Proposed DeviceStreamline Express forFresenius 2008 SeriesHemodialysis SystemsPredicate DeviceStreamline Airless SystemSet with Pre-AttachedDialyzer for B. Braun DialogSeries Hemodialysis Systems(K113023)Predicate Device.Blood Tubing SetStreamline Airless SystemSet with LockSiteNeedless Access Site(K080807)
Blood Tubing Set
Patient connections (Male luerlock)Same as K113023 & K080807YesYes
Blood tubing diameters (I.D.)Same as K113023 & K0808074.55 - 4.8 mm4.55 - 4.8 mm
Pump section I.D.Same as K113023 & K0808078.0 mm8.0 mm
Pump segment connectorsSame as K113023 & K080807YesYes
Heparin. line I.D.Same as K113023 & K0808070.75 mm0.75 mm
Saline lineSame as K113023 & K080807YesYes
Access lineSame as K113023 & K080807YesYes
Female luer connectorSame as K113023 & K080807YesYes
Female luer capSame as K113023 & K080807YesYes
Y-SiteSame as K113023 & K080807YesYes
Needless injection siteSame as K113023 & K080807YesYes
ClampsSame as K113023 & K080807YesYes
Priming volume201 cc212 cc109 - 119 cc
ReversoReverso is cleared as s bloodubing set accessory K994306The Reverso access flowreversing valve will be includedas standardAvailable for purchaseseparately

Page 26

{5}------------------------------------------------

H. Summary of Non-Clinical Test/Performance Testing - Bench

NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indication for use. Performance, verification and validation testing was conducted following the FDA's Guidance document entitled: Hemodialysis Blood Tubing Sets - Premarket Notification [510(k)] Submissions issued on: April 23, 2008 to characterize performance of the proposed device and the predetermined acceptance criteria was met. Results of this testing have documented that the proposed device is substantially equivalent to the predicate devices and is suitable for the labeled indication for use.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 4, 2014

NxStage Medical, Inc. Mary Lou Stroumbos Director, Regulatory Affairs 350 Merrimack Street Lawrence, MA 01843

· Re: K132602

Trade/Device Name: NxStage® Streamline® Express for Fresenius 2008 Series Hemodialysis Systems Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI, FJK, KOC Dated: December 20, 2013 Received: December 23, 2013

Dear Mary Lou Stroumbos,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{7}------------------------------------------------

Page 2 - Mary Lou Stroumbos

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132602

Device Name

NxStage® Streamline® Express for Fresenius 2008 Series Hemodialysis Systems

Indications for Use (Describe)

The single use dialyzer with pre-attached blood tubing set is indicated for use with the Fresenius 2008 Series Hemodialysis systems for the treatment of acute and chronic renal failure.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

...

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/8/Picture/11 description: The image shows the name "Herbert P. Lerner" in bold font. Below the name, the date and time "2014.02.04 07:08:35" are displayed, followed by "-05:00". The text is clear and easily readable.

FORM FDA 3881 (1/14)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”