K Number
K132474

Validate with FDA (Live)

Date Cleared
2014-01-30

(176 days)

Product Code
Regulation Number
870.2870
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.

Device Description

The Rapid Exchange (RXi) System and Navvus Catheter are intended to provide hemodynamic measurement information for use in the diagnosis and treatment of coronary or peripheral artery disease. When used in catheterization and related cardiovascular specialty laboratories, the Rapid Exchange (RXi) System and Navvus Catheter will compute and display physiological parameters based on the output of an optically-based pressure measuring sensor.

The Rapid Exchange (RXi) System and Navvus Catheter consist of a single patient use, catheter-based sensing device, a hardware system containing embedded software, a user interface touchscreen, and electronics for converting measured pressure signals into Fractional Flow Reserve (FFR) measurements. The intravascular blood pressure sensor is optically-based and adhered to the end of a fiber optic. The sensor and fiber optic are integrated into a monorail catheter, which connects to the system to deliver the measured pressure data.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (ACIST Rapid Exchange (RXi) System and Navvus Catheter)
Primary Objectives:
Valid FFR measurement by Navvus Catheter when Radi takes valid FFR100% successful in taking valid FFR measurements in all lesions where Radi took valid FFR measurements.
Secondary Criteria (from Bland-Altman, Kolmogorov-Smirnov, Passing-Bablok analysis):
Bias (Bland-Altman)-0.01
95% Limits of Agreement (Bland-Altman)-0.13 to 0.10
Linearity (Kolmogorov-Smirnov test, p-value > 0.1)Cumulative p-value of 0.72 (met success criteria)
Intercept (Passing-Bablok linear regression, 95% CI containing 0)95% CI of -0.15 to 0.19 (met success criteria)
Slope (Passing-Bablok linear regression, 95% CI containing 1)95% CI of 0.76 to 1.17 (met success criteria)
AUC (ROC, vs. first Radi FFR)87.4%
Rate of drift (mean)0.02 ± 0.02
Rate of clinically relevant drift13%
Safety ProfileVery good safety profile, no serious adverse events or adverse events related to the RXi system.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 58 subjects.
  • Data Provenance: The study was a prospective, observational, multi-center study. The text does not explicitly state the country of origin, but given the submitter's address (Eden Prairie, MN) and the FDA filing, it's highly likely to be a US-based study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth was established by comparison to a predicate device, the Radi Analyzer® and PressureWire® Sensor, within the same subjects. Therefore, the "ground truth" for the test set was not established by a panel of human experts in the traditional sense, but rather through paired measurements against an already legally marketed and accepted device. The study collected FFR measurements directly, so the "experts" were the interventional cardiologists performing the procedures and interpreting the real-time FFR data from both devices. Their qualifications are implicitly that they are clinicians capable of performing these procedures.

4. Adjudication Method for the Test Set

There was no explicit adjudication method described for the test set in the context of expert review. The study design involved directly comparing the Navvus Catheter's FFR measurements to those obtained concurrently from the Radi PressureWire Sensor within the same subject. The performance measures (Bland-Altman, Passing-Bablok, etc.) are statistical comparisons between these two measurements, not an adjudication of discrepancy by a separate expert panel.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a measurement tool (a catheter and system for obtaining FFR), not an AI-powered diagnostic imaging tool that assists human readers. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The study described is essentially a standalone performance assessment of the device (RXi System and Navvus Catheter) in generating FFR measurements compared to a predicate device. While a human (the physician) is involved in positioning the catheter and viewing the output, the FFR measurement itself is derived by the device's embedded software. The clinical study directly evaluated the performance of this system against the predicate, which can be considered its standalone performance relative to the established gold standard (the predicate device for FFR measurement).

7. The Type of Ground Truth Used

The ground truth used was the Fractional Flow Reserve (FFR) measurements obtained from the Radi Analyzer® and PressureWire® Sensor (the predicate device). This is a comparative effectiveness study where the predicate device acts as the reference standard for FFR measurement.

8. The Sample Size for the Training Set

The document does not mention a specific training set size. This appears to be a 510(k) submission for a medical device that measures physiological parameters. Training sets are typically relevant for machine learning algorithms. While the device contains "embedded software" for converting signals to FFR, the submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical and clinical testing, rather than an AI/ML development and validation process that would typically involve distinct training and test sets as primary components of the submission. The "training" for such a system would typically involve engineering calibration and validation against known physical standards.

9. How the Ground Truth for the Training Set Was Established

As no specific training set for a machine learning algorithm is detailed, the concept of establishing ground truth for a training set in this context is not directly applicable. If the embedded software involved parameters that were "trained," this would likely have been done through internal engineering testing and calibration against physical pressure standards and potentially in vivo or ex vivo models, verified against known true values. The documentation provided focuses on the clinical and non-clinical validation against the predicate device.

{0}------------------------------------------------

JAN 30 2014

Section 6

510(k) Summary [21 CFR 807.92]

Date the Summary was Prepared
January 28, 2014
510(k) Submitter's Name and Address
ACIST Medical Systems7905 Fuller RoadEden Prairie, MN 55344952-253-4519 (telephone)952-941-4648 (fax)matt.stepanek@acistmedical.com
Owner/Operator
ACIST Medical Systems7905 Fuller RoadEden Prairie, MN 55344952-995-9300 (telephone)952-941-4648 (fax)matt.stepanek@acistmedical.com
Primary Contact
Matthew D. StepanekPrincipal Regulatory Affairs SpecialistACIST Medical Systems7905 Fuller RoadEden Prairie, MN 55344952-253-4519 (telephone)952-941-4648 (fax)matt.stepanek@acistmedical.com
Alternative Contact
Dennis CraneDirector of Regulatory and Clinical AffairsACIST Medical Systems7905 Fuller RoadEden Prairie, MN 55344(952) 253 4515 (telephone)952-941-4648 (fax)dennis.crane@acistmedical.com

{1}------------------------------------------------

Sterilization Facility
Sterigenics US, Inc.7775 S Quincy St.Willowbrook, IL 60527
Establishment Registration Number: 1450293
Console Manufacturing Facility
ACIST Medical Systems, Inc.7905 Fuller RoadEden Prairie, MN 55344
Establishment Registration Number: 2134243
Catheter Manufacturing Facility
MedVenture2301 Centennial Blvd.Jeffersonville, IN 47130
Establishment Registration Number: 1222140
Device Classification Names
Transducer, Pressure, Catheter Tip and Catheter, Pressure Monitoring, Cardiac
Device Common/Usual Name
Catheter Tip Pressure Transducer and Diagnostic Intravascular Catheter
Device Trade/Proprietary Name
Rapid Exchange (RXi) System and Navvus Catheter
Product Codes
DXO and OBI
Classification of Device
Class II
21 CFR 870.2870 and 870.1200

{2}------------------------------------------------

Predicate Device(s)
Device NameSubmission Number.Clearance Date
Radi Analyzer® and PressureWire® Sensor(RADI Medical Systems, St. Jude Medical, Minnesota, USA)K062769December 4, 2006

Device Description

The Rapid Exchange (RXi) System and Navvus Catheter are intended to provide hemodynamic measurement information for use in the diagnosis and treatment of coronary or peripheral artery disease. When used in catheterization and related cardiovascular specialty laboratories, the Rapid Exchange (RXi) System and Navvus Catheter will compute and display physiological parameters based on the output of an optically-based pressure measuring sensor.

The Rapid Exchange (RXi) System and Navvus Catheter consist of a single patient use, catheter-based sensing device, a hardware system containing embedded software, a user interface touchscreen, and electronics for converting measured pressure signals into Fractional Flow Reserve (FFR) measurements. The intravascular blood pressure sensor is optically-based and adhered to the end of a fiber optic. The sensor and fiber optic are integrated into a monorail catheter, which connects to the system to deliver the measured pressure data.

Proposed Intended Use & Indication for Use

The Intended Use/Indication for Use of the Rapid Exchange (RXi) System and Navvus Catheter is as follows:

The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.

{3}------------------------------------------------

Summary of the Technological Characteristics to the Predicate Device(s)

The Rapid Exchange (RXi) System and Navvus Catheter employs many of the same technology characteristics as the predicate device to perform coronary pressure and fractional flow reserve (FFR) measurements. The Rapid Exchange (RXi) System and Navvus Catheter includes an intracoronary pressure measurement system that would allow the physician to use the best guidewire for the patient's anatomy and would facilitate access to coronary lesions. Using the Rapid Exchange (RXi) System and Navvus Catheter could avoid multiple guidewire exchanges if the patient went on to PCI after the FFR measurement, because the physician could use the same quidewire for the PCI.

The Rapid Exchange (RXi) System and Navvus Catheter and the predicate are intended for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Rapid Exchange (RXi) System. Navvus Catheter and predicate are designed to fit inside a percutaneous guiding catheter for the purpose of directing the catheter through a vessel. The signal output from the sensor is used for calculation and presentation of any physiological parameters, functions or indices based on pressure e.g. Fractional Flow Reserve (FFR).

The predicate devices are 0.014" guidewires with an integrated pressure sensor, together with a detachable cable for connection to a diagnostic computer.

The devices are for single use and are not to be re-sterilized.

The Rapid Exchange (RXi) System and Navvus Catheter device modifications are deemed substantially equivalent to the RAD! PressureWire Sensor, as there are no differences in the product performance, device indications for use/intended use and/or device functional scientific technology. The subject device uses the same surgical approach procedures as the predicate device.

ACIST Medical Systems considers the Rapid Exchange (RXi) System and Navvus Catheter product performance to be significantly equivalent to the predicate device, Radi Analyzer and PressureWire Sensor.

ACIST Medical Systems Traditional 510(k) Notification Rapid Exchange (RXi) System and Navvus Catheter Page 4 of 9

{4}------------------------------------------------

CharacteristicRadi Analyzer® and PressureWire®(RADI Medical Systems, St. JudeMedical, Minnesota, USA) (K062769)ACIST Rapid Exchange (RXi) System andNavvus Catheter
Intended UsePressureWire is designed to fit inside apercutaneous catheter for the purpose ofdirecting the catheter through a vessel. Thesignal output from the sensor is used forcalculation and presentation of anyphysiological parameters, functions orindices based on pressure or temperature,e.g. Fractional Flow Reserve (FFR).The ACIST Rapid Exchange (RXi) System isindicated for obtaining intravascular pressuremeasurements for use in the diagnosis andtreatment of coronary and peripheral arterydisease. The Navvus Catheter is intended foruse with the ACIST RXi System.
Conditions ofUse• Operating temperature 15°C to 30°C(59°F to 86°F)• Transport temperature -40°C to 70°C (-40°F to 158°F)• Storage temperature: Room temperature• Relative humidity operating 30-75%• Operating temperature 18°C to 30°C• Transport temperature -29°C to 60°C• Storage temperature: Room temperature• Relative humidity operating 10-95% non-condensing
SpecificationsandProperties• Pressure Accuracy: +/-1mmHg plus +/-1%of reading (over the pressure range -30 to50mmHg); +/-3% of reading (over therange 50 to 300mmHg)• Pressure range: -30 to 300mmHg• Frequency response: 0-25 Hz• Zero thermal effect: 0.3 mmHg/degC• Sensitivity thermal effect: 0.3%/degC• Zero drift: < 7 mmHg/h• Pressure Accuracy: +/- 3% of reading or +/- 3mmHg, whichever is greater• Pressure range: -30 to 300mmHg• Frequency response: Not specified• Zero thermal effect: <0.4 mmHg/degC• Sensitivity thermal effect: Not specified• Zero drift: < 7 mmHg/h
DesignStrain gauge type pressure sensorembedded in a guidewire.Fiber optic type pressure sensor embedded in a0.014" compatible rapid exchange catheter.
MethodsDevice is introduced as a guidewire andplaced in the vessel of interest.Device is introduced over a guidewire placed inthe vessel of interest.
Principles ofOperationCalculates FFR based on simultaneousacquisition of pressure from twotransducers:• Distal pressure from the sensor.• Aortic pressure from a transducer readingpressure in the ostium.Calculates FFR based on simultaneousacquisition of pressure from two transducers:• The flexible distal tip of the catheter thatcontains the optical pressure sensor isadvanced over the proximal end of the0.014 in (0.36 mm) guidewire that is alreadyin place in the distal vasculature.• The Navvus catheter is advanced over the wireand positioned with the radiopaque markerin the coronary ostium, 3 mm outside the tipof the guide catheter.

{5}------------------------------------------------

• Equalization must take place with the Navvus Catheter distal tip just outside the tip of the guide catheter. Pressure is measured less than 3 mm proximal from the marker band visible under fluoroscopy.
• The Navvus catheter sensor is advanced past the lesion under study.. To measure FFRmin or make other measurements using the RXi system, refer to the RXi System User's Guide for further instruction regarding the use of the RXi System for recording FFR values using the Navvus Catheter.

Systems Page 6 of 9

{6}------------------------------------------------

Summary of Non-Clinical Testing

The non-clinical testing included the assessment of the physical properties of the Rapid Exchange (RXi) System and Navvus Catheter and its ability to achieve its intended use and is substantially equivalent to the predicate device. Radi Analyzer and PressureWire Sensor. Bench testing of the device confirmed the suitability of the device for its intended use.

Animal testing provided evidence to confirm the same technology characteristics as the predicate device to perform coronary pressure and fractional flow reserve (FFR) measurements.

Biocompatibility assessment was to ensure that the biocompatibility has been established for the Rapid Exchange (RXi) System and Navvus Catheter; based on the similarity of the materials as the commercialized predicate device.

The following table is a summary, not limited to, the non-clinical testing conducted in accordance to industry standards/requlations:

TestIndustry Standards
DesignVerificationQuality System Requirements FDA 21 CFR Part 820.20 EN ISO 13485: 2003; DesignControls FDA 21 CFR Part 820.30
DevicePerformanceQuality System Requirements FDA 21 CFR Part 820.20 EN ISO 13485: 2003; DesignControls FDA 21 CFR Part 820.30
Animal ModelFDA 21 CFR: Part 58 Good Laboratory Practice Regulations
PackagingISO 11607: 2006 Part 1 and Part 2: Packaging for Terminal Sterilized Medical Devices
Shelf LifeISO 11607: 2006 Part 1, 6.4 Stability Testing
BiocompatibilityISO 10993-1: 2009 Biological Evaluation of Medical Devices.
SterilizationISO 11135: Medical Devices - Validation and Routine Control of Ethylene OxideSterilization and EN ISO 13485.

The proposed Rapid Exchange (RXi) System and Navvus Catheter device performance and fundamental scientific technology remains unchanged to the predicate device. The proposed modifications do not change to the coronary pressure and fractional flow reserve (FFR) measurements in the legally marketed predicate device.

{7}------------------------------------------------

Summary of Clinical Testing

A prospective, observational, multi-center study was conducted. The primary objectives of the study were to assess the relationship in FFR measurements between the Navyus catheter and the Radi, including rate of device success defined as a valid FFR measurement by the Navvus Catheter where a valid FFR measurement was taken using the Radi. The study used paired measurements from the Navvus catheterand Radi within subjects as controls. Pressure measurements were assessed intra-operatively; subjects were evaluated for safety events through hospital discharge.

Three centers participated and enrolled 58 subjects. Patients with single or multi-vessel coronary artery disease indicated for FFR and a vessel reference diameter of ≥ 2.5 mm were eligible.

The average subject was a sixty-five (65) year old male (80%), ranging in age from 42-86 years. The majority of subjects had a history of dyslipidemia (86%), hypertension (74%) and angina (64%),

Study procedures were performed through the femoral (32%) artery. Reference vessel diameters ranged between 2.5 and 5.5 mm (average 3.2 ± 0.6 mm) with percent diameter stenosis ranging from 40 - 90% (average 57.2 ± 10.0%).

Results of Clinical Investigation:In summary, the results of this study are as follows:The RXI system was 100% successful in taking valid FFR measurements in all lesions in which the Radi was able to take valid FFR measurements. The overall bias, as determined from a Bland-Altman analysis, was -0.01. Using a Bland-Altman analysis, the observed 95% limits of agreement was -0.13 to 0.10. The Kolmogorov-Smirnov test for linearity showed a cumulative p-value of 0.72, meeting success criteria of being > 0.1. The intercept determined by a Passing-Bablok linear regression between FFR measurements made using RXi system and that made using Radi had a 95% confidence interval of -0.15 to 0.19, meeting success criteria of the interval containing 0. The slope determined by a Passing-Bablok linear regression between FFR measurements made using RXi system and that made using Radi showed a 95%. confidence interval of 0.76 to 1.17, meeting success criteria of the interval containing 1. Receiver Operator Curves, using the first measured Radi FFR measurement as the standard, created for both the Navvus catheter measured FFR and the last Radi FFR demonstrated very similar areas under the curve (87.4% and 88.1%, respectively). The RXi System demonstrated a statistically significant lower rate of drift compared to the Radi system, both for mean drift (0.02 ± 0.02 for RXi vs. 0.06 ± 0.12 for Radi) and clinically relevant drift (13% vs. 33% rate of occurrence). This clinical study also demonstrated a very good safety profile. Although serious adverse events occurred during five subject's hospitalization none of these events were related to the RXi system. In a similar fashion, none of the 17 reported adverse events were device related. Eleven of the 17 were related to the procedure, including: hematomas, mild hemorrhage, elevated cardiac enzymes, nausea, vasovagal reaction, headache, hypokalemia, temporary rhythm disturbances and wheeze. All of the adverse events were mild to moderate in nature.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{8}------------------------------------------------

Substantial Equivalence

The proposed Rapid Exchange (RXi) System and Navvus Catheter device uses the same surgical approach procedures as the predicate device.

The proposed Rapid Exchange (RXi) System and Navvus Catheter device and the predicate device have similar indications for use/intended use, similar implant materials, similar sterilization methods; and similar delivery characteristics.

The device performance and fundamental scientific technology remains unchanged. The differences between the proposed device and the predicate device do not have any negative effect on the safety and effectiveness of the device.

The non-clinical test and clinical results conclude that the Rapid Exchange (RXi) System and Navyus Catheter to be substantially equivalent to the predicate device, Radi Analyzer and PressureWire Sensor.

Conclusion

Based on ACIST Medical Systems assessment of the Rapid Exchange (RXi) System, Navus Catheter and predicate device, we conclude that the Rapid Exchange (RXi) System and Navvus Catheter contains technological features that do not differ significantly from the currently marketed device. Bench testing completed according to the Verification and Validation Test Plans confirms that the design specifications for the Rapid Exchange (RXi) System and Navvus Catheter have been met and that the device meets the applicable requirements for the safety and performance standards cited. Therefore, ACIST Medical Systems concludes that the Rapid Exchange (RXi) System and Navvus Catheter are substantially equivalent to the legally marketed predicate device subject to this 510(k) based on design, performance, characteristics, and indication for use and does not introduce any new questions regarding safety or effectiveness of the product for its proposed intended use.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, with three wavy lines extending upwards.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 30, 2014

Acist Medical Systems Matthew Stepanek Principal Regulatory Affairs Specialist 7905 Fuller Road Eden Prairie, MN 55344 US

Re: K132474

Trade/Device Name: Rapid Exchange (RXi) System and Navvus Catheter Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer, Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DXO, OBI Dated: December 19, 2013 Received: December 20, 2013

Dear Mr. Stepanek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{10}------------------------------------------------

Page 2 - Matthew Stepanek

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Owen P. Faris -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Section 5

Statement of Indications for Use

510(k) Number: K132474

Device Name: Rapid Exchange (RXi) System and Navvus Catheter

Indications for Use: The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.

Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR

Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by Owen P.
Faris-S
Date: 2014.01.30 15:14:14
-05:00

ACIST Medical Systems Traditional 510(k) Notification Rapid Exchange (RXI) System and Navvus Catheler Page 1 of 1

§ 870.2870 Catheter tip pressure transducer.

(a)
Identification. A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.(b)
Classification. Class II (performance standards).