K Number
K130870

Validate with FDA (Live)

Date Cleared
2013-05-02

(34 days)

Product Code
Regulation Number
862.1145
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The S TEST Reagent Cartridge Calcium (CA) is intended for the quantitative determination of calcium in serum, lithium heparinized plasma, and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Calcium is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Device Description

The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.

The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28.mm (D) × 20.2 mm (H).

System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for Reagent Cartridge Calcium.

Chemistry reactions: Calcium in the sample combines with o-cresolphthalein complexone to form a purple-red complex. The concentration of calcium can be determined by measuring the absorbance of the resulting purple-red color. (With this method, the influence of magnesium is excluded by the addition of 8-hydroxyquinoline.)

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Hitachi S TEST Reagent Cartridge Calcium, based on the provided text:

Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" for each performance characteristic. Instead, it presents the results of various studies (precision, linearity, method comparison, interference, matrix comparison) and implies that these results meet regulatory and performance expectations for substantial equivalence. For the purpose of this analysis, I will infer the implied acceptance by comparing the device performance to the predicate device where applicable, or noting the performance demonstrated without a direct comparative statement if no specific target is given.

Performance CharacteristicImplied Acceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance
Analytical Sensitivity (LOD)Quantitation limit similar to predicate (0.4 mg/dL) and appropriate for clinical use.LOD: 0.19 mg/dL; Quantitation Limit: 0.4 mg/dL
LinearityRange similar to or exceeding predicate (0.4 to 20.0 mg/dL) and sufficient for clinical use.0.4 to 17.1 mg/dL
Reportable RangeSimilar to or exceeding predicate (0.4 to 20.0 mg/dL) and sufficient for clinical use.0.5 to 15.0 mg/dL
In-house Precision (%CV)Comparable to predicate (%CVs range from 0.3% to 1.4%).%CVs range from 0.7% to 1.9% (Total Precision)
External Site Precision (%CV)Acceptable for physician's office laboratory (POL) setting. (No specific %CV range given as acceptance criteria)%CVs range from 1.4% to 7.2% (Total Precision across sites and levels)
InterferenceRecoveries between 90% and 110% of neat value with noted interfering substances.Unconjugated bilirubin, Lipemia, Ascorbic acid, Hemoglobin showed no interference at tested levels.
Method Comparison (n=97)Good correlation (r close to 1), slope close to 1, and y-intercept close to 0 when compared to a standard laboratory system.r = 0.976; Slope = 0.99 (95% CI: 0.94 to 1.03); y-intercept = -0.25 (95% CI: -0.68 to 0.18).
Matrix Comparison (Plasma)Good correlation (r close to 1), slope close to 1, and y-intercept close to 0 when compared to serum.Na Citrate Plasma: r = 0.993; Slope = 0.98; y-intercept = -0.14. Heparinized Plasma: r = 0.988; Slope = 0.96; y-intercept = -0.02.
External Method Comparison (POL Sites)Good correlation (r close to 1), slope close to 1, and y-intercept close to 0 when compared to a comparative method.Site 1: r=0.986, y=0.96x-0.2. Site 2: r=0.984, y=0.97x-0.1. Site 3: r=0.987, y=0.99x-0.1.

Study Details

This device is an in vitro diagnostic (IVD) chemistry analyzer for calcium, not an AI/ML device. Therefore, several of the requested sections regarding AI-specific studies (experts, adjudication, MRMC, training set) are not applicable.

  1. Sample size used for the test set and the data provenance:

    • Analytical Sensitivity (LOD): Not explicitly stated, but the study followed CLSI EP17-A.
    • Linearity: Not explicitly stated, but the study followed CLSI EP-6A.
    • 20-day In-house Precision: Three levels of samples, each tested in two runs, twice a day, for 20 days. (Total of 120 measurements per level).
    • Interference Testing: Two serum pools.
    • In-house Method Comparison: 97 clinical specimens (spanning 1.2 to 14.8 mg/dL).
    • Matrices Comparisons: 44 matched serum/plasma samples (spanning 0.9 to 14.7 mg/dL).
    • External Site Precision (Clinical Data): Three blinded serum samples (low, middle, high calcium concentrations) tested at each of three external POL sites. Each sample was assayed 6 times per day for 5 days, resulting in 30 results per level per site.
    • External Method Comparison (Clinical Data): Approximately 55 serum specimens (ranging from 1.3 to 14.6 mg/dL) at each of three external POL sites.
    • Data Provenance: Not explicitly stated for specific samples, but the studies were conducted by "Hitachi Chemical Diagnostics, Inc." (in-house) and at "three external POL-type sites." The specimens are clinical specimens (serum and plasma). The country of origin is not specified but implicitly the US given the FDA submission. All data appears to be prospective collection for the validation of this device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. This is an IVD device for quantitative chemical analysis; ground truth is established by the analytical reference methods or comparative methods used.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is an IVD device for quantitative chemical analysis.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is an IVD device for quantitative chemical analysis, not an AI-assisted diagnostic imaging device involving human readers.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, implicitly. All studies (precision, linearity, method comparisons, interference, matrix comparisons) evaluate the performance of the device itself (Hitachi E40 Clinical Analyzer with S TEST Reagent Cartridge Calcium) in an automated or semi-automated standalone fashion, generating quantitative results. Human intervention is limited to sample loading, programming, and result interpretation, not in the analytical process itself.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Method Comparison studies, "ground truth" is established by a "standard laboratory system" or "comparative method" (predicate or other established analyzer), generating quantitative calcium values.
    • For Analytical Sensitivity, Linearity, Precision, Interference, and Matrix Comparison studies, "ground truth" refers to the expected or known concentrations of calcium in control samples, calibrators, or spiked samples, verified through established analytical methods.
  7. The sample size for the training set:

    • Not applicable. This is an IVD device for quantitative chemical analysis, not an AI/ML device requiring a training set in the typical sense. The "training" of the device is its calibration and quality control procedures, which are part of standard IVD practices.
  8. How the ground truth for the training set was established:

    • Not applicable. (See above.)

{0}------------------------------------------------

SECTION 8 510(k) SUMMARY

MAY 2 2013

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K130870.

807.92 (a)(1): Name: Hitachi Chemical Diagnostics 630 Clyde Court Address: Mountain View, CA 94043

Phone: (650) 961 5501 (650) 969 2745 FAX: Contact: Mr. Charles Tsou charlest@hcdiagnostics.com

and

Name:Ammirati Regulatory Consulting
Phone:(650) 949-2768
FAX:(650) 949-5347
Contact:Ms. Erika Ammirati
erikaba@pacbell.net

807.92 (a)(2): Device name- trade name and common name, and classification

Trade name:

Hitachi S TEST Reagent Cartridge Calcium (CA) .

Common Name: Routine chemistry analyzer for calcium

Regulation: 21 CFR § 862.1145 Calcium test system

Device Classification: Class 2

807.92 (a)(3): Identification of the legally marketed predicate devices

Cobas c systems Calcium (CA) (Roche Diagnostics, Inc., Indianapolis, IN)-K100853

807.92 (a)(4): Device Description

The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and Page 1 of 6

O Hitachi Chemical Dlagnostics, Inc.

630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
www.hcdiagnostics.com
1994 (1992) 1999-1999 (1992) 1992 (1992) 1992 1992 1992 1992 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 1994 199

{1}------------------------------------------------

displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.

The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating. The dimensions of the reagent cartridges are: 13.5 mm (W) × 28.mm (D) × 20.2 mm (H).

System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions, it calculates the concentration of analyte in the sample. The test system can measure analytes in serum or plasma and results are available in approximately 15 minutes per test. This submission is for Reagent Cartridge Calcium.

Chemistry reactions: Calcium in the sample combines with o-cresolphthalein complexone to form a purple-red complex. The concentration of calcium can be determined by measuring the absorbance of the resulting purple-red color. (With this method, the influence of magnesium is excluded by the addition of 8-hydroxyquinoline.)

807.92 (a)(5): Intended Use

The S TEST Reagent Cartridge Calcium (CA) is intended for the quantitative determination of calcium in serum, lithium heparinized plasma, and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Calcium is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Page 2 of 6

O Hitachi Chemical Diagnostics, Inc. 630 Clyde Court, Mounlain View, CA 94043-2239 Tel: 800 233 6278

www.hcdiagnostics.com

i.e. the set of all elements of either set or of both sets. Other set operations include set intersection (denoted ∩), set difference (denoted ), and set complement (denoted ′).

{2}------------------------------------------------

807.92 (a)(6): Technological Similarities and Differences to the Predicate

The following chart describes similarities and differences between the calcium test systems.

CharacteristicHitachi S TEST SystemsPREDICATE
Instrument PlatformHitachi Clinical Analyzer(originally cleared under K111753)Roche cobas c systems - K100853
CalciumK number- pendingRoche K number- K100853
Device Class, Regulation CodeClass II, 21 CFR 862.1145Same
Classification Product CodeCICSame
Intended UseQuantitative determination of CASame
Testing EnvironmentPhysician office or clinical labClinical lab
Test PrincipleCalcium in the sample combineswith o-cresolphthalein complexoneto form a purple-red complex. Theconcentration of calcium can bedetermined by measuring theabsorbance of the resulting purple-red color. (With this method, theinfluence of magnesium is excludedby the addition of 8-hydroxyquinoline.)Calcium ions react with o-cresolphthalein complexone underalkaline conditions to form a violet-colored complex. The color intensityof the formed complex is directlyproportional to the calciumconcentration and is measuredphotometrically. The addition of 8-hydroxyquinoline preventsinterference by magnesium and iron.
Specimen TypeHuman serum or plasmaSame
Reportable Range0.5 to 15.0 mg/dL0.4 to 20.0 mg/dL
Detection Wavelength570/660 nm700/600 nm
Detection Limit0.4 mg/dL0.4 mg/dL
Linearity0.5 to 15.0 mg/dL0.4 to 20.0 mg/dL
Precision%CVs range from 1.0% to 1.9%%CVs range from 0.3% to 1.4%(from product labeling)

Page 3 of 6

ത്ര Hitachi Chemical Diagnostics, Inc.

630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

ком седантика се при седента се седения се седения представления се пристементации и пристементации продобразования продобразования продобразования продобразования продобразо
www.hcdiagnostics.com
ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーwww.ba.us.often.org/lower/responses/articles/a-c-a status and the contract and consisted on the commend of the program and the program and the many of the

{3}------------------------------------------------

807.92 (b)(1): Brief Description of Nonclinical Data

A series of studies were performed that evaluated the following nonclinical performance characteristics for each analytical sensitivity (limits of detection), linearity, 20-day in-house precision, interference testing, in-house method comparisons, and matrices comparison between serum and various plasma types.

Analytical Sensitivity (Limits of Detection)- CA

The study followed CLSI EP17-A, and the limit of detection was found to be 0.19 mg/dL. The quantitation limit was found to be 0.4 mg/dL

Linearity- CA

The study followed CLSI EP-6A, and the range of linearity was 0.4 to 17.1 mg/dL. The reportable range is 0.5 to 15.0 mg/dL.

20-day In-house Precision- CA

The studies followed CLSI EP5-A2, where three levels of samples were each tested in two runs, twice a day, for 20 days. The results were as follows:

Precision Summary:

Calcium- Low, Level 1, Summary

CalciumWithin-RunTotal
Mean (mg/dL)5.215.21
SD (mg/dL)0.070.10
%CV1.3%1.9%

Calcium- Middle, Level 2, Summary

CalciumWithin-RunTotal
Mean (mg/dL)10.0610.06
SD (mg/dL)0.110.14
%CV1.1%1.4%

Calcium- High, Level 3, Summary

CalciumWithin-RunTotal
Mean (mg/dL)13.0313.03
SD (mg/dL)0.090.14
%CV0.7%1.0%

Page 4 of 6

O Hitachl Chemical Diagnostics, Inc.

630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

Controller controller controllation of the management construction of the manufacturer in the consistence in the may and
www.hcdiaanostics.com
Con all controllection of the comments of any of any of any of any of any of the final of the final of the final

{4}------------------------------------------------

Interference Testing (per CLSI EP7-A2)

The data demonstrated that the calcium test system was not affected by high levels of the following substances at the levels noted:

Unconiugated bilirubin: no interference up to 50 mg/dL Lipemia: no interference up to 500 mg/dL Ascorbic acid: no interference up to 50 mg/dL Hemoglobin: no interference up to 500 mg/dL

Lack of interference was defined as recoveries between 90% and 110% of the neat value, and assay performance claims were established on the HITACHI Clinical Analyzer by testing two serum pools containing approximately 4 and 11 mg/dL calcium.

Method Comparison

A total of 97 clinical specimens spanning the dynamic range (1.2 to 14.8 mg/dL), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system and a standard laboratory system. The comparative data were analyzed by linear regression and are shown below. (CI = confidence interval)

Regression Statistics:

nrSlope(95% CI)y-intercept(95% CI)X meanY mean
970.9760.99(0.94 to 1.03)-0.25(-0.68 to 0.18)9.4 mg/dL9.0 mg/dL

Matrices Comparisons

A study was performed to validate the use of two plasma types as an alternative to serum for the Hitachi Clinical Analyzer with S TEST Reagent Cartridge Calcium. The plasma types were sodium citrate and lithium heparin. Forty-four (44) matched serum/plasma samples that spanned the dynamic range (0.9 to 14.7 mg/dL) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.

N = 44

Range (serum) = 0.9 to 14.7 mg/dL calcium

Na Citrate PlasmaHeparinized Plasma
Slope (95% CIs)0.98 (0.95 to 1.02)0.96 (0.91 to 1.00)
y-intercept (95% CIs)-0.14 (-0.48 to 0.19)-0.02 (-0.44 to 0.41)
r0.9930.988

807.92 (b)(2): Brief Description of Clinical Data

Studies for precision and method comparison (accuracy) were performed at three external POL-type sites to evaluate the Hitachi E40 Clinical Analyzer with S TEST Reagent Cartridge Calcium in one of its targeted intended use environments, the physician's office laboratory.

Page 5 of 6

O Hitachi Chemical Dlagnostics, Inc.

630 Clyde Court, Mountain Vlew, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

1. Partil Pro Pro Production Comments of Children Program Program Program Program Program Program Program Programment Programment Programment Programment Programment Programm
www.hcdiaonostics.com
1 (2011) 11:12 PM (2017) 11:12 PM (2017) 11:11) 11:11 PM (2019) 11:10 PM (2017) 11:11 PM (2016) 11:10 PM (2016) 11:10 PM (2014) 11:11

{5}------------------------------------------------

For the external site precision study, each site received three blinded serum samples (the Precision Panel, labeled A, B, and C) that were chosen to represent low, middle, and high concentrations of calcium. Each sample was assayed six times per day for five days, reporting 30 results per level. Precision estimates for total precision were as follows:

SiteSampleMeanWithin-run PrecisionTotal Precision
SD (U/L)%CVSD (U/L)%CV
1A5.30.071.30.112.1
2A5.40.061.20.224.1
3A5.10.224.40.295.6
1B9.20.111.20.141.5
2B9.40.091.00.262.7
3B8.70.252.80.414.7
1C12.30.110.90.181.4
2C12.60.141.10.252.0
3C11.60.453.90.847.2

CA (mg/dL) n = 30 replicates per sample per site

For the external method comparison studies, a series of approximately 55 serum specimens with calcium values ranging from 1.3 to 14.6 mg/dL, were assayed on the Hitachi E40 Clinical Analyzer at three sites using S TEST Reagent Cartridge Calcium (y) and a comparative method as the reference method (x). Linear regression analyses (least squares) yielded the following results:

Site #nRange(mg/dL)RegressionEquation"r"SE(mg/dL)CI*SlopeCI Intercept
1571.3 to 14.6y = 0.96x -0.20.9860.40.91 to 1.00-0.5 to 0.3
2551.4 to 14.0y = 0.97x -0.10.9840.50.91 to 1.03-0.6 to 0.5
3551.3 to 13.6y = 0.99x -0.10.9870.40.95 to 1.04-0.6 to 0.3

POL ACCURACY DATA SUMMARY- CA (mg/dL)

*95% Confidence Interval

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

Nonclinical and clinical testing was performed for the Hitachi Clinical Analyzer E40 with Reagent Cartridge Calcium. The test system was shown to be safe and effective for its intended use.

Page 6 of 6

O Hitachi Chemical Dlagnostics, Inc.

630 Clyde Court, Mountain Vlew, CA 94043-2239 Tel: 800 233 6278

www.hcdiagnostics.com"

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the agency's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2, 2013

Hitachi Chemical Diagnostics, Inc. C/O Ms. Erika Ammirati Ammirati Regulatory Consulting 575 Shirlynn Court LOS ALTOS CA 94022

Re: K130870

Trade/Device Name: S TEST Reagent Cartridge Calcium Regulation Number: 21 CFR 862.1145 Regulation Name: Calcium test system

  • Regulatory Class: II Product Code: CIC Dated: March 28, 2013 Received: March 29, 2013
    Dear Ms. Ammirati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol G. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE

K130870 510(k) Number (if Known):

Device Name:

S TEST Reagent Cartridge Calcium

Indications for Use:

The S TEST Reagent Cartridge Calcium (CA) is intended for the quantitative determination of calcium in serum, lithium heparinized plasma, and sodium citrate plasma using the HITACHI Clinical Analyzer E40. The S TEST Reagent Cartridge Calcium is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Yung MD Chan -S

Division Sign-Off "Office of In Vitro Diagnostics and Radiological Health (OIR)

510(k) K130870

Page 1 of 1

§ 862.1145 Calcium test system.

(a)
Identification. A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).(b)
Classification. Class II.