K Number
K131080

Validate with FDA (Live)

Date Cleared
2014-06-13

(445 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended to emit energy in the infrared spectrum to provide topical heating for the purposes of elevating tissue temperature when heat is indicated for the temporary relief of minor muscle and joint pain and stiffness, for the temporary relief of minor joint pain associated with arthritis, to temporarily increase local blood circulation where applied and the relaxation of muscles. This is a prescription only device.

Device Description

Not Found

AI/ML Overview

I am unable to extract information about the acceptance criteria and the study that proves the device meets the acceptance criteria from the provided document. The document appears to be an FDA 510(k) clearance letter for a medical device (Physiolaser Olympic) and an associated Indications for Use form.

This type of document typically confirms that the device is substantially equivalent to a legally marketed predicate device, but it does not usually contain the detailed technical studies, acceptance criteria, or performance data that would be part of the 510(k) submission itself.

Therefore, I cannot provide the requested information, including:

  1. A table of acceptance criteria and the reported device performance
  2. Sample size used for the test set and the data provenance
  3. Number of experts used to establish the ground truth for the test set and their qualifications
  4. Adjudication method for the test set
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and its effect size
  6. If a standalone performance study was done
  7. The type of ground truth used
  8. The sample size for the training set
  9. How the ground truth for the training set was established

The document primarily focuses on regulatory clearance, general controls, and labeling requirements, along with the device's intended use for temporary relief of minor pain, stiffness, and to increase local blood circulation through topical heating.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 13, 2014

・・・ . :.

Reimers & Janssen Medizintechnik GmbH Josephus Janssen Fabrikstr. 22, Waldkirch, Baden Wurthemberg, Germany, 79183

Re: K131080

Trade/Device Name: Physiolaser Olympic: Single Probes CW; Cluster Probes CW Regulation Number: 21 CFR 890.5500 Regulatory Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: April 29, 2014 Received: May 01, 2014

Dear Mr. Janssen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

Page 2 – Josephus Janssen

device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

.

Sincerely yours.

Binita S. Ashar -S

2014.06.13 16:43:30 -04'00'

Binita S. Ashar, MD, MBA, FACS Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and

Radiological Health

cc: DMC 510(k) Staff Division D.O.

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K131080

Device Name

Physiolaser Olympic, Single Probes CW, Cluster Probes CW

Indications for Use (Describe)

Intended to emit energy in the infrared spectrum to provide topical heating for the purposes of elevating tissue temperature when heat is indicated for the temporary relief of minor muscle and joint pain and stiffness, for the temporary relief of minor joint pain associated with arthritis, to temporarily increase local blood circulation where applied and the relaxation of muscles. This is a prescription only device.

. :

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Neil R Ogden -S 2014.06.13 15:31:25 -04′00″

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

N/A