K Number
K131388

Validate with FDA (Live)

Date Cleared
2013-09-26

(135 days)

Product Code
Regulation Number
868.5450
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.

Device Description

The Transcend Heated Humidifier is a humidifier that is designed to humidify the air delivered to the airway during positive airway pressure (PAP) therapy with a compatible Transcend PAP device.

AI/ML Overview

The provided document describes the acceptance criteria and supporting studies for the Transcend Heated Humidifier, a medical device for humidifying air during positive airway pressure (PAP) therapy.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Transcend Heated Humidifier are primarily based on compliance with recognized consensus standards and direct comparison to a predicate device (ResMed HumidAire 2i™). The document presents a comparison of various characteristics, some of which serve as performance criteria.

Acceptance CriterionSpecifics/StandardReported Device Performance (Transcend Heated Humidifier)Comparison to Predicate / Compliance
Standards Compliance
Electrical SafetyIEC 60601-1CompliantCompliant
Electromagnetic CompatibilityIEC 60601-1-2Compliant with IEC 60601-1-2Identical (Predicate also compliant)
UsabilityIEC 60601-1-6Not explicitly stated, but implied compliance through substantial equivalenceImplied compliance
Home Health Care EnvironmentIEC 60601-1-11Not explicitly stated, but implied compliance through substantial equivalenceImplied compliance
Respiratory HumidifiersISO 8185:2007Compliant (Stated for multiple parameters)Confirmed compliance, predicate's compliance unknown
Biological EvaluationISO 10993-1Biocompatible: YesIdentical
Risk ManagementISO 14971Implied compliance through risk assessmentsImplied compliance
Performance Characteristics (compared to predicate or ISO 8185)
Indications for UseHumidification of air for compatible Transcend PAP device, single patient re-use in home/hospitalThe Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment.Similar (Transcend is single-patient vs. predicate's multi-patient option).
Intended Population of UseAdultAdultIdentical
Compatible DevicesTranscend PAP device and associated accessoriesTranscend PAP device and associated accessoriesSimilar (both are manufacturer-specific)
DisinfectionNone (single patient re-use only)Differs from predicate (which should be disinfected between patients)
CleaningDistilled water using a mild detergentDistilled water using a mild detergentIdentical
SterilizationNot sterilizedNot sterilizedIdentical
Pressure RegulationDetermined by compatible PAP deviceDetermined by compatible PAP deviceIdentical
Working Pressure Range4-20 cm H₂O4-20 cm H₂OIdentical
Humidifier Settings1-5Similar (Predicate has 1-6 settings)
Inspiratory / Expiratory Pressure DropMeets ISO 8185 requirements0.375 cmH₂O at 50 L/minMeets ISO 8185
Gas Leak at Max Operating PressureMeets ISO 8185 requirements4.85 L/minMeets ISO 8185
Sound Power Level at 10 cm H₂O static pressureMeets ISO 8185 requirements37 dBMeets ISO 8185
Maximum Heater Plate TemperatureMeets ISO 8185 requirements131 °FMeets ISO 8185 (Predicate: 167°F)
Maximum Output Humidity>95%>95%Similar (Predicate: 95%)
Maximum Gas TemperatureMeets ISO 8185 requirements77 °FMeets ISO 8185
Water CapacityEnough water to last a minimum of 8 hours325 mlSimilar (Predicate: 400ml; both intended for 8+ hours)
Operating TemperatureBased on compatible PAP device+41°F to +95°FSimilar (Predicate: +41°F to +104°F)
Operating HumidityBased on compatible PAP device10 - 80% relative humidity, non-condensingSimilar (Predicate: 10-95% RH)
Shipping/Storage Temperature-4° F to +140° FIdentical
Gas Inlet Temperature RangeBased on compatible PAP device+41°F to +95°FSimilar (Predicate: +41°F to +104°F)
Shipping/Storage Humidity10 - 90% relative humidity, non-condensingSimilar (Predicate: 15-95% RH)
Power Supply100-240 VAC, 50/60 Hz100-240 VAC, 50/60 HzIdentical
IEC 60601-1 ClassificationType BF or CF (ISO 8185 section 6.8.2-14)Type BFSubject device is compliant

2. Sample Size Used for the Test Set and Data Provenance

The document does not detail specific sample sizes for particular performance tests. The studies conducted are described as "Device design was qualified through the following tests and assessments" which implies these were bench and lab-based tests, not human studies. Therefore, typical "sample sizes" (as in patient data) and "data provenance" (e.g., country of origin, retrospective/prospective) are not applicable in the context of this device's premarket submission, which focuses on engineering, safety, and performance characteristics rather than clinical diagnostic accuracy. The studies performed are:

  • Electrical Safety
  • Electromagnetic Compatibility
  • Biocompatibility Assessment
  • Cleaning Validation
  • Software Validation
  • Packaging and Shipping
  • Performance: physical (cycling), resistance to flow, spillage, flow sensing, sound power level, humidification output, out of water shutdown, temperature, runtime, warm-up duration, and reservoir gas leak

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable. The device relies on physical and electrical performance data, and compliance with engineering standards (e.g., ISO 8185, IEC 60601-1). Ground truth is established by the specifications defined in these standards and by direct laboratory measurements, not by expert interpretation of patient data.

4. Adjudication Method for the Test Set

This section is not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple experts evaluate ambiguous cases to reach a consensus for ground truth. This submission focuses on engineering and performance testing, where outcomes are determined by objective measurements against predefined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers (e.g., radiologists) interpret images or data, and their performance with and without AI assistance is compared. The Transcend Heated Humidifier is a therapeutic accessory, not a diagnostic device involving human interpretation of complex data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This section is not applicable in the typical sense of AI algorithms. The "device" itself operates "standalone" in its function as a humidifier, based on its internal programming and hardware. However, it's not an AI algorithm in the context of data analysis or diagnostic support where a "human-in-the-loop" would be relevant. The software validation mentioned ensures the device's internal algorithms and controls function correctly, which is a standalone performance test.

7. The Type of Ground Truth Used

The ground truth for the device's performance is based on:

  • Engineering Standards and Specifications: Metrics defined by standards such as ISO 8185 (e.g., maximum output humidity, pressure drop, gas leak, sound power level, temperatures).
  • Predicate Device Characteristics: Direct comparison of physical, functional, and safety attributes against a legally marketed predicate device.
  • Laboratory Measurements: Objective measurements of the device's performance against the specified criteria (e.g., humidification output, temperature, water capacity).

8. The Sample Size for the Training Set

This section is not applicable. This device is not an AI/machine learning model where a "training set" of data would be used to develop an algorithm. Its design and validation are based on traditional engineering principles and testing.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable as there is no "training set" in the context of this device.

{0}------------------------------------------------

31388

SEP 2 6 2013

Somnetics International Inc.

Traditional 510(k) Premarket Submission

Transcend Heated Humidifier

5 510(k) Summary

[As required by 21 CFR 807.92]

510(k) Number: K131388

Date Prepared: June 3, 2013

Submitter/ManufacturerSomnetics International, Inc.33 5th Ave NW, Suite 500New Brighton, MN 55112
Establishment Registration #3008770104
Contact PersonMelinda SwansonRegulatory ConsultantTelephone: 651-621-1800Email: mswanson@somnetics.com
Trade NameTranscend Heated Humidifier
Common/Usual NameRespiratory Humidifier
Classification21 CFR 868.5450, product code BTTHumidifier, Respiratory Gas, (Direct Patient Interface)
Product CodeBTT
Predicate DeviceResMed HumidAire 2iTM, K080797

Device Description

The Transcend Heated Humidifier is a humidifier that is designed to humidify the air delivered to the airway during positive airway pressure (PAP) therapy with a compatible Transcend PAP device.

Indications for Use

The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.

{1}------------------------------------------------

:

Substantial Equivalence and Summary of Studies

The Transcend Heated Humidifier was tested and shown to be compliant with the following standards.

Document NumberTitle
IEC 60601-1Medical Electrical Equipment, Part 1: General Requirements for Safety,Ed. 3
IEC 60601-1-2Medical Electrical Equipment – Collateral Standard: Electromagneticcompatibility - Requirements and tests
IEC 60601-1-6Medical Electrical Equipment - Part 1-6: General requirements for basicsafety and essential performance - Collateral Standard: Usability
IEC 60601-1-11Medical Electrical Equipment – Collateral Standard: Requirements forhome health care environment
ISO 8185:2007Respiratory tract humidifiers for medical use -- Particular requirementsfor respiratory humidification systems
ISO 10993-1Biological evaluation of medical devices
ISO 14971Medical devices -- Application of risk management to medical devices

The Transcend Heated Humidifier is substantially equivalent to the predicate device based on comparisons of indications for use and technological characteristics.

Summary of Predicate Comparisons
CharacteristicTranscend HeatedHumidifierResMed LtdHumidAire 2iTMComparison
Indications for use:The Transcend HeatedHumidifier is indicated forthe humidification of the airdelivered from a compatibleTranscend positive airwaypressure therapy device.The Humidifier is intendedfor single patient re-use inthe home environment andin a hospital/institutionalenvironment. TheHumidifier is for use only asrecommended by aphysician.The HumidAire 2i isindicated for thehumidification of the airdelivered from a ResMedcompatible CPAP therapydevice.The HumidAire 2i isintended for single patientre-use in the homeenvironment and multi-patient re-use in a hospital /institutional environment.The HumidAire 2i is for useonly as recommended by aphysician.Similar. The TranscendHeated Humidifier isintended for use only by asingle patient.
Intended Population of UseAdultAdultIdentical
Summary of Predicate Comparisons
CharacteristicTranscend HeatedHumidifierResMed LtdTMHumidAire 2iTMComparison
Compatible DevicesTranscend PAP device andassociated accessoriesResMed PAP device andassociated accessoriesSimilar. The humidifiers aremade to be used withmanufacturers' existing PAPdevices.
Dimensions9 in X 5.5 in X 4.7 in4.8 in X 7.2 in X 8.6 inSimilar. Overall dimensionsdo not effect substantialequivalence determination.
Weight2.2 lbs2 lbSimilar. Overall weight doesnot affect substantialequivalence determination.
DisinfectionNoneShould be disinfected whenused between patients.The Transcend HeatedHumidifier is for singlepatient re-use only.
CleaningDistilled water using a milddetergent.Warm water using a milddetergent.Identical
SterilizationNot sterilizedNot sterilizedIdentical
BiocompatibleYesYesIdentical
Pressure RegulationDetermined by compatiblePAP device.Determined by compatiblePAP device.Identical
Working pressure range4-20 cm H₂O4-20 cm H₂OIdentical
Humidifier Settings1-51-6Similar. Both devices delivera range of humidity basedon setting level.
Inspiratory / ExpiratoryPressure Drop0.375 cmH₂O at 50 L/minNot reportedSubject device meetsrequirements of ISO 8185.
Gas Leak at Max OperatingPressure4.85 L/minNot reportedSubject device meetsrequirements of ISO 8185.
Sound power level at 10 cmH₂O static pressure37 dBNot reportedSubject device meetsrequirements of ISO 8185.
Maximum Heater PlateTemperature131 °F167 °FSubject device meetsrequirements of ISO 8185.
Maximum output humidity>95%95%Similar
Maximum gas temperature77 °FNot reportedSubject device meetsrequirements of ISO 8185.
Water Capacity325 ml400 mlSimilar. Devices areintended to have enoughwater to last a minimum of8 hours (equivalent to a fullnight of sleep)
Operating Temperature+41°F to +95°F+41°F to +104°FSimilar. Operatingtemperature is based oncompatible PAP device.
Operating Humidity10 - 80% relative humidity,non-condensing10-95% relative humidity,non-condensingSimilar. Operating humidityis based on compatible PAPdevice.
Summary of Predicate Comparisons
CharacteristicTranscend HeatedHumidifierResMed LtdHumidAire 2i™Comparison
Shipping/StorageTemperature-4° F to +140° F-4°F to +140°FIdentical
Gas inlet temperaturerange+41°F to +95°F+41°F to +104°FSimilar. Inlet temperature isbased on compatible PAPdevice.
Shipping/Storage Humidity10 - 90% relative humidity,non-condensing15 - 95% relative humidity,non-condensingSimilar
Power Supply100-240 VAC, 50/60 Hz100-240 VAC, 50/60 HzIdentical
IEC 60601-1 ClassificationType BFType CFSubject device is compliantwith ISO 8185 Section 6.8.2-14 which states the deviceshould be either BF or CF.
ElectromagneticcompatibilityCompliant with IEC 60601-1-2Compliant with IEC 60601-1-2Identical
ISO 8185 compliantYesUnknown - not stated in510k summary or UsersManualConsensus standards can beused to determinesubstantial equivalence to alegally marketed device.

{2}------------------------------------------------

Somnetics International Inc.

.

.

Traditional 510(k) Premarket Submission

Transcend Heated Humidifier

י

"

·

: .

:

{3}------------------------------------------------

Somnetics International Inc.

Transcend Heated Humidifier Traditional 510(k) Premarket Submission

The device design was qualified through the following tests and assessments:

  • . Electrical Safety
  • . Electromagnetic Compatibility
  • Biocompatibility Assessment .
  • . Cleaning Validation
  • . Software Validation
  • Packaging and Shipping .
  • Performance: physical (cycling), resistance to flow, spillage, flow sensing, sound . power level, humidification output, out of water shutdown, temperature, runtime, warm-up duration, and reservoir gas leak

These tests and assessments did not raise new safety or efficacy questions.

Conclusion

The Transcend Heated Humidifier is substantially equivalent to the HumidAire 2i (K080797). The subject and predicate devices are used in conjunction with CPAP devices to humidify air. They are equivalent in terms of technology and intended use. Risk assessments, biocompatibility evaluation, software, electromagnetic compatibility and electrical safety, bench testing, and compliance with recognized standards demonstrate that any differences do not raise new

{4}------------------------------------------------

Somnetics International Inc.

Transcend Heated Humidifier

Traditional 510(k) Premarket Submission

.

.

.

:

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

:

. .

.

・・

:

.

.

.

.

:

:

.

.

questions of safety or effectiveness. The Transcend Heated Humidifier is, therefore, substantially equivalent to the predicate HumidAire 2i device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized eagle with three wavy lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 26, 2013

Somnetics International, Incorporated Ms. Melinda Swanson Regulatory Consultant 33 510 Avenue NW, Suite 500 NEW BRIGHTON MN 22112

Re: K131388

Trade/Device Name: Transcend Heated Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Il Product Code: BTT Dated: August 22, 2013 Received: August 26, 2013

Dear Ms. Swanson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kwame O. Ulmer -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if Known): どころは38

Device Name: Transcend Heated Humidifier

Indications for Use:

The Transcend Heated Humidifier is indicated for the humidification of the air delivered from a compatible Transcend positive airway pressure therapy device. The Humidifier is intended for single patient re-use in the home environment and in a hospital/institutional environment. The Humidifier is for use only as recommended by a physician.

Prescription Use _ X _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anya C. Harry -S

Division Sign-Off) wision of Anesthesiology, General Hospital .ection Control, Dental Devices

.i0(k) Number:_ K1313888

§ 868.5450 Respiratory gas humidifier.

(a)
Identification. A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.(b)
Classification. Class II (performance standards).