K Number
K131583

Validate with FDA (Live)

Date Cleared
2013-10-03

(125 days)

Product Code
Regulation Number
876.5630
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the ThermoChem™ HT-2000 System is to raise the core temperature of the peritoneum to a desired target temperature by continuously lavaging the peritoneum with circulating and warmed Lactated Ringer's Solution, U.S.P., or another physiologically compatible sterile solution.

Device Description

The ThermoChem HT-2000 System is a next-generation design of the currently marketed ThermoChem HT-1000 System and consists of the ThermoChem unit, a touchscreen monitor and an Intraperitoneal Hyperthermia (IPH) Procedure Kit, or "Disposable." The ThermoChem unit houses an internal heating unit that warms distilled or sterile water to a desired temperature. This water is circulated around a single-use heat exchanger (included in the IPH Procedure Kit) where it warms Lactated Ringer's (or another physiologically compatible sterile solution) for infusion into the patient's peritoneal cavity. The IPH Procedure Kit contains two temperature probes, two suction guards, and a pressure sensor as well as a Perfusion Kit consisting of a heat exchanger, a fluid reservoir, and tubing set. The warmed Lactated Ringer's (or other physiologically compatible sterile solution) is pumped through the tubing to the patient and then returns to the fluid reservoir. The IPH Procedure Kit is a convenience kit of previously 510(k)-cleared devices that are repackaged / relabeled by ThermaSolutions, Inc. The ThermoChem HT-2000 unit has been redesigned to physically reduce the size of the unit and to allow use of components that are more commercially available. The software and touchscreen monitor user interface have been redesigned to make the device easier to use. The redesigned ThermoChem unit will continue to use the currently-marketed IPH Procedure Kit for use with the ThermoChem HT-1000 System.

AI/ML Overview

The ThermoChem™ HT-2000 System underwent formal verification testing to demonstrate its performance and substantial equivalence to its predicate device, the ThermoChem HT-1000 System.

Here's an analysis of the acceptance criteria and the study conducted:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes categories of performance testing that were conducted and that the device "met acceptance criteria." However, it does not explicitly state the specific numerical acceptance criteria for each test, nor does it provide the detailed quantitative results ("reported device performance") beyond the general statement that criteria were met.

Acceptance Criteria CategoryReported Device Performance
Bench testing of pump settings (pressure)Met acceptance criteria
Bench testing of pump settings (flow rate)Met acceptance criteria
Water bath heat timeMet acceptance criteria
Water bath temperaturesMet acceptance criteria
Operating conditionsMet acceptance criteria
Fluid ingressMet acceptance criteria
Overfill testsMet acceptance criteria
Spill testsMet acceptance criteria
Software validationMet acceptance criteria
Regression testing (software)Met acceptance criteria
Usability validation testingMet acceptance criteria

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for any of the performance tests. The data provenance is described as "bench testing," meaning the tests were conducted in a laboratory setting, not with human patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable. The performance testing described (bench testing, software validation, usability validation) does not involve establishing ground truth from expert consensus on clinical data. The tests focus on the device's functional and operational performance.

4. Adjudication Method for the Test Set

Not applicable, as the tests described are technical performance evaluations rather than clinical assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

No, an MRMC comparative effectiveness study was not done. The submission is for a Traditional 510(k) for a medical device (a heat exchanger system), not an AI/imaging diagnostic device that typically undergoes MRMC studies. The focus is on demonstrating that the new device (HT-2000) is substantially equivalent to a previously cleared device (HT-1000) based on functional and design changes.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is a medical system with hardware and software, not a standalone algorithm in the typical sense of AI/diagnostic software. The performance testing focuses on the integrated system's functionality. Therefore, a "standalone (algorithm only)" performance study as defined for AI models is not applicable in this context.

7. The Type of Ground Truth Used

For the functional and performance tests, the "ground truth" would be the engineering specifications and design requirements of the device. For example, for pump settings, the ground truth would be the target pressure and flow rate values as designed. For water bath temperatures, the ground truth would be the specified temperature range. No expert consensus, pathology, or outcomes data were used as ground truth for these tests.

8. The Sample Size for the Training Set

Not applicable. The ThermoChem™ HT-2000 System is a physical medical device with integrated software, not an AI model that undergoes machine learning training with a "training set" of data in the common understanding of that term. While software development involves testing and iteration, the concept of a "training set" for an AI algorithm does not apply here.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of an AI model for this device.

{0}------------------------------------------------

ThermoChem™ HT-2000 System

Traditional 510(k) Summary

Submitted by:ThermaSolutions, Inc.
1889 Buerkle Road
St. Paul, MN 55110
Contact Person:Steve Stonelake
Phone:651-209-3900
Fax:651 209 3903

OCT 03 2013

  • 29 September 2013 Date of Summary:
    ThermoChem™ HT-2000 System and Device Trade Name: Intraperitoneal Hyperthermia (IPH) Procedure Kit Common or Usual Warmer, Peritoneal Dialysate Name:

Classification: 876.5630

Product Code: MLW

Predicate Device(s) ThermoChem HT-1000 System, K993330

{1}------------------------------------------------

K131583

pg 2 of 3

Device Description: The ThermoChem HT-2000 System is a next-generation design of the currently marketed ThermoChem HT-1000 System and consists of the ThermoChem unit, a touchscreen monitor and an Intraperitoneal Hyperthermia (IPH) Procedure Kit, or "Disposable."

The ThermoChem unit houses an internal heating unit that warms distilled or sterile water to a desired temperature. This water is circulated around a single-use heat exchanger (included in the IPH Procedure Kit) where it warms Lactated Ringer's (or another physiologically compatible sterile solution) for infusion into the patient's peritoneal cavity.

The IPH Procedure Kit contains two temperature probes, two suction guards, and a pressure sensor as well as a Perfusion Kit consisting of a heat exchanger, a fluid reservoir, and tubing set. The warmed Lactated Ringer's (or other physiologically compatible sterile solution) is pumped through the tubing to the patient and then returns to the fluid reservoir. The IPH Procedure Kit is a convenience kit of previously 510(k)-cleared devices that are repackaged / relabeled by ThermaSolutions, Inc.

The ThermoChem HT-2000 unit has been redesigned to physically reduce the size of the unit and to allow use of components that are more commercially available. The software and touchscreen monitor user interface have been redesigned to make the device easier to use. The redesigned ThermoChem unit will continue to use the currently-marketed IPH Procedure Kit for use with the ThermoChem HT-1000 System.

  • The intended use of the ThermoChem™ HT-2000 System is to raise Intended Use: the core temperature of the peritoneum to a desired target temperature by continuously lavaging the peritoneum with circulating and warmed Lactated Ringer's Solution, U.S.P., or another physiologically compatible sterile solution.
    Technological The ThermoChem HT-2000 is a next-generation design of the Characteristics: currently-marketed ThermoChem HT-1000 system intended to integrate more-commercially-available components and incorporate changes to make the device easier to use, including a more userfriendly graphical user interface. In addition, the software was upgraded to incorporate currently-supported operating systems (QNX) and drivers. There was no change to the essential mode of operation of the device.

{2}------------------------------------------------

K131583

Performance Testing The ThermoChem HT-2000 System underwent and met acceptance criteria of formal verification testing including bench testing of pump settings (pressure, flow rate); water bath heat time and temperatures; operating conditions; fluid ingress, overfill, and spill tests; software validation and regression testing; and usability validation testing.

Substantial Based on device specifications and test results, the ThermoChem Equivalence: HT-2000 can be considered to be substantially equivalent to its predicate device, the ThermoChem HT-1000 system. Both devices share the same intended use and incorporate the same technological characteristics. Both devices are intended to be used with the ThermoChem IPH Procedure Kit. The ThermoChem HT-2000 is a next-generation design of the currently-cleared ThermoChem HT-1000 system intended to integrate morecommercially-available components and incorporate changes to make the device easier to use, including a more user-friendly graphical user interface.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2013

ThermaSolutions, Inc. % M.W. (Andy) Anderson, Ph.D., RAC Senior Principal Advisor Regulatory and Clinical Research Institute, Inc. (RCRI) 5353 Wayzata Boulevard, Suite 505 Minneapolis, MN 55416

Re: K131283

Trade/Device Name: ThermoChem™ HT-2000 System and Intraperitoneal Hyperthermia (IPH) Procedure Kit Regulation Number: 21 CFR$ 876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: MLW Dated: August 21. 2013 Received: August 22. 2013

Dear M.W. (Andy) Anderson,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - M.W. (Andy) Anderson, Ph.D., RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for Beniamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

ThermoChem™ HT-2000 System Traditional 510(k) Premarket Notification

Indications for Use Statement

510(k) Number:K131583
Device Name:ThermoChem™ HT-2000 System andIntraperitoneal Hyperthermia (IPH) Procedure Kit
Intended Use:The intended use of the ThermoChem™ HT-2000 System is to raisethe core temperature of the peritoneum to a desired targettemperature by continuously lavaging the peritoneum with circulatingand warmed Lactated Ringer's Solution, U.S.P., or anotherphysiologically compatible sterile solution.
Prescription Use:YES AND/OR Over-the-Counter Use: NO
(Part 21 CFR 801 Subpart D)(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner - S

2013.10.03 11:21:24 -04'00'

§ 876.5630 Peritoneal dialysis system and accessories.

(a)
Identification. (1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments.
(3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles.
(4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
(b)
Classification. Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.