K Number
K131637

Validate with FDA (Live)

Date Cleared
2013-07-12

(38 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RMST Staple is intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

Device Description

RMST Staple is an absorbable polymer strap with cleat tips. RMST Staple is composed of absorbable synthetic polyester derived from lactic acid and dyed with D&C Violet #2. The RMST Staple is used in conjunction with an orthopedic manual staple driver from Rotation Medical.

AI/ML Overview

The Rotation Medical Soft Tissue Staple (RMST Staple) was evaluated through a series of in vitro and in vivo tests to establish its safety and substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

Test CategoryAcceptance Criteria/DescriptionReported Device Performance
BiocompatibilityCompliance with FDA Blue Book Memorandum G95-1 and ISO 10993-1 for biological evaluation.The device passed all applicable FDA Blue Book Memorandum G95-1 and ISO 10993-1 testing. Specific tests performed included: - Cytotoxicity - Sensitization - Intercutaneous reactivity - Acute systemic toxicity - Pyrogenicity - Muscle implantation - Subchronic toxicity - Hemolysis
Product Characterization & Bench TestingEvaluation of general physical properties, including tensile strength and retention strength (pull-out), to demonstrate substantial equivalence to the predicate device.The RMST Staple underwent in vitro product characterization studies and bench testing. These included assessments of: - Tensile strength - Retention strength (pull-out) - General physical properties (e.g., size, surface area, weight, and strength retention) The results demonstrated that the RMST Staple is substantially equivalent to the predicate device, Ethicon SecurestrapTM 5mm Absorbable Strap Fixation Device (K093845).
Animal Efficacy StudyNot explicitly stated as acceptance criteria, but performed to demonstrate safety and equivalence.An animal efficacy study was performed. The results from this study, along with the in vitro and in vivo biocompatibility studies and bench testing, demonstrate that the RMST Staple is safe and substantially equivalent to the predicate devices.

2. Sample size used for the test set and the data provenance:

  • The document does not explicitly state specific sample sizes for each individual test (e.g., cytotoxicity, tensile strength). The phrasing "a number of in vitro and in vivo tests" suggests a sufficient number were conducted to meet regulatory requirements.
  • The data provenance is not explicitly stated in terms of country of origin.
  • The studies were likely prospective in nature, as they were conducted specifically for the purpose of demonstrating safety and efficacy for regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided in the document. The studies described are primarily laboratory and animal-based, not involving human expert interpretation for "ground truth" in the way a diagnostic AI might.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. The studies described are pre-clinical (in vitro, in vivo animal) and bench testing, not reliant on human adjudication of a specific output like in a clinical diagnostic study.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This device is a surgical staple, not a diagnostic imaging device or an AI-assisted diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The device is a physical surgical staple, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the biocompatibility studies, the "ground truth" would be established by standard scientific and regulatory methods for assessing biological response (e.g., cellular observation for cytotoxicity, macroscopic/microscopic tissue evaluation for implantation studies).
  • For the product characterization and bench testing, the "ground truth" would be objective measurements and engineering standards (e.g., force required for tensile strength, displacement for pull-out).
  • For the animal efficacy study, the "ground truth" would be pre-defined clinical and histological endpoints observed in the animal models.

8. The sample size for the training set:

  • Not applicable. This device is a physical product, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • Not applicable, as no training set for an AI model was involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for "Rotation Medical". The logo consists of a semi-circle of black dots above the word "rotation" in a stylized font. Below "rotation" is the word "MEDICAL" in a smaller, sans-serif font. There is a line above the word "MEDICAL".

K131637
page 1/2

510(k) Summary

JUL 1 2 2013

Applicant Information

Applicant Name: Applicant Address:

Telephone: Fax: Contact Person: Rotation Medical, Inc. 15350 25th Avenue North, Suite 100 Plymouth, MN 55447 763-746-7521 763-746-7501 Gail Schroeder Director, Quality Assurance and Operations May 31, 2013

Date Prepared:

Name of Device

Device Common Name: Device Trade Name: Device Classification Name: Absorbable Soft Tissue Staple Rotation Medical Soft Tissue Staple (RMST Staple) Staple, Implantable 878.4750 Class II GDW

Legally Marketed Devices to Which Substantial Equivalence is Claimed

Predicate Device(s):

Ethicon SecurestrapTM 5mm Absorbable Strap Fixation Device, K093845; Ethicon, Inc.

Description of the Device

RMST Staple is an absorbable polymer strap with cleat tips. RMST Staple is composed of absorbable synthetic polyester derived from lactic acid and dyed with D&C Violet #2. The RMST Staple is used in conjunction with an orthopedic manual staple driver from Rotation Medical. (Note: Rotation Medical's orthopedic manual staple driver is a Class I exempt device pursuant to 21 CFR 888.4540 and is not the subject of this submission).

15350 25th Avenue No · Suite 100 · Plymouth MN 55447 · 763.746.7500

RMST Staple 510(k) Application, page 14

{1}------------------------------------------------

K131637
page 2/2

Image /page/1/Picture/1 description: The image shows the logo for Rotation Medical. The logo consists of a semi-circle of dots above the word "rotation" in a bold, sans-serif font. Below "rotation" is the word "MEDICAL" in a smaller, sans-serif font.

Intended Use

The RMST Staple is intended for fixation of prosthetic material to soft tissues in various minimally rne Kins 1 Stapie is international such as the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

Summary/Comparison of Technical Characteristics

RMST Staple and the predicate device have similar indications for use, technological characteristics, and material composition. Both devices are comprised of a polymer strap with barbed ends which engage in soft tissue. RMST Staple and the currently marketed Ethicon Securestrap (predicate device) are both indicated for the fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures. Both devices are composed of similar absorbable synthetic polyester and contain the same D&C Violet #2 dye.

RMST Staple has been evaluated in a number of in vitro and in vivo tests to assess its safety/biocompatibility, including: tensile strength, retention strength (pull-out), cytotoxicity, sensitization, intercutaneous reactivity, acute systemix toxicity, pyrogenicity, muscle implantation, subchronic toxicity, and hemolysis. RMST Staple and its predicate have been implantedially for general physical properties including size, surface area, weight, and strength retention to demonstrate substantial equivalence.

Conclusion

The device passed all applicable FDA Blue Book Memorandum G95-1 and ISO 10993-1 testing for the biological evaluation of medical devices. The results of the in vitro product characterization studies, bench testing and in vitro and in vivo biocompatibility studies, as well as the animal efficacy study demonstrate that the RMST Staple is safe and substantially equivalent to the predicate devices.

15350 25th Avenue No ・ Suite 100 ・ Plymouth MN 55447 ・ 763.746.7500

RMST Staple 510(k) Application, page 15

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble human figures or flowing lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Rotation Medical, Inc. % Ms. Gail Schroeder Director, Quality Assurance and Operations 15350 25th Avenue North, Suite 100 Plymouth, Minnesota 55447

July 12, 2013

Re: K131637

Trade/Device Name: Rotation Medical Soft Tissue Staple (RMST Staple) Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: May 31, 2013 Received: June 04, 2013

Dear Ms. Schroeder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Gail Schroeder

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Confidential Rotation Medical, Inc.

Indications for Use

K131637 510(k) Number (if known): _

Rotation Medical Soft Tissue Staple (RMST Staple) Device Name: _

Indications for Use:

The RMST Staple is intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not WRITE BELOW THIS LINE - Continue on ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krause -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K131637

Page 1 of _1

RMST Staple 510(k) Application, page 13

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.