K Number
K131052

Validate with FDA (Live)

Date Cleared
2013-08-16

(123 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.

The Clearing Stem Models are indicated for use as follows:

  • NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
  • NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
  • NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
  • NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
  • NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).
  • NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
  • NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
  • G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
  • G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
  • G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 -- 18 Fr and have a length of 25 cm (10 in).
  • G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).
  • G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
  • G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
Device Description

The Proposed Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NE-1036, NE-1042, NE-1043, NE-1045, NE-1048, NE-1050, NE-1055, G-1008, G-1009, G-1010, G-1012, and G-1014. The Control Box is reusable and the Clearing Stems are single use. The Operator manually inserts the Clearing Stem into the feeding and/or decompression tube (i.e., Tube) and directs the Clearing Stem's progression along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear reciprocating motion. The linear reciprocating motion is transferred to the proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Mire. The motion at the Wire Tip mechanically acts on the occlusion and restores Tube patency.

AI/ML Overview

The provided document details the 510(k) summary for the TubeClear device, specifically focusing on the new Clearing Stem Models. The submission claims substantial equivalence to its predicate device, also named TubeClear. The core of the device is to clear occlusions in feeding and decompression tubes using a reciprocating wire.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format for each individual criterion. Instead, it describes various tests and states whether they were "successfully completed," "passed acceptance criteria," or found to be "equivalent" to the predicate device.

Below is a synthesized table based on the descriptions provided:

Test CategorySpecific TestAcceptance Criteria (Implied)Reported Device Performance
Technical TestingProduct Specifications VerificationMeet product specifications"Confirmed that the Clearing Stem Models met product specifications passing all acceptance criteria."
Transportation VibrationPass acceptance criteria for US highway truck"Transportation testing successfully passed acceptance criteria for US highway truck."
Shelf LifePass 24-month storage"Shelf life was tested via accelerated life testing and passed 24 month storage."
Efficacy TestingUsabilitySuccessfully reviewed by licensed healthcare practitioners"Usability testing by licensed healthcare practitioners was successfully completed."
Effectiveness (Simulated Use)Equivalent to Predicate Clearing Stem Models"Found to be of equivalent effectiveness" to predicate devices.
Safety TestingTube Damage (scratches, nicks, tears, abrasions, punctures)Equivalent to Predicate Clearing Stem Models"Found to be equivalent for the Proposed and Predicate Devices."
Tube Heating during OperationNegligible and equivalent to Predicate Clearing Stem Models"Negligible and equivalent for the Proposed and Predicate Devices, thus the Devices have equivalent safety in terms of heating."
Tube Movement during OperationSame magnitude as Predicate Clearing Stem Models"Measured and found to be of the same magnitude, thus equivalent for both Devices."
Control Box TestingElectrical Safety (IEC 60601-1 3rd Edition)Compliance with IEC 60601-1 3rd Edition"Successfully completed in December, 2012."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes used for the test sets in any of the described tests (Technical, Efficacy, Safety). It mentions "Clearing Stem Models" in plural, implying multiple units were tested, but no concrete numbers.

The data provenance is from bench testing performed by the manufacturer (Actuated Medical, Inc.) and a third-party lab (Intertek Group for electrical safety). All testing appears to be retrospective in the sense that it's performed on manufactured devices for regulatory submission. There is no mention of country of origin for the data; however, the testing entity (Intertek Group, Boxborough, MA) suggests the testing itself occurred in the US.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

For the "Usability testing," it states "Usability testing by licensed healthcare practitioners was successfully completed." However, it does not specify the number of these practitioners or their specific qualifications (e.g., years of experience, specialty). Therefore, the number and qualifications of experts for establishing ground truth are not provided.

4. Adjudication Method for the Test Set

The document does not describe any formal adjudication method (e.g., 2+1, 3+1, none) for the test results. The assessment of equivalence and success appears to be based on direct measurement and comparison to the predicate device or established standards, rather than expert consensus on individual cases that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly conducted or reported. This submission primarily focuses on the technical and safety equivalence of the device itself through bench testing, not on the improvement of human readers' performance with or without the device. The device is a tool used by practitioners, not an AI diagnostic aid for human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The device described, "TubeClear," is a physical medical device that requires direct human operation to insert and advance the clearing stem. It is not an algorithm or AI system that can perform without human-in-the-loop. Therefore, a standalone (algorithm-only) performance study is not applicable and was not performed. The "Control Box" provides electromechanical actuation, but the device's function is inherently human-assisted.

7. The Type of Ground Truth Used

The ground truth for the efficacy and safety tests appears to be based on:

  • Product Specifications: For technical verification.
  • Simulated Use Conditions: For efficacy testing, where successful clearing of occlusions in simulated tubes would be the "ground truth."
  • Measurement and Observation: For safety tests like tube damage, heating, and movement, where direct measurement of these parameters against acceptable ranges or predicate device performance constitutes the "ground truth."
  • Licensed Healthcare Practitioner Evaluation: For usability, where their feedback on the device's ease of use and performance in a simulated environment serves as the "ground truth."

There is no mention of pathology, biological outcomes data, or patient outcomes data being used as ground truth for this submission, as it explicitly states "No clinical data was collected."

8. The Sample Size for the Training Set

Since "No clinical data was collected" and the device is a mechanical tool rather than an AI/ML algorithm, there is no concept of a "training set" in the context of this 510(k) submission.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI/ML algorithm, the question of how its ground truth was established is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Actuated Medical. The logo consists of a stylized medical symbol to the left of the word "ACTUATEDMEDICAL" in all caps. Below the company name is the tagline "Innovative motion · Positive outcomes".

pg 1 of 7 310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355.0003 + f (814) 355 1532 + w ActuatedMedical com

Certified

  • ISO 13485 & 14971
    • 9001 Compliant
  • WBF

SECTION 5: 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR807.92.

510(k) Number: K131052

5.1: APPLICANT INFORMATION

Date Prepared:August 9, 2013
Name and Address:Actuated Medical, Inc.310 Rolling Ridge DriveBellefonte, PA 16823Ph: 814-355-0003Fax: 814-355-1532AUG 16 2013
Contact Person:Victoria A. Kellogg, Ph.D., CRNP, M.B.A., Esq.Director Medical AffairsPh: 814-355-0003 x118Fax: 814-355-1532

Email: Victoria.kellogg@actuatedmedical.com

5.2: DEVICE INFORMATION

Classification:KNT
Trade Name:TubeClear
Common Name:In Patient Tube Clearing System
Classification Name:Tubes, Gastrointestinal and Accessories21 C.F.R. §876.5980

{1}------------------------------------------------

5.3: PREDICATE DEVICE

510(K)NumberTrade or ProprietaryNameManufacturerComment
K121571TubeClearActuated Medical, Inc.Original andTraditional510KCleared6/18/12
K123659TubeClearActuated Medical, Inc.Special510K1Cleared12/20/12

The legally marketed device to which substantial equivalence is being claimed:

The Predicate Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NG-1036 and NG-1043.

5.4: DEVICE DESCRIPTION

The Proposed Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NE-1036, NE-1042, NE-1043, NE-1045, NE-1048, NE-1050, NE-1055, G-1008, G-1009, G-1010, G-1012, and G-1014. The Control Box is reusable and the Clearing Stems are single use. The Operator manually inserts the Clearing Stem into the feeding and/or decompression tube (i.e., Tube) and directs the Clearing Stem's progression along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear The linear reciprocating motion is transferred to the reciprocating motion. proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Mire. The motion at the Wire Tip mechanically acts on the occlusion and restores Tube patency.

1 Special 510K submitted to address a change to the Clearing Stem. The change was the removal of pad-printed markings from the Clearing Stem Sheath.

{2}------------------------------------------------

Thirteen (13) new Clearing Stem Models are proposed. The additional Clearing Stem Models expand the specific indications for use. 5.5: INTENDED USE

The Proposed Clearing Stem Models are substantially equivalent to the Predicate Clearing Stem Models in regards to intended use and therapeutic effect. The intended use for both the Proposed and Predicate Clearing Stems is to clear occlusions / clogs from feeding and decompression tubes (i.e., Tubes). The therapeutic effect is restoring patency to the Tube and alleviating the need for Tube replacement.

All Models of Clearing Stem are disposable, single use devices. All Models are intended for use by Licensed Healthcare Practitioners. 'All Models are intended for use in hospitals, in long term care facilities, and in homes (serviced by licensed healthcare practitioners). Because all Clearing Stem Models enter a Tube that is within the Patient, no Model makes direct contact with the Patient. All models of Clearing Stems are placed into a Tube that is in the Patient's gastro-intestinal (GI) system. Sterile conditions are not required for the intended purpose.

5.6: INDICATIONS FOR USE

TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.

The Clearing Stem Models are indicated for use as follows:

  • NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
  • NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
  • NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
  • NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
  • NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).

{3}------------------------------------------------

  • NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
  • NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
  • G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
  • G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
  • G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 -- 18 Fr and have a length of 25 cm (10 in).
  • G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).
  • G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
  • G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).

5.6.1: Specific Indications for Use of Predicate Clearing Stem Models

Clearing Stem Model NG-1036 is indicated for use in Nasogastric tubes that are of size 10 - 14 French and have a length of 91 - 108 cm (36 - 42 in).

Clearing Stem Model NG-1043 is indicated for use in Nasogastric tubes that are of size 10 - 14 French and have a length of 109 - 127 cm (43 - 50 in).

The specific indications for use vary from Clearing Stem Model to Clearing Stem Model; however, the differences are not critical to the intended therapeutic use of the device, which is restoring patency to the Tube and alleviating the need for Tube replacement. The differences do not affect the safety and effectiveness of the Proposed Clearing Stem Models.

{4}------------------------------------------------

Address Macked Coro

K131052 pg 5 of 7

5.7: TECHNOLOGICAL CHARACTERISTICS

The technology is the same for the Proposed and Predicate Clearing Stems. The operating principle, Operator interaction with the Device, and the mode of actuation of the Clearing Stem Wire is identical (i.e., Control Box Model 101). Both the Proposed and Predicate Clearing Stems are manually inserted and advanced into the Patient's Tube. Both the Proposed and Predicate Clearing Stem Models physically interact with the occlusion material inside a Tube. Both the Proposed and Predicate Clearing Stem Models mechanically clear occlusions. The Wire inside the Sheath is linearly actuated by the Motor inside the Control Box.

The design features of the Proposed Clearing Stem Models and the Predicate Clearing Stem Models are very similar. For both the Proposed and Predicate Clearing Stem Models, the Sheath is the same material (i.e., PTFE extruded tubing) and length. Each Model has a Collar specifically positioned to limit the depth that the Clearing Stem can be inserted into the Tube. The Collar position determines the Working Length of the Clearing Stem and is specific to each Model. The main difference between the Proposed Clearing Stem Models and the Predicate Clearing Stem Models is the Wire inside the Sheath of the Clearing Stem; specifically the Wire Tip. The Proposed Clearing Stem Design incorporates a more flexible Wire Tip than the Predicate Clearing Stem Design. The Proposed Wire Tip Design is safe and effective for the expanded indications for use.

The Wire used in the Proposed Clearing Stems is 203.340 cm (80.055 in) in length; 304V stainless steel; and a Lunderquist Style Mandrel Guide Wire. It has a stiff, fixed Core and a flexible Tip. Beginning at 109.22 cm (43.000 in) from the distal end, a flat Coil Wire is wound over the core wire creating an outer diameter of 0.089 ± 0.008 cm (0.035 ± 0.003 in). Ten (10) cm (3.937 in) from the distal end, the core wire gradually tapers over a 5.000 ± 0.500 cm (1.969 ± 0.197 in) length to 0.015 ± 0.013 cm (0.006 ± 0.005 in). The distal 5.000 ± 0.500 cm (1.969 ± 0.197 in) of the core wire are a constant 0.015 ± 0.013 cm (0.006 ± 0.005 in) in diameter. The Coil Wire is brazed onto the Core wire. The distal end of the wire is rounded and smooth with a radius of .046 cm (0.018 in).

The Wire used in the Predicate Clearing Stems is 203 cm (80 in) in length; flexible twisted 1x7 strand 304V stainless steel. The Wire diameter is 0.069 ± 0.008 cm (0.027 ± 0.003 in). The ends of the Wire are fused and rounded.

{5}------------------------------------------------

Actuato Madizal Com

K131052 pg 6 of 7

5.8: NON-CLINICAL PERFORMANCE DATA

As the same Control Box (i.e., Model 101) will be used, testing focused predominantly on the Clearing Stem. Testing for the Control Box Model 101 consisted of upgrading electrical safety testing to IEC 60601-1 30 Edition; Medical Electrical Equipment, Part 1: General requirements for basic safety and essential performance. Testing was performed by Intertek Group (Boxborough, MA) and was successfully completed in December, 2012.

Bench Testing of the Clearing Stem focused in three areas: Technical, Efficacy, Technical testing included verification of product specifications, and Safety. transportation vibration, and shelf life. Efficacy testing included usability and effectiveness. Safety testing evaluated the potential for Tube damage, Tube heating and Tube movement during use of TubeClear.

5.8.1: Technical Testing

Verification confirmed that the Clearing Stem Models met product specifications passing all acceptance criteria. Transportation testing successfully passed acceptance criteria for US highway truck. Shelf life was tested via accelerated life testing and passed 24 month storage. Additional testing for longer shelf life is ongoing.

5.8.2: Efficacy Testing

Usability testing by licensed healthcare practitioners was successfully completed. Evaluation of effectiveness during simulated use was performed for the Proposed Clearing Stem Models and compared to the effectiveness of the Predicate Clearing Stem Models. They were found to be of equivalent effectiveness.

5.8.3: Safety Testing

Safety testing in terms of Tube damage (e.g., scratches, nicks, tears, abrasions, punctures) to the inner tube surface after device operation was found to be equivalent for the Proposed and Predicate Devices. Heating during device operation was negligible and equivalent for the Proposed and Predicate Devices, thus the Devices have equivalent safety in terms of heating. Tube movement was measured and found to be of the same magnitude, thus equivalent for both

{6}------------------------------------------------

Actuata Madical com

K131052 pg 7 of 7

Devices. The potential for Tube damage, Tube heating and Tube movement during use of TubeClear was found to be equivalent for the Proposed and Predicate Clearing Stems.

5.9: CLINICAL PERFORMANCE DATA

No clinical data was collected; therefore, no clinical data is presented in this submission.

5.10: CONCLUSIONS

The scientific data demonstrates that the Proposed Clearing Stem Models are substantially equivalent to the Predicate Clearing Stem Models. The 510(k) Substantial Equivalence Decision Making Process Flow Chart was used by Actuated Medical, Inc. to determine Substantial Equivalence.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 16, 2013

Actuated Medical, Inc. % Jeffrey Rongero Senior Project Engineer UL LLC 12 Laboratory Drive Research Triangle Park, NC 27709-3995

Re: K131052

Trade/Device Name: TubeClear™ Regulation Number: 21 CFR §876.5980 Regulation Name: Tubes, Gastrointestinal and Accessories Regulatory Class: Class II Product Code: KNT, FPD Dated: July 31, 2013 Received: August 1, 2013

Dear Jeffrey Rongero,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{8}------------------------------------------------

Page 2 - Jeffrey Rongero

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

4.0 Indications for Use

510(k) Number (if known): K131052

Device Name: TubeClear

Indications for Use Statement

TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.

The Clearing Stem Models are indicated for use as follows:

  • NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
  • NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
  • NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
  • NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
  • NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).
  • NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
  • NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
  • G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
  • G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
  • G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 25 cm (10 in).
  • G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).

Page 1 of 2

4-1

{10}------------------------------------------------

  • G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
  • G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).

Prescription Use __X (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin R. Fisher -S 2013.08.16 15:02:28 -04'00'

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K131052 510(k) Number

Page 2 of 2

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.