K Number
K132122

Validate with FDA (Live)

Date Cleared
2014-01-22

(196 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kestrel Posterior Cervical Fixation System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput - T3). The system is intended for posterior, cervical, non-pedicle fixation, or for posterior, noncervical pedicle fixation for the following indications:

  • Degenerative disc disease (DDD) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies
  • Spondylolisthesis
  • Trauma (i.e fracture or dislocation)
  • Spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and /or lordosis)
  • Tumor
  • Pseudoarthrosis
  • Failed previous fusion

Occipital bone screws are limited to occipital fixation only. Pedicle bone screws are limited to placement in the upper thoracic spine (TI, T3) when anchoring the OCT construct only. Pedicle screws are not intended to be placed in the cervical spine. Hooks and wires (not pedicle screws) are used to achieve cervical fusion for the occipital/cervical loop

Device Description

The proposed SPINE360 Kestrel system consists of screws, locking caps, rods, cross links, hooks, and instruments.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Important Note: The provided document is a 510(k) summary for a medical device (Kestrel Posterior Cervical Fixation System), which focuses on demonstrating substantial equivalence to a predicate device rather than outright proving clinical efficacy or specific performance metrics against pre-defined acceptance criteria in the way a clinical trial might. Therefore, many of the requested fields related to clinical study design (sample size of test set, number of experts, adjudication, MRMC studies, standalone performance, training set details) are not applicable (N/A) in this context as this document describes a mechanical testing study.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (ASTM F2706-08)Reported Device Performance
Mechanical TestingStatic Compression Bending (as per ASTM F2706-08)Kestrel performed comparably or superior to the Anatomica predicate system
Mechanical TestingStatic Torsion (as per ASTM F2706-08)Kestrel performed comparably or superior to the Anatomica predicate system
Mechanical TestingDynamic Compression Bending (as per ASTM F2706-08)Kestrel performed comparably or superior to the Anatomica predicate system
Mechanical TestingDynamic Torsion (as per ASTM F2706-08)Kestrel performed comparably or superior to the Anatomica predicate system
Material CompositionTitanium 6Al 4V ELI (ISO 5832/3, ASTM F136-02a)Meets this criterion (Titanium 6Al 4V ELI)
Indications for UseMatch predicate device: DDD, Spondylolisthesis, Trauma, Spinal Stenosis, etc.Matches predicate device (listed verbatim)
Design FeaturesScrews, connectors, hooks, wires, rods to stabilize vertebraeMatches predicate device (listed verbatim)

Study Proving Acceptance Criteria:

The study referenced is Mechanical Testing performed in accordance with ASTM F2706-08 "Standard Test Methods for Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified in the provided text. Mechanical testing standards typically require a certain number of samples for each test type, but the exact count is not given here.
  • Data Provenance: N/A for this type of mechanical testing. It's a laboratory study of physical device properties, not patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • N/A. This was a mechanical engineering study comparing the physical properties of the device to a standard and a predicate device, not a study requiring expert clinical opinion for ground truth.

4. Adjudication Method for the Test Set

  • N/A. As above, no clinical adjudication was involved. The performance was measured against established ASTM standards and compared to a predicate device's performance.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • N/A. This is not an AI-enabled diagnostic device, but a surgical implant. No MRMC study was conducted.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • N/A. This is a surgical implant, not an algorithm.

7. The Type of Ground Truth Used

  • For mechanical testing: ASTM F2706-08 Standard. The "ground truth" is defined by the rigorous methodologies and performance benchmarks specified in this industry standard for spinal implant constructs, and comparison against the predicate device.
  • For material composition and design features: Regulatory standards (ISO, ASTM) and direct comparison to the predicate device.

8. The Sample Size for the Training Set

  • N/A. There is no "training set" in the context of mechanical testing for a spinal implant. This is not a machine learning model.

9. How the Ground Truth for the Training Set Was Established

  • N/A. No training set was used.

{0}------------------------------------------------

.

SPINE
360

5000 Plaza on the Lake, Suite 305 Austin, TX 78746 512-327-6400 800-640-6045 Fax · Dave Lamb dlamb@spine360.com

.

: .

January 17, 2014

CLASS OF DEVICE

.

RE: 510(k)
Premarket Notification - Kestrel Posterior Cervical Fixation System K132122

Traditional 510(k) Summary

ESTABLISHMENT REGISTRATION NUMBER3005841736
CONTACT PERSONPrimaryDave LambQuality and Regulatory AffairsPhone: 512-327-6400 ext. 24Fax: 800-640-6045
DATE PREPAREDJune 30, 2013
CLASSIFICATION NAMEKWP 888.3050 - Spinal Interlaminal Fixation Orthosis
COMMON NAMEPosterior Occipital Cervico-thoracic Spinal System
TRADE NAMEKestrel Posterior Cervical Fixation System
PREDICATE DEVICEAnatomica Posterior Cervical Fixation System K061943

Class II

.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "SPINE" in bold, stylized letters. To the left of the word, there is a number 70. To the right of the word, there are three circles. The image has a grainy, textured appearance, giving it a vintage or distressed look.

DEVICE DESCRIPTION

The proposed SPINE360 Kestrel system consists of screws, locking caps, rods, cross links, hooks, and instruments.

INTENDED USE

The Kestrel Posterior Cervical Fixation System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput - T3). The system is intended for posterior, cervical, non-pedicle fixation, or for posterior, noncervical pedicle fixation for the following indications:

  • Degenerative disc disease (DDD) defined as back pain of discogenic origin with ● degeneration of the disc confirmed by history and radiographic studies
  • Spondylolisthesis
  • Trauma (i.e fracture or dislocation)
  • Spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and /or lordosis)
  • Tumor
  • Pseudoarthrosis
  • Failed previous fusion

Occipital bone screws are limited to occipital fixation only. Pedicle bone screws are limited to placement in the upper thoracic spine (TI, T3) when anchoring the OCT construct only. Pedicle screws are not intended to be placed in the cervical spine. Hooks and wires (not pedicle screws) are used to achieve cervical fusion for the occipital/cervical loop

TECHNOLOGICAL CHARACTERISTICS

The Kestrel Posterior Cervical System implants are manufactured from titanium alloy, Ti - A16 -4V (ISO 5832/3), which conforms to ASTM F136-02a (Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI Alloy for Surgical Implant Applications) material standards. The associated Class I instruments are made primarily of surgical grade stainless steel (ISO 7153/I).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the word "SPINE" in a stylized font. The word is part of a larger design that includes what appears to be the number 360. The overall design has a somewhat distressed or textured appearance, as if it were printed on a rough surface or has been weathered.

SUBSTANTIAL EQUIVALENCE

AnatomicaKestrel
Indications for use• Degenerative discdisease (DDD)• Spondylolisthesis• Trauma• Spinal stenosis,curvatures• Tumor• Pseudoarthrosis• Failed previous fusion• Degenerative discdisease (DDD)• Spondylolisthesis• Trauma• Spinal stenosis,curvatures• Tumor• Pseudoarthrosis• Failed previous fusion
MaterialTitanium 6Al 4V ELITitanium 6Al 4V ELI
Design featuresScrews, connectors, hooks,wires, rods connected tostabilize the vertebrae duringfusionScrews, connectors, hooks,wires, rods connected tostabilize the vertebrae duringfusion

PERFORMANCE TESTING TO ESTABLISH SUBSTANTIAL EQUIVALENCE

Mechanical Testing was performed in accordance with ASTM F2706-08 "Standard Test Methods for Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model" including Static Compression Bending, Static Torsion, Dynamic Compression Bending, Dynamic TorsionThe results of the testing indicate that the Kestrel performed comparably or superior to the Anatomica predicate system.

CONCLUSIONS

The Kestrel Posterior Cervical System demonstrated substantial equivalence to the predicate Anatomica System (K061943).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, black font. The text is likely part of a document or letterhead. The words are all capitalized and evenly spaced. The text is horizontally oriented.

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 22, 2014

Omni Surgical, LLC (dba. Spine 360) Mr. David Lamb 5000 Plaza on the Lake, Suite 305 Austin, Texas 78746

Re: K132122

Trade/Device Name: Kestrel Posterior Cervical Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminar fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: December 11, 2013 Received: December 12, 2013

Dear Mr. Lamb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. David Lamb

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins

  • for
    Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K132122 - Page 1 of 1

Image /page/5/Picture/2 description: The image shows the word "SPINE" in a stylized font. The word is slanted upwards from left to right. The word is surrounded by a design that includes curved lines and a circle. The image is in black and white.

5000 Plaza on the Lake, Suite 305 Austin, TX 78746 512-327-6400

RE: 510(k) K132122 Premarket Notification - Kestrel Posterior Cervical Fixation System Indications for Use

The Kestrel Posterior Cervical Fixation System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput - T3). The system is intended for posterior, cervical, non-pedicle fixation, or for posterior, noncervical pedicle fixation for the following indications:

  • Degenerative disc disease (DDD) defined as back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies
  • Spondylolisthesis .
  • Trauma (i.e fracture or dislocation) ●
  • Spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and /or lordosis) ●
  • Tumor .
  • Pseudoarthrosis .
  • Failed previous fusion ●

Occipital bone screws are limited to occipital fixation only. Pedicle screws are limited to placement in the upper thoracic spine (T1-T3), when anchoring the OCT construct only. Pedicle screws are not intended to be placed in the cervical spine. Hooks and wires (not pedicle screws) are used to achieve cervical fusion for the occipital/cervical loop.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Zane W. Wyatt-S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132122

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.