K Number
K131478
Date Cleared
2013-08-15

(85 days)

Product Code
Regulation Number
870.1650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Osprey Medical AVERT Contrast Modulation System is intended to be used for the controlled infusion of radiopaque Iodixanol 270 mg/ml contrast media for angiographic procedures.

Device Description

The Osprey Medical AVERT Contrast Modulation System (AVERT System) consists of a Linear Slide Apparatus, Pressure Dampening Plates, and a Wheeled Stand. The nonsterile, reusable stainless steel dampening plates are attached to a non-sterile, reusable aluminum linear slide, positioned above a 6 cc syringe. The plates apply a constant force (using gravity) on the 6 cc syringe plunger acting as pressure dampener. The number of plates can be changed depending on the injection requirements. The Linear Slide Apparatus is attached to the Wheeled Stand, near the patient outside of the sterile field. The AVERT System is used in conjunction with standard disposable, sterile, off-the-shelf items including a 6 cc syringe, a 4-way stopcock, and an extension line that are provided in a Convenience Kit with the system (CK-100).

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Osprey Medical AVERT Contrast Modulation System, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific Criteria/Tests ConductedReported Device Performance
Performance CharacteristicsFlow RateMet established performance criteria
Peak Pressure ReductionMet established performance criteria
Flow Rate AdjustabilityMet established performance criteria
Mechanical Cycle TestingMet established performance criteria
Contrast ReductionMet established performance criteria

Note: The document states that "All test results demonstrated the materials, manufacturing processes, and design of the Osprey Medical AVERT System met the established performance criteria, will perform as intended and in a manner that is substantially equivalent to the predicated devices cited."

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated. The document refers to "in vitro bench testing" and "all test results."
    • Data Provenance: In vitro bench testing. The country of origin is not specified but is implicitly the location of Osprey Medical (Minnetonka, MN, USA). The study was prospective as it involved newly conducted bench tests.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable. This device underwent in vitro bench testing, not clinical studies requiring expert ground truth for patient data.
  3. Adjudication Method for the Test Set:

    • Not applicable for in vitro bench testing. Performance criteria were likely measured against predefined engineering specifications.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. An MRMC study was not conducted as this was in vitro (bench) testing, not a clinical study involving human readers or cases.
  5. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

    • Yes, in spirit. The performance tests described (Flow Rate, Peak Pressure Reduction, etc.) assess the device's inherent mechanical and fluid dynamics performance in a controlled environment, without direct human intervention in the 'reading' or interpretation of results beyond setting up and running the tests. It's not an "algorithm" in the typical software sense, but rather the standalone function of the hardware.
  6. Type of Ground Truth Used:

    • Pre-established engineering specifications and performance criteria. The bench tests measured objective physical parameters (flow rate, pressure, etc.) against known or desired values for safe and effective operation, and for comparison to predicate devices.
  7. Sample Size for the Training Set:

    • Not applicable. This is a mechanical device, not an AI/machine learning algorithm that requires a "training set" of data in the conventional sense. The "training" for the device's design and manufacturing would come from engineering principles, predicate device analysis, and iterative design/testing.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable for a mechanical device. The "ground truth" for its design and manufacturing would be based on established engineering standards, regulatory requirements, and the functional characteristics of the predicate devices it aims to be substantially equivalent to.

{0}------------------------------------------------

AUG 1 5 2013

8. 510(K) SUMMARY

. も

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: _______________________________________________________________________________________________________________________________________________________________ · · ·

Applicant Information:

Date Prepared:July 12, 2013
Name:Osprey Medical
Address:5600 Rowland Road, Suite 250Minnetonka, MN 55343Phone: 651-270-0572Fax: 952-955-8171
Contact Person:Jill Munsinger
Phone Number:651-270-0572
E-mail:jmunsinger@visi.com

Device Information:

Trade NameCommon NameClassification NameClass
AVERT Contrast ModulationSystemAngiographic InjectorAngiographic InjectorClass II

Predicate Devices:

The Osprey Medical AVERT Contrast Modulation System is substantially equivalent in intended use and/or method of operation and technical aspects to the following predicate devices:

DeviceReference 510(k) NumberIndication for Use
Acist Angiographic Injection SystemK993774, K991103,K000013, and K052744...intended to be used forthe controlled infusion ofradiopaque contrast mediafor angiographic procedures
Medline Angiographic Control SyringeK093830An angiographic syringe is adevice that consists of asyringe which is used toinject contrast material intothe heart, great vessels, andcoronary arteries duringangiographic or CTprocedures.

{1}------------------------------------------------

Device Description:

The Osprev Medical AVERT Contrast Modulation System (AVERT System) consists of a Linear Slide Apparatus, Pressure Dampening Plates, and a Wheeled Stand. The nonsterile, reusable stainless steel dampening plates are attached to a non-sterile, reusable aluminum linear slide, positioned above a 6 cc syringe. The plates apply a constant force (using gravity) on the 6 cc syringe plunger acting as pressure dampener. The number of plates can be changed depending on the injection requirements. The Linear Slide Apparatus is attached to the Wheeled Stand, near the patient outside of the sterile field. The AVERT System is used in conjunction with standard disposable, sterile, off-the-shelf items including a 6 cc syringe, a 4-way stopcock, and an extension line that are provided in a Convenience Kit with the system (CK-100).

Intended Use:

The Osprey Medical AVERT Contrast Modulation System is intended to be used for the controlled infusion of radiopaque Iodixanol 270 mg/ml contrast media for angiographic procedures.

Comparison to Predicate Device(s):

The design of the AVERT System is comparable to the Acist Angiographic Injection System and the Medline Angiographic Control Syringe. The indication for use statement of the Acist Angiographic Injection System and the AVERT System are similar with the A VERT System having a more specific indication of contrast media type. The indication for use statement of the Medline Angiographic Control Syringe is similar in intention to the AVERT System. The Medline indication relies on the user's ability to inject fluid to control the infusion. While this is different from the AVERT System, it is not considered a critical different in that they are both designed to inject materials into blood vessels/arteries.

The AVERT System is manually adjusted via addition/subtraction of plates to divert injection pressure and control infusion. Whereas the Acist system is a powered injector adjusted via the user interface and the Medlin syringe is manually controlled.

Performance Data:

The Osprey Medical AVERT System has been evaluated using the following in vitro bench testing to confirm the performance characteristics:

  • . Flow Rate

  • Peak Pressure Reduction .

  • · Flow Rate Adjustability

  • . Mechanical Cycle Testing

  • Contrast Reduction .

{2}------------------------------------------------

All test results demonstrated the materials, manufacturing processes, and design of the Osprey Medical AVERT System met the established performance criteria, will perform as intended and in a manner that is substantially equivalent to the predicated devices cited.

Summary:

Based upon the intended use and descriptive information provided in this pre-market notification, the Osprey Medical AVERT System has been shown to be substantially equivalent to the currently marketed predicate devices.

{3}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 15, 2013

Osprey Medical C/O Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K131478

Trade/Device Name: AVERT Contrast Modulation System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class; II Product Code: DXT Dated: June 26, 2013 Received: June 27, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Mark Job

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuglierman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

7. INDICATIONS FOR USE STATEMENT

510(k) Number: K131478

Device Name: AVERT Contrast Modulation System

Indications For Use:

The Osprey Medical AVERT Contrast Modulation System is intended to be used for the controlled infusion of radiopaque lodixanol 270 mg/ml contrast media for angiographic procedures.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

2013

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.