K Number
K131662

Validate with FDA (Live)

Date Cleared
2013-10-10

(125 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMMULITE® ACTH Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration and reportable range of the IMMULITE ACTH assay on the IMMULITE 2000 systems

Device Description

The Calibration Verification Material (CVM) contains one set of four vials, 3 mL each. LACCVM1 contains processed bovine protein matrix with preservatives. LACCVM2, LACCVM3 and LACCVM4 contain low, intermediate and high levels of ACTH respectively, in processed bovine protein based matrix with preservatives.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the IMMULITE® 2000 ACTH Calibration Verification Material, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CVM LevelAcceptance Criteria (Guideline Criteria % difference to assigned dose)Acceptable Dose Range (pg/mL)Reported Performance (Implied from meeting criteria)
LACCVM1≤5.00≤5.00Met criteria (as study claims success)
LACCVM2±1019.62 - 23.98Met criteria (as study claims success)
LACCVM3±10214.20 - 261.80Met criteria (as study claims success)
LACCVM4±101092.6 - 1335.40Met criteria (as study claims success)
Review Limits (Secondary Acceptance Criteria): Controls are within 2SD of target from stability calibrator curve.

Note: The document states "Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use." This implies that the reported device performance met these acceptance criteria, though specific numerical results within the ranges are not explicitly provided.

Study that Proves the Device Meets the Acceptance Criteria:

The study conducted was a stability study designed to validate the shelf life claim for the IMMULITE® 2000 ACTH Calibration Verification Material (CVM).

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Description: The document refers to the CVM study protocols being run as part of calibrator stability testing. The CVMs (test samples) were run in duplicate (as a minimum) at specified time points.
  • Sample Size for Stability Testing:
    • For the stability study, CVM levels (LACCVM1, LACCVM2, LACCVM3, LACCVM4) were tested at time points: Day 1, Day 182, Day 548, and Day 730. The quantity of "samples" at each time point is "duplicate (as a minimum)".
    • For the Expected Values/Reference Range establishment, which involved testing of each CVM level to define the target mean and SD, there were "a total of 27 replicates; 9 runs and 3 replicates per run."
  • Data Provenance: The data is internally generated from a non-clinical performance testing which makes it retrospective in nature. No country of origin is explicitly mentioned other than the submitter's address in Tarrytown, NY, USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This is not applicable as the "ground truth" for this device, a calibration verification material, is established through a precise, gravimetrically prepared reference standard and an internal value assignment process, not through expert human interpretation.

4. Adjudication Method for the Test Set:

Not applicable. The "ground truth" (assigned dose) is determined through instrumental and gravimetric methods, not through human adjudication. The acceptance criteria include two parts: a guideline percentage difference to the assigned dose and a review limit of controls within 2SD. If the result is not within the acceptable dose range of ±10%, then additional data review is performed using the part 2 acceptance criterion. This implies a hierarchical review process, but not expert adjudication in the traditional sense of diagnostic imaging or clinical assessment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance:

No, an MRMC comparative effectiveness study was not conducted. This device is a calibration verification material, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is a chemical assay control material, not a software algorithm. Its "performance" is its ability to maintain its intended value and characteristics over time, which is assessed through laboratory testing.

7. The Type of Ground Truth Used:

The ground truth for the IMMULITE® 2000 ACTH Calibration Verification Materials is based on traceability to internal assigned reference calibrators prepared using ACTH antigen stock solution that are gravimetrically prepared. This represents a form of measurement standard/reference material ground truth.

8. The Sample Size for the Training Set:

The concept of a "training set" is not directly applicable in the conventional sense for this type of device (calibration verification material). However, the value assignment process for the CVMs, which establishes their target values, involved extensive testing:

  • The CVMs were tested on 27 replicates in total (comprised of nine runs and three replicates per run) on eight systems and four different reagent kit lots.
  • Additionally, two levels of commercially available controls, and 40 patient samples (5 normal patients samples and 35 spiked normal patients samples) are used to validate CVM value assignments.

This extensive testing serves a similar purpose to establishing robust targets in a training phase.

9. How the Ground Truth for the Training Set Was Established:

As noted above, "training set" isn't a direct fit. However, the ground truth for the CVM's assigned values (which could be considered the "ground truth" for its utility) was established through:

  • Traceability: The CVMs are traceable to internal assigned reference calibrators.
  • Gravimetric Preparation: These reference calibrators are prepared using ACTH antigen stock solution that is gravimetrically prepared (i.e., measured by weight for high accuracy).
  • Internal Value Assignment Procedures: The IMMULITE calibrators (and thus CVMs) are value-assigned using these assigned reference calibrators.
  • Validation Studies: Value assignments are validated using commercially available controls and 40 patient samples (5 normal, 35 spiked normal).
  • Extensive replicate testing used to establish the target mean and standard deviation for each CVM level.

{0}------------------------------------------------

SIEMENS

510(k) Summary

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.

The assigned 510(k) Number:

    1. Submitter Mailing Address:
      Contact Person:

Phone Number: Fax Number: E-mail Address: Date Prepared:

    1. Device Name Proprietary Name: Measurand: Type of Test:
      Regulation Section: Classification: Products Code:

Panel:

    1. Predicate Device Name Predicate 510(k) No:
    1. Device Description:
    1. Intended Use: Indication for Use:

Special Conditions for Use Statement(s): Special Instrument Requirements:

Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591

Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com August 30, 2013

IMMULITE® 2000 ACTH Calibration Verification Material Quality Control materials for IMMULITE® 2000 ACTH assay Calibration Verification Material (CVM) for IMMULITE® 2000 ACTH assay 21 CFR 862.1660, Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassaved) Clinical Chemistry (75)

Elecsys ACTH CalCheck K060585

The Calibration Verification Material (CVM) contains one set of four vials, 3 mL each. LACCVM1 contains processed bovine protein matrix with preservatives. LACCVM2, LACCVM3 and LACCVM4 contain low, intermediate and high levels of ACTH respectively, in processed bovine protein based matrix with preservatives.

See Indications for Use Statement below The IMMULITE® ACTH Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration and reportable range of the IMMULITE ACTH assay on the IMMULITE 2000 systems

For prescription use only

IMMULITE® 2000 Systems

OCT 1 0 2013

{1}------------------------------------------------

    1. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
      A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 ACTH Calibration Verification Material (CVM) is substantially equivalent to the predicate device, Elecsys ACTH CalCheck, as summarized in Table 1.
Table 1: Substantial Equivalence Comparison
SIMILARITIES
Candidate DeviceIMMULITE 2000 ACTH CVMPredicate DeviceElecsys ACTH CalCheck
Intended UseThe IMMULITE® ACTH CalibrationVerification Material (CVM) is for in vitrodiagnostic use in the verification of calibrationand reportable range of the IMMULITEACTH assay on the IMMULITE 2000 systemsFor use in the verification of thecalibration established by the ElecsysACTH reagent on the indicated Elecsysand cobas e immunoassay analyzers.
AnalyteACTHSame
FormLyophilizedSame
TraceabilityStandardized gravimetricallySame
DIFFERENCES
StabilityStable until the expiration date when storedfrozen.Stable until the expiration date whenstored refrigerated.
Storage-20°C2-8°C
MatrixBovine protein based matrix withpreservatives.Buffered equine serum with preservatives.
UseSingle Use OnlyNot for Single Use

7. Non-Clinical Performance Testing

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.

7.1 Stability Summary:

The stability study was conducted to validate the shelf life claim for the IMMULITE® 2000 ACTH Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platform throughout the established shelf life of the CVM. The IMMULITE® 2000 ACTH Calibration Verification Materials (CVMs) are stable up to 2 years when stored frozen at -20℃ prior to opening.

7.1.1 Stability Protocol Summary:

The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) at the time points shown in Table 2 and the dose value determined from the reference calibrator curve.

CVM LevelTime-Points(Days)
LACCVM11182548730
LACCVM21182548730
LACCVM31182548730
LACCVM41182548730

Table 2: Stability Time Points:

{2}------------------------------------------------

7.1.2 Stability Acceptance Criteria Summary:

The Acceptance Criteria for the IMMULITE 2000 ACTH Calibration Verification Material (CVM) consists of 2 parts. Part 1 consists of the Guideline Acceptance Criteria which requires the dose value stability CVM to fall between ±10% of the assigned dose. Part 2 consists of the Review Limits Acceptance Criteria which requires the dose value of the controls to be within 2SD of the control target value generated from the stability calibrator curve. If the result is not within the acceptable dose range of ±10%, then additional data review is performed using the part 2 acceptance criterion. The acceptance criteria is summarized in Table 3.

CVM LevelAssignedDose (pg/mL)Guideline Criteria %difference to assigned doseAcceptable doserange (pg/mL)Review Limits
LACCVM10.00≤5.00≤5.00Not Applicable
LACCVM221.8±1019.62 - 23.98Controls are within 2SD
LACCVM3238±10214.20 - 261.80of target from stability
LACCVM41214±101092.6 - 1335.40calibrator curve

Table 3: Stability Acceptance Criteria for IMMULITE 2000 ACTH CVM

7.2 Traceability:

The IMMULITE® 2000 ACTH Calibration Verification Materials are traceable to internal assigned reference calibrators prepared using ACTH antigen stock solution and are traceable to internal material which is gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures.

7.3 Value Assignment:

ACTH CVMs are 4 level materials which are subset of 8 level ACTH calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of ACTH reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The assigned reference calibrators are prepared using recombinant human ACTH 1-39 antigen stock and are traceable to internal material which has been gravimetrically prepared. Two levels of commercially available controls, and 40 patient samples (5 normal patients samples and 35 spiked normal patients samples) are used to validate CVM value assignments.

The CVMs are manufactured using qualified materials and measurement procedures. The CVMs were tested on 27 replicates in total comprised of nine runs and three replicates per run on eight systems and four different reagent kit lots. The CVMs dose values are generated using. Surve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall, within their target ranges

{3}------------------------------------------------

SIEMENS

7.4 Expected Values/Reference Range:

Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run. 5 different reagent kit lots and 8 different instruments were used to gain 27 replicates. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected values are provided in the IMMULITE® 2000 CVM Calibration verification Material lot-specific value card. The expected assay range is 5 - 1250 pg/mL. The target values in Table 4 can be considered as guidelines.

Analyte target levelsLevelTarget Mean(pg/mL)StandardDeviation(SD)Guideline ±2SD Range(pg/mL)
10.000.00≤5.00
221.82.217.426.2
323822.5193283
412149110321396
Assay Range5 -1250 pg/mL

Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.

Standard/Guidance Documents Referenced:

  • CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
  • Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic . Calibrators
  • Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control . Material

Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10

8. Conclusion:

The IMMULITE® 2000 ACTH Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Elecsys ACTH CalCheck. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, Elecsys ACTH CalCheck, The IMMULITE® 2000 ACTH Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. The text is in all capital letters and is smaller than the eagle symbol. The image is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002

October 10, 2013

Siemens Healthcare Diagnostics, Inc. c/o Garo Mimaryan 511 Benedict Ave TARRYTOWN NY 10591-5097

Re: K131662

Trade/Device Name: Immulite 2000 ACTH Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I, reserved Product Codc: JJX Dated: September 10, 2013 Received: September 11, 2013

Dear Garo Mimaryan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misJeading,

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1 050.

{5}------------------------------------------------

Page 2—Garo Mimaryan

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/McdicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): kl31662

Device Name: IMMULITE®2000 ACTH Calibration Verification Material

Indication for Use:

The IMMULITE® ACTH Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration and reportable range of the IMMULITE ACTH assay on the IMMULITE 2000 systems

Prescription Use_X_ (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use_ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Yung-W.Chan -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K131662

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.