K Number
K131408

Validate with FDA (Live)

Date Cleared
2013-08-29

(106 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.

Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.

Device Description

Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. Principles of operation and technological characteristics of the new and predicate devices are largely the same as other computed radiography systems:

Phosphor coated imaging plates and cassettes for image capture. Laser digitizer for generating the electronic image. NX workstation for image previewing, processing and routing.

AI/ML Overview

ACCEPTANCE CRITERIA AND STUDY ANALYSIS FOR AGFA CR 12-X DIGITIZER (K131408)

The Agfa CR 12-X Digitizer received 510(k) clearance based on its substantial equivalence to predicate devices (Agfa CR 10-X Digitizer (K121948) and 3Disc America's Fire CR+ Digitizer (K102619)). The primary method for proving substantial equivalence was through technical comparisons and phantom image quality tests, as clinical images were deemed unnecessary given the modifications from the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative sense for performance metrics like sensitivity or specificity. Instead, the approach is based on demonstrating that the CR 12-X Digitizer performs equivalently to the predicate devices, particularly the CR 10-X, regarding image quality and functional specifications. The "acceptance" is implied by demonstrating substantial equivalence, meaning the device meets the same safety and effectiveness standards as its predicate.

Feature/MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (CR 12-X Digitizer)
Image QualityEquivalent to CR 10-X Digitizer (K121948)Image quality tests comparing CR 12-X to CR 10-X predicate using phantom images showed equivalence.
Scanning Resolution100µm (CR 10-X), 100µm & 200µm (FireCR+)100µm, 150µm, 200µm selectable resolutions
Throughput (35x43 cm plates/hr)34 @ 100µm (CR 10-X); 20 @ 200µm, 70 @ 100µm (FireCR+)60 @ 200µm; 72 @ 150µm; 90 @ 100µm
Image Plate PhosphorsBaSrFBrI:EuSame as both predicates (BaSrFBrI:Eu)
Scanning TechnologyPoint at-a-timeSame as both predicates (Point at-a-time)
Dynamic Range (acquisition)16 bit, sq. root compressedSame as both predicates (16 bit, sq. root compressed)
Image ProcessingMUSICA, MUSICA2 (CR 10-X)Same as CR 10-X (MUSICA, MUSICA2)
Safety & EffectivenessConformance to product safety, radiology, and imaging standards (e.g., IEC 60601-1, IEC 60601-1-2)Device completed verification and validation to confirm it meets specifications and operates as planned; risk assessment concluded medical risk is not higher than previous released systems.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document states that phantom images were used for evaluation. However, the specific number of phantom images or the sample size of the test set is not explicitly provided.
  • Data Provenance: The data used for testing was generated internally by Agfa as part of their verification and validation process. The type of data (phantom images) means the provenance is synthetic/controlled rather than real patient data from a specific geographical region. All testing was presumably conducted in a controlled lab environment. The study is prospective in the sense that the tests were designed and executed to evaluate the new device, but the "data" itself (phantom images) is not from a retrospective or prospective clinical cohort.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: The document states that "internal and experts" were involved in comparing the CR 12-X to the CR 10-X predicate for image quality tests. The exact number of experts is not specified.
  • Qualifications of Experts: The specific qualifications (e.g., years of experience, subspecialty) of these "experts" are not detailed in the provided text.

4. Adjudication Method for the Test Set:

  • The document implies a comparison was made by "internal and experts." However, the adjudication method (e.g., majority vote, consensus, blinded review, 2+1, 3+1) is not described.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

  • No Multi Reader Multi Case (MRMC) comparative effectiveness study was mentioned or performed. The evaluation focused on substantial equivalence through technical comparisons and phantom image quality tests. No assessment of human reader improvement with or without AI assistance was conducted or reported, as this is neither an AI device nor an MRMC study.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:

  • This device is a Computed Radiography system (a digitizer), not an algorithm or AI solution. Therefore, the concept of "standalone (algorithm only)" is not applicable to this submission. The device performs image acquisition and initial processing before human review.

7. The Type of Ground Truth Used:

  • For the image quality tests, the ground truth was based on the image characteristics of the predicate device (CR 10-X Digitizer) using phantom images. The goal was to show that the CR 12-X produced images of equivalent diagnostic quality. Since phantom images were used, the "ground truth" refers to the expected image characteristics as defined by the predicate and established imaging standards.

8. The Sample Size for the Training Set:

  • This submission does not discuss a "training set" in the context of machine learning or AI. The device is a CR digitizer, and its performance is based on hardware and firmware design, not a trained algorithm in the AI sense. Therefore, the concept of a training set is not applicable.

9. How the Ground Truth for the Training Set Was Established:

  • As the concept of a "training set" is not applicable, the establishment of its ground truth is also not applicable.

{0}------------------------------------------------

K 131408
Page 1 of 5

Computed Radiography (CR) Systems with CR 12-X Digitizer

510(K) SUMMARY Agfa Computed Radiography (CR) Systems with CR 12-X Digitizer

Common Name: Computed Radiography System Classification Name: Solid State X-Ray System Regulatory Classification: 21 CFR 892.1680 Product Code: MOB Proprietary Name: Computed Radiography (CR) Systems with CR 12-X Digitizer Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Koen Vervoort. Prepared: May 15, 2013 Telephone: +32-34444-7368

AUG 2 9 2013

LEGALLY MARKETED PREDICATE DEVICES A.

This is a 510(k) for Agfa's Computed Radiography (CR) System with CR 12-X Digitizer, a solid state x-ray imaging device. It is substantially equivalent to systems with Agfa's CR 10-X Digitizer (K121948) and 3Disc America's Fire CR+ Digitizer (K102619).

DEVICE DESCRIPTION B.

Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. Principles of operation and technological characteristics of the new and predicate devices are largely the same as other computed radiography systems:

Phosphor coated imaging plates and cassettes for image capture. Laser digitizer for generating the electronic image. NX workstation for image previewing, processing and routing.

C. INTENDED USE

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.

Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.

The intended use has not changed as a result of any labeling modifications.

{1}------------------------------------------------

D. SUBSTANTIAL EQUIVALENCE SUMMARY

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement identical to the statement for the predicate devices, K121948 and K102619. Intended uses are the same. The devices have the same technological characteristics. Descriptive characteristics and performance data are adequate to ensure equivalence.

Differences in devices do not alter the intended therapeutic/diagnostic effect.

PRODUCT COMPARISON TABLE
Systems with CR 12-XDigitizer(New Device)Systems with CR 10-X Digitizer(PREDICATE-K121948)3Disc: FireCR+ Digitizer(PREDICATE-K102619)
CommunicationsSame as both predicatesDICOMDICOM
Cassette and Image Plates
Cassettes and ImagePlate SizesSame as CR 10-X35x43 cm35x43 cm &24x30 cm
Image PlatePhosphorsSame as both predicatesBaSrFBrl:EuBaSrFBrl: Eu
Digitizer
Scanning technologySame as both predicatesPoint at-a-timePoint at-a-time
Light collectionSame as CR 10-XSolid, high efficiencyUnavailable
Scanning resolution100µm, 150µm, 200µm100µm100µm and 200µm
Active matrixSame as CR 10-X3420 x 4218Unavailable
Dynamic range(acquisition)Same as both predicates16 bit, sq. root compressed16 bit, sq. root compressed
Throughput(35x43 cm plates/hr.)60 @ 200µm72 @ 150µm90 @ 100µm34 @ 100µm20 @ 200µm70 @ 100µm
Power SupplySame as CR 10-X100-240V 50/60 Hz110-240V 50/60 Hz
Mobile installationsSame as CR 10-XMobile mounting optionSolid Transport case
Image stitchingSame as CR 10-XFull Leg Full SpineN/A
Workstation
Image processingSame as CR 10-XMUSICA, MUSICA2QuantorMed+ ImagingSoftware
Dynamic range(display)Same as CR 10-X12 bit32 bit & 64 bit
Operating systemSame as both predicatesWindows 7Windows 7
Power SupplySame as CR 10-X100 - 240V 50/60 Hz110-240V 50/60 Hz
Display SystemSame as CR 10-XStandard PC display or separatelycleared medical display1,280x900 pixel, DVIinterface

{2}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS E.

Agfa's Computed Radiography (CR) Systems with CR 12-X Digitizer is a solid state x-ray imaging device. It uses phosphor coated imaging plates to capture an x-ray exposure. Upon scanning of the image plates by the CR-12-X digitizer, the x-ray image is read and transmitted to the NX workstation where the image can be previewed, processed and transmitted to other devices for viewing, printing or storage.

The CR 12-X digitizer is identical to the CR 10-X with the exception of the following:

60 plates/hr with scanning resolution of 200um

72 plates/hr with added scanning resolution of 150um

Increase to 90 plates/hr with scanning resolution of 100um

Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options. The remaining differences would not be expected to impact safety and effectiveness.

F. TESTING

The device has completed verification and validation testing to confirm it meets specifications and operates as planned. Tests included image quality tests with internal and experts comparing the CR 12-X to the CR 10-X predicate. Clinical Images were unnecessary to evaluate modifications from the predicate device; therefore, phantom images were used to evaluate substantial equivalence with the CR 10-X predicate device (K121948).

The change in the design of the CR 10-X to include the image processing was verified and validated per the following:

Exhibit 3-2, Software Design Input Output Verification (DIOV) Report - Arcus Digitizer ..

Exhibit 5-1, Verification Plan - Consolidated Requirement & Testing (incl. SW), Arcus Solution.

Exhibit 5-2, Verification Report - Consolidated Requirement & Testing (incl. SW), Arcus Solution.

Exhibit 5-3, Validation Test Plan - Arcus Solution

Exhibit 5-4, Design Validation Report - Arcus Solution

Exhibit 5-5, Validation Test Report - Arcus Solution

Exhibit 5-6, Final Design Review for the CR 12-X Digitizer.

Exhibit 5-7, Design Complete Acceptance for the CR 12-X Digitizer

Exhibit 5-8, CRB (Change Review Board) Arcus Software Version ARC 1404 (This is the maintenance release for digitizer software version ARC 1404).

{3}------------------------------------------------

The product, manufacturing and development processes have been shown to conform to product safety, radiology, and imaging standards including:

PRODUCT STANDARDS

IEC 60601-1: 2005 Medical Electrical Equipment - Part 1: General Requirements for Safety

IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests

ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM)

MANAGEMENT STANDARDS

ISO 14971:2007 Application of Risk Management to Medical Devices

ISO 13485:2003 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes

G. RISK ASSESSMENT AND MANAGEMENT SUMMARY

During the final risk analysis meeting, the risk management team concluded that the medical risk is not higher than previous released system to the field.

After review of these risks with the risk analysis team. it was decided that this risk could not be further mitigated and that for this risk, the benefits of the system outweigh the risks.

There are a total of 22 risks in the broadly acceptable region and 2 riskd in the ALARP region. Zero risks were identified in the not acceptable region.

The residual risk of the device is acceptable.

  • · the risk control measures are effective
  • · usability acceptance criteria for safety-related functions have been met (IEC 62366)
  • · no residual risk is situated in the intolerable region
  • · the residual risks are controlled as far as reasonably practicable

Of the 72 identified medical risks, 7 are in the ALARP region. The residual risks cannot be reduced any further with reasonable effort. Therefore is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.

The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the device has been determined to be moderate.

Refer to Section 16 Software, for more information regarding risk management.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a handwritten document with the text "K131408 Page 5 of 5". The text appears to be a document identifier and page number. The handwriting is legible, but the style is casual and not formal.

H. CONCLUSIONS

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer has an Indications For Use statement is identical to the statement for the predicate devices. K121948 and K 102619. Intended uses are the same. The devices have the same technological characteristics in that they use the same imaging phosphor and the same laser diode for stimulation light operating with the same spot size as both predicates. Descriptive characteristics and performance data are adequate to ensure equivalence.

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is substantially equivalent to its CR 10-X System (K121948) in that it uses the same technology to capture and transmit images, use the NX user workstation, use the same MD1.0 image plate and cassette with manual cassette insertion, includes the full leg full spine function, provides a mobile mounting option, and uses 100-240V 50/60 Hz power supply.

Agfa CR 12-X and 3Disc America's FireCR+(K102619) both enable selection of resolution options from 100um to 200um.

The CR 12-X is equivalent to the predicate devices in regards to the plate speed.

This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 29, 2013

AGFA HealthCare N.V. % Ms. ShaeAnn Cavanagh Regulatory Affairs Specialist NA AGFA HealthCare 10 South Academy Street GREENVILLE SC 29601

Re: K131408

Trade/Device Name: CR 12-X Digitizer Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 1, 2013 Received: August 7, 2013

Dear Ms. Cavanagh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2-Ms. Cavanagh

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/RcportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K131408

Device Name: CR 12-X Digitizer

Indications For Use:

Agfa's Computed Radiography (CR) System with CR 12-X Digitizer is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The system may be used wherever conventional screen-film systems are used.

Agfa's Computed Radiography (CR) System With CR 12-X Digitizer is not indicated for use in mammography.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Smh7)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.