K Number
K132146

Validate with FDA (Live)

Date Cleared
2013-08-30

(50 days)

Product Code
Regulation Number
886.5925
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Saview-Aqua 58 UV (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.

The Saview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.50 diopters that does not interfere with visual acuity.

The Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity and require add power of up to +3.25 diopters.

The Saview-Aqua 58 UV, Saview-Aqua 58 UV Toric and Saview-Aqua 58 UV Multifocal lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
The lenses are to be prescribed for single-use daily disposable wear and not intended to be cleaned or disinfected and should be discarded after a single use.

Device Description

The lens material (etafilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate and methacrylic acid crosslinked with 1, 1, 1-trimethylol propane trimethacrylate. A UV absorbing monomer, 2-[3-(2H-Benzotriazol-2y1)-4hydroxyphenyl]] ethyl methacrylate, is incorporated into the lens polymer and used to block UV radiation. The UV blocker for Saview-Aqua 58 UV (etafilcon A) Visibility tinted Soft (Hydrophilic) Contact Lens averages 85.063 % in the UVA range of 316 nm to 380 nm and 97.984 % in the UVB range of 315 nm to 280 nm. The lens contains 58% water by weight and each lens is supplied sterile in a blister container containing SYS TO.2 % hyaluronic acid polymer in saline solution. The lens is visibility tinted Pigment Blue 15(Copper phthalocyanine) to make the lens more visible for handling.

The Saview-Aqua 58 UV (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available as a single vision lens.

The Saview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available in a double slab-off back surface design. The lens design incorporates a cylinder and base curve. From the bi-curve reduced optic front surface, there exists a slab-off of the upper and lower half of the lens. This makes both sides thicker at the horizontal level on the front surface to keep the axis stable.

The Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available as an aspherical multifocal lens.

AI/ML Overview

The provided document is a 510(k) summary for contact lenses and does not describe AI/ML device performance, acceptance criteria, or a study that evaluates such a device. Since the request pertains to AI/ML device performance and related studies, and the document is not about an AI/ML device, I cannot fulfill the request based on the provided text.

The document focuses on the substantial equivalence of new contact lenses to previously approved predicate devices, based on material properties, chemical composition, and non-clinical testing (stability, toxicology, physical/chemical testing). It explicitly states: "Clinical data is not required for this submission."

Therefore, I cannot provide information on:

  1. A table of acceptance criteria and reported device performance for an AI/ML device.
  2. Sample sizes for test sets, data provenance, number of experts for ground truth, or adjudication methods for an AI/ML device.
  3. Multi-reader multi-case (MRMC) comparative effectiveness studies or standalone performance of an AI/ML algorithm.
  4. Type of ground truth for an AI/ML device.
  5. Sample size for the training set or how ground truth for the training set was established for an AI/ML device.

{0}------------------------------------------------

St, Shine Optical Co., Ltd

.

r

05 Summary of 510(k) Submission

Submitter Information:

Address:St. Shine Optical Co., Ltd.4,5F No. 276-2, Sec. 1, Ta Tung Rd.Hsi Chih Dist., 221, New Taipei CityTaiwan R.O.C.
Registration No.:9617499
Contact Person:Ella Lee,Product Manager, R&D Div.
Telephone:886-2-2647-9356
Fax:886-2-8691-6776
Date Prepared:June 28, 2013

AUG 30 2013

Device:

Common Name:Soft (Hydrophilic) Contact Lens
Trade/Proprietary Name:Saview-Aqua 58 UV (etafilcon A) Visibility Tinted Soft(Hydrophilic) Contact LensSaview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact LensSaview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens
Classification Name:Soft (Hydrophilic) Contact Lens (daily wear)
Device Classification:Class II (21 CFR 886.5925)
Product Code:LPL, MVN
Panel:Ophthalmic

Predicate Devices:

The predicate devices are Saview 58 UV, Saview 58 UV Toric, and Saview 58 UV Multifocal (etafilcon A) Soft (hydrophilic) Lens for Daily Wear covered under 510(k) K092852 and Saview-Aqua 55, Saview-Aqua 55 UV Toric and Saview-Aqua 55 UV Multifocal UV (methafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens covered under 510(k) K121201.

Description of Devices:

{1}------------------------------------------------

The lens matcrial (ctafilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate and methacrylic acid crosslinked with 1, 1, 1-irimethylol propane trimethacrylate. A UV absorbing monomer, 2-13-(2H-Benzotriazol-2y1)-4hydroxypheny]] ethyl methacrylate, is incorporated into the lens polymer and used to block UV radiation. The UV blocker for Saview-Aqua 58 UV (etafilcon A) Visibility tinted Soft (Hydrophilic) Contact Lens averages 85.063 % in the UVA range of 316 nm to 380 nm and 97.984 % in the UVB range of 315 nm to 280 nm. The lens contains 58% water by wcight and each lens is supplied sterile in a blister container containing SYS TO.2 % "hyaluronic acid polymer in saline solution. The lens is visibility innelli Pigment Blue 15(Copper phthalocyanine) to make the lens more visible for handling.

The Saview-Aqua 58 UV (ctafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available as a single vision lens.

The Saview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available in a double slab-off back surface design. The lens design incorporates a cylinder and base curve. From the bi-curve reduced optic front surface, there exists a slab-off of the upper and lower half of the lens. This makes both sides thicker at the horizontal level on the front surface to keep the axis stable.

The Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is available as an aspherical multifocal lens.

Indication for Use:

The Saview-Aqua 58 UV (ctafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hypcropia) in aphakic or not-aphakic persons with non-diseased cycs that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.

The Saview-Aqua 58 UV Toric (ctafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lcns is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic or not-aphakic persons with non-discased cycs that may exhibit refractive astigmatism up to 2.50 diopters that docs not interfere with visual acuity.

The Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia

{2}------------------------------------------------

(myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity and require add power of up to +3.25 diopters.

The Saview-Aqua 58 UV, Saview-Aqua 58 UV Toric and Saview-Aqua 58 UV Multifocal lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The lenses are to be prescribed for single-use daily disposable wear and not intended to be cleaned or disinfected and should be discarded after a single use.

Non-Clinical Testing:

The following tests were conducted as recommended by the Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, revised May 1994.

  • . Stability testing
  • . Toxicology testing
    • Cytotoxicity I
    • Ocular Irritation ■
    • 트 Acute Systemic Injection
  • . Physical/Chemical Testing

Clinical Testing:

Clinical data is not required for this submission.

Description of Safety and Substantial Equivalence:

Information submitted in the 510(k) cstablishes that the Saview-Aqua 58 UV lenses have comparable physicochemical properties to the predicate devices and do not raise questions of safety and effectiveness. Shelf life testing has shown the lenses remain sterile and that lens properties do not change before the expiration date. Results of biocompatibility tests showed the Saview-Aqua 58 UV lenses are substantially equivalent to the predicate device in safety and biocompatibility. Therefore, the Saview-Aqua 58 UV, Savicw-Aqua 58 UV Toric, and Saview-Aqua 58 UV Multifocal (ctafflcon A) Soft (Hydrophilic) Contact Lens are substantially equivalent to the predicate devices.

{3}------------------------------------------------

.

;

.

.

.

:

:

.

:

·

s,

Table 1 Comparison Chart
Saview-Aqua 58 UV,Saview-Aqua 58 UV58 UV, 58UVMultifocal, and 58Saview-Aqua 55 UV,Saview-Aqua 55 UV
Toric, and Saview-AquaUV Toric SoftToric, and Saview-Aqua
58 UV Multifocal Soft(Hydrophilic)55 UV Multifocal Soft
Device(Hydrophilic) ContactLensContact Lens( K092852)(Hydrophilic) ContactLens (K121201)
etafilcon Aetafilcon Amethafilcon A
Material (Classification)(Group 4)(Group 4)(Group 4)
myopia, hyperopia,myopia, hyperopia,myopia, hyperopia,
Indication for usepresbyopia andpresbyopia andpresbyopia and
astigmatismastigmatismastigmatism
Water content58 %58 %55 %
Visible light
transmittance(381 nm ~ 780 nm)91.752%91.752%96.78 %
UV Transmittance
UVA(316 nm ~ 380 nm)14.937 %14.937 %12.370 %
UVB(280 nm ~ 315 nm)2.016 %2.016 %1.025 %
Dk (35° C)$26.3 \times 10^{-11}$$26.3 \times 10^{-11}$$21.4 \times 10^{-11}$
Powers+12.00 to -20.00D+12.00 to -12.00D+12.00 to -20.00D
ColorPigment blue 15Pigment blue 15Pigment blue 15
Refractive index1.3992 (wet)1.3992 (wet)1.404 (wet)
Method of manufactureMouldedMouldedMoulded
Package Storage salinesolutionSaline solution containinghyaluronic acid polymerSaline solutionSaline solution containinghyaluronic acid polymer

·

:

:

.

Table 1 Comparison Chart

Saview-Aqua 58 UV (etafilcon A) Soft (Hydrophilic) Contact Lens 510(k)

:

{4}------------------------------------------------

Conclusion:

A series of pre-clinical tests were performed to demonstrate the safety and cffectivencss of the Saview-Aqua 58 UV, Saview-Aqua 58 UV Toric, and Saview-Aqua 58 UV Multifocal (hefilcon A) Soft (Hydrophilic) Contact Lens and to establish substantial equivalence to the predicate devices. Information submitted in the 510(k) also establishes that the Saview-Aqua 58 UV lenses do not raise questions of safety and effectiveness. Therefore, the Saview-Aqua 58 UV, Saview-Aqua 58 UV Toric, and Saview-Aqua 58 UV Multifocal (etafilcon A) Soft (Hydrophilic) Contact Lens are substantially equivalent to the predicate devices.

Saview-Aqua 58 UV (etafilcon A) Soft (Hydrophilic) Contact Lens 510(k)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread. The eagle is facing to the right.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 30, 2013

St Shine Optical Co., Ltd. % Ms. Ella Lee Product Manager, R & D Div. . 4,5F No. 276-2, Sec. 1, Ta Tung Rd. His Chih Dist., 221 New Taipei City Taiwan R.O.C.

Re: K132146

Trade/Device Name: Saview-Aqua 58 UV (etallicon A) Visibility Tinted Soft (Hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (I-lydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: July 25, 2013 Received: July 26, 2013

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate . commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{6}------------------------------------------------

Page 2 - Ms. Ella Lee

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Y. Alexander -S

for Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear. Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{7}------------------------------------------------

Indications for Use

510(k) Number: K132146

Device Name:

Saview-Aqua 58 UV (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens, Saview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens, Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens

Indications For Use:

The Saview-Aqua 58 UV (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.

The Saview-Aqua 58 UV Toric (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.50 diopters that does not interfere with visual acuity.

The Saview-Aqua 58 UV Multifocal (etafilcon A) Visibility Tinted Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity and require add power of up to +3.25 diopters.

The Saview-Aqua 58 UV, Saview-Aqua 58 UV Toric and Saview-Aqua 58 UV Multifocal lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. The lenses are to be prescribed for single-use daily disposable wear and not intended to be cleaned or disinfected and should be discarded after a single use.

Prescription Use (part 21 CFR 801 Subpart D)X
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Jeffrev M. Brocious -S 2013.08.28 10:54:15 -04'00' (Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices

510(k) Number:________________________________________________________________________________________________________________________________________________________________

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.