K Number
K131141

Validate with FDA (Live)

Device Name
EARPROBE
Manufacturer
Date Cleared
2013-06-27

(65 days)

Product Code
Regulation Number
882.1900
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ear Probe is intended to be an accessory to auditory testing equipment that use evoked responses (e.g. DPOAE, ABR) to assess hearing function.
· The Ear Probe is capable of both generating and recording sounds in the ear canal.
· The Ear Probe is intended for use in all ages.
The interface to the patient's ear is provided by means of disposable standard eartips, which are available as separate consumables.

Device Description

The Earprobes EP-DP and EP-TE are passive Transducers, which are used to covert electrical stimulus into acoustical stimulus. The electrical stimuli are provided by otoacoustic emissions devices or auditory evoked responses stimulators (afterwards referred as interfacing instruments). The acoustical stimulus is then coupled into the patient's ear.
The probe comprises of four sections:
a) Electrical transmission path
b) speaker and microphone
c) acoustic transmission path
d) transducer case

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Earprobe EP-DP and EP-TE, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance:

The 510(k) summary does not explicitly state numerical "acceptance criteria" for the Earprobe EP-DP and EP-TE that were established prior to testing. Instead, the study conducted was a performance comparison against a legally marketed predicate device (Echo-Screen TDA probe). The "reported device performance" is essentially the finding of substantial equivalence to the predicate in acoustical and electrical parameters.

Acceptance Criteria (Implicit)Reported Device Performance (Findings from Studies)
Electrical Interface Performance (between probe and instrument)"In this 510 (k) application, the probe without instrument and without eartips shall be compared to the substantial equivalent probe of echo-screen with respect to the electrical interface." (Implies meeting the electrical interface performance of the predicate)
Acoustical Parameters (sound generation and recording)"Substantial equivalence to the EchoScreen is based on non clinical performance testing of the acoustical and electrical parameters." Specifically, for the study: - Earprobe EP connected to AccuScreen (equivalent to EchoScreen TDA) showed "proof of substantial equivalence in performance" when compared to the original AccuScreen probe connected to AccuScreen.
Overall Performance/Functionality (e.g., generating/recording sounds, compatibility with auditory testing equipment)The device was certified with a CE mark together with the Sentiero device, indicating it met required performance standards for that certification for "preschooler's hearing screening." This implies successful demonstration of generating and recording sounds as an accessory to auditory testing equipment.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not explicitly stated as a numerical value (e.g., number of patients or specific measurements). The studies mention testing in "different hospitals in Germany and Danmark" and comparisons to predicate devices.
  • Data Provenance:
    • Country of Origin: Germany and Danmark.
    • Retrospective or Prospective: The description indicates that "performance studies were conducted," implying a prospective nature for these tests, where the Earprobe EP was actively tested in various configurations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the 510(k) summary. The studies described are performance comparisons against a predicate device and certification efforts, rather than studies requiring expert adjudication of data for establishing ground truth in the traditional sense (e.g., clinical diagnosis). The "ground truth" here is the performance of the predicate device, or established engineering specifications for electrical and acoustical parameters.

4. Adjudication Method for the Test Set:

This information is not applicable/provided as the studies described are performance comparisons and certification processes, not clinical studies requiring expert adjudication of patient outcomes or interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. The device described is a passive transducer (ear probe) used as an accessory for auditory testing equipment. It is not an AI-powered diagnostic or interpretive device, nor does it involve "human readers" in the context of interpretation that would necessitate an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. The device is a physical accessory (ear probe), not an algorithm. Its performance is tested in conjunction with the interfacing instruments it connects to. However, the 510(k) application focuses on the probe itself "without instrument and without eartips" for the electrical interface comparison, which could be considered a form of "standalone" testing for its intrinsic electrical properties.

7. The Type of Ground Truth Used:

The ground truth used for evaluating the Earprobe EP-DP and EP-TE was primarily:

  • Predicate Device Performance: The performance of the "original AccuScreen probe connected to AccuScreen" (which is equivalent to the EchoScreen TDA probe) served as the benchmark for demonstrating "substantial equivalence."
  • Engineering Specifications/Standards: For the CE mark certification and the comparison of electrical and acoustical parameters, the device's performance was measured against established technical specifications or standards relevant to auditory transducers.

8. The Sample Size for the Training Set:

Not applicable. The Earprobe EP-DP and EP-TE is a passive transducer and does not involve machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As the device does not employ machine learning, there is no training set or corresponding ground truth establishment process in this context.

{0}------------------------------------------------

K131141

510K Summary

JUN 2 7 2013

Date:

Submitted by:

Telephone Fax:

Contact person

April 18th, 2013

PATH medical GmbH Landsberger Str. 63 82110 Germering Germany 0049 89 80076502 0049 89 80076503

Johann Oswald Director oswald@pathme.de

Name of device

Common name

Classification name

Predicative Device

Earprobe EP-TE and EP-DP

Evoked Response Auditory Stimulator

Accessory to Evoked Response Auditory Stimulator (per 21 CFR section 882.1900) and otoacoustic emissions instruments

Probe only of Echo-Screen T series (510 (k)# K013977)

{1}------------------------------------------------

Description of the Device

The Earprobes EP-DP and EP-TE are passive Transducers, which are used to covert electrical stimulus into acoustical stimulus. The electrical stimuli are provided by otoacoustic emissions devices or auditory evoked responses stimulators (afterwards referred as interfacing instruments). The acoustical stimulus is then coupled into the patient's ear.

The probe comprises of four sections:

a) Electrical transmission path

b) speaker and microphone

c) acoustic transmission path

d) transducer case

a) The electrical transmission path is connected to the interfacing instrument by a connector, which is connected to a shielded cable. The other end of the cable is attached to the microphone and speaker system.

Within the connector, an EEPROM is placed to store calibration data of the probe and to enable an interfacing device to identify the attached probe.

b) Each of the two speakers (in case of EP-DP) or the only speaker (in case of EP-TE) converts the electrical stimulus into an acoustic stimulus. The acoustic stimulus is delivered to the patient's ear by means of the acoustic transmission path.

One microphone in the probe is used to pick up the acoustical stimulus (which is delivered by the acoustical transmission path) and convert it to an electrical stimulus.

c) The acoustic transmission path is separated for each speaker and microphone. It consists of a silicone sealing which contains the speaker and microphones and connects to the acoustic ducts. which are implemented within the probe tip. The ducts are separated from each other and are used to smooth the acoustic output at the probe tip. Ear tips, which is not part of the probe, are put onto the tip of the probe tip in order to interface to the patient's ear.

d) The transducer case is the housing of the speaker and acts as the bridge between electrical and acoustical transmission path. It also provides means of handling.

Intended use of the Device

The Ear Probe is intended to be an accessory to auditory testing equipment that use evoked responses (e.g. DPOAE, ABR) to assess hearing function.

· The Ear Probe is capable of both generating and recording sounds in the ear canal.

· The Ear Probe is intended for use in all ages.

The interface to the patient's ear is provided by means of disposable standard eartips. which are available as separate consumables.

{2}------------------------------------------------

Comparison to Predicative Device:

The Earprobe EP-DP and EP-TE (from now on referred to as EP) were designed as accessories for different otoacoustic emissions instruments and auditory response stimulators. In comparison to the probe of the Echo-Screen T series (510 (k)# K013977), the Earprobe EP share the same technological characteristics.

The Earprobe EP and Echo-Screen probe comprise of a cable with connector on one end and probe on the other end. The only difference is the cable manufacturer, cable color and dimensions and connector type.

In both probes, within the connector, an EEPROM is placed to store calibration data of the probe and to enable an interfacing device to identify the attached probe

The microphone and the loudspeaker in the Earprobe EP and Echo-Screen probe are housed in a transducer case. The manufacturer of the loudspeaker and microphones are the same. Only the type of the speakers are different together with the physical dimensions, color and shape of the housing.

Similar to the Echo-Screen probe, the Earprobe EP has a removable probetip with separate ducts to deliver the acoustical stimuli and pick up the acoustical response. The only difference is the shape, dimension and color.

The intended use of both the Earprobe EP and the Echo-Screen probe is the same. Both are intended to be used as transducers to convert electrical stimuli delivered by an otoacoustic emissions instrument or evoked response stimulator, into acoustical stimuli, which is then delivered to the patient's ear.

In both cases, the Earporbe EP and the Echo-Screen probe, a disposable eartip will be used to interface to the patients ear. The eartips to be used with the Earprobe EP are standard earlins and not part of this 510 (k).

Substantial Equivalence Performance Metrics

Substantial equivalence to the EchoScreen is based on non clinical performance testing of the acoustical and electrical parameters.

The Earprobe EP DP and EP-TE are already certified with CE mark together with the Sentiero device (from PATH medical for preschooler's hearing screening). To reach certification for the instrument together with the Earprobe EP, the following performance studies were conducted:

  • [2] Earprobe EP connected to AccuScreen (equivalent to EchoScreen TDA, GN Otometrics, Danmark). This study was conducted in comparison to the orignial AccuScreen probe connected to AccuScreen and is a proof of substantial equivalence in performance.
  • [8] Earprobe EP connected to Otobox (a clinical research platform of Technische Universitaet Muenchen, Munich, Germany.)
  • [7] Earprobe EP connected to Sentiero (PreSchool Hearing Screening Instrument using TEOAE and DPOAE, PATH medical GmbH, Germany)

Performance and safety testing was conducted in different hospitals in Germany and Danmark. As the probe itself is passive, the firmware of the active instrument to which the probe connects to and the used eartips are crucial components for performance testing and substantial equivalence.

In this 510 (k) application, the probe without instrument and without eartips shall be compared to the substantial equivalent probe of echo-screen with respect to the electrical interface.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human form or a caduceus, rendered in black. The overall design is simple and conveys a sense of official government authority.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2013

PATH medical GmbH Dr. Johann Oswald Lansdsberger Str 63 Germering, Bavaria 82110 GERMANY

Re: K131141

Trade/Device Name: Earprobe EP-DP and EP-TE Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ Dated: May 21, 2013 Received: May 21, 2013

Dear Dr. Oswald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Dr. Johann Oswald

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K131141 510(k) Number (if known):

Device Name: Ear Probe EP-DP and EP - TE

Indications for Use:

The Ear Probe is intended to be an accessory to auditory testing equipment that use evoked responses (e.g. DPOAE, ABR) to assess hearing function.

· The Ear Probe is capable of both generating and recording sounds in the ear canal.

· The Ear Probe is intended for use in all ages.

The interface to the patient's ear is provided by means of disposable standard eartips, which are available as separate consumables.

Prescription Use × Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Eric A. Mann -S -04'00

Page of

§ 882.1900 Evoked response auditory stimulator.

(a)
Identification. An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.(b)
Classification. Class II (performance standards).