K Number
K132542

Validate with FDA (Live)

Date Cleared
2013-11-15

(94 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TransGate™ Guide Catheters are intended to facilitate the placement of interventional devices into the coronary, peripheral, and neuro vasculature.

Device Description

Stryker Neurovascular's TransGate Guide Catheters are constructed with polytetrafluoroethylene (PTFE) on the inner lumen for lubricity, stainless steel wire reinforcement within the wall for trackability. torque transmission and strength, and polymer materials along the length of the catheter for support and flexibility. The catheter has an atraumatic tip, a hub for device connectivity and device handling, and a strain relief at the hub for kink resistance. The distal segment of the device shaft is radiopacified and. in addition, a radiopaque marker is embedded at the distal tip of the device to aid visualization under fluoroscopy. .

The Stryker Neurovascular TransGate Guide Catheter has straight (ST) and pre-shaped distal tips, available in 40 (40), and Multipurpose (MP) shapes. The TransGate Guide Catheters are offered in both 90 and 100 cm lengths and 5 and 6 French diameters.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Stryker Neurovascular TransGate™ Guide Catheter.

Important Note: The provided document is a 510(k) Summary for a medical device – a guide catheter specifically. These types of devices do not typically involve AI components or software algorithms that would require the typical "AI/ML acceptance criteria" found in software as a medical device (SaMD) clearances. Therefore, many of your requested points regarding AI/ML studies (like standalone algorithm performance, MRMC studies, training set details, ground truth for training) are not applicable to this particular device and document.

The "acceptance criteria" here refer to meeting performance standards for a physical medical device.


1. Table of Acceptance Criteria and Reported Device Performance

Performance TestAcceptance Criteria (Implicit)Reported Device Performance
Kink DistanceEstablished criteriaMet established criteria
Torque Degrees Rotation to KinkEstablished criteriaMet established criteria
Tip Shape RetentionEstablished criteriaMet established criteria
Distal Shaft FlexibilityEstablished criteriaMet established criteria
PushabilityEstablished criteriaMet established criteria
Tensile StrengthEstablished criteriaMet established criteria
Chemical CompatibilityEstablished criteriaMet established criteria
Corrosion ResistanceEstablished criteriaMet established criteria
ParticulateEstablished criteriaMet established criteria
Liquid LeakageEstablished criteriaMet established criteria
Air LeakageEstablished criteriaMet established criteria
Unscrewing TorqueEstablished criteriaMet established criteria
Resistance to OverridingEstablished criteriaMet established criteria
Stress CrackingEstablished criteriaMet established criteria
Ease of AssemblyEstablished criteriaMet established criteria
Luer GaugingEstablished criteriaMet established criteria
Luer Separation ForceEstablished criteriaMet established criteria
Freedom from leakage and damage under high static pressureEstablished criteriaMet established criteria
RadiopacityEstablished criteriaMet established criteria
Shelf Life Testing (Product and Packaging)Established criteriaMet established criteria
Distribution / Shipping Challenge Conditioning and TestingEstablished criteriaMet established criteria
Packaging Verification (sterile barrier, protection)EN ISO 11607-1 and -2 requirementsComplies and met established criteria
Biocompatibility (Hemolysis, Sensitization, etc.)EN ISO 10993-1, relevant sub-partsSuccessfully passed all tests
Sterilization (EO and ECH residuals)EN ISO 10993-7:2008 limitsMeets limits
Sterility Assurance Level (SAL)10^-6Demonstrated 10^-6

Note on "Established Criteria": For physical medical devices, "established criteria" typically refer to internal design specifications, industry standards (like ISO standards), and recognized test methods that demonstrate the device is safe and effective for its intended use. The document explicitly mentions adherence to international standards like EN ISO 10555-1, EN ISO 10555-2, EN 1707, EN ISO 11607-1 and -2, EN ISO 10993-1, EN ISO 10993-7, EN ISO 11135-1, and EN 556-1. These standards define many of the specific acceptance criteria for these types of tests.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each specific performance test (e.g., how many catheters were tested for kink distance, tensile strength, etc.). It generally states that "Testing was conducted" and "Test samples consisted of 2X EtO sterilized devices (entire device)" for the SAL measurement.

Data Provenance: This is a physical device testing, not a clinical data study. The "provenance" would be from laboratory testing conducted by Stryker Neurovascular. There is no mention of country of origin for data beyond the company's location (Fremont, CA, USA) and adherence to international (EN ISO) standards. The testing is prospective for the device's design verification and validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. This is not an AI/ML device that generates diagnostic outputs requiring expert ground truth for a test set. The "ground truth" for these tests is defined by the physical or chemical properties measured against predefined scientific and engineering standards and specifications.

4. Adjudication Method for the Test Set

N/A. Adjudication methods like 2+1 or 3+1 are used for clinical image interpretation or diagnostic performance evaluation, which is not relevant to the design verification and validation of a guide catheter.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

N/A. MRMC studies are specific to evaluating diagnostic technologies, often AI-powered, where human readers interpret cases. This is not applicable to a physical medical device like a guide catheter. There is no AI component in this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

N/A. This device does not involve an algorithm or AI.

7. The Type of Ground Truth Used

The "ground truth" for this device's safety and performance studies is based on:

  • Engineering and Design Specifications: The device must meet its own internal specifications for physical properties.
  • International Standards: Adherence to recognized international standards (e.g., EN ISO 10555 for catheters, EN ISO 10993 for biocompatibility, EN ISO 11135 for sterilization). These standards define acceptable ranges and methodologies.
  • Scientific Principles: Basic principles of material science, mechanics, and biology for biocompatibility.

8. The Sample Size for the Training Set

N/A. This is a physical medical device, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

N/A. As there is no training set, there is no ground truth to establish for it.

{0}------------------------------------------------

jK132542

510(k) Summary

Summary Date:August 9. 2013
Submitter Name andAddress:Stryker Neurovascular47900 Bayside ParkwayFremont, CA. 94538
Contact:Anjali Atal-GuptaSenior Regulatory Affairs SpecialistPhone: 510 413-2573Fax: 510 413-2588Email: anjali.atal-gupta@stryker.com
Trade Name:TransGate™ Guide Catheter
Common Name: .Guide Catheter
Classification Name:Percutaneous Catheter /Class II device under 21 CFR 870.1250(Product Code: DQY).

Legaliy Marketed Predicate Devices:

Reference (Clearance Date)Device
K093184 (Nov 6, 2009)Codman & Shurtleff, Inc. Envoy® GuideCatheter
K010853 (Apr 18, 2001)Boston Scientific Guider Softip™ XF 5F GuideCatheter
K980453 (Aug 11, 1998)Boston Scientific Guider Softip™ XF 6F GuideCatheter

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

.

.

{1}------------------------------------------------

Device Description:

Stryker Neurovascular's TransGate Guide Catheters are constructed with polytetrafluoroethylene (PTFE) on the inner lumen for lubricity, stainless steel wire reinforcement within the wall for trackability. torque transmission and strength, and polymer materials along the length of the catheter for support and flexibility. The catheter has an atraumatic tip, a hub for device connectivity and device handling, and a strain relief at the hub for kink resistance. The distal segment of the device shaft is radiopacified and. in addition, a radiopaque marker is embedded at the distal tip of the device to aid visualization under fluoroscopy. .

The Stryker Neurovascular TransGate Guide Catheter has straight (ST) and pre-shaped distal tips, available in 40 (40), and Multipurpose (MP) shapes. The TransGate Guide Catheters are offered in both 90 and 100 cm lengths and 5 and 6 French diameters.

Indications for Use / Intended Use:

TransGate™ Guide Catheters are intended to facilitate placement of interventional devices into the coronary. peripheral and neurovasculature.

Accessories:

There are no accessories to the TransGate™ Guide Catheters.

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

{2}------------------------------------------------

510(k) Summary Of Safety And Effectiveness (cont.)

Comparison to Predicate Device:
TransGate GuideCatheter (SubjectDevice)Codman Envoy GuidingCatheterBoston Scientific GuiderSoftip XF 5F and 6FGuide Catheters
510(k)NumberTBDK093184K980453, K010853
ClassificationClass II, DQYClass II, DQYClass II, DQY
IndicationTransGate GuideCatheters areintended to facilitatethe placement ofinterventionaldevices into thecoronary, peripheral,andneurovasculature.The ... is intended for usein the peripheral,coronary, andneurovasculature for theintravascular introductionofinterventional/diagnosticdevices.The ... is intended tofacilitate the placement ofinterventional devicesinto the coronary,peripheral, andneurovascular systems.
Materials
-ShaftMaterialsPTFE lined.nylon/thermoplasticelastomer withstainless steel braidPTFE linednylon/polyurethane withstainless steel braid1PTFE linednylon/thermoplasticelastomer/thermoplasticpolyester with stainlesssteel braid
-Proximal EndConfigurationHub with female luerconical lock fittingHub with female luerconical lock fittingHub with female luerconical lock fitting
-RadiopacityRadiopaque tip withdistal markerRadiopaque distal tipRadiopaque distal tip
-PackagingCatheter attached toSBS packaging cardinsideNylon/PE/Tyvekpouch, inside SBScartonCatheter attached topackaging card insidepouch, inside cartonCatheter attached to SBSpackaging card, insidePET/LDPE/Tyvek pouch,inside SBS carton
-SterilizationEthylene OxideEthylene OxideEthylene Oxide
Dimensions
-EffectiveLength5F: 90 cm, 100 cm6F: 90 cm, 100 cm5F: 90 cm, 100 cm6F: 90 cm, 100 cm5F: 90 cm, 100 cm6F: 90 cm, 100 cm
-Outer5 F: 1.75 mm5F: 1.65 mm5F: 1.70 mm
TransGate GuideCatheter (SubjectDevice)Codman Envoy GuidingCatheterBoston Scientific GuiderSoftip XF 5F and 6FGuide Catheters
Diameter6F: 2.1 mm6F: 2.0 mm6F: 2.0 mm
Inner Diameter5F: .057 in5F: .056 in5F: .053 in
6F: .072 in6F: .070 in6F: .064 in
Tip Shapes40, MP, StraightMPD*, MPC*, Straight40. MP. Straight

Compurison to Predicate Device:

1 Materials used for Envoy are not known with certainty

·

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

r

August 2013

{3}------------------------------------------------

*MPD is similar to 40 tip shape and MPC is similar to MP shape

Design verification of the TransGate™ Guide Catheters consisted of:

Performance TestResult
Kink DistanceMet established criteria
Torque Degrees Rotation to KinkMet established criteria
Tip Shape RetentionMet established criteria
Distal Shaft FlexibilityMet established criteria
PushabilityMet established criteria
Tensile StrengthMet established criteria
Chemical CompatibilityMet established criteria
Corrosion ResistanceMet established criteria
ParticulateMet established criteria
Liquid LeakageMet established criteria
Air LeakageMet established criteria
Unscrewing TorqueMet established criteria
Resistance to OverridingMet established criteria
Stress CrackingMet established criteria
Ease of AssemblyMet established criteria
Luer GaugingMet established criteria
Luer Separation ForceMet established criteria
Freedom from leakage and damage under high static pressureMet established criteria
Flow Rate TestingCharacterization test
RadiopacityMet established criteria

Testing was conducted in accordance with EN ISO 10555-1, EN ISO 10555-2, and EN 1707.

Shelf Life Testing (Product and Packaging) and Distribution / Shipping Challenge Conditioning and Testing were performed and the devices met established criteria.

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

{4}------------------------------------------------

Packaging Verification Testing complies with EN ISO 11607-1 and -2 and assessed the ability of finished packages to withstand the effects of anticipated hazards of the distribution environment on essential packaging characteristics and the ability of packaging to protect the device and to maintain sterility (sterile barrier testing).

Biocompability and Sterilization testing were conducted in accordance with EN ISO 10993-1 and EN ISO 11135-1. TransGate Guide Catheters have been classified according to EN ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing as follows:

Category: Externally Communicating Contact type: Circulating blood Contact Duration: Limited exposure (< 24 hours)

Based on this classification, tests relevant to the device described within this premarket notification were selected and conducted in accordance with EN ISO 10993-1 and its applicable sub-parts.

The TransGate Guide Catheter devices meet ethylene oxide (EO) and ethylene chlorohydrin (ECH) residuals limits specified in EN ISO 10993-7:2008. Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. All process and product sterilization validation activities have been conducted and completed in accordance with EN ISO 11135-1:2007, Sterilization of health care products – Ethylene oxide - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices. The sterility of the devices conforms to the requirements of EN 556-1:2001, Sterilization of medical devices - Requirements for medical devices to be designated 'STERILE' - Part 1 : requirements for terminally sterilized medical devices.

A sterility assurance level (SAL) of 10th has been demonstrated. Test samples consisted of 2X EtO sterilized devices (entire device).

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

{5}------------------------------------------------

· Testing showed the device, including its packaging, to be biocompatible for its intended use as an externally communicating, circulating blood contacting device as classified under EN ISO 10993-1:2009. The TransGate Guide Catheter successfully passed all of the following biocompatibility lests:

TestMethod
HemolysisMEM Elution Assay
SensitizationKligman Maximization
Intracutaneous Reactivity(Irritation)Intracutaneous Injection Test
Systemic Toxicity (Acute)ISO Acute Systemic InjectionTest
HemocompatibilityComplement Activation
Hemolysis
Inactivated Partial
Thromboplastin Time Test
In vitro hemocompatibility
Physicochemical TestsUSP <661> Aqueous
PlasticsExtraction
Fourier Transform Infrared(FTIR) ScanFTIR
LatexASTM D6499-07
PyrogenicityUSP Material Mediated RabbitPyrogen Test
EtO ResidualsEthylene oxide and Ethylenechlorohydrins residuals

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

{6}------------------------------------------------

Conclusion:

Because intended use and indications for use are the same as for the predicate devices, and because there is no difference in the fundamental scientific technology of the devices, and because risk assessments and successful verification testing, including testing to EN ISO 10555-1 and EN ISO 10555-2, raise no new questions of safety and effectiveness. Stryker Neurovascular has determined the TransGate™ Guide Catheters to be substantially equivalent to the current legally marketed predicate devices.

Premarket Notification, Stryker Neurovascular TransGate™ Guide Catheter

August 2013

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a black and white logo. The logo consists of two parts: a circular text element on the left and a stylized abstract symbol on the right. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" and is arranged along the circumference of the circle. The abstract symbol is composed of three curved lines that resemble a stylized human figure or a flowing ribbon.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 15, 2013

Stryker Neurovascular % Ms. Anjali Atal-Gupta Senior Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, CA 94538

Re: K132542

Trade/Device Name: TransGate™ Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 16, 2013 Received: August 19, 2013

Dear Ms. Atal-Gupta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I outher and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{8}------------------------------------------------

Page 2 - Ms. Anjali Atal-Gupta

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joyce M. Whang -S

Victor Krauthamer, Ph.D. for Acting Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K132542

Device Name: TransGate™ Guide Catheter

· Indications For Use:

TransGate™ Guide Catheters are intended to facilitate the placement of interventional devices into the coronary, peripheral, and neuro vasculature.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Joyce M. Whang -S

Page 1 of 1

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).