K Number
K130963

Validate with FDA (Live)

Date Cleared
2013-05-06

(28 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Liquichek Pediatric Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Device Description

Liquichek Pediatric Control is prepared from bovine serum with added chemicals. therapeutic drugs, preservatives and stabilizers. The control is provided in liquid form for convenience.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Liquichek Pediatric Control device:

The provided K130963 document is a 510(k) Premarket Notification for a quality control material, not a diagnostic device that performs clinical measurements or generates diagnoses. Therefore, the typical acceptance criteria and study designs (like standalone performance, MRMC studies, expert ground truth for clinical decision-making, large training sets, etc.) for AI/ML-driven diagnostic devices are not applicable.

This document describes a substantial equivalence claim for a laboratory control. The "performance" being evaluated is its stability and its ability to serve as a reliable control material over time.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are implicitly related to the stability claims, ensuring the control material remains consistent and effective for its stated duration. The "performance" is the successful demonstration of these stability periods.

Acceptance Criteria (Implicit)Reported Device Performance (Claims)
Maintained stability when thawed and opened in use.Thawed and Opened Stability: 14 days at 2 to 8°C
Maintained stability when thawed and unopened.Thawed and Unopened Stability: 45 days at 2 to 8°C or on-board Dimension Vista at 2 to 8°C
Maintained stability/shelf life in long-term storage.Shelf Life Stability: 3 years at -20 to -50°C
Equivalent characteristics to the predicate device.Demonstrated similarity in intended use, form, matrix, and thawed & opened stability to K872227. Differences in fill volume, thawed & unopened stability duration, storage temperature, and specific analytes were noted but deemed acceptable for substantial equivalence due to the stability studies.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "replicate analyses" and "representative sampling of this lot of product" for value assignment and stability studies. However, specific sample sizes (number of vials, number of runs, etc.) for the stability studies are NOT provided in this public summary.

  • Data Provenance: Not explicitly stated regarding country of origin, but the testing was performed by "the manufacturer and/or independent laboratories." The studies are prospective in nature, as they are "real-time stability studies" and "accelerated stability studies" specifically conducted to establish the claims.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not applicable in the typical sense for a diagnostic device. For this quality control material, the "ground truth" for analytes is established through value assignment and stability testing using validated analytical methods. This involves laboratory instruments and established chemical/biological testing procedures, not interpretation by human clinical experts. The "experts" would be analytical chemists or clinical laboratory scientists operating the testing equipment.

4. Adjudication Method for the Test Set

  • Not applicable. This isn't a scenario requiring adjudication between human readers or algorithms in a diagnostic context. The "adjudication" would be based on the established analytical methods meeting predefined laboratory performance criteria for stability, precision, and accuracy compared to the predicate and internal specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No such study was done. MRMC studies are used for diagnostic devices where human interpretation is involved, often to assess the impact of AI assistance. This device is a quality control material and does not involve human readers making clinical interpretations.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Not applicable in the typical sense. This device is a material used within laboratory testing procedures, not a standalone algorithm. The "performance" is its physical and chemical stability over time, as assessed by analytical instruments, not an AI algorithm's output.

7. Type of Ground Truth Used

  • The "ground truth" for this device relates to the assigned values for the analytes and the stability of these values over time under various storage conditions. This is established through:
    • Validated analytical methods: The analytes' concentrations are determined using standard laboratory instruments and assays.
    • Reference materials/calibrators: Primary calibrators or reference materials would be used to ensure the accuracy of the assigned values.
    • Stability protocols: Real-time and accelerated stability studies are conducted according to predefined protocols to determine how long the analyte values remain within acceptable limits.

8. Sample Size for the Training Set

  • Not applicable. This device is a quality control material, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no AI/ML algorithm and no training set, this question is not relevant to the provided submission.

{0}------------------------------------------------

K130963

510(k) Summary Liquichek Pediatric Control

1.0 Submitter

Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1557 Fax:

Contact Person

Suzanne Parsons Regulatory Affairs Manager Telephone: (949) 598-1467

Date of Summary Preparation

April 2, 2013

2.0 Device Identification

Product Trade Name: Liquichek Pediatric Control Common Name: Multi-Analyte Controls, All Kinds (Assayed) Classifications: Class I, Reserved Product Code: JJY 21 CFR 862.1660 Regulation Number:

3.0 Device to Which Substantial Equivalence is Claimed

Liquichek Pediatric Control Bio-Rad Laboratories Irvine, California

510 (k) Number: K872227

4.0 Description of Device

Liquichek Pediatric Control is prepared from bovine serum with added chemicals. therapeutic drugs, preservatives and stabilizers. The control is provided in liquid form for convenience.

5.0 Value Assignment

The mean values and the corresponding ±3SD ranges printed in this insert were derived from replicate analyses and are specific for this lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer supported reagents and a representative sampling of this lot of product. It is recommended that each laboratory establish its own acceptable ranges and use those provided only as guides. Laboratory established ranges may vary from those listed during the life of this control. Variations over time and between laboratories may be caused by differences in laboratory technique, instrumentation and reagents, or by manufacturer test method modifications.

MAY 0 6 2013

{1}------------------------------------------------

6.0 Intended Use

Liquichek Pediatric Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

7.0 Comparison of the new device with the Predicate Device

Liquichek Pediatric Control claims substantial equivalence to the Liquichek Pediatric Control currently in commercial distribution (K872227). Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.

Table 1: Similarities and Differences Between the New and Predicate Device

CharacteristicsLiquichek Pediatric Control(New Device)Liquichek Pediatric Control(Predicate Device, K872227)
Similarities
Intended UseThis product is intended for use as anassayed quality control serum to monitor theprecision of laboratory testing procedures forthe analytes listed in the package insert.This product is intended for use as anassayed quality control serum to monitor theprecision of laboratory testing proceduresfor the analytes listed in the package insert.
FormLiquidLiquid
MatrixBovine SerumBovine Serum
Thawed & OpenedStability14 days at 2 to 8°C14 days at 2 to 8°C
Differences
Fill Volume2.5 mL4 mL
Thawed andUnopened Stability45 days at 2 to 8°C or on-boardDimension Vista at 2 to 8°C3 months at 2 to 8°C
Storage Unopened(Shelf life)At -20 to -50°C until the expiration dateAt -20 to -70°C until the expiration date
AnalytesContains:Bilirubin, DirectBilirubin, TotalCaffeineCalciumChlorideGlucoseMagnesiumPotassiumSodiumTheophyllineDoes not contain:Bilirubin, IndirectBilirubin, NeonatalPhenylalanineContains:Bilirubin, DirectBilirubin, IndirectBilirubin, NeonatalBilirubin, TotalCaffeineCalciumChlorideGlucoseMagnesiumPhenylalaninePotassiumSodiumTheophylline

{2}------------------------------------------------

8.0 Statement of Supporting Data

Real-time stability studies were conducted to establish the thawed stability claims (opened-vial and unopened-vial). Accelerated stability studies were conducted to establish the shelf-life claims at -20 to -50°C. Product claims are as follows:

Thawed and Opened Stability: Thawed and Unopened Stability: Shelf Life Stability: 14 days at 2 to 8°C 45 days at 2 to 8°C 3 years at -20 to -50°C

9.0 Conclusion

.

Based on the performance characteristics indicated above, the Bio-Rad Liquichek Pediatric Control is substantially equivalent to the predicate device, K872227.

All supporting data are retained on file at Bio-Rad Laboratories.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Mav 6. 2013

Bio-Rad Laboratories C/O Ms. Suzanne Parsons 9500 Jeronimo Road IRVINE CA 92618

Re: K130963

Trade/Device Name: Liquichek Pediatric Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: I, reserved Product Code: JJY Dated: April 02, 2013 Received: April 08, 2013

Dear Ms. Parsons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

· Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K130963

Device Name: Liquichek Pediatric Control

Liquichek Pediatric Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

YungWDChan-S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k):

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.