K Number
K132090

Validate with FDA (Live)

Date Cleared
2013-08-08

(27 days)

Product Code
Regulation Number
862.3200
Age Range
18 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Randox Benzodiazepine Calibrator Set: The Randox Benzodiazepine Calibrator set consists of liquid calibrators containing Oxazepam. There are 5 levels of calibrator. They have been developed for use in the calibration of the RX series analysers in human urine, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Benzodiazepine Controls Level 1 & 2: The Randox Benzodiazepine Controls are liquid controls containing Oxazepam. There are 2 levels of controls. They have been developed for use in the quality control and validation of Benzodiazepine assay on the RX series analysers in human urine, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

Device Description

The Benzodiazepine Calibrator Set contains 5 levels of calibrator containing the specific drug Oxazepam. The Calibrators are spiked at 4 different levels which assess below, at and above the cutoff of 200 ng/ml.

The Benzodiazepine Controls are manufactured at two levels, one below the cutoff and one above the cutoff of 200 ng/ml.

The base matrix used for the manufacture of Randox Benzodiazepine Calibrators and Controls is Drug Free Human Urine.

The Benzodiazepine Calibrators and Controls contain the specific drug Oxazepam. The Oxazepam is supplied by Cerilliant Corporation the accuracy of which is ensured by purity determinations (GC/FID, HPLC and NMR) and gravimetric preparation using balances calibrated with NIST traceable weights.

Randox Benzodiazepine Calibrators and Controls have been designed for in vitro diagnostic use only. They should not be pipetted by mouth and the normal precautions for handling laboratory reagents should be applied. Randox Benzodiazepine Calibrators and Controls contain sodium azide at 0.05%.

AI/ML Overview

The provided document describes the Randox Benzodiazepine Calibrator Set and Randox Benzodiazepine Controls Level 1 and 2. This device is a set of calibrators and controls used for the calibration and quality control of Benzodiazepine assays on Randox RX series analyzers in human urine.

Here's an analysis of the acceptance criteria and study data provided:

1. A table of acceptance criteria and the reported device performance

The document details two types of stability studies, each with specific acceptance criteria and reported performance:

Study TypeAcceptance CriteriaReported Device Performance
Open Vial StabilityThe percentage deviation of the fresh vial compared to the open vial at each time point (Day 7, 21, 28) should be less than or equal to 10%.Calibrators (Batches 242516, 51110, 61110): - Day 7: Pass - Day 21: Pass - Day 28: Pass Controls (Batches 250-251DA, 51110, 61110): - Day 7: Pass - Day 21: Pass - Day 28: Pass Conclusion: The Benzodiazepine Calibrators and Controls for three batches are stable for 28 days when opened, recapped and stored at +2°C to +8ºC.
Real-Time Shelf Life (Unopened)Calibrators: Calibration absorbance achieved at each time point should be 100% ±10% when compared to the fresh assessment. Controls: (1) For Control 1, the concentration should be less than the cutoff of 200 ng/ml. (2) For Control 2, the concentration should be greater than the cutoff of 200 ng/ml.Calibrators (Batch 023-027DA): - 1 Month: Pass - 3 Months: Pass - 6 Months: Pass - 9 Months: Pass - 12 Months: Pass Conclusion: The Benzodiazepine calibrators were stable for 12 months at +2℃ to +8℃ when stored unopened. Controls (Batches 482DA, 483DA): - 3 Months: Pass - 4 Months: Pass - 6 Months: Pass - 9 Months: Pass - 12 Months: Pass Conclusion: The Benzodiazepine Controls are stable for at least 12 months at +2ºC to +8ºC when stored unopened.
Accelerated Shelf Life (Unopened)Storage at +37℃ for 5 days is equivalent to 12 months shelf life. (Implicit acceptance criteria: device remains stable at the specified time points under accelerated conditions, indicating stability for the claimed shelf life).Calibrators (Batch 023-027DA): - 1 Day: Pass - 3 Days: Pass - 5 Days: Pass - 7 Days: Pass - 10 Days: Pass - 12 Days: Pass Conclusion: The Benzodiazepine calibrators are stable for at least 12 months at +2ºC to +8ºC when stored unopened.
Value Assignment (Nest Testing)Precision measured by the CV should be less than or equal to 15%. Recovery error of the master lot should be ± 10% for all calibrator and control levels.The document states "Please see summary table below" and then lists the assigned concentrations for each calibrator and control level (e.g., Calibrator Level 1: 100 ng/ml). It implies that these values were successfully assigned according to the criteria but does not explicitly show the CVs or recovery errors for the tested batches. The passage states that the "concentration is calculated for each new batch... by nest testing," and the table presents the final calculated concentrations, suggesting the criteria were met.

2. Sample sizes used for the test set and the data provenance

  • Open Vial Stability:
    • Test set: Three batches of calibrators (242516, 51110, 61110) were tested, and three batches of controls (250-251DA, 51110, 61110) were tested. For each batch, a set of calibrators/controls was opened and assessed at Day 7, 21, and 28 against a freshly opened set.
    • Data Provenance: The studies were conducted by Randox Laboratories Limited, United Kingdom. The data is prospective, generated specifically for this submission.
  • Real-Time Shelf Life:
    • Calibrators: One batch (023-027DA) was tested, with 5 calibrator sets set aside and assessed at 1, 3, 6, 9, and 12 months.
    • Controls: Two batches (482DA, 483DA) were tested, with 5 sets of controls set aside and assessed at 3, 4, 6, 9, and 12 months.
    • Data Provenance: The studies were conducted by Randox Laboratories Limited, United Kingdom. The data is prospective.
  • Accelerated Shelf Life:
    • Calibrators: One batch (023-027DA) was tested, with six sets stored at +37°C and assessed at 1, 3, 5, 7, 10, and 12 days.
    • Data Provenance: The studies were conducted by Randox Laboratories Limited, United Kingdom. The data is prospective.
  • Value Assignment (Nest Testing):
    • Test set: Ten replicates of the test calibrators/controls were assessed against a master lot. The specific number of batches tested for value assignment is not explicitly stated beyond "each new batch."
    • Data Provenance: The studies were conducted by Randox Laboratories Limited, United Kingdom. The data is prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is Not Applicable to this device. This device is a calibrator and control set for an in vitro diagnostic assay. Its performance is measured by chemical and analytical stability and accuracy against a known standard, not by expert interpretation of clinical images or data. The "ground truth" for the test set (e.g., the concentration of Oxazepam) is established by the manufacturer (Randox) based on gravimetric preparation using NIST traceable weights and purity determinations.

4. Adjudication method for the test set

This information is Not Applicable. Clinical adjudication methods (like 2+1, 3+1) are used for evaluating diagnostic performance involving human interpretation, especially in imaging studies. For a chemical calibrator/control, performance is determined by meeting pre-defined analytical specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is Not Applicable. This device is not an AI algorithm or a diagnostic tool that involves human readers interpreting cases. It is a calibrator/control for an automated assay.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is Not Applicable. This device is a calibrator/control, not a standalone algorithm. Its performance is evaluated through analytical stability and value assignment studies in conjunction with the RX series analyzers and Benzodiazepine Assay.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the Randox Benzodiazepine Calibrator Set and Controls is the known concentration of Oxazepam. This is established through:

  • Gravimetric preparation: Oxazepam is spiked into the calibrators and controls at various concentrations.
  • Traceability to NIST traceable weights: The accuracy of the Oxazepam is ensured by Cerilliant Corporation through purity determinations (GC/FID, HPLC, and NMR) and gravimetric preparation using balances calibrated with NIST traceable weights.
  • Drug-free human urine: The base matrix for the calibrators and controls is drug-free human urine, ensuring a clear matrix for spiking.

8. The sample size for the training set

This information is Not Applicable. This device is not an AI/ML algorithm that requires a training set. The "samples" used in its development and validation are the manufactured batches of calibrators and controls themselves, which are then subjected to the stability and value assignment studies.

9. How the ground truth for the training set was established

This information is Not Applicable, as there is no "training set" in the context of an AI/ML algorithm for this device. The equivalent of "ground truth" for its manufacturing and quality control, as described in point 7, is established through precise chemical formulation, gravimetric preparation, and analytical verification of the Oxazepam concentration and purity traceable to NIST standards.

{0}------------------------------------------------

K132090

1. SAFETY AND EFFECTIVENESS AS REQUIRED BY 21 CFR 807.92 STATEMENT

AUG 8 2013

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807.92.

2. SUBMITTER NAME AND ADDRESS

Name: Dr Clodagh Finnegan clodagh.finnegan@randox.com Randox Laboratories Limited

Address: 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom.

Telephone: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912 E-mail: marketing@randox.com

3. 510k NUMBER, DEVICE PROPRIETARY NAME, COMMON NAME, PURPOSE FOR SUBMISSION. REGULATORY CLASSIFCATION. PANEL. PRODUCT CODE AND 21 CFR NUMBER

510k No: -

Device Proprietary Name: Randox Benzodiazepine Calibrator Set and Randox Benzodiazepine Controls Level 1 and 2

Common Name:

Benzodiazepine Calibrators Set (Levels 0, 1, 2, 3 and 4) Benzodiazepine Controls Levels 1 and 2

Purpose for Submission: New Device

Regulatory Classification:

Calibrators Drug Specific Clinical Toxicology Control Material

Panel: Toxicology

Product Code: DLJ & LAS

21 CFR Number: 21 CFR 862.3200 & 21CFR 862.3280

{1}------------------------------------------------

4. PREDICATE DEVICE PROPRIETARY NAMES AND 510 (k) NUMBERS

Predicate Device Proprietary Name:

Lin-Zhi International, Inc, LZI Multiple Analyte Urine Drugs of Abuse Calibrators and Controls

510 (k) Number: K051088

5. INTENDED USE

Randox Benzodiazepine Calibrator Set

The Randox Benzodiazepine Calibrator set consists of liquid calibrators containing Oxazepam. There are 5 levels of calibrator. They have been developed for use in the calibration of the RX series analysers in human urine, which includes the RX Davtona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Benzodiazepine Controls Level 1 & 2

The Randox Benzodiazepine Controls are liguid controls containing Oxazepam. There are 2 levels of controls. They have been developed for use in the quality control and validation of Benzodiazepine assay on the RX series analysers in human urine, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

6. DEVICE DESCRIPTION

The Benzodiazepine Calibrator Set contains 5 levels of calibrator containing the specific drug Oxazepam. The Calibrators are spiked at 4 different levels which assess below, at and above the cutoff of 200 ng/ml.

The Benzodiazepine Controls are manufactured at two levels, one below the cutoff and one above the cutoff of 200 ng/ml.

The base matrix used for the manufacture of Randox Benzodiazepine Calibrators and Controls is Drug Free Human Urine.

The Benzodiazepine Calibrators and Controls contain the specific drug Oxazepam. The Oxazepam is supplied by Cerilliant Corporation the accuracy of which is ensured by purity determinations (GC/FID, HPLC and NMR) and gravimetric preparation using balances calibrated with NIST traceable weights.

Randox Benzodiazepine Calibrators and Controls have been designed for in vitro diagnostic use only. They should not be pipetted by mouth and the normal precautions for handling laboratory reagents should be applied. Randox Benzodiazepine Calibrators and Controls contain sodium azide at 0.05%.

{2}------------------------------------------------

7. PREDICATE DEVICE COMPARISON TABLE

Oxazepam

DRUGS

Similarities
CHARACTERISTICSRANDOX BENZODIAZEPINECALIBRATORS AND CONTROLSLIN-ZHI DRUG MIXTURECALIBRATOR ANDCONTROLS K051088
FORMATLiquid calibrators and controlsSame
MATRIXUrine matrix for calibrators and controlsSame
Differences
CHARACTERISTICSRANDOX BENZODIAZEPINECALIBRATORS AND CONTROLSLIN-ZHI DRUG MIXTURECALIBRATOR ANDCONTROLS K051088
INTENDED USEThe Randox Benzodiazepine Calibrator setconsists of liquid calibrators containing Oxazepam.There are 5 levels of calibrator. They have beendeveloped for use in the calibration of the RX seriesanalysers in human urine, which includes the RXDaytona and the RX Imola. This in vitro diagnosticdevice is intended for prescription use only.The Randox Benzodiazepine Controls are liquidcontrols containing Oxazepam. There are 2 levels ofcontrols. They have been developed for use in thequality control and validation of Benzodiazepineassay on the RX series analysers in human urine,which includes the RX Daytona and the RX Imola.This in vitro diagnostic device is intended forprescription use only.Intended for in vitro diagnosticuse for the calibration andvalidation of LZI DAU enzymeimmunoassays to detectmethamphetamine, opiate,phencyclidine, benzoylecgonine,barbiturates, methadone, andpropoxyphene in human urine
LEVELSCalibrators: there are a total of 5 levels, includingnegativeControls: there are two levelsTotal of 7 levels, includingnegative
STORAGE(Unopened)Calibrators and Controls: when not in use bottlesshould be capped at all times and refrigerated at2-8°C until expiration date.Calibrators and Controls whennot in use bottles should becapped at all times andrefrigerated at 2-8°C.
OPEN VIAL CLAIMCalibrators and Controls are stable for 28 daysafter opening capped and stored at 2-8°CUnknown
Calibrators and Controls:BenzodiazepineMultianalyteMethamphetamine, Opiate

Phencyclidine, Benzolecgonine Benzodiazepines, Barbiturates Methadone, Propoxyphene

{3}------------------------------------------------

8. SUMMARY OF STABILITY STUDIES

Open vial stability of the Benzodiazepine Calibrators and Controls level 1 and 2 were assessed by opening a set of calibrators and controls, replacing the caps and storing at +2℃ to +8℃ for 28 days. At day 7, 21 and 28 an aliquot was removed for assessment. The recovery of each calibrator and control was compared to a freshly opened set of calibrators and controls at day 7, 21 and 28. The acceptance criteria was the percentage deviation of the fresh vial compared to the open vial at each time point should be less than or equal to 10%. Table 1 below shows a summary of the open vial stability for Benzodiazepine calibrators and controls. The instrument used for this assessment was the Randox Imola (K052914). The assay used was the Benzodiazepine Assay (K092274).

Table 1
Batch LotMaterialTime point
Day 7Day 21Day 28
242516Calibrators 1-4PassPassPass
51110Calibrators 1-4PassPassPass
61110Calibrators 1-4PassPassPass
250-251DAControls 1-2PassPassPass
51110Controls 1-2PassPassPass
61110Controls 1-2PassPassPass

The data demonstrates that the Benzodiazeoine Calibrators and Controls for three batches are stable for 28 days when opened, recapped and stored at +2°C to +8ºC.

Shelf life for Benzodiazepine Calibrators was assessed by real time stability studies and accelerated stability studies. Once a new batch of calibrators are manufactured, 5 calibrator sets are set aside and stored at +2℃ to +8℃. After 1. 3, 6, 9 and 12 months as set of calibrators are assessed and compared to day 0 calibrators. The acceptance criteria states the calibration absorbance achieved at each time point should be 100% ±10% when compared to the fresh assessment. The instrument used for this assessment was the Randox Imola (K052914). The assay used was the Benzodiazepine Assay (K092274).

1ableC
Batch LotMaterialTime point in months
136912
023-027DACalibrators1-4PassPassPassPassPass

The data demonstrates in table 2 that the Benzodiazepine calibrators were stable for 12 months at +2℃ to +8℃ when stored unopened.

{4}------------------------------------------------

Accelerated stability studies were used to predict the shelf life of the Benzodiazepine calibrator set. Six sets of Benzodiazepine calibrators were stored at +37°C for 1, 3, 5, 7, 10 and 12 days. At each time point the calibration absorbance's achieved are compared to those achieved at Day 0. Storage at +37℃ for 5 days is equivalent to 12 months shelf life.

Table 3

Batch LotMaterialTime point in Days
13571012
023-027DACalibrators 1-4PassPassPassPassPassPass

The data demonstrates in table 3 that the Benzodiazepine calibrators are stable for at least 12 months at +2ºC to +8ºC when stored unopened.

time stability testing for Benzodiazepine Controls. Following Real the manufacture of Benzodiazepine controls 5 sets are set aside and stored at +2°C to +8℃. At 3. 4. 6. 9 and 12 months a set of controls are assessed. The acceptance criteria for Benzodiazepine control 1 states the concentration should be less than the cutoff of 200 ng/ml and for Benzodiazepine control 2 the concentration should be greater than the cutoff of 200ng/ml.

Batch LotMaterialTime point in Months
346912
482DAControls 1-2PassPassPassPassPass
483DAControls 1-2PassPassPassPassPass

The data demonstrates that the Benzodiazepine Controls are stable for at least 12 months at +2ºC to +8ºC when stored unopened.

9. SUMMARY OF VALUE ASSIGNMENT

A concentration is calculated for each new batch of Randox Benzodiazepine Calibrators and Controls by nest testing. Nest testing involves assessment of the new lot of controls against a master lot of controls or calibrators. Ten replicates of the test calibrators/controls are assessed and the mean and CV calculated. The recovery of the master lot is also measured. The acceptance criteria states the precision measured by the CV should be less than or equal to 15%. The recovery error of the master lot is also measured and should be ± 10% for all calibrator and control levels. The instrument used for this assessment was the Randox Imola (K052914). The assay used was the Benzodiazepine Assay (K092274).

{5}------------------------------------------------

Please see summary table below.

Calibrator / Control LevelLot #Concentration of Oxazepam (ng/ml)
Calibrator Level 1584DA100
Calibrator Level 2585DA200
Calibrator Level 3586DA300
Calibrator Level 4587DA1000
Control Level 1596DA150
Control Level 2597DA250

10. TRACEABILITY

The Benzodiazepine Calibrators and Controls are prepared using drug free human urine as the base material. Oxazepam is spiked into the calibrators and controls at various concentrations. The Oxazepam is supplied by Cerillant Corporation the accuracy of which is ensured by purity determinations (GC/FID, HPLC and NMR) and gravimetric preparation using balances calibrated with NIST traceable weights.

11. CONCLUSION

Testing results indicate that the proposed device is substantially equivalent to the predicate devices.

{6}------------------------------------------------

EPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains an image of an eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the perimeter of the circle.

Public Health Service

Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G6 Silver Spring, MD 20993-0002

August 8, 2013

Randox Laboratories Limited C/O Dr. Pauline Armstrong 55 Diamond Road, Crumlin County Antrim, BT29 4QY UNITED KINGDOM

Re: K132090

Trade/Device Name: Randox Benzodiazepine Calibrator Set Randox Benzodiazepine Controls Level 1 and 2 Regulation Number: 21 CFR 862.3200 Regulation Name: Clinical toxicology calibrator Regulatory Class: II Product Code: DLJ, LAS Dated: July 3, 2013 Received: July 15, 2013

Dear Dr. Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, I isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must CER Port 807), labeline (21) Connements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2-Dr. Armstrong

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please the Colliness. It a gov/AboutFDA/CentersOffices/CDRH/CDRHORTics://cm115809.html for the Center for Devices and Radiological Health's (CDRH) s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

l1110//WWW.Ida.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollit the number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K132090

Device Name: Randox Benzodiazepine Calibrator Set Randox Benzodiazepine Controls Level 1 and 2

Indications for Use:

Randox Benzodiazepine Calibrator Set

The Randox Benzodiazepine Calibrator set consists of liquid calibrators containing Oxazepam. There are 5 levels of calibrator. They have been developed for use in the calibration of the RX series analysers in human urine, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Benzodiazepine Controls Level 1 &2

The Randox Benzodiazepine Controls are liquid controls containing Oxazepam. There are 2 levels of controls. They have been developed for use in the quality control and validation of Benzodiazepine assay on the RX series analysers in humar winro, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use __ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnson-lyles -S 2013.08.08 07:32:10 -04'00'

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K132090

§ 862.3200 Clinical toxicology calibrator.

(a)
Identification. A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.