K Number
K130859

Validate with FDA (Live)

Date Cleared
2013-07-25

(119 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for small bone fusion, fractures and inter-digital fusion of the fingers, toes and small bones.

Device Description

The Arthrosurface Inc's Hammertoe Correction System consists of two intramedullary bone screws, a taper lock pin and a set of instruments used for implant site preparation and delivery. The taper lock pin provides a press fit connection between the two screws with light contact pressure. The implant components are manufactured using implant grade titanium alloy and cobaltchrome alloy.

AI/ML Overview

The provided text is a 510(k) summary for the Arthrosurface Inc's Hammertoe Correction System. This document focuses on demonstrating substantial equivalence to existing devices through comparative mechanical testing and cadaveric testing, rather than an independent study proving device performance against acceptance criteria with a focus on AI or algorithmic performance.

Therefore, many of the requested categories (especially those related to AI algorithm performance, ground truth, experts, and training/test sets) are not applicable to this type of regulatory submission as they are for a surgical implant, not an AI software medical device.

I will provide information relevant to the acceptance criteria and supporting studies as much as possible from the given text.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred)Reported Device Performance (Inferred)
Mechanical Equivalence:
- Comparable static cantilever bending strength"Results from static and dynamic cantilever bending, torque to failure, pull-out force, insertion-removal torque and axial disassembly force tests along with comparative dimensional analyses were used to support equivalence to predicate devices." The implication is that the device performed comparably to the predicate devices in these mechanical tests.
- Comparable dynamic cantilever bending strength
- Comparable torque to failure
- Comparable pull-out force
- Comparable insertion-removal torque
- Comparable axial disassembly force
Functional Equivalence:
- Performance in anatomical simulation (cadaveric testing)"Data from cadaveric testing were also reported." The implication is that the device performed acceptably in a simulated clinical setting, comparable to its predicates.
Material Equivalence:
- Use of implant-grade materials"The implant components are manufactured using implant grade titanium alloy and cobalt-chrome alloy." (This is a design specification, aligning with predicate devices).
Design/Intended Use Equivalence:
- Substantially similar intended use, design features, materials, and application to predicate devices.The FDA determined the device is "substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices." The text explicitly states: "The intended use, materials, design features and application of the Proposed Device are substantially equivalent to the following previously cleared and commercially marketed devices."

Study Proving Device Meets Acceptance Criteria:

The study proving the device meets the acceptance criteria is a Comparative Mechanical Testing study and Cadaveric Testing. The primary goal of these studies, in the context of a 510(k) submission, is to demonstrate substantial equivalence to predicate devices already on the market, not necessarily to meet pre-defined standalone performance metrics against an absolute standard.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated for specific tests. The text mentions "Comparative Mechanical Testing was performed per relevant recognized standards" and "Data from cadaveric testing were also reported." The sample sizes for these tests are generally dictated by the chosen standards (e.g., ISO, ASTM) or internal protocols for cadaveric work, but the exact numbers are not provided in this summary.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, given that this is a 510(k) submission to the FDA for a US market, it is reasonable to infer the testing was conducted in facilities compliant with US regulatory requirements (e.g., GLP for non-clinical testing). The studies would be considered prospective as they were conducted specifically to support this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is Not Applicable in this context. The "ground truth" for a mechanical implant device is primarily established through objective physical measurements and material science standards, not by expert consensus in a diagnostic sense. For cadaveric testing, the "ground truth" would be the observed biomechanical performance or failure modes. There is no mention of human interpretation being the primary ground truth.

4. Adjudication Method for the Test Set

This section is Not Applicable. Adjudication methods like 2+1 or 3+1 refer to how discrepant expert opinions are resolved, which is relevant for diagnostic or AI performance evaluations, not for mechanical testing of an implant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This section is Not Applicable. This request pertains to AI software performance evaluation. The submitted device is a physical surgical implant, not an AI or diagnostic imaging device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This section is Not Applicable. The device is a surgical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is primarily:

  • Mechanical Performance Data: Objective measurements obtained from standardized mechanical tests (static and dynamic cantilever bending, torque to failure, pull-out force, insertion-removal torque, axial disassembly force).
  • Material Specifications: Adherence to implant-grade material standards (titanium alloy, cobalt-chrome alloy).
  • Biomechanical Observation/Data: Observations and measurements from cadaveric testing that simulate in-vivo conditions.
  • Comparative Dimensional Analysis: Measurement and comparison of device dimensions to predicate devices.

8. The Sample Size for the Training Set

This section is Not Applicable. There is no "training set" in the context of mechanical testing for a physical implant. The device design and materials are based on established engineering principles and prior knowledge.

9. How the Ground Truth for the Training Set Was Established

This section is Not Applicable. As there is no training set for a mechanical implant, there is no ground truth to establish for it.

{0}------------------------------------------------

Section 5 510(k) Summary

510(k) Owner:Arthrosurface, Inc.28 Forge ParkwayFranklin, MA 02038Tel: 508.520.3003Fax: 508.528.4604JUL 2 5 2013
Contact:Dawn WilsonVP, Quality & Regulatory
Date of Preparation:March 26, 2013
Trade Name:Hammertoe Correction System
Common Name:Intramedullary Bone Screw
Device:Screw, Fixation, Bone
Regulation Description:Smooth or threaded metallic bone fixationfastener
Regulation Number:888.3040
Device Class:Class II
Review Panel:Orthopedic
Product Code:HWC

Intended Use

Indicated for small bone fusion, fractures and inter-digital fusion of the fingers, toes and small bones.

Device Description

The Arthrosurface Inc's Hammertoe Correction System consists of two intramedullary bone screws, a taper lock pin and a set of instruments used for implant site preparation and delivery. The taper lock pin provides a press fit connection between the two screws with light contact pressure. The implant components are manufactured using implant grade titanium alloy and cobaltchrome alloy.

{1}------------------------------------------------

Substantial Equivalency

The intended use, materials, design features and application of the Proposed Device are substantially equivalent to the following previously cleared and commercially marketed devices:

• Nextremity Solutions, LLCFlexFusion™ Fixation ImplantK110445
• Wright Medical Technology, Inc.PRO-TOE™ VO Hammertoe Implant SystemK101165
  • K990804 Tornier, Inc. . . StayFuse
    Comparative Mechanical Testing was performed per relevant recognized standards. Results from static and dynamic cantilever bending, torque to failure, pull-out force, insertion-removal torque and axial disassembly force tests along with comparative dimensional analyses were used to support equivalence to predicate devices. Data from cadaveric testing were also reported.

The fundamental scientific technology of the proposed device has not changed relative to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized symbol that resembles an abstract human figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 25, 2013

Arthrosurface, Incorporated % Ms. Dawn Wilson Vice President, Quality & Regulatory 28 Forge Parkway Franklin, Massachusetts 02038

Re: K130859

Trade/Device Name: Hammertoe Correction System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: June 6, 2013 Received: June 11, 2013

Dear Ms. Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Dawn Wilson

forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/dcfault.htm.

Sincerely yours,

Erin I. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4 Indications for Use Statement $ 1

510(k) Number (if known): _K130859

Device Name: Hammertoe Correction System

Indications for Use:

Indicated for small bone fusion, fractures and inter-digital fusion of the fingers, toes and small bones.

Prescription Use(Part 21 CFR 801 Subpart D)________________________________
-----------------------------------------------------------------------------------

AND/OR

Over-The-Counter(21 CFR 8)
----------------------------------

Counter Use 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lori A. Wiggins

Division of Orthopedic Devices

Page 1 of 1

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.