K Number
K131033

Validate with FDA (Live)

Date Cleared
2014-02-12

(306 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).
How the device functions: PVC films form a barrier to body fluids and bloodborne Pathogens
Scientific concepts that form the basis for the device The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.
Physical and performance characteristics such as design, materials and physical properties: PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

AI/ML Overview

{"acceptance_criteria_and_performance": "| Features &Description | Acceptance Criteria (Predicate Device) | Reported Device Performance (Subject Device) | Result ofComparison |\n|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|\n| Device Name | Powder Free Vinyl PatientExamination Gloves, Clear(Non-colored) | Powder Free Vinyl PatientExamination Gloves, Clear(Non-colored) | Same |\n| Product Code | LYZ | LYZ | Same |\n| Size | Small/ Medium/Large/X large | Small/ Medium/Large/X large | Substantiallyequivalent |\n| Intended Use | Powder free Vinyl PatientExamination Gloves, Clear(Non-colored)is a disposable deviceintended for medical purposesthat is worn on the examiner'shand or finger toprevent contamination betweenpatient and examiner. | Powder free Vinyl PatientExamination Gloves, Clear(Non-colored) is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand or finger toprevent contamination betweenpatient and examiner. | Substantiallyequivalent |\n| Device Descriptionand Specifications | Meets ASTM D5250-06(Reapproved 2011) | Meets ASTM D5250 -06(Reapproved 2011) | Substantiallyequivalent |\n| Dimensions-- Length | Meets ASTM D5250-06(Reapproved 2011)≥230mm min. | 230mm min for all sizes | Substantiallyequivalent |\n| Dimensions-- Width (Small) | 80-90 mm | 80-85 mm | Substantiallyequivalent |\n| Dimensions-- Width (Medium) | 90-100mm | 95-97 mm | Substantiallyequivalent |\n| Dimensions-- Width (Large) | 100-110mm | 102-108mm | Substantiallyequivalent |\n| Dimensions-- Width (X-Large) | 110-120 mm | 114-118 mm | Substantiallyequivalent |\n| -- Thickness (Finger) | 0.05mm min. | 0.05mm min. | Substantiallyequivalent |\n| -- Thickness (Palm) | 0.08mm min. | 0.08mm min. | Substantiallyequivalent |\n| Physical Properties (Elongation) | ≥300% | ≥300% | Substantiallyequivalent |\n| Physical Properties (Tensile Strength) | ≥14MPa | ≥14MPa | Substantiallyequivalent |\n| Freedom from Pinholes | Meets ASTM D5151-06(Reapproved 2011) and 21 CFR 800.20Inspection Level 1, AQL 2.5 | Meets ASTMD5151-06(Reapproved 2011)Inspection Level 1, AQL 2.5 | Substantiallyequivalent |\n| Residual Powder | Meets ASTM D 6124-06(Reapproved 2011) | Results generated values below2mg of residual powder | Substantiallyequivalent |\n| Materials used | PVC | PVC | Substantiallyequivalent |\n| Dusting or DonningPowder (substance) | PU | PU | Substantiallyequivalent |\n| Dusting or DonningPowder (name) | PU | Surface Coating Agent | Substantiallyequivalent |\n| Performance Data | Meets ASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reapproved 2011) | Meets ASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reapproved 2011) | Substantiallyequivalent |\n| Single Patient Use | Single Patient Use | Single Patient Use | Substantiallyequivalent |\n| Biocompatibility | SKIN IRRITATION DERMALand SENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006 | The test article was a non-irritant or non- sensitizer.SKIN IRRITATION DERMALand SENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006 | Substantiallyequivalent |\n| Labeling | -Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot | -Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot | Substantiallyequivalent |\n\nStudy proving device meets acceptance criteria:\nThe information provided is a 510(k) Summary for a medical device (Powder Free Vinyl Patient Examination Gloves). The study described is a nonclinical assessment for substantial equivalence to a predicate device. \n\n2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):\nThe document does not specify a separate "test set" sample size in the conventional sense of a clinical or imaging study. The "testing" refers to nonclinical performance tests conducted to demonstrate compliance with recognized consensus standards (ASTM and ISO). The data provenance is not explicitly stated in terms of country of origin of the data, but the submitter is Decent Plastic Co., Ltd. from China. The study is a nonclinical assessment.\n\n3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):\nThis information is not applicable. The device is a patient examination glove, and the evaluation is based on nonclinical performance testing against established standards, not on expert interpretation of observational data.\n\n4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:\nThis information is not applicable, as there is no traditional "test set" and expert adjudication in the context of this nonclinical assessment.\n\n5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:\nThis information is not applicable. The device is a patient examination glove, not an AI diagnostic tool, so an MRMC study is not relevant.\n\n6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:\nThis information is not applicable. The device is a patient examination glove, not an algorithm.\n\n7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):\nThe "ground truth" for this device, in the context of demonstrating substantial equivalence, is compliance with established international and national consensus standards. Specifically, the device was evaluated against:\n* ASTM D5250-06 (Reaffirmation 2011) for physical and performance characteristics.\n* ASTM D5151-06 (Reaffirmation 2011) for freedom from pinholes.\n* ASTM D6124-06 (Reaffirmation 2011) for residual powder.\n* 21 CFR 800.20 for water leak test on pinhole AQL.\n* ISO 10993-10: 2002/Amd. 1:2006(E) for biocompatibility (skin irritation and sensitization studies).\n\n8. The sample size for the training set:\nThis information is not applicable. The device is not an AI algorithm that requires a training set.\n\n9. How the ground truth for the training set was established:\nThis information is not applicable, as the device is not an AI algorithm and does not have a training set.", "additional_information": "The statement explicitly mentions that "Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.""}

{0}------------------------------------------------

Section C

510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is:___ K131033__" (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared:

Submitter's name :Decent Plastic Co., Ltd.
Submitter's address :No.44, Dajing St., Shijiazhuang, 050000, China
Phone number :(86) 311-67699616
Fax number :(86) 311-67699906
Name of contact person:Ms. Qi Fang
Date the summary was prepared:2013-12-31

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)
Proprietary/Trade name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)

Product Code: LYZ [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06 (Reaffirmation 2011).

Predicate device: Powder free Vinyl Patient Examination Gloves, Clear (Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908.

{1}------------------------------------------------

[(a)(4)| A description of the device

Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).

-- How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

-- Scientific concepts that form the basis for the device The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

-- Physical and performance characteristics such as design, materials and physical properties:

PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

[(a)(5)| The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or linger to prevent contamination between patient and examiner.

[{a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

Features &DescriptionPredicate DeviceSubject DeviceResult ofComparison
CompanyShijiazhuang Fuguan PlasticProducts Co., Ltd.Decent Plastic Co., Ltd.--
510(K) NumberK032908K131033--
Product namePowder Free Vinyl PatientExamination Gloves, Clear(Non-colored)Powder Free Vinyl PatientExamination Gloves, Clear(Non-colored)Same
Product CodeLYZLYZSame
SizeSmall/ Medium/Large/X largeSmall/ Medium/Large/X largeSubstantiallyequivalent
Intend for usePowder free Vinyl PatientExamination Gloves, Clear(Non-colored)is a disposable deviceintended for medical purposesthat is worn on the examiner'shand or finger toprevent contamination betweenpatient and examiner.Powder free Vinyl PatientExamination Gloves, Clear(Non-colored) is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand or finger toprevent contamination betweenpatient and examiner.Substantiallyequivalent
Device Descriptionand SpecificationsMeets ASTM D5250-06(Reapproved 2011)Meets ASTM D5250 -06(Reapproved 2011)Substantiallyequivalent
Dimensions-- LengthMeets ASTM D5250-06(Reapproved 2011)≥230mm min.230mm min for all sizesSubstantiallyequivalent
Dimensions-- WidthMeets ASTM D5250-06(Reapproved 2011)Substantiallyequivalent
Small 80-90 mmMedium 90-100mmLarge 100-110mmX large 110-120 mmSmall 80-85 mmMedium 95-97 mmLarge 102-108mmX large 114-118 mm
-- Thickness(Reapproved 2011)
Finger 0.05mm min.Palm 0.08mm min.Finger 0.05mm min.Palm 0.08mm min.
Physical PropertiesMeets ASTM D5250-06(Reapproved 2011)Before aging/after agingSubstantiallyequivalent
Before aging/after agingElongation ≥300%Tensile Strength≥14MPaElongation ≥300%Tensile Strength≥14MPa
FreedomfromMeetsMeets ASTMD5151-06(Reapproved 2011)Substantiallyequivalent
Pinholes• 21 CFR 800.20• ASTM D5250-06(Reapproved 2011)• ASTM D 5151-06(Reapproved 2011)HolesInspection Level 1AQL2.5
Residual PowderMeets ASTMD 6124-06(Reapproved 2011)D 6124-06(Reapproved 2011)Substantiallyequivalent
Results generated values below2mg of residual powder
Compare allmaterials used tofabricate the devicesPVCPVCSubstantiallyequivalent
Dusting or DonningPowder:PUPUSubstantiallyequivalent
Dusting or DonningPowder: namePUSurface Coating AgentSubstantiallyequivalent
Compareperformance datasupportingsubstantialequivalenceMeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reapproved 2011)MeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reapproved 2011)Substantiallyequivalent
Single Patient UseSingle Patient UseSingle Patient UseSubstantiallyequivalent
BiocompatibilitySKIN IRRITATION DERMALand SENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006The test article was a non-irritant or non- sensitizer.SKIN IRRITATION DERMALand SENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006Substantiallyequivalent
Labeling for thelegally marketeddevice to whichsubstantialequivalence isclaimed-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- LotSubstantiallyequivalent

{2}------------------------------------------------

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)meet requirements per ASTM D5250-06(Reaffirmation 2011), per ASTM D6124-06(Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10: 2002/Amd. 1:2006(E).

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

Section C(rev.02)

{3}------------------------------------------------

((b)(3)) The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims and the Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is as safe, as effective, and performs as well as the predicate device, Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 .

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2014

Decent Plastic Co., Ltd. C/O Mr. Chu Xiaoan Official Correspondent Room 1606 Bldg.1 Jianxiang Yuan No.209 Bei Si Huan Zhong Road Haidian District Beijing, 100083 CHINA

Re: K131033

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: December 31, 2013 Received: January 8, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/5 description: The image shows a logo with the letters "FDA" in a stylized font. Next to the logo is the text "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID". Below the logo is the word "FOR".

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K 131033

Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)

Indications for Use (Describe)

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

[x] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/13 description: The image shows the name "Elizabeth F. Claverie -S" in bold, black font. Below the name is the date "2014.02.11" followed by the time "16:48:21" and timezone "-05'00'" also in bold, black font. The text appears to be extracted from a document or record.

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.