K Number
K131694

Validate with FDA (Live)

Date Cleared
2013-08-08

(59 days)

Product Code
Regulation Number
880.5965
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NMI Port II with and without PASV Valve Technology is indicated for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

When used with a power injectable needle, the NMI Port II is indicated for power injection of contrast media. The maximum recommended infusion rate is 5 ml/sec with a 19G or 20G noncoring power injectable needle or 2 ml/sec with a 22G non-coring power injectable needle.

Device Description

The NMI Port II with and without PASV Valve Technology is a subcutaneous implantable venous access device with one reservoir and is designed for optional power injection of contrast media, CECT. The ports are designed to be accessed using a non-coring Huber needle introduced through the skin into the self-sealing silicone septum covering the reservoir.

NMI Port II is available in plastic or titanium single lumen and valved or non-valved configurations. The ports are available with either silicone filled or non-filled suture fixation holes. Ports with non-filled suture fixation holes are generally utilized based on clinical need to anchor the port to the subcutaneous tissue; whereas ports with filled suture holes, designed to prevent tissue in-growth to the suture holes, are generally utilized when not anchoring the port to the subcutaneous tissue. If needed, filled suture holes are accessed through the silicone. All port configurations have a radiopaque identifier (CT mark) to identify the port as power injectable. The radiopaque catheter has graduated marks at 1 centimeter intervals and can be cut to the desired length by the clinician. Ports are provided with a variety of procedural accessories.

The catheter shaft incorporates Endexo polymer for improved resistance to thrombus accumulation and/or formation on the catheter.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the NMI Port II, based on the provided text:

Acceptance Criteria and Device Performance for NMI Port II (K131694)

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary does not explicitly list quantitative acceptance criteria for each performance test. Instead, it states that the device "successfully passed relevant testing per the above Guidance, standards, and pre-established acceptance criteria." The identified performance tests and the reported outcome are:

Performance Test CategoryAcceptance Criteria (Explicitly Stated)Reported Device Performance
General ComplianceAdherence to:Successfully passed
- FDA Guidance (Oct 1990)- FDA guidance on 510(k) Submissions
- EN ISO 10555-1:2009- Standard requirements
- EN ISO 10555-3:1997 COR 2002- Standard requirements
- Biocompatibility (ISO 10993-1)- Standard requirements
Specific Performance TestsPre-established acceptance criteriaSuccessfully passed
- Internal Product Specification(Not explicitly defined in document)
- Power Injection(Not explicitly defined in document)
- Valve Integrity(Not explicitly defined in document)
- Catheter Compatibility(Not explicitly defined in document)
- Port Septum Testing(Not explicitly defined in document)
- Chemical / Vesicant Compatibility(Not explicitly defined in document)
- Thromboresistance(Not explicitly defined in document)

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not provide details on the sample sizes used for any of the performance tests (Internal Product Specification, Power Injection, Valve Integrity, etc.).
It also does not specify the data provenance (e.g., country of origin, retrospective or prospective nature) for these tests. Based on the context of a 510(k) submission, these tests would typically be conducted by the manufacturer in a controlled laboratory setting (i.e., prospective engineering and bench testing) rather than involving human subject data from a specific country.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The disclosed information does not involve a "test set" in the context of expert-reviewed data (e.g., medical images). The "tests" mentioned are primarily engineering and bench performance evaluations of the device itself. Therefore, the concept of "experts used to establish ground truth" with specific qualifications like radiologists with years of experience is not applicable to the studies described. The ground truth for these engineering tests would be established by the physical and chemical properties of the device and its interaction with test fluids and equipment, according to established industry standards and internal specifications.

4. Adjudication Method for the Test Set

Since the studies described are engineering and bench tests rather than studies involving expert review of data, the concept of an "adjudication method" (like 2+1 or 3+1) is not applicable. The outcome of these tests would be determined by whether the device met the pre-defined engineering specifications and performance metrics.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The provided document describes bench and engineering tests, not studies involving human readers or comparative effectiveness of AI (or any other technology) with or without human assistance.

6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable as the device is a physical medical device (an implanted port catheter), not an algorithm or AI software. Therefore, there is no "standalone algorithm" performance to evaluate.

7. The Type of Ground Truth Used

For the performance tests mentioned, the "ground truth" would be established by:

  • Engineering Specifications: Internal product specification requirements.
  • International Standards: The pass/fail criteria defined within EN ISO 10555-1, EN ISO 10555-3, and ISO 10993-1.
  • Regulatory Guidance: The requirements outlined in FDA's "Guidance on 510(k) Submissions for Implanted Infusion Ports."
  • Bench Test Results: Measurable outcomes from laboratory tests (e.g., pressure resistance during power injection, flow rates, material compatibility, integrity assessments).

8. The Sample Size for the Training Set

This question is not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set for a physical medical device.

{0}------------------------------------------------

K131694

510(k) Summary for the NMI Port II

Date prepared: 07-June-2013

AUG 0 8 2013

A. Sponsor

Navilyst Medical, Inc 26 Forest Street Marlborough, MA 01752

B. Contact

Brandon M. Brackett Specialist, Global Regulatory Affairs 508-658-7984

OR

Lorraine M. Hanley Vice President, Global Regulatory Affairs 508-494-1129

C. Device Name

Trade Name: Common/Usual name: Classification Name:

Classification Panel:

NMI Port II

Implanted Port Catheter, Subcutaneous. Implanted, Intravascular Infusion Port and Catheter 21CFR§880.5965, Class II General Hospital

D. Predicate Device(s)

Trade Name: Common/Usual name: Classification Name:

Classification Panel:

Premarket Notification:

Trade Name: Common/Usual name: Classification Name:

Classification Panel:

Premarket Notification:

NMI Port

Implanted Port Catheter, Subcutaneous, Implanted, Intravascular Infusion Port and Catheter

21CFR§880.5965, Class II General Hospital

K122767

NMI PICC III

Peripherally Inserted Central Catheter (PICC) Short and Long-Term Intravascular Catheter

21CFR§880.5970, Class II General Hospital

K121089

{1}------------------------------------------------

E. Device Description

The NMI Port II with and without PASV Valve Technology is a subcutaneous implantable venous access device with one reservoir and is designed for optional power injection of contrast media, CECT. The ports are designed to be accessed using a non-coring Huber needle introduced through the skin into the self-sealing silicone septum covering the reservoir.

NMI Port II is available in plastic or titanium single lumen and valved or non-valved configurations. The ports are available with either silicone filled or non-filled suture fixation holes. Ports with non-filled suture fixation holes are generally utilized based on clinical need to anchor the port to the subcutaneous tissue; whereas ports with filled suture holes, designed to prevent tissue in-growth to the suture holes, are generally utilized when not anchoring the port to the subcutaneous tissue. If needed, filled suture holes are accessed through the silicone. All port configurations have a radiopaque identifier (CT mark) to identify the port as power injectable. The radiopaque catheter has graduated marks at 1 centimeter intervals and can be cut to the desired length by the clinician. Ports are provided with a variety of procedural accessories.

The catheter shaft incorporates Endexo polymer for improved resistance to thrombus accumulation and/or formation on the catheter.

F. Indication for Use

The NMI Port II with and without PASV Valve Technology is indicated for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

When used with a power injectable needle, the NMI Port II is indicated for power injection of contrast media. The maximum recommended infusion rate is 5 ml/sec with a 19G or 20G noncoring power injectable needle or 2 ml/sec with a 22G non-coring power injectable needle.

G. Summary of Similarities and Differences in Technological Characteristics and Performance

The proposed device has similar materials, design and components and technological characteristics as predicate devices. Both the NMI Port II and predicate ports are, in brief, intended for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products; available in single lumen configurations; plastic or titanium port body available with either a 6F or 8F outside diameter catheter; rated for maximum power injector settings up to 300 psi with maximum power injection flow rate up to 5 ml/second based on model; and available kitted with a variety of procedural accessories.

{2}------------------------------------------------

H. Performance Data

The performance evaluation of the NMI Port II included testing conducting in accordance with the following FDA guidance documents and international standards:

  • . FDA's "Guidance on 510(k) Submissions for Implanted Infusion Ports dated October 1990.
  • . EN ISO 10555-1:2009, Sterile, Single Use Intravascular Catheters - Part 1: General Requirements
  • EN ISO 10555-3:1997 COR 2002, Sterile, Single Use Intravascular Catheters -- Part 3: . Central Venous Catheters
  • . Biocompatibility per ISO 10993-1

The proposed NMI Port II successfully passed relevant testing per the above Guidance, standards, and pre-established acceptance criteria, including:

  • . Internal Product Specification Requirements
  • Power Injection .
  • t Valve Integrity
  • Catheter Compatibility with Procedural Aid Devices .
  • Port Septum Testing .
  • Chemical / Vesicant Compatibility .
  • . Thromboresistance

I. Conclusion

Based on successful results of testing and on responses to questions posed in FDA's 510(k) Decision Making Tree, the proposed device is determined to be substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

August 08, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MIJ 20993-0002

Navilyst Medical, Incorporated C/O Mr. Brandon M. Brackett Specialist, Global Regulatory Affairs 26 Forest Street Marlborough, MA 01752

Re: K131694

Trade/Device Name: NMI Port II Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutancous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: June 7, 2013 Received: June 12, 2013

Dear Mr. Brackett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fdagov/MedicalDevices/Resourcesfor You/Industrvidefault.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kwame O. Ulmer -S

Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

×131694 510(k) Number (if Known):

Device Name: NMI Port II

Indications for Use:

The NMI Port II with and without PASV Valve Technology is indicated for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

When used with a power injectable needle, the NMI Port II is indicated for power injection of contrast media. The maximum recommended infusion rate is 5 ml/sec with a 19G or 20G noncoring power injectable needle or 2 ml/sec with a 22G non-coring power injectable needle.

Prescription Use
(21 CFR 801 Subpart D)

ズ And/Or AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Richard C. Chapman 2013.08.07 17:01:31 -04'00'

(Division Sign-Off) (Division Sign-On)
Division of Anesthestology, General Hospital Division of Ancolnocionology, Dental Devices

510(k) Number;

4-2

§ 880.5965 Subcutaneous, implanted, intravascular infusion port and catheter.

(a)
Identification. A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.