K Number
K131047

Validate with FDA (Live)

Date Cleared
2013-10-16

(184 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fingertip Pulse Oximeter MD300CB3 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent, child and infant patient in hospital.

Device Description

The applicant device of Fingertip Pulse Oximeter MD300CB3 is a battery powered fingertip device, which can detect and display the measured %SpO2 and pulse rate value, pulse bar graph and SpO2 waveform. The device is normally applied to adult, adolescent, child and infant patient in hospital. The applicant device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit. The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The applicant devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products. The devices are software -driven and the software validation is provided in Software.

AI/ML Overview

This document describes the FDA 510(k) submission for the Fingertip Pulse Oximeter MD300CB3. The acceptance criteria and the study used to meet them are detailed below based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Predicate)Reported Device Performance (MD300CB3)
SpO2 Accuracy (Arms)70%~99%: ±3%70%~100%: 1.94
70%~80%: 2.32
80%~90%: 1.95
90%~100%: 1.57
0%-69% no definition0%-69% no definition
SpO2 Resolution1% (implicitly matched)1%
PR Display RangeNot explicitly stated for predicate but matched0-254bpm
PR Measurement RangeNot explicitly stated for predicate but matched30~235bpm
PR AccuracyNot explicitly stated for predicate but matched±2bpm (30-99bpm) and 2% (100-235bpm)
PR Resolution1bpm (implicitly matched)1bpm

Note: The predicate device's specific SpO2 accuracy is given as a single value (±3% for 70-99%). The applicant device provides a breakdown across different ranges, all of which fall below or within the ±3% range, indicating superior or equivalent accuracy in those segments. The "no definition" for 0-69% SpO2 is a shared characteristic. Other performance metrics for the predicate (PR accuracy, ranges, resolutions) are not explicitly stated with numerical values in the "Comparison with the Predicate Device" table for direct comparison, but the document states "The applicant device has similar device specifications as the predicate device," implying they meet equivalent performance standards.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a number of subjects. The document mentions "clinical test for device accuracy is conducted in the Yue Bei people's Hospital." The specific number of patients or recordings used to determine the SpO2 accuracy (Arms) values (1.94, 2.32, 1.95, 1.57) is not provided in this summary.
  • Data Provenance: Prospective clinical study conducted in Yue Bei people's Hospital, P.R. China.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the given text. The ground truth ("Golden Standard" Co-Oximeter) is directly measured by an instrument, not established by human experts.

4. Adjudication Method for the Test Set

  • This information is not applicable as the ground truth is established by a "Golden Standard" Co-Oximeter, not through human expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • A MRMC comparative effectiveness study was not performed. This device is a standalone measurement device (pulse oximeter), not an AI-assisted diagnostic tool that would involve human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone performance evaluation was conducted for the device. The clinical study directly assessed the device's accuracy against a "Golden Standard" Co-Oximeter without human interpretation being part of the measurement process. The device's SpO2 and PR measurements are automatically generated.

7. The Type of Ground Truth Used

  • Instrumental Ground Truth: The ground truth for the clinical accuracy test was established using a "Golden Standard" Co-Oximeter, as stated: "accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter."

8. The Sample Size for the Training Set

  • This device does not appear to employ machine learning or AI models that would require a separate "training set" in the conventional sense. It is a traditional medical device based on established physical principles. Therefore, this information is not applicable/not provided.

9. How the Ground Truth for the Training Set was Established

  • As noted above, there is no mention of a training set for machine learning. The device's operation relies on direct physiological measurement and signal processing algorithms, not a learned model from a training set. Therefore, this information is not applicable/not provided.

{0}------------------------------------------------

Section 3 510(k) Summary

(As required by 21 CFR 807.92)

This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

There is no prior submission for the devices.

The assigned 510(K) number is K131047

3.1 Date of Submission: Sept. 10, 2013

OCT 1 6 2013

3.2 Sponsor Information

Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 320, West Building 4, No.83 Fuxing Road, Beijing 100039, P.R.China

Contact Person:

Mr. Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No.9 Shuangyuan Road, Badachu Hi -tech Zone, Shijingshan District, Beijing. P.R. China, 100041 Phone: +86-10-88798300 Ext 6020 Fax: 215-4052545 Email: cc@choicemmed.com

3.3 Proposed Device Information

Device Common or Usual Name: Pulse Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300CB3 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700 Panel: Anesthesiology Class: Il Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Intended Use : The Fingertip Pulse Oximeter MD300CB3 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and

{1}------------------------------------------------

pulse rate of adult, adolescent, child and infant patient in hospital.

3.4 Predicate Device

510(k) Number: K070371 Common Name: Oximeter Device Trade or Proprietary Name: Fingertip Pulse Oximeter Model: MD300C Classification Name: Oximeter Device Class: II Product Code: DQA Regulation Number: 870.2700 Review Panel: Anesthesiology Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Intended Use : Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check,

oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.

3.5 Device Description

The applicant device of Fingertip Pulse Oximeter MD300CB3 is a battery powered fingertip device, which can detect and display the measured %SpO2 and pulse rate value, pulse bar graph and SpO2 waveform. The device is normally applied to adult, adolescent, child and infant patient in hospital.

The applicant device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.

The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The applicant devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products.

The devices are software -driven and the software validation is provided in Software.

{2}------------------------------------------------

3.6 Intended Use

The Fingertip Pulse Oximeter MD300CB3 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent, child and infant patient in hospital.

3.7 Contraindication

The Fingertip Pulse Oximeter MD300CB3 is not for continuous monitoring.

{3}------------------------------------------------

Premarket Notification 510(k) Submission -- Sec 3 510(k) Summar

3.8 Comparison with the Predicate Device

{4}------------------------------------------------

remarket Notification 510(k) Submission - Sec 3 510(k) Summary
Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
MeasurementRed660±2nm
wavelengthInfrared940±10nm
Comparison StatementThe proposed devices have the same design principle and similar components.
Device SpecificationDisplay TypeOLED
Working timeWork for 8 hours continuously
Brightness of backlightAdjustable
User Interface6 directions for display
Power supply1*AAA alkaline battery
Display DataSpO2、PR
SpO2 display range0~100%
SpO2 measurement range70~100%
70%~100%: 1.94;
70%~80%: 2.32;
SpO2 Arms Accuracy80%~90%: 1.95;
90%~100%: 1.57;
0%~69% no definition
Display TypeOLED
Working timeApproximately 30 hours of continuous operation
Brightness of backlightAdjustable
User Interface6 directions for display
Power supply2*AAA alkaline battery
Display DataSpO2、PR
SpO2 display range0~99%
SpO2 measurement range70~99%
70%~99%: ±3%
0%-69% no definition

n - Sec 3 510(k) Su Premarket Notification 510(k) Sub

{5}------------------------------------------------

market Notification 510(k) Submission - Sec 3 510(k) Summary
SpO2 resolution1%
PR Display Range0-254bpm
PR Measurement Range30~235bpm
PR Accuracy$\pm$ 2bpm (30-99bpm) and 2% (100-235bpm)
PR resolution1bpm
Operating temperature5°C~40°C
Relative humidity≤80%, no condensation (operating)≤93% no condensation (storage)
Atmosphere pressure86kPa~106kPa
Comparison StatementThe applicant device has similar device specifications as the predicate device.
Battery coverABS
ContactingMaterialFingertipMedical Silicon gel
CushionMedical Silicon gel
EnclosureABS
Comparison StatementThe contacting materials of applicant device are similar to that of the predicate device.
Performance TestingBench TestThe bench tests include Test for Pulse signal strengthIndicator Bar、Pulse rate and SpO2 accuracy test after 3000cycles disinfection, Test for Random vibration, wide band Meet the requirements of FDA Guidanceand Test according to ISO9919. All the bench test results areprovided in Performance Testing-BenchConformed to ISO 9919
Clinical TestClinical test for device accuracy is conducted in the Yue Beipeople's Hospital. The clinical test report and protocol areprovided in Performance Tasting ClinicalConformed to ISO 9919

Premarket Notification 510(k) Submission – Sec 3 510(k) Summary

ﺑﺮ

{6}------------------------------------------------

Electrical SafetyElectromagneticCompatibilitySoftwareBiocompatibilityLabel and Labeling
Conformed to IEC60601-1.The test results are provided in ElectromagneticCompatibility and Electrical SafetyConformed to IEC60601-1-2.The EMC test reports were provided in ElectromagneticCompatibility and Electrical SafetyModerate Level of ConcernMedical silicone gelConformed to IEC60601-1
Compliance with FDA Guidance for the Content ofPremarket Submissions for Software Contained in MedicalDevices.Conformed to IEC60601-1-2.
Risk Management in Compliance with IEC 60601-1-4Moderate Level of Concern
In Vitro CytotoxicityCompliance with FDA Guidance for the Content ofPremarket Submissions for Software Contained in MedicalDevices.
Skin Irritation TestRisk Management in Compliance with IEC 60601-1-4
Animal skin irritationtestIn Vitro Cytotoxicity
No cytotoxic potential
No evidence of causingsensitization
No evidence of significantirritation from the test extract torabbits
Compliance with FDA guidance
Compliance with the Guidance of pulse oximeter-premarketnotification submission issued on March 4,2013
Skin Irritation Test
No cytotoxic potential
No evidence of causingsensitization
No evidence of significantirritation from the test extractto rabbits
Animal skin irritation test

remarket Notification 510(k) Submission – Sec 3 510(k) Summar

{7}------------------------------------------------

3.9 Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1:1988 + A1(1991) + A2(1995), Medical Electrical Equipment - Part1: General requirements for safety.

IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.

ISO 9919:2005. Medical electrical equipment - Particular requirements for basic safety and essential performance of pulse oximeter equipment for medical use.

ISO10993-5: 2009, Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity.

ISO10993-10: 2010, Biological evaluation of medical device - Part 10: Tests for irritation and delayed-type hypersensitivity.

The Clinical Test of MD300CB3 following ISO 9919:2005, Annex EE.4 was conducted in Yue Bei people's Hospital. The study protocol was subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE2 were adopted. It can be determined from the result of the study that the accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter.

3.9 Substantially Equivalent Conclusion

The proposed device, Fingertip Pulse Oximeter MD300CB3, is determined to be Substantially Equivalent (SE) to the predicate device, Fingertip Pulse Oximeter MD300C, K070371, in respect of safety and effectiveness.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2013

Beijing Choice Electronic Technology Co., Ltd Mr. Lei Chen 3270 Alpine Road North Building 3F, No.9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District. Beijing, P.R. China, 100041

Re: K131047

Trade/Device Name: Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: September 13. 2013 Received: September 16, 2013

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{9}------------------------------------------------

Page 2 - Mr. Lei Chen

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/Mcdicall.evices/Safety/ReportalProblem/default.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.

Sincerely yours,

Image /page/9/Picture/8 description: The image contains a logo or signature on the left side, which appears to be stylized and possibly handwritten. To the right of the logo, there is some text, including the words "Tejashri" and "Clinical D". Below that, the word "DAGRID" is written in a larger font size.

Teiashri Purohit-Sheth, M.D. Clinical Deputy Director

FOR

Kwame Ulmer, M.S. Acting Director Division of Anesthesiology, Respiratory, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Section 2 Indications for Use Statement

Indications for Use

510(k) Number (if known): K131047

Device Name: Fingertip Pulse Oximeter MD300CB3

Indications for Use:

The Fingertip Pulse Oximeter MD300CB3 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpOz) and pulse rate of adult, adolescent, child and infant patient in hospital.

V Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

(21 CFR 801 Subpart C)

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nayan J. Patel -S 2013.10.15 13:01:56 -04'00'

Page_1_of_1

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).