K Number
K132224

Validate with FDA (Live)

Date Cleared
2013-10-08

(83 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Coated Electrosurgical Electrodes are intended for the following indications: Cutting of Soft Tissue, Coagulation of Soft Tissue.

Device Description

Not Found

AI/ML Overview

The provided 510(k) summary for K132224 describes the Unimed Surgical Coated Electrosurgical Electrodes. This document does not detail a study involving AI performance, human readers, or image analysis, as it pertains to a physical medical device (electrosurgical electrodes) rather than an AI-powered diagnostic tool. Therefore, many of the requested categories related to AI performance, expert adjudication, MRMC studies, and ground truth for AI models are not applicable.

Here's an analysis of the provided information based on the sections that are relevant to the device:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Minimum Coating Adhesion ≥ 3.5 lbs.Compliant (Implied by statement "Verification testing of the electrodes was performed sufficient to confirm that the products meet their specifications.")
Minimum 10-lbs Pull Force (Attachment to Pencil compliant with IEC 60601-2-2)Compliant (Implied by statement "Verification testing of the electrodes was performed sufficient to confirm that the products meet their specifications.")
Coating Performance and integrity for aged devices at Normal and Extreme use conditionsConfirmed (Implied by statement "Performance and integrity of the coatings of aged devices were confirmed at Normal and Extreme use conditions.")
Dielectric Strength (per IEC 60601-2-2)Compliant (Implied by statement "Verification testing of the electrodes was performed sufficient to confirm that the products meet their specifications.")
Hipot Testing (per IEC 60601-2-2)Compliant (Implied by statement "Verification testing of the electrodes was performed sufficient to confirm that the products meet their specifications.")
Biocompatibility meets requirements of ISO 10993-1Meets requirements of ISO 10993-1
Color Additives are GRAS and non-toxic (per Code of Federal Regulations)Listed as GRAS and are non-toxic in the Code of Federal Regulations
Shelf Life established using Arrhenius ModelEstablished (Implied by statement "Shell Life date using the Arrhenius Model.")

2. Sample size used for the test set and the data provenance

The document does not specify a distinct "test set" sample size in the way it might for an AI model. For physical testing of the electrodes, the sample sizes for each specific test (e.g., coating adhesion, pull force, electrical testing) are not provided in this summary. The data provenance is internal to Unimed Surgical Products, Inc. for the verification testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a physical device, and the ground truth for its performance is established through engineering and biological testing against predefined specifications and standards (e.g., IEC 60601-2-2, ISO 10993-1). No human experts are described as "establishing ground truth" in the context of diagnostic performance for this type of device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable, as this relates to human expert review for establishing ground truth in diagnostic studies, which is not relevant here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an electrosurgical electrode and does not involve AI assistance for human readers or diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an electrosurgical electrode and does not involve an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance is based on established engineering specifications, international standards (e.g., IEC 60601-2-2 for electrical safety, ISO 10993-1 for biocompatibility), and regulatory requirements. For example:

  • Physical Testing: Measured values (e.g., pull force, adhesion strength) compared against predefined minimums.
  • Electrical Testing: Compliance with specified electrical characteristics according to industry standards.
  • Biocompatibility: Adherence to the requirements of ISO 10993-1 for biological safety.

8. The sample size for the training set

Not applicable. This is a physical device, and the concept of a "training set" is not relevant.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this physical device.

{0}------------------------------------------------

K132224

510(K) SUMMARY 8.0

1

r

Date Prepared: July 3, 2013

8.1SUMMARY OF SAFETY AND EFFECTIVENESS
Submitted By:Lee Alexander
President
Unimed Surgical Products, Inc
8.2 Trade/Proprietary Name:Unimed Surgical Coated Electrosurgical Electrodes
8.3 Common/Usual Name:Electrosurgical Electrodes
8.4 Classification Name:Electrosurgical, Cutting & Coagulation & Accessories
8.5 Manufacturing Facility Address:Unimed Surgical Products, Inc. 1041 Belcher Road Largo, FL 33777
Establishment Registration Number:1058746
8.6 Sterilization Facility Address:Food Technology Service, Inc. 502 Prairie Mine Road Mulberry, FL 33860
Establishment Registration Number:1054811
8.7 Classification:Class: II

OCT 0 8 2013Classification: 8.1

Panel: General and Plastic Surgery Product Code: GEI Cite: 21 CFR 878.4400

Substantial Equivalence 8.8

Unimed Surgical Coated Electrosurgical Electrodes are substantially equivalent to the Unimed Unimed Surgical Coated Electrosungliar Electronic (19770066) and the Unimed
Coated Laparoscopic/ Arthroscopy/ Suction-Imgation Electrodes (K970066) Coated Laparoscopics Anthroscopyr Suction Institute/Coated Ball Electrodes (K962935)

  • Technological Characteristics 8.9 Technological Characteristics
    Unimed Surgical Coated Electrosurgical Electrodess have the same technological characteristics as the current marketed product.

Performance Data 8.10

Penformance Data
Verification testing of the electrodes was performed sufficient to confirm that the products meet their specifications.

8.10.1 Physical Testing

  • Minimum Coating Adhesion ≥ 3.5 Ibs. 8.10.1.1
  • Minimum Coating Adriesion E 0.0 150.
    Pull Test Attachment to Pencil compliant with IEC 60601-2-2 -Minimum 8.10.1.2 10-lbs Pull Force
  • 8.10.2 Coating Testing
    • ling
      Performance and integrity of the coatings of aged devices were confirmed at 8.10.2.1 Normal and Extreme use conditions.
  • 8.10.3 Electrical Testing per IEC 60601-2-2
    • Dielectric Strength 8.10.3.1
    • Hipot Testing ' 8.10.3.2

{1}------------------------------------------------

8.10.4 Biological Specifications

8.10.4.1 Biocompatibility

o. 10.4. I meet the requirements of ISO 10993-1

Color Additives 8.10.4.2

o: 10.4.2 - Solor Additives
Color additives are listed as GRAS and are non-toxic in the Code of Federal Regulations.

8.10.5 Shelf Life

o. 10.5 Shell Life date using the Arrhenius Model.

Conclusion 8.11

Conclusion
Unimed Surgical concludes, based on the information presented, that the US Onlined Ourgical concludes, based on the the trently marketed legally in the US.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

October 8, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Unimed Surgical Products, Incorporated Ms. Lee Alexander President 1041 Belcher Road Largo, Florida 33777

, Florida 337.

Re: K132224

Trade/Device Name: Coated Electrosurgical Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 30, 2013 Received: October 4, 2013

Dear Ms. Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 1 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Lee Alexander

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K132224

Coated Electrosurgical Electrodes Device Name:

The Coated Electrosurgical Electrodes are intended for the following Indications for Use: indications:

  • Cutting of Soft Tissue ♥
  • Coagulation of Soft Tissue .

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

!:

Over-The-Counter Use (21 CFR 801 Subpart C)

(Please Do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Long H. Chen States and out only come to commons of Prople, costong K. Ch ん
Q B.2.34Z_1920D100.100.100.1.11ー) J00344054
Dete: 2013.10.00 13:31.02 -04-00 -A . .

for MXM

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132224

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.