K Number
K130998

Validate with FDA (Live)

Date Cleared
2014-01-03

(268 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

syngo.via WebViewer is a software-only device indicated for reviewing medical images from syngo via. It supports interpretation and evaluations within healtheare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images.

syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healtheare institution IT infrastructure where the customer will insure HIPAA compliance. The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM.

The system is not intended for the display of digital mammography images for diagnosis.

Device Description

The syngo.via WebViewer is a software-based Picture Archiving and Communications System (PACS) used with the syngo.via system. The syngo.via WebViewer provides secure access to rendered medical image data and basic image manipulation through web browsers and mobile devices within the reach of the hospital network.

It extends the syngo.via WebViewer software application previous cleared under K111079. New image types supported are PET and X-Ray images. It also provides functionality for displaying images via a mobile application on an iPad.

AI/ML Overview

The Siemens syngo.via WebViewer (K130998) is a PACS viewing software. The provided document does not contain acceptance criteria or a study that directly proves the device meets specific performance criteria through metrics like sensitivity, specificity, or accuracy for diagnostic tasks. Instead, the submission focuses on establishing substantial equivalence to a predicate device (syngo.via WebViewer K111079) based on its intended use, technical characteristics, and the results of non-clinical software verification and validation.

1. Table of Acceptance Criteria and Reported Device Performance:

As the device is a PACS viewing software, the acceptance criteria are not typically expressed in terms of diagnostic performance metrics (e.g., sensitivity, specificity) but rather in terms of functional performance, adherence to standards, and safety. The document states that "software verification and validation (Unit Test Level, Integration Test Level and System Test Level) was performed for all newly developed components and the complete system according to the following standards." The table below summarizes the implied acceptance criteria from the non-clinical tests and the device's adherence:

Acceptance Criterion (Implied from Standards & V&V)Reported Device Performance
Adherence to DICOM StandardSoftware verification and validation performed
Adherence to ISO/IEC 15444-1:2005+TC 1:2007 (JPEG 2000)Software verification and validation performed
Adherence to ISO/IEC 10918-1:1994 + TC 1:2005 (JPEG)Software verification and validation performed
Adherence to HL7 [2006]Software verification and validation performed
Adherence to IEC 62304:2006 (Medical device software)Software verification and validation performed
Adherence to IEC 62366:2007 (Usability)Software verification and validation performed
Adherence to ISO 14971:2007 (Risk Management)Software verification and validation performed; Risk analysis performed to identify potential hazards
Adherence to IEC 60601-1-4:2000 (Safety)Software verification and validation performed
Secure access to rendered medical image dataEnsured via syngo.via WebViewer Data Management
Basic image manipulation functionalityProvided via web browsers and mobile devices
Compatibility with supported image types (CT, MR, CR, DR, DX, PET)DICOM formatted images supported
Compatibility with connected medical IT systems via standard interfaces (e.g., DICOM, HL7)Communication via standard interfaces
System safety and effectivenessInstructions for use, cautions, and warnings in labeling; Risk management process followed
Substantial equivalence to predicate deviceConfirmed through comparison of intended use and technical characteristics

2. Sample Size Used for the Test Set and Data Provenance:

The document does not describe a clinical study or a test set of medical images with a specific sample size used to evaluate diagnostic performance. The validation mentioned is "software verification and validation (Unit Test Level, Integration Test Level and System Test Level)", which refers to engineering and software quality assurance testing rather than a clinical performance study using patient data. Therefore, there is no information on sample size or data provenance (e.g., country of origin, retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This level of detail is not provided as the submission focuses on software validation and substantial equivalence, not a clinical study involving ground truth establishment by experts for diagnostic performance.

4. Adjudication Method for the Test Set:

Not applicable, as no clinical test set for diagnostic performance requiring expert adjudication is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned or appears to have been performed for this 510(k) submission. The document focuses on showing substantial equivalence to a predicate device rather than demonstrating a performance improvement with or without AI assistance.

6. Standalone Performance Study:

No standalone (algorithm only without human-in-the-loop performance) study is described, as the device is a medical image viewing software, not an AI diagnostic algorithm. The "software-only device" refers to its deployment model, not its functionality as an autonomous diagnostic tool.

7. Type of Ground Truth Used:

Ground truth as understood in the context of diagnostic accuracy studies (e.g., pathology, expert consensus) is not mentioned. The "ground truth" for the software validation activities would be the expected functional behavior and adherence to standards, checked against specified requirements.

8. Sample Size for the Training Set:

Not applicable. The device is a viewing software, not an AI/ML algorithm that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as no training set is relevant for this type of device.

{0}------------------------------------------------

K 130998
JAN - 3 2014

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

I. GENERAL INFORMATION

Establishment:

Address:Siemens AG, Medical SolutionsHenkestrasse 127D-91052 ErlangenGermany
Registration Number:3002808157
Contact Person:Joerg TeicheTelephone: +49 (9131) 84-4687Telefax: +49 (9131) 84-8691

Device Name and Classification:

• Trade Name:syngo.via WebViewer
• Classification Name:Picture Archiving and Communications System
• Classification Panel:Radiology
• CFR Section:21 CFR §892.2050
• Device Class:Class II
• Product Code:LLZ

Date of submission:

.

.

April 2013

510(k) for syngo®.via WebViewer

December 31, 2013 - '

Page 1 of 5

.

{1}------------------------------------------------

II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION

Indications for Use:

syngo.via WebViewer is a software-only device indicated for reviewing medical images from syngo via. It supports interpretation and evaluations within healtheare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images.

syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healtheare institution IT infrastructure where the customer will insure HIPAA compliance. The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM.

The system is not intended for the display of digital mammography images for diagnosis.

Device Description:

The syngo.via WebViewer is a software-based Picture Archiving and Communications System (PACS) used with the syngo.via system. The syngo.via WebViewer provides secure access to rendered medical image data and basic image manipulation through web browsers and mobile devices within the reach of the hospital network.

It extends the syngo.via WebViewer software application previous cleared under K111079. New image types supported are PET and X-Ray images. It also provides functionality for displaying images via a mobile application on an iPad.

syngo.via WebViewer Data Management:

...ensures all authorized personnel fast and continuous access to radiological data. It's main functionality ranges from availability of images with regard to data sceurity, open interfaces.

Integration:

The Workflow Management enables by integration of any HL7- / DICOM-compatible RIS (IHE Year 5) to the syngo product family consistent workflow within the healthcare organization.

Technological Characteristics:

syngo.via WebViewer server part is a "software only"-system, which will be delivered on CD-ROM / DVD to be installed on common 1T hardware. This hardware has to fulfil the defined requirements. The Software will be installed by Siemens service engineers only.

The backend communication and storage solution is based on Windows 2008 operating system. The client machines are based on standard computer (Mac / Windows PC / Linux PC). The web

510(k) for syngo®.via WebViewer

December 31, 2013

Page 2 of 5

{2}------------------------------------------------

and mobile client application runs in a standard web browser - such as but not limited to Internet Explorer, Firefox, Safari - please refer to the specification for the complete list of supported web browsers.

Any hardware platform, which complies to the specified minimum hardware and software requirements and with successful installation verification activities, can be supported.

The herewith described syngo.via WebViewer supports DICOM formatted images (CT, MR. CR, DR, DX, PET) and objects (SC, pdf).

The syngo.via WebViewer will be marketed as a software only solution for the end-user (with recommended hardware requirements). The server part will be installed by trained service engineers only. Any special needs such as integration in a specific environment and undates / upgrades will be covered by individual service contract and fulfilled by special trained service technicians.

Summary of Non-Clinical Tests

The software verification and validation (Unit Test Level. Integration Test Level and System Test Level) was performed for all newly developed components and the complete system according to the following standards:

  • DICOM Standard [2011] -
  • ISO/IEC 15444-1:2005+TC 1:2007 -
  • ISO/IEC 10918-1:1994 + TC 1:2005 -
  • HL7 [2006] -
  • IEC 62304:2006 -
  • -IEC 62366:2007
  • -ISO 14971:2007
  • -IEC 60601-1-4:2000

General Safety and Effectiveness Concerns:

The device labelling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize hazards, Siemens adheres to recognized and established industry practice and standards.

Substantial Equivalence:

The syngo.via Web Viewer, addressed in this premarket notification, is substantially equivalent to the following commercially available device:

December 31, 2013

Page 3 of 5

{3}------------------------------------------------

ManufacturerPredicate Device NameFDA ClearanceNumber
Siemenssyngo.via WebViewerK111079

The syngo.via Web Viewer described in this 510(k) has similar intended use and similar technical characteristics as the device listed below in regard to the specific functionalities.

Functionalitysyngo.via WebViewer VAI1syngo.via WebViewer VA10A
Manufactureridenticalidentical
Intended usesyngo.via WebViewer is a software-only device indicated for reviewingmedical images from syngo.via.It supports interpretation and evalu-ation of examinations withinhealthcare institutions, for example,in Radiology, Nuclear Medicineand Cardiology environments (sup-ported Image types: CT, MR, CR,DR, DX, PET). It is not intendedfor storage or distribution of medi-cal images.syngo.via WebViewer is an optionfor the syngo.via system and cannotbe run without it. It is client serverarchitecture and the client is in-tended to run on web clients whichare connected to the healthcare in-stitution IT infrastructure where thecustomer will insure HIPAA com-pliance.The communication of syngo.viaWebViewer with connected medi-cal IT systems will be done viastandard interfaces such as but notlimited to DICOM.The system is not intended for thedisplay of digital mammographyimages for diagnosis.syngo.via WebViewer is intended tobe a software-only solution for re-viewing medical images from syn-go.via.The system cannot be used as stand-alone device. It is intended to be anoption for syngo.via system only.syngo.via WebViewer is not intendedfor storage or distribution of medicalimages from one medical device toanother.syngo.via WebViewer is a clientserver architecture and the client isintended to run on web clients whichare connected to the healthcare insti-tution IT infrastructure where thecustomer has to ensure HIPAA com-pliance.syngo.via WebViewer supports inter-pretation and evaluation of examina-tions within healthcare institutions,for example, in Radiology, NuclearMedicine and Cardiology environ-ments.The communication of syngo.viaWebViewer with connected medicalIT systems will be done via standardinterfaces such as but not limited toDICOM.The system is not intended for the

Device Comparison Table between new device and predicate device:

510(k) for syngo®.via WebViewer

December 31, 2013

{4}------------------------------------------------

Functionalitysyngo.via WebViewer VA11syngo.via WebViewer VA10A
displaying of digital mammographyimages for diagnosis in the U.S
Imagecom-municationidenticalidentical
ImageProcessing andEvaluationidenticalidentical
Supported Im-age TypesMR, CT, SC, PDF, PET, CR, DXMR, CT, SC, PDF
Imagedatacompressionidenticalidentical
User admini-strationUI is syngo®-based web client andmobile clientUI is syngo®-based web client
UserInterfaceidenticalidentical
HardwareClient: PC with Windows Systemor MAC with Mac iOS Server:Windows Server 2008 SP1iPad (with iOS 5.0 or 5.1)Client: PC with Windows System,MAC with Mac OS Server: WindowsServer 2008
Imaging Algo-rithmsMPR, MIP, VRT, PET/CT, SUVMPR, MIP, VRT
QuantitativeAlgorithmsidenticalidentical

In summary, Siemens is of the opinion that syngo.via WebViewer does not introduce any new significant potential safety risks and is substantially equivalent to and performs as well as the predicate device.

.

:

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract shapes or lines, possibly representing human figures or abstract forms.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

January 3, 2014

Siemens AG, Medical Solutions % Ms. Sabine Schroedel Regulatory Affairs Manager Henkestrasse 127 Erlangen, Bavaria 91052 GERMANY

Re: K130998

Trade/Device Name: syngo.via WebViewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 19, 2013 Received: December 20, 2013

Dear Ms. Schroedel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K130998

Device Name

syngo®.via WebViewer

Indications for Use (Describe)

syngo.via WebViewer is a software-only device indicated for reviewing medical innages from syngo.via. It supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments (supported Image types: CT, MR, CR, DR, DX, PET). It is not intended for storage or distribution of medical images.

syngo.via WebViewer is an option for the syngo.via system and cannot be run without it. It is client server architecture and the client is intended to run on web clients which are connected to the healtheare institution IT infrastructure where the customer will insure IIIPAA compliance.

The communication of syngo.via WebViewer with connected medical IT systems will be done via standard interfaces such as but not limited to DICOM.

The system is not intended for the display of digital manimography images for diagnosis.

Type of Use (Select one or both, as applicable)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Smh.7)

FORM FDA 3881 (9/13)

PSC Publishing Services (101) 461-4740 EG

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{8}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

N/A