K Number
K131303

Validate with FDA (Live)

Date Cleared
2013-06-11

(35 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Circular Stapler has application throughout the alimentary tract for end to-end, end-to-side and side-to-side anastomoses.

Device Description

The disposable circular stapler is preloaded with two staggered rows of 26mm titanium staples in both inner and outer circles. After the deploying of the instrument, the circular knife blade inside the instrument can cut off excess tissue automatically to create a circular anastomosis. The instrument is fired by squeezing the trigger handle firmly as far as it will go. The size of the anastomosis site is determined by the diameter of the selected circular knife blade.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Disposable Circular Stapler, YH-26) and focuses on establishing substantial equivalence to a predicate device. It does not describe an AI/ML device or a study that evaluates its performance against acceptance criteria in the typical sense of AI model metrics (like AUC, sensitivity, specificity).

Therefore, I cannot extract the information required for an AI/ML device as per your request related to:

  • A table of acceptance criteria and reported device performance (in terms of AI metrics)
  • Sample sizes and data provenance for test sets
  • Number and qualifications of experts for ground truth
  • Adjudication method
  • MRMC comparative effectiveness study
  • Standalone performance
  • Type of ground truth used (for AI)
  • Sample size for training set
  • How ground truth for the training set was established

However, I can describe the acceptance criteria and study that proves the device meets its acceptance criteria based on the provided document, which is a regulatory submission for a traditional medical device.


1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Tests/StandardsReported Device Performance
Material BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11Successfully passed all tests.
Sterilization & PackagingISO 11137-1:2006, ISO 11137-2:2006, ISO 11737-1:2006, ISO 11737-2:2009, ISO 11607-1:2006, ASTM F1980-07, ASTM F88/F88M-09, ASTM D3078-02 (2008)Products successfully passed all tests, ensuring sterility for a 3-year shelf life.
Functional Equivalence (Staple Performance)Bench testing based on predicate device performance and standards: ISO 5832-2:1999 (Unalloyed Titanium for surgery) & ASTM F67-06 (Unalloyed Titanium for Surgical Implant Applications)Energy to deliver and remove staples was consistent with the predicate device. Stated "no appreciable differences in the energy needed to insert or extract the staples" despite staple size difference.

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes for the bench testing, biocompatibility testing, or sterility testing. It describes these as "bench testing," "biocompatibility testing," and "sterility testing" without detailing the number of units or samples used in each test.

The data provenance is from laboratory and bench tests conducted by or for Changzhou City Zhiye Medical Devices Institute Co., Ltd. (China). The data is prospective in the sense that these tests were performed specifically for this 510(k) submission to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable as the study did not involve human experts establishing "ground truth" in the context of diagnostic interpretation. The "ground truth" for this device focused on objective engineering, material, and biological safety standards. The "experts" would be the scientists, engineers, and microbiologists conducting and evaluating the tests against established international and ASTM standards.

4. Adjudication method for the test set

Not applicable. The tests performed are objective measurements against defined standards, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for diagnostic imaging devices or AI tools that assist human interpretation, not for mechanical surgical staplers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a mechanical surgical instrument, not an algorithm or AI. Its performance is inherent to its design and manufacturing, tested objectively, and always used by a human surgeon ("human-in-the-loop" in a different sense).

7. The type of ground truth used

The "ground truth" for this device was established based on:

  • Pre-defined international standards (e.g., ISO, ASTM) for materials, biocompatibility, and sterilization.
  • Predicate device performance characteristics for functional aspects like staple insertion/extraction energy.
  • Objective measurements and laboratory assays.

8. The sample size for the training set

Not applicable. This device is not an AI/ML model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable. As there is no training set for an AI/ML model, this question is irrelevant to the provided document.

{0}------------------------------------------------

JUN 1 1 2013

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: April 22, 2013

Company making the submission:

Name - Changzhou City Zhiye Medical Devices Institute Co., Ltd.

Address - No. 127 Xiacheng Road, Changzhou City, Jiangsu Province, China, 518054 Tel: +86-519-86707185

Fax: +86-519-86707186

Contact - Mrs. Huang Xinying

Email -charliemack@irc-us.com

Correspondent:

Name: Charles Mack

Address: 12226 Washington Lane, Parker, Arizona, 85344

Telephone: 931-625-4938

Email: charliemack@irc-us.com

Device:

Trade/proprietary names: Disposable Circular Stapler, YH-26 Common Name: Implantable Staple Classification Name: Staple, implantable

{1}------------------------------------------------

K131303 pa

page 4/3

Predicate Device:

Circular Stapler:

Endopath ILs Endoscopic Circular Stapler Ethicon, Inc. 4545 Creek Rd. Cincinnati, OH 45242 283 K920752

Classifications Names & Citations: 21CFR 878.4750. GDW, Implantable staple, Class II

Description:

General

Disposable Circular Stapler

The disposable circular stapler is preloaded with two staggered rows of 26mm titanium staples in both inner and outer circles. After the deploying of the instrument, the circular knife blade inside the instrument can cut off excess tissue automatically to create a circular anastomosis. The instrument is fired by squeezing the trigger handle firmly as far as it will go. The size of the anastomosis site is determined by the diameter of the selected circular knife blade.

Indications For Use

The Disposable Circular Stapler has application throughout the alimentary tract for end to-end, end-to-side and side-to-side anastomoses.

Technical Characteristics:

The disposable circular stapler is similar in construction and operation to the submitted predicate device. The only difference in the submitted device is the size of the staples. The staples in the submitted device is 26 mm height and is 1.5-2.0 mm compressed, where the predicate staples are offered in 25 and 29 mm heights and when compressed they are 1.5-2.5 mm. The staple differences are not significant and festing of the staples during insertion and extraction showed that there are no appreciable differences in the energy needed to insert or extract the staples.

42

{2}------------------------------------------------

K131303 page 3/3

Test Summary:

In order to establish substantial equivalence to the identified predicate device, we performed the following tests:

Bench testing was performed to demonstrate the energy to deliver staples and also to remove staples was consistent with the predicate device. ISO 5832-2:1999 Implants for surgery - Metallic materials - Part 2: Unalloyed Titanium and ASTM F67-06 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700).

Biocompatibility testing was performed to ensure the stapler and cutters were constructed of material which passes the ISO 10993 benchmarks. The devices successfully passed tests for ISO 10993-1, ISO 10993-5, ISO 10993-5, ISO 10993-6, ISO 10993-10 and ISO 10993-11.

As these products are delivered sterile, sterility testing was performed to ensure the sterilization methodology was sufficient and that the packaging kept the products sterile for the stated 3 year shelf life. Testing was performed to ISO 11137-1:2006, ISO 11137-2:2006, ISO 11737-1:2006 and ISO 11737-2:2009, ISO 11607-1:2006, ASTM F1980-07, ASTM F88/F88M-09 and ASTM D3078-02 (2008) with the products successfully passing all tests.

Conclusions:

In accordance with the Federal Food, Druq and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Changzhou City Zhiye Medical Devices Institute Co., Ltd. concludes that Disposable Circular Stapler, Model YH-26 is substantially equivalent to predicate device as described herein.

END

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with three overlapping, curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Changzhou City Zhiye Medical Devices Institute Co., Ltd. % IRC Mr. Charlie Mack 12226 Washington Lane Parker, Arizona 85344

June 11, 2013

Re: K131303

Trade/Device Name: Disposable Circular Stapler, YH-26 Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: May 07, 2013 Received: May 07, 2013

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal-Food -- Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Charlie Mack

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours,

For

Peter D. Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health"

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K131303

Device Name: Disposable Circular Stapler, YH-26

Indications For Use:

The Disposable Circular Stapler has application throughout the alimentary tract for end to-end, end-to-side and side-to-side anastomoses.

Prescription Use X (Part 21 CFR 801 Subpart D)

i

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page of

David Krailise -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K131303

40

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.