K Number
K131386

Validate with FDA (Live)

Date Cleared
2014-03-14

(304 days)

Product Code
Regulation Number
868.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EZ-OX Plus Generation II is an integrated portable oxygen delivery system intended to provide supplemental oxygen to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the EZ-OX Plus Generation II is for emergency use only. For all other medical applications, the device is Rx only.

Device Description

The EZ-OX Plus Generation II portable oxygen system is a solution for supplying Oxygen USP using a device comprised of an integrated valve-regulator, flow meter and medical D and E-Oxygen aluminum cylinder with handle and shroud all integrated into a single unit. A range of user-selectable flow setting is available with the user being able to control the flowrate, including low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides standard 40 L/min oxygen delivery. When administered by properly trained personnel for oxygen deficiency and resuscitation, the EZ-OX Plus Generation II is for emergency use only. For all other medical applications, the device is Rx only.

Key specifications include hose barb connection, protective shroud, carrying handle, easy to read content gauge, indexed flow meter, integrated valve-regulator and usage chart label with safety instructions. This design allows medical personnel the ability to provide patient care and treatment sooner without delays caused by the need to mount a conventional regulator.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the EZ-OX Plus Generation II Portable Oxygen System, based on the provided document:

Acceptance Criteria and Device Performance

This device is not an AI/ML device, so many of the common fields for such devices (e.g., number of experts, adjudication methods, MRMC studies, ground truth for training data) are not applicable. The provided document details the testing performed to demonstrate the safety and effectiveness of the updated medical device in comparison to its predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Device: EZ-OX Plus Generation II Portable Oxygen System

Acceptance CriteriaStandard/RequirementReported Device Performance
AccuracyISO 10524-3:2005Demonstrates accuracy, all flow rates met acceptance criteria, proposed device more accurate than predicate.
PrecisionStandard Deviation ≤ 2.0 (USP & ICH community supported)All obtained results were passing, improved precision over predicate device.
Safety TestingISO 10524-3 (Outlet flow limit, Safety valve test, Leakage rate, Adiabatic compression test)Successfully completed.
Safety TestingEN 15996 (Characteristics of functional at reception, Tightness at reception, RPV performance test, Tightness after endurance, Characteristics of functional after endurance)Successfully completed.
Safety TestingASTM G175-03 (Promoted ignition test)Successfully completed.
Cylinder Conformance49 CFR 178.46, Specification seamless aluminum cylindersAluminum cylinders conform.

Note: The document focuses on demonstrating that the updated device meets established safety and performance standards for portable oxygen systems, and also shows improvements over its predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not specify a distinct "test set" sample size in terms of number of devices tested or data points collected for the performance bench testing. It states that testing was "performed across the entire range of user configurable flow control settings."
  • Data Provenance: Not explicitly stated (e.g., country of origin). The testing seems to be internal validation by Air Liquide Healthcare America Corporation. The document does not specify if the testing was retrospective or prospective, but as it's for a new device's 510(k) submission, it would inherently be prospective testing for this specific device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This question is not applicable to this type of medical device submission. The "ground truth" for a portable oxygen system's performance is established by objective engineering and safety standards (e.g., flow rate measurements, leakage tests, material strength), not by expert interpretation.

4. Adjudication Method for the Test Set

Not applicable. As described above, the evaluation relies on objective measurements against engineering and safety standards. There is no subjective interpretation that would require an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically for image-based diagnostic or prognostic AI/ML devices where reader performance is a key metric. This device is a mechanical portable oxygen system.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, this question is not applicable. The device is a physical medical device, not a software algorithm. Its performance is measured directly through bench testing.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is based on:

  • Established Industry Standards: Specifically, ISO 10524-3:2005 for accuracy and EN 15996 and ASTM G175-03 for safety.
  • USP (United States Pharmacopeia) and ICH (International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use) guidelines: Supported standard deviation for precision.
  • Engineering Specifications: Conformance to 49 CFR 178.46 for cylinders.

These standards and specifications define what constitutes acceptable "truth" in terms of device function and safety.

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical device, and there is no "training set" in the context of machine learning or AI. The design modifications and improvements (e.g., excess flow device, brass orifice plate, vent holes, non-shuttling RPV) are based on traditional engineering design, analysis, and validation, not on a machine learning training process.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device. The improvements and design changes were driven by engineering analysis to increase safety and performance, address potential failure modes (even if not yet observed), and enhance user experience (e.g., eliminating audible vibration).

{0}------------------------------------------------

IR LIQUIDE

510 (k) Summary EZ-OX Plus Generation II Portable Oxygen System 510(k) Number: __ K131386

Submitted in accordance with the requirements of Safe Medical Device Act (SMDA) 1990 and 21 CFR 807.92

1.APPLICANT'S/SUBMITTER'S INFORMATION
Air Liquide Healthcare America Corporation
12800 West Little York Road
Houston, TX 77041
Direct Phone:800-624-8000
Internet:www.us.airliquide.com
Establishment Registration No:3003764448
Contact:Angie Beyer
Contact's Phone:713-624-8268
Contact's Fax:713-803-1246
    1. DATE May 31, 2013

3. DEVICE INFORMATION

Trade/Proprietary Name: EZ-OX Plus - Generation 2

Common Name:Portable Oxygen Delivery System
Device Name:Cylinder, Compressed Gas, and Integrated Valve-Regulator
Classification Panel:Cardiovascular and Respiratory Devices
Classification Number:Unclassified
Product Nomenclature:Cylinder, Compressed Gas, and Integrated Valve-Regulator
Product Code(s):ECX
Prior Submission :No prior submission for the subject device.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a stylized "A" in a black square on the left, followed by the words "AIR LIQUIDE" in bold, black letters. The logo is enclosed in a rectangular box.

DEVICE CLASSIFICATION 4.

Empty Compressed gas cylinders and compressed gas cylinder with valve assemblies are unclassified devices under product code ECX and reviewed by the Anesthesiology and Respiratory Devices Branch, Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices.

Individually, gas cylinder pressure regulators and gas pressure gauges are Class 1 devices and exempted from pre-market notification, but are part of the above mentioned assembly.

PREDICATE DEVICE(s) న.

The predicate device for the EZ-OX Plus Generation II is the EZ-OX Plus, which was approved under 510(k) K053117.

DEVICE DESCRIPTION રે.

The EZ-OX Plus Generation II portable oxygen system is a solution for supplying Oxygen USP using a device comprised of an integrated valve-regulator, flow meter and medical D and E-Oxygen aluminum cylinder with handle and shroud all integrated into a single unit. A range of user-selectable flow setting is available with the user being able to control the flowrate, including low flows that may be clinically appropriate for certain classes of patients. An additional DISS-1240 connection provides standard 40 L/min oxygen delivery. When administered by properly trained personnel for oxygen deficiency and resuscitation, the EZ-OX Plus Generation II is for emergency use only. For all other medical applications, the device is Rx only.

Key specifications include hose barb connection, protective shroud, carrying handle, easy to read content gauge, indexed flow meter, integrated valve-regulator and usage chart label with safety instructions. This design allows medical personnel the ability to provide patient care and treatment sooner without delays caused by the need to mount a conventional regulator.

{2}------------------------------------------------

AIR LIQUIDE

Very similar to the original EZ-OX Plus, the newly designed EZ-OX Plus Generation II has a few modifications to increase the safety and performance of the device. These changes include:

  • The first level of filtration (out of three levels) was replaced with an excess . flow device.
    Image /page/2/Figure/3 description: The image shows a technical diagram of a device, with a cross-sectional view on the left and a schematic diagram on the right. The cross-sectional view shows the internal components of the device, with various parts labeled with numbers such as 41 and 87. The schematic diagram illustrates the flow of fluids or gases through the device, including components such as filters, valves, and a pressure regulator. The diagram also includes text labels such as 'SAFETY RELIEF VALVE' and 'PRESSURE REGULATOR'.

Figure 1: EZ-OX Plus with no Flow Limiter showing the inlet filter that was removed and replaced with the excess flow device.

Figure 2: Flow Limiter Subassembly

{3}------------------------------------------------

Image /page/3/Figure/0 description: This image is a technical drawing of a valve. The drawing includes labels with numbers such as 33, 28, 17, 30, 29, 16, and 22. The drawing shows the internal components of the valve and how they fit together.

Figure 2: EZ-OX Plus Generation II with Flow Limiter

By adding the excess flow device, this will assure the body does not become a projectile if the valve is sheered. Air Liquide has not received any complaints of this nature, but the firm feels this proactive design will increase the safety of the device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of the company's name, "AIR LIQUIDE," in a bold, sans-serif font. To the left of the name is a stylized symbol, possibly representing the company's focus on gases or industrial processes. The logo is presented in a simple, clean design.

  • The multiple piece nickel orifice plate was replaced with a one piece brass @ orifice plate.
    Image /page/4/Figure/2 description: This image shows a technical drawing of a mechanical part with various dimensions and annotations. The drawing includes dimensions such as 0.925, 0.915, 0.470, 0.496, 0.066, 0.060, 0.822, 0.816, 0.239, 0.233, 0.302, and 0.295. It also specifies a minimum finish length of 150 and an angle of 27 to 31 degrees. The drawing includes tolerance specifications such as a diameter of 0.241 to 0.239 with a tolerance of 0.002 relative to datum B.

Figure 3: EZOX Plus Generation II Orifice Plate drawing

Image /page/4/Picture/4 description: The image shows two different views of a mechanical part. The top view shows a solid disk attached to a cylindrical shaft with a keyway cut into it. The bottom view shows a similar disk, but with a series of holes drilled around its circumference, also attached to a cylindrical shaft.

Figure 4: EZOX Plus Generation II Orifice Plate model

Concerning the design change to the orifice plate, the new design allows for an increase in performance which translates into a more effective device. Utilizing a one piece design reduces the opportunity for failure at a connection point of a multiple piece assembly. By utilizing this one piece orifice plate design, the firm was also able to eliminate the nickel plate and incorporate it in to the existing brass plate. This change eliminated any potential risk of the nickel slipping out of alignment.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of the company's name, "AIR LIQUIDE", in bold, sans-serif font. To the left of the name is the company's symbol, which appears to be an abstract representation of a gas cylinder or a stylized "A". The logo is enclosed in a rectangular border.

  • Two vent holes were added to the low pressure safety (piston guide) to � relieve pressure in the o-ring cavity.
    Image /page/5/Picture/2 description: This image shows a cross-sectional view of a mechanical assembly, possibly a valve or pump mechanism. The components are rendered in grayscale, highlighting the internal structure and interaction of parts. The text "Shear Force" is visible, indicating a critical area where shear stress is a factor in the design or operation of the mechanism.

Figure 5: EZOX Plus Low Pressure safety

Image /page/5/Picture/4 description: This image shows a diagram of a piston guide with vent holes. The piston is in the center of the image and is surrounded by the guide. The vent holes are located on the side of the piston guide. The text in the image labels the vent holes as "Piston Guide Vent Holes."

Figure 6: EZOX Plus Generation II Low Pressure safety

To increase the safety and effectiveness of the device, two vent holes were added to the low pressure safety (piston guide) to relieve pressure in the o-ring cavity. This relief in pressure will stop the piston guide from sufficiently moving upwards, therefore, preventing the o-ring from extruding and preventing an oxygen leak greater then what a cylinder would otherwise leak.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of a stylized letter "A" in a square to the left of the words "AIR LIQUIDE" in a sans-serif font. The logo is simple and modern, and the colors are muted.

  • Changed from a shuttling residual pressure valve design to a non-shuttling residual pressure valve to eliminate audible vibration and noise.
    Image /page/6/Picture/2 description: This image shows a diagram of a non-shuttling RPV poppet in the open position. The diagram shows the internal components of the poppet, including a spring and a cylinder. The text in the image indicates that the cylinder gas pressure holds the RPV poppet open, and that the RPV poppet is in the open position.

Figure 7: EZOX Plus Generation II RPV

The original design included a shuttling residual pressure valve (PRV). With a shuttling RPV design, the flow of gas from the cylinder through the RPV to the regulator drives intermittent opening and closing of the RPV poppet assembly as the pressure differential across the poppet fluctuates. With this original design, it was observed that an audible vibration or humming was occurring in the EZ-OX Plus units during flowing conditions when it was installed on cylinders. In order to eliminate the humming, the shuttling RPV was replaced with a non-shuttling RPV. In the non-shuttling RPV design, the RPV poppet spring overcomes the force on the poppet due to cylinder gas pressure when the cylinder pressure is depleted below approximately 2.5 bar and forces the RPV poppet to close. After shutting, the RPV poppet interrupts gas flow from the cylinder to the regulator, thereby preventing further cylinder depletion.

7. INDICATIONS FOR USE

The EZ-OX Plus is an integrated portable oxygen delivery system intended to provide supplemental oxygen to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the EZ-OX Plus is for emergency use only. For all other medical applications, the device is Rx only.

{7}------------------------------------------------

AIR LIQUIDE

DEVICE FUNCTION 8

The EZ-OX Plus Generation II is a compressed gas cylinder with an adjustable valve to control the flow of Oxygen, USP to the patient.

9. TECHNOLOGICAL CHARACTORISTICS

A summary comparison of technological characteristics, including design and materials is provided in the table below:

ParameterEZ-OX PlusEZ-OX Plus Gen II
Valve/Regulator
Low Flow SettingsYes (≥ .5L)Yes (≥ .5L)
Flow Between SettingsNoNo
Cylinder on/offNoNo
Filling PortActiveActive
Contents GaugeActiveActive
Filters32
Pressure Design3,000 psi (max)3,000 psi
Single stage piston styleYes.Yes
Guard
Hand Grip1 grip1 grip
Access PortsYesYes
Flow selector/ hose barb/YesYes
gauge aligned
ColorGreenGreen
Height (guard andintegrated valve-regulator)7"7"
Cylinder
Cylinder SizesD, ED, E
Weight (E) (product)950 gr950gr
Materials/constructionAluminumAluminum

The manufacturer believes that the technological characteristics of the EZ-OX Plus Generation II portable oxygen system are substantially similar to those of the predicate device. The predicate device have been granted marketing clearance by FDA following the submission of a 510(k) (510(k) number for EZ-OX Plus is K053117). The modified device has the same intended use and indications for use as the predicate device.

10. PERFORMANCE DATA

The aluminum cylinders conform to the requirements of 49 CFR 178.46, Specification seamless aluminum cylinders.

11. SAFETY TESTING

Safety testing was successfully completed in accordance with the following standards (see Attachment A):

  • . ISO 10524-3 (Pressure regulators for use with medical gases -- Part 3: Pressure regulators integrated with cylinder valves)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of the company's initials, "AL", in a stylized, blocky font on the left side of the image. To the right of the initials is the company's name, "AIR LIQUIDE", in a bold, sans-serif font.

  • Outlet flow limit $6.3 o
  • Safety valve test §6.3 o
  • Leakage rate §6.4 o
  • O Adiabatic compression test $6.6
  • EN 15996 (Gas cylinders Residual pressure valves -- General requirements and type testing)
    • Characteristics of functional at reception $5.4.1 O
    • Tightness at reception $5.4.2 o
    • RPV performance test §5.4.3 o
    • Tightness after endurance $5.4.2 O
    • Characteristics of functional after endurance §5.4.1 o
  • . ASTM G175-03 (Standard Test Method for Evaluating the Ignition Sensitivity and Fault Tolerance of Oxygen Regulators Used for Medical and Emergency Applications)
    • o Promoted ignition test $8.2.8

12. DESCRIPTION OF GAS OUTPUT PERFORMANCE BENCH TESTING

The performance bench testing that was performed demonstrated documented evidence that all flow rates and Standard deviations for the proposed EZ-OX Plus Generation II device met acceptance criteria and all obtained results were passing. The acceptance criteria used was per ISO 10524-3:2005 for demonstrating accuracy and a Standard Deviation of ≤2.0, which is supported by the USP and ICH community, for demonstrating precision. The testing was performed across the entire range of user configurable flow control settings. Additionally, the flow data obtained for the proposed device was compared to that of the predicate device across different test pressures. All flow rate testing results indicated that the proposed device was more accurate and had improved precision over the predicate device.

13. STATEMENT OF SUBSTANTIAL EQUIVALENCE

Based upon the safety and performance testing and compliance with voluntary standards, the manufacturer believes that the EZ-OX Plus Generation II portable oxygen delivery system is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Air Liquide. The logo consists of the company's name, "AIR LIQUIDE", in a bold, sans-serif font. To the left of the name is a stylized symbol, possibly representing the letter "A" or a graphic element associated with the company's brand. The logo is simple and clean, with a focus on readability and brand recognition.

14. EZ-OX PLUS GENERATION II PHOTOGRAPH

The photo below is the assembled device with the regulator head and aluminum cylinder.

Image /page/9/Picture/3 description: In the image, there is a close-up of a glass bottle. The bottle is dark in color and has a rounded shape. The bottle appears to be hanging from something, possibly a tree branch. The bottle is partially obscured by the background.

The illustration below shows the details of the regulator head of the assembly.

Image /page/9/Figure/5 description: The image shows a medical device with various labeled parts. On the left side, the labels point to the flow selector knob, flow setting (LPM), delivery/use port, auxiliary port, cylinder contents gauge, and cylinder connection. On the right side, the labels indicate the product identification label, shut-off valve, CGA 540 fill port, and high-pressure safety feature.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 14, 2014

Air Liquide Healthcare America Corporation Ms. Angie Beyer Compliance Specialist 2700 Post Oak Boulevard, Suite 325 Houston, TX 77056

Re: K131386

Trade/Device Name: EZ-OX Plus - Generation II Regulation Number: Unclassified Regulation Name: Cylinder, Compressed Gas, And Valve Regulatory Class: Unclassified Product Code: ECX Dated: February 20, 2014 Received: February 24, 2014

Dear Ms. Beyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part

807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

https://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Teiashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID' FEDA, I

Sincerely yours,

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

FOR

Enclosure

{12}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: EZ-OX Plus Generation II Portable Oxygen Delivery System

Indications For Use:

The EZ-OX Plus Generation II is an integrated portable oxygen delivery system intended to provide supplemental oxygen to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the EZ-OX Plus Generation II is for emergency use only. For all other medical applications, the device is Rx only.

AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

K131386

Image /page/12/Picture/8 description: The image shows two lines of text. The first line says "Anya C." and the second line says "Harry -S". The text is written in a simple, sans-serif font and is black against a white background. The text is slightly blurry.

Digitally signed by Anya C. Harry -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Anya C. Harry -S. cheAnya C. Hany -5,
0.9.2342.19200300.100.1.1=0011315 Page 1 of _f
500 590 Date: 2014.03.14 04:36:11 -04/00'

§ 868.2700 Pressure regulator.

(a)
Identification. A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.