K Number
K132259

Validate with FDA (Live)

Date Cleared
2013-08-07

(19 days)

Product Code
Regulation Number
866.3328
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal and nasal swabs of symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influenza A viral antigens can be distinquished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A neqative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions. The test is not intended to detect influenza C antigens.

Device Description

The BD Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in respiratory specimens. The patient specimen is mixed in a prefilled unitized tube containing RV Reagent D and added to the test device. RV Reagent D contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Flu A+B test device.

The specimen is mixed and added to the test device where influenza A or influenza B viral antigens bind to anti-influenza antibodies conjugated to detector particles on the BD Flu A+B test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of influenza A or B viral antigens.

The BD Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone addresses non-specific signal generation and is not labeled on the test device. The remaining zone is used to measure the assay background and is also not labeled.

The BD Flu A+B assay incorporates an active negative control feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eve is unable to accurately perform the subtraction of the nonspecific signal.

AI/ML Overview

The provided text is a 510(k) summary for the BD Veritor™ System Flu A+B assay POC kit. It describes the device, its intended use, and indicates its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or a detailed study report that proves the device meets those criteria with performance metrics such as sensitivity and specificity from a clinical trial.

The document mentions "Performance characteristics for influenza A and B were established during January through March of 2011..." but does not elaborate on these characteristics, the study design, sample sizes, ground truth establishment, or expert involvement. The primary change described in this 510(k) submission is the addition of analytical sensitivity (LOD) reactivity data for A/Anhui/1/2013 H7N9 and Strain Reactivity tables to the labeling, not a new clinical performance study for the original indications.

Therefore, many of the requested details cannot be extracted from this specific document.

Here's an attempt to answer based only on the provided text, highlighting what is not available:


Acceptance Criteria and Device Performance Study for BD Veritor™ System Flu A+B assay POC kit (K132259)

This 510(k) submission primarily addresses the addition of analytical sensitivity data for a new strain (A/Anhui/1/2013 H7N9) to the labeling of an already legally marketed device. It does not provide the detailed clinical performance study for the original device or specific acceptance criteria met by a clinical study. It refers to "Performance characteristics for influenza A and B were established during January through March of 2011" but does not provide the details of those characteristics or the study that established them.

1. Table of Acceptance Criteria and Reported Device Performance

This information is not available in the provided 510(k) summary. The document states that performance characteristics were established but does not present them or the acceptance criteria they were measured against.

2. Sample size used for the test set and the data provenance

This information is not available in the provided 510(k) summary. The text mentions "Performance characteristics for influenza A and B were established during January through March of 2011," but does not detail the sample size or provenance of the data from this establishment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not available in the provided 510(k) summary.

4. Adjudication method for the test set

This information is not available in the provided 510(k) summary.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

This information is not available in the provided 510(k) summary. The device is described as read by a "System Reader" which uses a proprietary algorithm, suggesting it's not a human-interpreted visual read in the final determination.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The device description implies a standalone algorithm's role, as "The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive." This suggests the interpretation of the test strip is done by the algorithm. However, no specific "standalone performance study" details (e.g., sensitivity, specificity of the algorithm alone) are provided.

7. The type of ground truth used

The intended use statement suggests that for negative results, confirmation by "viral culture or an FDA-cleared influenza A and B molecular assay" is recommended. This implies that for the original performance characteristics, such methods (viral culture or molecular assays) would likely have been used as the ground truth. However, this is an inference, and the document does not explicitly state how ground truth was established for the performance studies it references.

8. The sample size for the training set

This information is not available in the provided 510(k) summary.

9. How the ground truth for the training set was established

This information is not available in the provided 510(k) summary.


Summary of What is Known from the Provided Text:

  • Device: BD Veritor™ System Flu A+B assay POC kit, a rapid chromatographic immunoassay for direct and qualitative detection of influenza A and B viral nucleoprotein antigens.
  • Intended Use: Aid in diagnosis of influenza A and B viral infections from nasopharyngeal and nasal swabs of symptomatic patients. It differentiates between Flu A and Flu B.
  • Key Technology: Uses an active negative control feature and a proprietary algorithm in the BD Veritor™ System Reader to interpret test results by subtracting non-specific signal.
  • Basis for 510(k): Substantial equivalence to the predicate device (K112277). The modification primarily concerns updating labeling with analytical sensitivity data for the A/Anhui/1/2013 H7N9 strain and strain reactivity tables.
  • Performance Reference: "Performance characteristics for influenza A and B were established during January through March of 2011" against specific prevalent strains (A/2009 H1N1, A/H3N2, B/Victoria lineage, and B/Yamagata lineage). Details of these characteristics and the studies are not included in this document.
  • H7N9 Data: While new analytical sensitivity data for A/Anhui/1/2013 H7N9 was added, the labeling includes a disclaimer that "performance characteristics of this device with clinical specimens that are positive for the novel avian influenza A(H7N9) virus have not been established." This means the new data is analytical (LOD) and not clinical performance.

{0}------------------------------------------------

510(K) SUMMARY K132259

SUBMITTED BY:BECTON, DICKINSON AND COMPANY10865 Road to the Cure, Suite 200San Diego, CA 92121Tel: (858) 795 7890Fax: (858) 795 7885
CONTACT NAME:Gregory P. Payne, RAC, Director Regulatory Affairs
DATE PREPARED:June 25, 2013
DEVICE TRADE NAME:BD Veritor TM System Flu A+B assay POC kit
DEVICE COMMON NAME:Influenza virus serological reagents
DEVICE CLASSIFICATION:21 CFR 866.3330
PREDICATE DEVICES :BD Veritor TM System for Rapid Detection of Flu A+BPOC (K112277)

AUG 07 2013#### INTENDED USE :

POC Kit

The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal and nasal swabs of symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influenza A viral antigens can be distinquished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A neqative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions. The test is not intended to detect influenza C antigens.

Performance characteristics for influenza A and B were established during January through March of 2011 when influenza viruses A/2009 H1N1, A/H3N2, B/Victoria lineage, and B/Yamagata lineage were the predominant influenza viruses in circulation according to the Morbidity and Mortality Weekly Report from the CDC entitled "Update: Influenza Activity-United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine." Performance characteristics may vary aqainst other emerging influenza viruses.

If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to the state or local health department for testing. Virus

{1}------------------------------------------------

culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.i

DEVICE DESCRIPTION :

The BD Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in respiratory specimens. The patient specimen is mixed in a prefilled unitized tube containing RV Reagent D and added to the test device. RV Reagent D contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Flu A+B test device.

The specimen is mixed and added to the test device where influenza A or influenza B viral antigens bind to anti-influenza antibodies conjugated to detector particles on the BD Flu A+B test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of influenza A or B viral antigens.

The BD Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone addresses non-specific signal generation and is not labeled on the test device. The remaining zone is used to measure the assay background and is also not labeled.

Image /page/1/Figure/5 description: The image shows a diagram of a test strip, possibly for influenza, with labels indicating the location of different control and test lines. The labels include "Positive Ctrl C", "Flu B test line L4", "Flu A test line L3", "Negative Ctrl L2", and "Background". To the right of the labels is a visual representation of the test strip with four dark lines, indicating a positive result for the control and both Flu A and Flu B.

The BD Flu A+B assay incorporates an active negative control feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eve is unable to accurately perform the subtraction of the nonspecific signal.

DEVICE COMPARISON:

ﮐﮯ

{2}------------------------------------------------

The modified device differs from the currently marketed BD Veritor™ System Flu A+ B in the following way:

The labeling has been changed to reflect the addition of Analytical Sensitivity (LOD) reactivity data for A/Anhui/1/2013 H7N9 and Strain Reactivity tables.

SUBSTANTIAL EQUIVALENCE:

The modified device BD Veritor™ System Flu A+ B is substantially equivalent to the current legally marketed device, BD Veritor™ System Flu A+B assay. Additions made to the labeling to add additional strain testing did not change the intended use of the device or the fundamental scientific technology.

Risk analysis was not conducted to add this analytical sensitivity information to the product insert as no new issues of safety and effectiveness were identified for this addition to the product insert.

Additions are as follows:

ChangePotential Impact of Change
Addition of data for Analytical Sensitivityand Strain reactivity tables forA/Anhui/1/2013 H7N9Change to package insert to reflect thischange along with addition of FDArequired disclaimer. ""Although this testhas been shown to detect the novel avianinfluenza A(H7N9) cultured virus, theperformance characteristics of this devicewith clinical specimens that are positive forthe novel avian influenza A(H7N9) virushave not been established. The BD VeritorSystem Flu A+B test can distinguishbetween influenza A and B viruses, but itcannot differentiate influenza subtypes."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle with three lines representing its wings, and a circular border containing the text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters. The eagle faces left and appears to be in flight.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-Godg Silver Spring, MI) 20993-0002

GREGORY P. PAYNE RAC, DIRECTOR REGULATORY AFFAIRS BECTON, DICKINSON AND COMPANY 10865 ROAD TO THE CURE, SUITE 200 SAN DIEGO, CA 92121

August 7, 2013

Re: K132259

Trade/Device Name: BD Veritor™ System Flu A+B assay POC kit Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza virus scrological reagents Regulatory Class: I Product Code: GNX Dated: July 17, 2013 Received: July 19, 2013

Dear Mr. Payne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Payne

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDeviccs/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Scherf -S for

Sally Hojvat. M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: K132259

BD Veritor™ System Flu A+B assay POC kit Device Name:

Indications For Use:

The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal and nasal swabs of symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influens can be distinguished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions. The test is not intended to detect influenza C antigens.

Performance characteristics for influenza A and B were established during January through March of 2011 when influenza viruses A/2009 H1N1, A/H3N2, B/Victoria lineage, and B/Yamagata lineage were the predominant influenza viruses in circulation according to the Mortality Weekly Report from the CDC entitled "Update: Influenza Activity-United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine." Performance characteristics may vary against other emerging influenza viruses.

If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to the state or local health department for testing. Virus culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Tamara V. Feldblyum -S 2013.08.02 12:27:00 -04′00″

Page 1 of 1

§ 866.3328 Influenza virus antigen detection test system.

(a)
Identification. An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.