K Number
K132397

Validate with FDA (Live)

Date Cleared
2014-05-07

(279 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GORE® BIO-A® Wound Matrix is intended for use in the management of wounds.

Wound types include: Partial and full-thickness wounds, pressure ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds (donor sites/grafts, post-Mob's surgery, post laser surgery, podiatic, wound dehiscence), trauma wounds (abrasions, second degree burns, skin tears) and draining wounds.

Device Description

GORE® BIO-A® Wound Matrix device is a tailorable, porous, bioabsorbable material matrix that provides a scaffold for cellular infiltration and vascularization. The device permits the ingress of cells and soft tissue formation into the defect space/wound bed. GORE® BIO-A® Wound Matrix is a porous fibrous structure composed solely of a synthetic copolymer comprised of polyglycolic acid and trimethylene carbonate (PGA:TMC).

AI/ML Overview

This document is a 510(k) Summary for the GORE® BIO-A® Wound Matrix. It does not describe an AI medical device or provide acceptance criteria and study data in the typical format for such a device.

Therefore, I cannot fulfill your request for an acceptance criteria table and study details as they pertain to an AI device. The document describes a traditional medical device (wound matrix) and its substantial equivalence to predicate devices based on pre-clinical (bench and animal) testing, not on AI algorithm performance.

Here's why the provided information doesn't fit the requested structure:

  • No AI Device: The device described is a biocompatible wound matrix, not an AI algorithm.
  • No "Acceptance Criteria" for AI Performance: The document discusses meeting "intended product specifications" and similarity to predicate devices, which are not the same as performance metrics like sensitivity, specificity, or AUC typically used for AI.
  • No "Study that Proves the Device Meets Acceptance Criteria" for AI: The "Performance Data" section details bench and animal studies (biocompatibility, wound closure, histological assessment) for the physical wound matrix, not studies evaluating an AI algorithm's diagnostic or predictive capabilities.
  • No "Test Set," "Training Set," "Ground Truth," "Experts," "MRMC Study," or "Standalone Performance" related to AI: These concepts are specific to AI/machine learning model evaluation and are not applicable to the GORE® BIO-A® Wound Matrix.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 15, 2022

W. L. Gore & Associates, Incorporated Barbara L. Smith, RAC Official Correspondent 301 Airport Rd. Elkton, Maryland 21921

Re: K132397

Trade/Device Name: Gore® Bio-A® Wound Matrix Regulatory Class: Unclassified Product Code: QSZ

Dear Barbara L. Smith, RAC:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 7, 2014. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSZ.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's emblem in the center. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three wavy lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the perimeter of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 7, 2014

W.L. Gore & Associates Incorporated % Ms. Barbara L. Smith, RAC Regulatory Associate 301 Airport Road Elkton, Maryland 21921

Rc: K132397

Trade/Device Name: GORE® BIO-A™ Wound Matrix Regulatory Class: Unclassified Product Code: FRO Dated: April 2. 2014 Reccived: April 7, 2014

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Mcdical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Barbara L. Smith, RAC

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Binita S. Ashar - S 2014.05.07 12:01:48 -04'00'

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement befow.

510(k) Number (if known) K 132397

Device Name

GORE® BIO-A® Wound Matrix

Indications for Use (Describe)

The GORE® BIO-A® Wound Matrix is intended for use in the management of wounds.

Wound types include: Partial and full-thickness wounds, pressure ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds (donor sites/grafts, post-Mob's surgery, post laser surgery, podiatic, wound dehiscence), trauma wounds (abrasions, second degree burns, skin tears) and draining wounds.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

STORE CONSTITUTION CORTEDATION SECONE CONCELLERS CONCELLERS CONCELLAND CONSECTION CONSULTION CONSECTION

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/3/Picture/15 description: The image shows the name "Cynthia Chang" in a bold, sans-serif font. The letters are black against a white background, with some patterned noise in the background. The name is presented in a straightforward, clear manner.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this Information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

PSC Publishing Servises (101) 441-6240

{4}------------------------------------------------

W. L. Gore & Associates, Inc. GORE® BIO-A® Wound Matrix Section 5. 510(k) Summary

MAY 0 7 2014 K132397 page 1 of 3

510(k) SUMMARY (Per 21CFR807.92)

Submission Information

Owner/Operator: W. L. Gore & Associates, Inc. 1505 N. Fourth Street Flagstaff, AZ 86004 USA

Regulatory Contact: W. L. Gore & Associates, Inc. 301 Airport Road Elkton, Maryland 21921 Attn: Barbara L. Smith, RAC Phone: 410-506-8189 Fax: 410-506-8221 E-mail: blsmith@wlgore.com

Date Summary Prepared

April 29, 2014

Device Names/Classification

Trade Name:GORE® BIO-A® Wound Matrix
Common Name:Wound Dressing
Classification Name:Unclassified
Product Code:FRO (Dressing, wound, drug)

Predicate Device

The predicate devices to which the GORE® BIO-A® Wound Matrix is being compared are:

  • K021792 INTEGRA™ Bilayer Matrix Wound Dressing (FRO) .
  • K022127 AVAGEN Wound Dressing (KGN) .
  • K090160 SUPRATHEL® Wound & Burn Dressing (FRO) .
  • K083266 GORE® BIO-A® Fistula Plug (FTL) .

GORE AND DESIGNS ARE TRADEMARKS OF W. L. GORE & ASSOCIATES, INC.

Integra is a trademark of INTEGRA Lifesciences Corporation. SUPRATHEL is a registered trademark of PolyMedics Innovations GmbH.

{5}------------------------------------------------

W. L. Gore & Associates, Inc. GORE® BIO-A® Wound Matrix Section 5. 510(k) Summary

Intended/Indications for Use

The GORE® BIO-A® Wound Matrix is intended for use in the management of wounds.

Wound types include: Partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, second degree burns, skin tears) and draining wounds.

Device Description

GORE® BIO-A® Wound Matrix device is a tailorable, porous, bioabsorbable material matrix that provides a scaffold for cellular infiltration and vascularization. The device permits the ingress of cells and soft tissue formation into the defect space/wound bed. GORE® BIO-A® Wound Matrix is a porous fibrous structure composed solely of a synthetic copolymer comprised of polyglycolic acid and trimethylene carbonate (PGA:TMC).

Summary of Similarities and Difference in Technological Characteristics, Performance and Intended Use

The primary difference between the subject Wound Matrix device and Integra (Avagen) predicate is in the material type i.e. synthetic polymer scaffold vs. a collagen scaffold.

Performance Data / Predicate Device Comparison

Pre-Clinical

Bench study: Testing of the GORE® BIO-A® Wound Matrix device demonstrated the performance of the GORE® BIO-A® Wound Matrix is capable of meeting intended product specifications which are similar to the predicate Gore device comprised of the same material construct.

Animal study: Testing of the GORE® BIO-A® Wound Matrix device also included biocompatibility testing in accordance with ISO 10993-1 and in vivo safety and performance studies. The performance of the GORE® BIO-A® Wound Matrix device and the collagen predicate control device was evaluated in both full and partial thickness wounds in a porcine model analysis including wound closure, macroscopic appearance, and microscopic evaluation via histological assessment. The results of the studies showed that the test and predicate control device exhibited no adverse tissue response, a similar inflammatory response, and no differences in time to wound closure. All wounds healed prior to device absorption, and the devices were no longer histologically detectable at 8 months.

Gore and Designs are trademarks of W. L. Gore & Associates, Inc.

Integra is a trademark of INTEGRA Lifesciences Corporation. SUPRATHEL is a registered trademark of PolyMedics Innovations GmbH.

{6}------------------------------------------------

W. L. Gore & Associates, Inc. GORE® BIO-A® Wound Matrix Section 5. 510(k) Summary

Clinical

No clinical evaluations of this product have been conducted.

Conclusion

W. L. Gore & Associates concludes the GORE® BIO-A® Wound Matrix is substantially equivalent to the predicate devices based upon similarities in product indications, operation, design, materials of construction, sterilization, labeling, shelf life, and performance.

GORE AND DESIGNS ARE TRADEMARKS OF W. L. GORE & ASSOCIATES, INC.

Integra is a trademark of INTEGRA Lifesciences Corporation. SUPRATHEL is a registered trademark of PolyMedics Innovations GmbH.

N/A