ENDOCHOICE HOT BIOPSY FORCEPS
K131991 · Endochoice, Inc. · KGE · Oct 15, 2013 · Gastroenterology, Urology
Device Facts
| Record ID | K131991 |
| Device Name | ENDOCHOICE HOT BIOPSY FORCEPS |
| Applicant | Endochoice, Inc. |
| Product Code | KGE · Gastroenterology, Urology |
| Decision Date | Oct 15, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.4300 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
EndoChoice Hot Biopsy Forceps are intended for use by trained medical professionals. The device is to be used endoscopically in conjunction with monopolar electrosurgical current to obtain gastrointestinal mucosal tissue biopsies and for removal of sessile polyps.
Device Story
Sterile, single-use endoscopic biopsy forceps; used by trained medical professionals in gastrointestinal tract. Device features handle-actuated jaws and electrical plug for connection to external monopolar electrosurgical unit. Operation: user inserts device through endoscope (min 2.8 mm channel); positions jaws against mucosal wall; closes jaws on tissue; applies high-frequency monopolar current to cauterize and resect; retrieves tissue sample. Enables simultaneous tissue biopsy and sessile polyp removal. Benefits include efficient tissue sampling and cauterization during endoscopic procedures.
Clinical Evidence
Bench testing only. Includes functional performance, biocompatibility, sterility assurance, shelf life, and electrical safety testing for high-frequency surgical accessories. No clinical data presented.
Technological Characteristics
Sterile, single-use endoscopic accessory. Working length 230.0 cm; compatible with 2.8 mm endoscopic channels. Monopolar electrosurgical energy source. Manual jaw actuation. No software or electronic processing components.
Indications for Use
Indicated for endoscopic use in conjunction with monopolar electrosurgical current to obtain gastrointestinal mucosal tissue biopsies and remove sessile polyps in patients requiring such procedures.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
- Radial Jaw 4TM Hot Biopsy Forceps (K101657)
Related Devices
- K972505 — RESUABLE HOT BIOPSY FORCEPS · United States Endoscopy Group, Inc. · Sep 18, 1997
- K101657 — RADIAL 4 HOT BIOPSY FORCEPS BOX 5; RADIAL JAW 4 HOT BIOPSY FORCEPS BOX 20; RADIAL JAW 4 HOT BIOPSY FORCEPS BOX 40 · Boston Scientific Corp · Oct 5, 2010
- K171973 — Captura® Disposable Hot Biopsy Forceps · Wilson-Cook Medical, Inc. · Mar 21, 2018
- K160625 — Disposable Hot Biopsy Forceps · Micro-Tech (Nanjing) Co., Ltd. · Jun 14, 2016
- K050136 — REPROCESSED HOT BIOPSY FORCEPS · Sterilmed, Inc. · Jul 20, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
EndoChoice Hot Biopsy Forceps
EndoChoice, Inc.
# 510(k) Summary EndoChoice Hot Biopsy Forceps
## 1. Company Identification
EndoChoice, Inc. 11800 Wills Road, Suite 100 Alpharetta, GA 30009 Telephone Fax (678) 567 8218 Establishment Registration: 300759133
**OCT 15 2013**
## 2. Contact Person
Daniel Hoefer Regulatory Affairs Manager
#### 3. Device Name
Commercial name: EndoChoice Hot Biopsy Forceps Classification name: Forceps, Biopsy, Electric
#### 4. Device Classification
Product Code: KGE Regulation Number: 876.4300 Class: II
#### 5. Intended Use
EndoChoice Hot Biopsy Forceps are intended for use by trained medical professionals. The device is to be used endoscopically in conjunction with monopolar electrosurgical current to obtain gastrointestinal mucosal tissue biopsies and for removal of sessile polyps.
#### 6. Device Description
EndoChoice Hot Biopsy Forceps is a sterile single use device designed for use in the gastrointestinal tract. A plug is located on the handle for connection to an electrosurgical unit, giving the capability to apply high frequency monopolar electrocautery during resection.
The device is compatible with endoscopic access channels of at least 2.8 mm and is provided with a working length of 230.0 cm.
Using the EndoChoice Hot Biopsy Forceps, the user can cauterize and remove tissue from the mucosal wall by using the handle to open the jaws, pressing the jaws against the site, closing, applying electrical current via a compatible electrosurgical unit (not included), and gently pulling the away from the site.
page I of 2
{1}------------------------------------------------
EndoChoice, Inc.
#### 7. Substantial Equivalence
The EndoChoice Hot Biopsy Forceps is substantially equivalent to the Radial Jaw 4TM Hot Biopsy Forceps (K101657) manufactured by Boston Scientific Corporation.
The devices are equivalent in terms of intended use, operating principle, technology, energy used, packaging, and materials.
## 8. Non-Clinical testing
Testing includes:
- Bench testing of the functional performance, .
- biocompatibility testing, .
- a sterility assurance validation,
- . shelf life testing,
- . electrical safety testing for high frequency surgical accessories.
Results demonstrate that the device is safe and effective in meeting user requirements in accordance with its intended use.
#### 9. Conclusion
The EndoChoice Hot Biopsy Forceps is substantially equivalent to the predicate device listed above.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 15, 2013
EndoChoice, Inc. % Daniel Hoefer Regulatory Affairs Manager 11810 Wills Road Alpharetta, GA 30009
Re: K131991
> Trade/Device Name: EndoChoice Hot Biopsy Forceps Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KGE Dated: August 19, 2013 Received: August 20, 2013
Dear Daniel Hoefer,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of
{3}------------------------------------------------
Page 2 - Daniel Hoefer
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131991
Device Name: EndoChoice Hot Biopsy Forceps
Indications for Use:
EndoChoice Hot Biopsy forceps are indicated for endoscopic use in in conjunction with monopolar electrosurgical current to obtain gastrointestinal mucosal tissue biopsies and for removal of sessile polyps.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Benjamin R. Fisher -S 2013.10.15 10:41:30 -04'00' Page 1 of 1