K Number
K132104

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-03-21

(256 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The force sensor is sterile. for single patient use.

Device Description

The device consists of one unit, containing the force sensor and an electronic display. The eLIBRA® Soft-Tissue Force Sensor for a total knee replacement is a single use battery powered device designed to show relative forces on the integrated display. The eLIBRA® Unit is a non-reusable battery powered device designed to receive an electronic signal from the integrated force sensor. The unit displays a number from 0-20 for both the medial and lateral compartments in the knee joint to aid the surgeon in balancing soft tissue structures during a total knee arthroplasty (TKA).

AI/ML Overview

Here's an analysis of the provided text regarding the Synvasive Technology, Inc. eLibra® Soft Tissue Force Sensor (K132104):

This submission is a **Special 510(k) for a device modification, not an entirely new device seeking initial clearance. Therefore, the depth of performance testing and ground truth establishment described is significantly different from what would be expected for a novel device. The primary claim of this submission is substantial equivalence to an existing predicate device (K070108 eLibra 2. Dynamic Knee Balancer).


Description of Acceptance Criteria and Study Findings

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Test DescriptionReported Device Performance
Mechanical TestingNot explicitly detailed, but implied to ensure structural integrity and correct function of the integrated components.Satisfactory. Revealed no concerns over safety and effectiveness as compared to the predicate device.
Force TestingNot explicitly detailed, but implied to ensure accurate and consistent force measurement by the sensor.Satisfactory. Revealed no concerns over safety and effectiveness as compared to the predicate device.
Electrical SafetyTesting to ensure the device meets electrical safety standards (e.g., preventing shock, electromagnetic compatibility).Satisfactory. Revealed no concerns over safety and effectiveness as compared to the predicate device.
EMC (Electromagnetic Compatibility)Testing to ensure the device operates correctly in its electromagnetic environment without interfering with or being affected by other devices.Satisfactory. Revealed no concerns over safety and effectiveness as compared to the predicate device.
Sterility TestingTesting to confirm the device maintains sterility until use (e.g., microbial ingress, package integrity).Satisfactory.
EO ResiduesTesting to ensure residual ethylene oxide (EO) levels are within safe limits if EO sterilization is used.Satisfactory.
Software ValidationTesting to confirm the software (for the integrated display) functions correctly and as intended without errors.Conducted. Results were satisfactory.
Risk AnalysisEvaluation of potential risks associated with the device modification and implementation of mitigation strategies.Conducted. Results were satisfactory, revealing no concerns.
Substantial EquivalenceDemonstrating the modified device is as safe and effective as the predicate device.The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner to the predicate device. The technical changes (integrated display) do not alter the indications for use or fundamental technological characteristics.

2. Sample Size for Test Set and Data Provenance

  • Sample Size: The document does not specify a distinct "test set" in terms of patient data or a specific number of devices tested for each category. Performance testing was conducted as "bench tests."
  • Data Provenance: Not applicable in the context of clinical data for a premarket submission focused on substantial equivalence through bench testing. The data originates from internal company testing (Synvasive Technology, Inc.) in a controlled laboratory environment. This is a retrospective submission based on the company's internal testing data.

3. Number of Experts for Ground Truth and Qualifications

  • Experts: Not applicable. For this type of Special 510(k), ground truth is established by engineering and quality assurance personnel through specified test protocols against established standards or internal company specifications derived from the predicate device's performance. Clinical expert consensus is not described as part of this evaluation.
  • Qualifications: While not explicitly stated, it's assumed that engineers, quality assurance personnel, and technical experts within Synvasive Technology, Inc., who are proficient in mechanical, electrical, software, and sterility testing, conducted and validated these tests.

4. Adjudication Method for Test Set

  • Adjudication Method: Not applicable in the traditional sense of clinical or image-based adjudication. Bench testing results are typically evaluated against pre-defined passing criteria according to test protocols. Any discrepancies or failures would lead to further investigation and re-testing until acceptance criteria are met, or the design is modified.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done?: No.
  • Effect Size: Not applicable. This device is an intraoperative mechanical tool with an integrated display, not an AI-assisted diagnostic or interpretative system that would involve human readers or image analysis. The comparison is against a predicate device, not against human performance.

6. Standalone (Algorithm Only) Performance Study

  • Was it done?: Not applicable. The "eLibra® Soft Tissue Force Sensor" is a direct measurement and display device, not an algorithm that performs a standalone diagnostic or predictive task. Its "software" refers to the embedded logic for the display, which was validated as part of software validation.

7. Type of Ground Truth Used

  • Ground Truth Type:
    • Bench Test Standards/Specifications: Performance was measured against internal engineering specifications, recognized industry standards (e.g., for electrical safety, EMC, sterility), and the known performance characteristics of the predicate device (K070108).
    • Expected Functional Output: For force testing, the "ground truth" would be calibrated force values, and the device's readings are compared against these known values. For sterility, the "ground truth" is absence of viable microorganisms.

8. Sample Size for the Training Set

  • Sample Size: Not applicable. This device does not use machine learning or AI models that require a separate training set. The "software validation" performed relates to the fixed logic of the embedded system controlling the display, not a learned model.

9. How Ground Truth for the Training Set Was Established

  • How Established: Not applicable, as there is no training set for an AI/ML model for this device.

{0}------------------------------------------------

MAR 2 1 2014

K132104 (page 1/2)

510(K) Summary K132104

Image /page/0/Picture/3 description: The image shows the logo for Synvasive Technology, Inc. The logo consists of a stylized "S" made of three curved lines, followed by the word "SYNVASIVE" in bold, sans-serif font. Below "SYNVASIVE" is the text "TECHNOLOGY, Inc." in a smaller font. Underneath the logo is the phrase "Operative Solutions to Enhance Reconstructive Surgery".

Synvasive Technology, Inc. 8690 Technology Way Reno, NV 89521 Phone: 916-939-3913 Contact: Michael G. Fisher, President Date prepared: September 4, 2013

  • Trade Name: eLibra® Soft Tissue Force Sensor 1. Common Name: Intraoperative orthopedic joint assessment aid Classification Name: Stereotaxic instrument., product code ONN, Regulation: 882.4560 Class of device: Class 2. Type of submission: Device Modification, Special 510(k).
  • The legally marketed device to which we are claiming equivalence [807.92(a)(3)] : K070108 eLibra 2. Dynamic Knee Balancer.
    1. Description of device: The device consists of one unit, containing the force sensor and an electronic display. The eLIBRA® Soft-Tissue Force Sensor for a total knee replacement is a single use battery powered device designed to show relative forces on the integrated display. The eLIBRA® Unit is a non-reusable battery powered device designed to receive an electronic signal from the integrated force sensor. The unit displays a number from 0-20 for both the medial and lateral compartments in the knee joint to aid the surgeon in balancing soft tissue structures during a total knee arthroplasty (TKA).
    1. Intended use: For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The unit is sterile, for single patient use. (Unchanged from our predicate devices)
    1. Technological characteristics: The technological characteristics are essentially identical to our predicate device with only one difference. The display was formerly reusable and wireless. The display is now integrated into the force sensor. The materials and construction techniques have not changed. The units are still battery operated. The sterilization and packaging techniques have not changed.

{1}------------------------------------------------

Image /page/1/Figure/6 description: The image shows a comparison between a predicate device labeled K070108 and a modified device. The predicate device features a rectangular unit with a display screen and buttons, along with separate right and left sensors. The modified device integrates the sensor into the display unit, as indicated by the text, and includes the word "SYNVASIVE" on the device.

    1. Performance testing: Bench tests were performed. Bench testing included mechanical testing, force testing, electrical safety, EMC, and sterility testing, including EO residues. Software validation and risk analysis was conducted. The results were satisfactory and revealed no concerns over safety and effectiveness as compared to our predicate device. The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner to the predicate device.
    1. Conclusion: The indications for use has not changed from our predicate, and the technical changes involve only integrating the formerly wireless display into the force sensing unit. Therefore the modified device is as safe and effective as our predicate device and is substantially equivalent to that device.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 21, 2014

Synvasive Technology Incorporated % Mr. Danel Kamın. P.E. Principal Engineer 8870 Ravello Court Naples, Florida 34114

Re: K132104

Trade/Device Name: eLibra Soft Tissue Force Sensor Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: II Product Code: ONN Dated: September 4, 2013 Received: February 26. 2014

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Daniel Kamm, P.E.

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

/

Sincerely yours,

Lori A. Wiggins

  • for Mark N. Melkerson Director Division of Orthopedic Devices
    Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K132104

Device Name: eLibra Soft Tissue Force Sensor

Indications For Use:

For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The force sensor is sterile. for single patient use.

Prescription Usc X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.

Page 1 of 1

Division

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).